Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
PCL retaining fixed-bearing TKA is a highly successful
operation with the need for more surgery occurring at the rate of
approximately 0.4% per year over the first 27 years. The most common
cause for revision surgery is related to polyethylene insert failure
and accounts for approximately 50% of re-operations. Late metastatic
infection is the next most frequent cause followed by patellar problems,
late instability and component loosening in decreasing frequency.
A myriad of rare miscellaneous problems can also occur.
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
The June 2015 Knee Roundup360 looks at: Cruciate substituting
Current analysis of unicondylar knee replacements
(UKRs) by national registries is based on the pooled results of medial
and lateral implants. Consequently, little is known about the differential
performance of medial and lateral replacements and the influence
of each implant type within these pooled analyses. Using data from
the National Joint Registry for England and Wales (NJR) we aimed
to determine the proportion of UKRs implanted on the lateral side
of the knee, and their survival and reason for failure compared
with medial UKRs. By combining information on the side of operation
with component details held on the NJR, we were able to determine
implant laterality (medial
We evaluated the incidence of heterotopic ossification
following total ankle replacement to determine whether the degree
of ossification was associated with the clinical outcome. We evaluated
90 ankles in 81 consecutive patients who underwent total ankle replacement,
and heterotopic ossification was assessed according to proportional involvement
of the ankle joint. Correlation analysis was used to investigate
the association between heterotopic ossification and outcome. No significant association was found between the formation of
heterotopic ossification and the clinical outcome. The degree of
heterotopic ossification in the posterior ankle joint was not significantly
correlated with posterior ankle pain (p = 0.929), the American Orthopaedic
Foot and Ankle Society score (p = 0.454) or range of movement (p
= 0.283). This study indicates that caution should be observed in attributing
symptoms and functional limitation to the presence of heterotopic
ossification in the posterior ankle joint when considering excision
of heterotopic bone after total ankle replacement.
The aim of our study was to describe the characteristics,
treatment, and outcomes of patients with periprosthetic joint infection
(PJI) and normal inflammatory markers after total knee arthroplasty
(TKA) and total hip arthroplasty (THA). In total 538 TKAs and 414 THAs underwent surgical treatment for
PJI and met the inclusion criteria. Pre-operative erythrocyte sedimentation
rate (ESR) and C-reactive protein level (CRP) were reviewed to identify
the seronegative cohort. An age- and gender-matched cohort was identified
from the remaining patients for comparison. Overall, 4% of confirmed
infections were seronegative (21 TKA and 17 THA). Of those who underwent
pre-operative aspiration, cultures were positive in 76% of TKAs
(n = 13) and 64% of THAs (n = 7). Cell count and differential were
suggestive of infection in 85% of TKA (n = 11) and all THA aspirates
(n = 5). The most common organism was coagulase-negative Cite this article:
As many as 25% to 40% of unicompartmental knee
replacement (UKR) revisions are performed for pain, a possible cause
of which is proximal tibial strain. The aim of this study was to
examine the effect of UKR implant design and material on cortical
and cancellous proximal tibial strain in a synthetic bone model.
Composite Sawbone tibiae were implanted with cemented UKR components
of different designs, either all-polyethylene or metal-backed. The tibiae
were subsequently loaded in 500 N increments to 2500 N, unloading
between increments. Cortical surface strain was measured using a
digital image correlation technique. Cancellous damage was measured
using acoustic emission, an engineering technique that detects sonic
waves (‘hits’) produced when damage occurs in material. Anteromedial cortical surface strain showed significant differences
between implants at 1500 N and 2500 N in the proximal 10 mm only
(p <
0.001), with relative strain shielding in metal-backed implants.
Acoustic emission showed significant differences in cancellous bone
damage between implants at all loads (p = 0.001). All-polyethylene implants
displayed 16.6 times the total number of cumulative acoustic emission
hits as controls. All-polyethylene implants also displayed more
hits than controls at all loads (p <
0.001), more than metal-backed
implants at loads ≥ 1500 N (p <
0.001), and greater acoustic
emission activity on unloading than controls (p = 0.01), reflecting
a lack of implant stiffness. All-polyethylene implants were associated
with a significant increase in damage at the microscopic level compared
with metal-backed implants, even at low loads. All-polyethylene
implants should be used with caution in patients who are likely
to impose large loads across their knee joint. Cite this article:
Minimally invasive total knee replacement (MIS-TKR)
has been reported to have better early recovery than conventional
TKR. Quadriceps-sparing (QS) TKR is the least invasive MIS procedure,
but it is technically demanding with higher reported rates of complications
and outliers. This study was designed to compare the early clinical
and radiological outcomes of TKR performed by an experienced surgeon
using the QS approach with or without navigational assistance (NA),
or using a mini-medial parapatellar (MP) approach. In all, 100 patients
completed a minimum two-year follow-up: 30 in the NA-QS group, 35
in the QS group, and 35 in the MP group. There were no significant
differences in clinical outcome in terms of ability to perform a
straight-leg raise at 24 hours (p = 0.700), knee score (p = 0.952),
functional score (p = 0.229) and range of movement (p = 0.732) among
the groups. The number of outliers for all three radiological parameters
of mechanical axis, frontal femoral component alignment and frontal
tibial component alignment was significantly lower in the NA-QS
group than in the QS group (p = 0.008), but no outlier was found
in the MP group. In conclusion, even after the surgeon completed a substantial
number of cases before the commencement of this study, the supplementary
intra-operative use of computer-assisted navigation with QS-TKR
still gave inferior radiological results and longer operating time,
with a similar outcome at two years when compared with a MP approach. Cite this article:
We report the clinical and radiological outcome of total ankle replacement performed in conjunction with hindfoot fusion or in isolation. Between May 2003 and June 2008, 60 ankles were treated with total ankle replacement with either subtalar or triple fusion, and the results were compared with a control group of 288 ankles treated with total ankle replacement alone. After the mean follow-up of 39.5 months (12 to 73), the ankles with hindfoot fusion showed significant improvement in the mean visual analogue score for pain (p <
0.001), the mean American Orthopaedic Foot and Ankle Society score (p <
0.001), and the mean of a modified version of this score (p <
0.001). The mean visual analogue pain score (p = 0.304) and mean modified American Orthopaedic Foot and Ankle Society score (p = 0.119) were not significantly different between the hindfoot fusion and the control groups. However, the hindfoot fusion group had a significantly lower mean range of movement (p = 0.009) and a higher rate of posterior focal osteolysis (p = 0.04). Both groups showed various complications (p = 0.131) and failure occurring at a similar rate (p = 0.685). Subtalar or triple fusion is feasible and has minimal adverse effects on ankles treated with total ankle replacement up to midterm follow-up. The clinical outcome of total ankle replacement when combined with hindfoot fusion is comparable to that of ankle replacement alone. Thus, hindfoot fusion should be performed in conjunction with total ankle replacement when indicated.
Revision knee arthroplasty presents a number
of challenges, not least of which is obtaining solid primary fixation
of implants into host bone. Three anatomical zones exist within
both femur and tibia which can be used to support revision implants.
These consist of the joint surface or epiphysis, the metaphysis
and the diaphysis. The methods by which fixation in each zone can
be obtained are discussed. The authors suggest that solid fixation
should be obtained in at least two of the three zones and emphasise
the importance of pre-operative planning and implant selection. Cite this article:
The February 2015 Foot &
Ankle Roundup360 looks at: Syndesmosis screw removal in randomised controlled trial; Diagnostic value of Hawkins sign; Chevron rules supreme?; Diabetes and ankle replacement; Fixed-bearing ankle replacement; Fusion for osteomyelitis of the ankle; ‘Reformed’ fallers.
The Oxford unicompartmental knee replacement gives good results in patients with symptomatic osteoarthritis of the medial compartment. Previous studies have suggested that the presence of radiolucent lines (RLLs) does not reflect a poor outcome in such patients. However, the reliability and validity of this assessment have not been determined. Our aim was to assess the intra- and interobserver reliability and the sensitivity and specificity of the assessment of RLLs around both tibial and femoral components using standard radiographs. Two reviewers assessed the radiographs of 45 patients who had loosening of the tibial or femoral component confirmed at revision surgery and compared them with those of a series of 45 asymptomatic patients matched for age and gender. The results suggested that, using standard radiographs, tibial RLLs were 63.6% sensitive and 94.4% specific and femoral RLLs 63.9% sensitive and 72.7% specific for loosening. Overall intra- and interobserver reliability was highly variable, but zonal analysis showed that lucency at the tip of the femoral peg was significantly associated with loosening of the femoral component. Fluoroscopically guided radiographs may improve the zonal reliability of the assessment of RLLs, but further independent and comparative studies are required. In the meantime, the innocence of the physiological RLLs detected by standard radiographs should be viewed with caution.
The purpose of this study was to examine the effect of posterior
cruciate ligament (PCL) retention, PCL recession, and PCL excision
during cruciate-retaining total knee replacement. A total of 3018 anatomic graduated component total knee replacements
were examined; 1846 of these retained the PCL, 455 PCLs were partially
recessed, and in 717 the PCL was completely excised from the back
of the tibia.Objectives
Methods
Revision total knee arthroplasty (TKA) is a complex
procedure which carries both a greater risk for patients and greater
cost for the treating hospital than does a primary TKA. As well
as the increased cost of peri-operative investigations, blood transfusions,
surgical instrumentation, implants and operating time, there is
a well-documented increased length of stay which accounts for most
of the actual costs associated with surgery. We compared revision surgery for infection with revision for
other causes (pain, instability, aseptic loosening and fracture).
Complete clinical, demographic and economic data were obtained for
168 consecutive revision TKAs performed at a tertiary referral centre
between 2005 and 2012. Revision surgery for infection was associated with a mean length
of stay more than double that of aseptic cases (21.5 Current NHS tariffs do not fully reimburse the increased costs
of providing a revision knee surgery service. Moreover, especially
as greater costs are incurred for infected cases. These losses may
adversely affect the provision of revision surgery in the NHS. Cite this article:
Bone defects are occasionally encountered during
primary total knee replacement (TKR) and cause difficulty in establishing
a stable well-aligned bone-implant interface. Between March 1999
and November 2005, 59 knees in 43 patients underwent primary TKR
with a metal block augmentation for tibial bone deficiency. In all,
six patients (eight knees) died less than four years post-operatively,
and four patients (five knees) were lost to follow-up leaving 46 knees
in 33 patients available for review at a mean of 78.6 months (62
to 129). The clinical results obtained, including range of movement,
American Knee Society and Oxford knee scores, and the Western Ontario
and McMaster Universities osteoarthritis index, were good to excellent,
with no failures. Radiolucent lines at the block-cement-bone interface
were noted in five knees (11%) during the first post-operative year,
but these did not progress. Modular rectangular metal augmentation for tibial bone deficiency
is a useful option. No deterioration of the block-prosthesis or
block-cement-bone interface was seen at minimum of five years follow-up.
Radiological evidence of post-traumatic osteoarthritis
(PTOA) after fracture of the tibial plateau is common but end-stage arthritis
which requires total knee arthroplasty is much rarer. The aim of this study was to examine the indications for, and
outcomes of, total knee arthroplasty after fracture of the tibial
plateau and to compare this with an age and gender-matched cohort
of TKAs carried out for primary osteoarthritis. Between 1997 and 2011, 31 consecutive patients (23 women, eight
men) with a mean age of 65 years (40 to 89) underwent TKA at a mean
of 24 months (2 to 124) after a fracture of the tibial plateau.
Of these, 24 had undergone ORIF and seven had been treated non-operatively.
Patients were assessed pre-operatively and at 6, 12 and >
60 months
using the Short Form-12, Oxford Knee Score and a patient satisfaction
score. Patients with instability or nonunion needed total knee arthroplasty
earlier (14 and 13.3 months post-injury) than those with intra-articular
malunion (50 months, p <
0.001). Primary cruciate-retaining implants
were used in 27 (87%) patients. Complication rates were higher in
the PTOA cohort and included wound complications (13% Total knee arthroplasty undertaken after fracture of the tibial
plateau has a higher rate of complications than that undertaken
for primary osteoarthritis, but patient-reported outcomes and satisfaction
are comparable. Cite this article:
The use of hinged implants in primary total knee
replacement (TKR) should be restricted to selected indications and mainly
for elderly patients. Potential indications for a rotating hinge
or pure hinge implant in primary TKR include: collateral ligament
insufficiency, severe varus or valgus deformity (>
20°) with necessary
relevant soft-tissue release, relevant bone loss including insertions
of collateral ligaments, gross flexion-extension gap imbalance,
ankylosis, or hyperlaxity. Although data reported in the literature
are inconsistent, clinical results depend on implant design, proper
technical use, and adequate indications. We present our experience
with a specific implant type that we have used for over 30 years
and which has given our elderly patients good mid-term results.
Because revision of implants with long cemented stems can be very
challenging, an effort should be made in the future to use shorter stems
in modular versions of hinged implants. Cite this article:
The lateral compartment is predominantly affected
in approximately 10% of patients with osteoarthritis of the knee. The
anatomy, kinematics and loading during movement differ considerably
between medial and lateral compartments of the knee. This in the
main explains the relative protection of the lateral compartment
compared with the medial compartment in the development of osteoarthritis.
The aetiology of lateral compartment osteoarthritis can be idiopathic,
usually affecting the femur, or secondary to trauma commonly affecting
the tibia. Surgical management of lateral compartment osteoarthritis
can include osteotomy, unicompartmental knee replacement and total
knee replacement. This review discusses the biomechanics, pathogenesis
and development of lateral compartment osteoarthritis and its management. Cite this article: