Previous classification systems of failure of
limb salvage focused primarily on endoprosthetic failures and lacked sufficient
depth for the effective study of the causes of failure. In order
to address these inadequacies, the International Society of Limb
Salvage (ISOLS) formed a committee to recommend revisions of the
previous systems. The purpose of this study was to report on their
recommendations. The modifications were prepared using an earlier,
evidence-based model with subclassification based on the existing
medical literature. Subclassification for all five primary types
of failure of limb salvage following endoprosthetic reconstruction
were formulated and a complementary system was derived for the failure
of biological reconstruction. An additional classification of failure
in paediatric patients was also described. Limb salvage surgery presents a complex array of potential mechanisms
of failure, and a complete and precise classification of types of
failure is required. Earlier classification systems lacked specificity,
and the evidence-based system outlined here is designed to correct
these weaknesses and to provide a means of reporting failures of
limb salvage in order to allow the interpretation of outcome following
reconstructive surgery. Cite this article:
Total knee replacement (TKR) smart tibial trials
have load-bearing sensors which will show quantitative compartment
pressure values and femoral-tibial tracking patterns. Without smart
trials, surgeons rely on feel and visual estimation of imbalance
to determine if the knee is optimally balanced. Corrective soft-tissue
releases are performed with minimal feedback as to what and how
much should be released. The smart tibial trials demonstrate graphically
where and how much imbalance is present, so that incremental releases
can be performed. The smart tibial trials now also incorporate accelerometers
which demonstrate the axial alignment. This now allows the surgeon
the option to perform a slight recut of the tibia or femur to provide
soft-tissue balance without performing soft-tissue releases. Using
a smart tibial trial to assist with soft-tissue releases or bone
re-cuts, improved patient outcomes have been demonstrated at one
year in a multicentre study of 135 patients (135 knees). Cite this article:
The August 2014 Knee Roundup360 looks at: re-admission following total knee replacement; out with the old and in with the new? computer navigation revisited; approach less important in knee replacement; is obesity driving a rise in knee replacements?; knee replacement isn’t cheap in the obese; cruciate substitution doesn’t increase knee flexion; and sonication useful diagnostic aid in two-stage revision.
We reviewed the long-term radiological outcome,
complications and revision operations in 19 children with quadriplegic
cerebral palsy and hip dysplasia who underwent combined peri-iliac
osteotomy and femoral varus derotation osteotomy. They had a mean
age of 7.5 years (1.6 to 10.9) and comprised 22 hip dislocations
and subluxations. We also studied the outcome for the contralateral
hip. At a mean follow-up of 11.7 years (10 to 15.1) the Melbourne
cerebral palsy (CP) hip classification was grade 2 in 16 hips, grade
3 in five, and grade 5 in one. There were five complications seen
in four hips (21%, four patients), including one dislocation, one
subluxation, one coxa vara with adduction deformity, one subtrochanteric
fracture and one infection. A recurrent soft-tissue contracture occurred
in five hips and ten required revision surgery. In pre-adolescent children with quadriplegic cerebral palsy good
long-term outcomes can be achieved after reconstruction of the hip;
regular follow-up is required.
Cite this article:
The June 2014 Spine Roundup360 looks at: spinal pedicle screws in paediatric patients; improving diagnosis in lumbar spine stenosis; back pain all in the head?; brace three patients, save one scoliosis operation; pedicle screws more often misplaced than one would think; and incidental dural tears usually no problem
We describe the results of a prospective case series of patients with spondylolysis, evaluating a technique of direct stabilisation of the pars interarticularis with a construct that consists of a pair of pedicle screws connected by a U-shaped modular link passing beneath the spinous process. Tightening the link to the screws compresses bone graft in the defect in the pars, providing rigid intrasegmental fixation. We have carried out this procedure on 20 patients aged between nine and 21 years with a defect of the pars at L5, confirmed on CT. The mean age of the patients was 13.9 years (9 to 21). They had a grade I or less spondylolisthesis and no evidence of intervertebral degeneration on MRI. The mean follow-up was four years (2.3 to 7.3). The patients were assessed by the Oswestry Disability Index (ODI) and a visual analogue scale (VAS). At the latest follow-up, 18 patients had an excellent clinical outcome, with a significant (p <
0.001) improvement in their ODI and VAS scores. The mean ODI score at final follow-up was 8%. Assessment of the defect by CT showed a rate of union of 80%. There were no complications involving the internal fixation. The strength of the construct removes the need for post-operative immobilisation.
Orthopaedic surgeons use stems in revision knee surgery to obtain
stability when metaphyseal bone is missing. No consensus exists
regarding stem size or method of fixation. This A custom test rig using differential variable reluctance transducers
(DVRTs) was developed to record all translational and rotational
motions at the bone–implant interface. Composite femurs were used.
These were secured to permit variation in flexion angle from 0°
to 90°. Cyclic loads were applied through a tibial component based
on three peaks corresponding to 0°, 10° and 20° flexion from a normal
walking cycle. Three different femoral components were investigated
in this study for cementless and cemented interface conditions.Objectives
Methods
The technical advances in arthroscopic surgery
of the hip, including the improved ability to manage the capsule
and gain extensile exposure, have been paralleled by a growth in
the number of conditions that can be addressed. This expanding list
includes symptomatic labral tears, chondral lesions, injuries of
the ligamentum teres, femoroacetabular impingement (FAI), capsular
laxity and instability, and various extra-articular disorders, including snapping
hip syndromes. With a careful diagnostic evaluation and technical
execution of well-indicated procedures, arthroscopic surgery of
the hip can achieve successful clinical outcomes, with predictable
improvements in function and pre-injury levels of physical activity
for many patients. This paper reviews the current position in relation to the use
of arthroscopy in the treatment of disorders of the hip. Cite this article:
We reviewed 31 consecutive patients with Friedreich’s
ataxia and scoliosis. There were 24 males and seven females with
a mean age at presentation of 15.5 years (8.6 to 30.8) and a mean
curve of 51° (13° to 140°). A total of 12 patients had thoracic
curvatures, 11 had thoracolumbar and eight had double thoracic/lumbar.
Two patients had long thoracolumbar collapsing scoliosis with pelvic
obliquity and four had hyperkyphosis. Left-sided thoracic curves in
nine patients (45%) and increased thoracic kyphosis differentiated
these deformities from adolescent idiopathic scoliosis. There were
17 patients who underwent a posterior instrumented spinal fusion
at mean age of 13.35 years, which achieved and maintained good correction
of the deformity. Post-operative complications included one death due
to cardiorespiratory failure, one revision to address nonunion and
four patients with proximal junctional kyphosis who did not need
extension of the fusion. There were no neurological complications
and no wound infections. The rate of progression of the scoliosis
in children kept under simple observation and those treated with bracing
was less for lumbar curves during bracing and similar for thoracic
curves. The scoliosis progressed in seven of nine children initially
treated with a brace who later required surgery. Two patients presented
after skeletal maturity with balanced curves not requiring correction.
Three patients with severe deformities who would benefit from corrective
surgery had significant cardiac co-morbidities.
We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip. The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° ( Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.
The spinal manifestations of neurofibromatosis include cervicothoracic kyphosis, in which scalloping of the vertebral body and erosion of the pedicles may render conventional techniques of fixation impossible. We describe a case of cervicothoracic kyphosis managed operatively with a vascularised fibular graft anteriorly across the apex of the kyphus, followed by a long posterior construct using translaminar screws, which allow segmental fixation in vertebral bodies where placement of the pedicle screws was impracticable.
Inherent disadvantages of reverse shoulder arthroplasty
designs based on the Grammont concept have raised a renewed interest
in less-medialised designs and techniques. The aim of this study
was to evaluate the outcome of reverse shoulder arthroplasty (RSA)
with the fully-constrained, less-medialised, Bayley–Walker prosthesis performed
for the treatment of rotator-cuff-deficient shoulders with glenohumeral
arthritis. A total of 97 arthroplasties in 92 patients (53 women
and 44 men, mean age 67 years (standard deviation ( The Bayley–Walker prosthesis provides reliable pain relief and
reasonable functional improvement for patients with symptomatic
cuff-deficient shoulders. Compared with other designs of RSA, it
offers a modest improvement in forward elevation, but restores external
rotation to some extent and prevents scapular notching. A longer
follow-up is required to assess the survival of the prosthesis and
the clinical performance over time. Cite this article:
In a randomised trial involving 598 patients
with 600 trochanteric fractures of the hip, the fractures were treated with
either a sliding hip screw (n = 300) or a Targon PF intramedullary
nail (n = 300). The mean age of the patients was 82 years (26 to
104). All surviving patients were reviewed at one year with functional
outcome assessed by a research nurse blinded to the treatment used.
The intramedullary nail was found to have a slightly increased mean
operative time (46 minutes ( In summary, both implants produced comparable results but there
was a tendency to better return of mobility for those treated with
the intramedullary nail.
A combination of hemivertebrae and diastematomyelia is rare. We have identified 12 such patients seen during a period of 11 years in the orthopaedic, spinal and neurosurgical units in Nottingham and analysed their treatment and outcome.
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
We performed a case–control study to compare
the rates of further surgery, revision and complications, operating time
and survival in patients who were treated with either an uncemented
hydroxyapatite-coated Corail bipolar femoral stem or a cemented
Exeter stem for a displaced intracapsular fracture of the hip. The
mean age of the patients in the uncemented group was 82.5 years
(53 to 97) and in the cemented group was 82.7 years (51 to 99) We used
propensity score matching, adjusting for age, gender and the presence
or absence of dementia and comorbidities, to produce a matched cohort
receiving an Exeter stem (n = 69) with which to compare the outcome of
patients receiving a Corail stem (n = 69). The Corail had a significantly
lower all-cause rate of further surgery (p = 0.016; odds ratio (OR)
0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further
surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating
time was significantly less for the Corail group than for the cemented Exeter
group (59 min [12 to 136] Cite this article:
We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm3) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations. In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.
This is the second of a series of reviews of registries. This review looks specifically at worldwide registry data that have been collected on knee arthroplasty, what we have learned from their reports, and what the limitations are as to what we currently know.
We present an illustrative case using a modification of the Gaines procedure for the surgical management of patients with spondyloptosis. It involves excision of the inferior half of the body of L5 anteriorly combined with posterior reduction and fusion.