Foreign-body reaction to polyglycolide (PGA) implants has been described in man. Many animal experiments have verified the mechanical properties of fixation devices made from PGA, but a significant foreign-body reaction has not been described. We studied the effect of PGA rods in 12 sheep with standardized osteochondral fractures of the medial femoral condyle fixed with uncoloured, self-reinforced PGA rods (Biofix). Radiographs were taken at intervals ranging from two weeks to two years, and the sheep were killed at intervals ranging from six to 24 months. All knees were examined histologically. Eleven of the 12 fractures healed radiologically and histologically. Moderate to severe
The aim of this study was to analyze the results of two radiocapitellar prostheses in a large case series followed prospectively, with medium-term follow-up. A total of 31 patients with a mean age of 54 years (27 to 73) were analyzed; nine had primary osteoarthritis (OA) and 17 had post-traumatic OA, three had capitellar osteonecrosis, and two had a fracture. Overall, 17 Lateral Resurfacing Elbow (LRE) and 14 Uni-Elbow Radio-Capitellum Implant (UNI-E) arthroplasties were performed. Pre- and postoperative assessment involved the Mayo Elbow Performance Score (MEPS), the Quick Disabilities of the Arm, Shoulder and Hand (Q-DASH) score, and the modified American Shoulder Elbow Surgeons (m-ASES) score.Aims
Patients and Methods
The aims of this study were to determine the clinical and radiographic
outcomes, implant survivorship, and complications of patients with
a history of poliomyelitis undergoing total hip arthroplasty (THA)
in affected limbs and unaffected limbs of this same population. A retrospective review identified 51 patients (27 male and 24
female, 59 hips) with a mean age of 66 years (38 to 88) and with
the history of poliomyelitis who underwent THA for degenerative
arthritis between 1970 and 2012. Immigrant status, clinical outcomes,
radiographic results, implant survival, and complications were recorded.Aims
Patients and Methods
A heavy infiltrate of foreign-body macrophages is commonly seen in the fibrous membrane which surrounds an aseptically loose cemented implant. This is in response to particles of polymethylmethacrylate (PMMA) bone cement and other biomaterials. We have previously shown that monocytes and macrophages responding to particles of bone cement are capable of differentiating into osteoclastic cells which resorb bone. To determine whether the radio-opaque additives barium sulphate (BaSO. 4. ) and zirconium dioxide (ZrO. 2. ) influence this process, particles of PMMA with and without these agents were added to mouse monocytes and cocultured with osteoblast-like cells on bone slices. Osteoclast differentiation, as shown by the presence of the osteoclast-associated enzyme tartrate-resistant acid phosphatase (TRAP) and lacunar bone resorption, was observed in all cocultures. The addition of PMMA alone to these cocultures caused no increase in TRAP expression or bone resorption relative to control cocultures. Adding PMMA particles containing BaSO. 4. or ZrO. 2. , however, caused an increase in TRAP expression and a highly significant increase in bone resorption. Particles containing BaSO. 4. were associated with 50% more bone resorption than those containing ZrO. 2. . Our results suggest that radio-opaque agents in bone cement may contribute to the bone resorption of aseptic loosening by enhancing macrophage-osteoclast differentiation, and that PMMA containing is BaSO. 4. likely to be associated with more
In previous studies, we identified multiple factors influencing the survivorship of hip resurfacing arthroplasties (HRAs), such as initial anatomical conditions and surgical technique. In addition, the University of California, Los Angeles (UCLA) activity score presents a ceiling effect, so a better quantification of activity is important to determine which activities may be advisable or detrimental to the recovered patient. We aimed to determine the effect of specific groups of sporting activities on the survivorship free of aseptic failure of a large series of HRA. A total of 661 patients (806 hips) representing 77% of a consecutive series of patients treated with metal-on-metal hybrid HRA answered a survey to determine the types and amounts of sporting activities they regularly participated in. There were 462 male patients (70%) and 199 female patients (30%). Their mean age at the time of surgery was 51.9 years (14 to 78). Their mean body mass index (BMI) was 26.5 kg/m2 (16.7 to 46.5). Activities were regrouped into 17 categories based on general analogies between these activities. Scores for typical frequency and duration of the sessions were used to quantify the patients’ overall time spent engaging in sporting activities. Impact and cycle scores were computed. Multivariable models were used.Aims
Patients and Methods
The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation.Objectives
Methods
Our aim was to determine if the serum levels of bone-resorbing cytokines (IL-1β, TNF-α, IL-6, GM-CSF) are altered in patients with aseptic loosening of a total hip prosthesis, and if such levels are influenced by the type of implant. We determined cytokine levels in sera from 35 patients before revision for failed total hip arthroplasty and compared them with those in 25 healthy donors. We also assessed the soluble receptor of interleukin-2 (sIL-2r) in serum as an indication of a specific immune reaction against the implant. Our findings showed that the sIL-2r and TNF-α serum level did not change. The IL-6 level was not significantly altered, but was higher in patients with TiAlV prostheses than in those with a CrCoMo implant and in patients with cemented prostheses. The IL-1β level was found to be higher in those with a TiAlV cemented prosthesis than in the control group (p = 0.0001) and other groups of patients (p = 0.003 v uncemented TiAlV, p = 0.01 v cemented CrCoMo, p = 0.001 v uncemented CrCoMo). The GM-CSF level significantly increased in patients compared with healthy subjects (p = 0.008), and it was higher in those with cemented than with uncemented implants (p = 0.01). Only patients with cementless CrCoMo prostheses had levels of GM-CSF similar to those of the control group. The highest GM-CSF concentrations were observed in patients treated with non-steroidal anti-inflammatory drugs (NSAIDs) in the last months before revision (p = 0.04). In addition, when massive
Ceramic bearings have several desirable properties, such as resistance to wear, hardness, and biocompatibility, that favour it as an articulating surface in hip arthroplasty. However, ceramic fracture remains a concern. We have reviewed the contemporary literature, addressing the factors that can influence the incidence of ceramic bearing surface fracture. Cite this article:
The Alloclassic and Endoplus femoral stems have the same grit-blasted surface and are hot forged from the same titanium alloy. Only the external form of the implants differs slightly. It was our aim to examine the differences in radiographic bone response between the Alloclassic (second generation) and the Endoplus (third generation) femoral stems. We compared 79 prostheses in 70 matched patients studied over a minimum of two years. Radiolucent lines, adaptive bone remodelling, subsidence, heterotopic bone formation and lysis were recorded in the Gruen zones. Radiolucencies were mainly found in zones 1 and 7 but to a greater extent in the Endoplus than in the Alloclassic group (p <
0.001 in zone 1, p <
0.05 in zone 7). We found lucent lines in three or more Gruen zones in seven patients all of whom were in the Endoplus group (p <
0.05). Zones 2 and 6 had a significantly higher rate of lucencies in the Endoplus group (p <
0.001). We encountered a combination of proximal lucent lines in zones 1 and 7 with distal hypertrophy of the cortical bone in zones 2, 3, 5 and 6 in eight patients, all from the Endoplus group (p <
0.05). In other patients bone atrophy (stress shielding) in zones 2 and 6 was seen more frequently in the Endoplus than in the Alloclassic group (p <
0.001). In neither group was there radiological evidence of
The pathogenesis of aseptic loosening of total joint prostheses is not clearly understood. Two features are associated with loosened prostheses, namely, particulate debris and movement of the implant. While numerous studies have evaluated the cellular response to particulate biomaterials, few have investigated the influence of movement of the implant on the biological response to particles. Our aim was therefore to test the hypothesis that excessive mechanical stimulation of the periprosthetic tissues induces an inflammatory response and that the addition of particulate biomaterials intensifies this. We allocated 66 adult Beagle dogs to four groups as follows: stable implants with (I) and without (II) particulate polymethylmethacrylate (PMMA) and moving implants with (III) and without (IV) particulate PMMA. They were then evaluated at 2, 4, 6, 12 and 24 weeks. The stable implants were well tolerated and a thin, fibrous membrane of connective tissue was observed. There was evidence of positive staining in some cells for interleukin-6 (IL-6). Addition of particulate PMMA around the stable implants resulted in an increase in the fibroblastic response and positive staining for IL-6 and tumour necrosis factor-alpha (TNF-α). By contrast, movement of the implant resulted in an immediate inflammatory response characterised by large numbers of histiocytes and cytokine staining for IL-1ß, TNF-α and IL-6. Introduction of particulate PMMA aggravated this response. Animals with particulate PMMA and movement of the implant have an intense inflammatory response associated with accelerated bone loss. Our results indicate that the initiation of the inflammatory response to biomaterial particles was much slower than that to gross mechanical instability. Furthermore, when there was both particulate debris and movement, there was an amplification of the adverse tissue response as evidenced by the presence of
We investigated the implant-bone interface around one design of femoral stem, proximally coated with either a plasma-sprayed porous coating (plain porous) or a hydroxyapatite porous coating (porous HA), or which had been grit-blasted (Interlok). Of 165 patients implanted with a Bimetric hip hemiarthroplasty (Biomet, Bridgend, UK) specimens were retrieved from 58 at post-mortem. We estimated ingrowth and attachment of bone to the surface of the implant in 21 of these, eight plain porous, seven porous HA and six Interlok, using image analysis and light morphometric techniques. The amount of HA coating was also quantified. There was significantly more ingrowth (p = 0.012) and attachment of bone (p >
0.05) to the porous HA surface (mean bone ingrowth 29.093 ± 2.019%; mean bone attachment 37.287 ± 2.489%) than to the plain porous surface (mean bone ingrowth 21.762 ± 2.068%; mean bone attachment 18.9411 ± 1.971%). There was no significant difference in attachment between the plain porous and Interlok surfaces. Bone grew more evenly over the surface of the HA coating whereas on the porous surface, bone ingrowth and attachment occurred more on the distal and medial parts of the coated surface. No significant differences in the volume of HA were found with the passage of time. This study shows that HA coating increases the amount of ingrowth and attachment of bone and leads to a more even distribution of bone over the surface of the implant. This may have implications in reducing stress shielding and limiting
Previous studies have shown that the activity of the cytostatic drug methotrexate (MTX) embedded in acrylic cement is not affected by thermal changes in the cement. MTX is slowly released from the cement for several months and remains biologically active throughout this period. Our aim was to determine whether MTX embedded in cement would control the local growth of a tumour. In 15 rabbits we injected 0.1 ml of VX. 2. tumour suspension into the proximal tibia. At 3, 5, 7, 10 and 14 days three animals were killed and the tibiae removed and examined histologically. With increasing growth of the VX. 2. carcinoma there was increased bone destruction and a rise in the numbers of osteoclasts, but after 14 days the numbers of osteoclasts had decreased. We then injected VX. 2. into the tibiae of another 45 rabbits. After 5 days most of the tumour was curetted out and the defect filled with cement containing either 0 g, 0.1 g, 0.5 g, 1.0 g or 2.0 g MTX/40 g cement. The rabbits were divided into three groups and killed at 3, 7 or 10 days after implantation of cement. The number of osteoclasts and the amount of bone destruction were measured in each tibia. In all three groups bone destruction and osteoclast proliferation were markedly decreased with higher doses of MTX, but bone destruction was not eliminated. Our findings show that in the higher doses used, which were not toxic to the animal, MTX-embedded cement may be of value in minimising the amount of tumour-induced
It is increasingly appreciated that coordinated regulation of angiogenesis and osteogenesis is needed for bone formation. How this regulation is achieved during peri-implant bone healing, such as osseointegration, is largely unclear. This study examined the relationship between angiogenesis and osteogenesis in a unique model of osseointegration of a mouse tibial implant by pharmacologically blocking the vascular endothelial growth factor (VEGF) pathway. An implant was inserted into the right tibia of 16-week-old female C57BL/6 mice (n = 38). Mice received anti-VEGF receptor-1 (VEGFR-1) antibody (25 mg/kg) and VEGF receptor-2 (VEGFR-2) antibody (25 mg/kg; n = 19) or an isotype control antibody (n = 19). Flow cytometric (n = 4/group) and immunofluorescent (n = 3/group) analyses were performed at two weeks post-implantation to detect the distribution and density of CD31hiEMCNhi endothelium. RNA sequencing analysis was performed using sorted CD31hiEMCNhi endothelial cells (n = 2/group). Osteoblast lineage cells expressing osterix (OSX) and osteopontin (OPN) were also detected with immunofluorescence. Mechanical pull-out testing (n = 12/group) was used at four weeks post-implantation to determine the strength of the bone-implant interface. After pull-out testing, the tissue attached to the implant surface was harvested. Whole mount immunofluorescent staining of OSX and OPN was performed to determine the amount of osteoblast lineage cells.Aims
Materials and Methods
We investigated the long-term performance of the Tripolar Trident acetabular component used for recurrent dislocation in revision total hip arthroplasty. We assessed: 1) rate of re-dislocation; 2) incidence of complications requiring re-operation; and 3) Western Ontario and McMaster Universities osteoarthritis index (WOMAC) pain and functional scores. We retrospectively identified 111 patients who had 113 revision tripolar constrained liners between 1994 and 2008. All patients had undergone revision hip arthroplasty before the constrained liner was used: 13 after the first revision, 17 after the second, 38 after the third, and 45 after more than three revisions. A total of 75 hips (73 patients) were treated with Tripolar liners due to recurrent instability with abductor deficiency, In addition, six patients had associated cerebral palsy, four had poliomyelitis, two had multiple sclerosis, two had spina bifida, two had spondyloepiphyseal dysplasia, one had previous reversal of an arthrodesis, and 21 had proximal femoral replacements. The mean age of patients at time of Tripolar insertions was 72 years (53 to 89); there were 69 female patients (two bilateral) and 42 male patients. All patients were followed up for a mean of 15 years (10 to 24). Overall, 55 patients (57 hips) died between April 2011 and February 2018, at a mean of 167 months (122 to 217) following their tripolar liner implantation. We extracted demographics, implant data, rate of dislocations, and incidence of other complications.Aims
Patients and Methods
There is little information regarding the risk of a patient developing prosthetic joint infection (PJI) after primary total knee arthroplasty (TKA) when the patient has previously experienced PJI of a TKA or total hip arthroplasty (THA) in another joint. The goal of this study was to compare the risk of PJI of primary TKA in this patient population against matched controls. We retrospectively reviewed 95 patients (102 primary TKAs) treated between 2000 and 2014 with a history of PJI in another TKA or THA. A total of 50 patients (53%) were female. Mean age was 69 years (45 to 88) with a mean body mass index (BMI) of 36 kg/m2 (22 to 59). In total, 27% of patients were on chronic antibiotic suppression. Mean follow-up was six years (2 to 16). We 1:3 matched these (for age, sex, BMI, and surgical year) to 306 primary TKAs performed in 306 patients with a THA or TKA of another joint without a subsequent PJI. Competing risk with death was used for statistical analysis. Multivariate analysis was followed to evaluate risk factors for PJI in the study cohort.Aims
Patients and Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods
The aim of this study was to analyze the effect of a lateral rim mesh on the survival of primary total hip arthroplasty (THA) in young patients, aged 50 years or younger. We compared a study group of 235 patients (257 hips) who received a primary THA with the use of impaction bone grafting (IBG) with an additional lateral rim mesh with a group of 306 patients (343 hips) who received IBG in the absence of a lateral rim mesh during the same period from 1988 to 2015. In the mesh group, there were 74 male and 183 female patients, with a mean age of 35 years (13 to 50). In the no-mesh group, there were 173 male and 170 female patients, with a mean age of 38 years (12.6 to 50). Cox regression analyses were performed to study the effect of a lateral rim mesh on acetabular component survival. Kaplan–Meier analyses with 95% confidence intervals (CIs) were performed to estimate the survival of the acetabular implant.Aims
Patients and Methods
During revision procedures for aseptic reasons, there remains a suspicion that failure may have been the result of an undetected subclinical infection. However, there is little evidence available in the literature about unexpected positive results in presumed aseptic revision spine surgery. The aims of our study were to estimate the prevalence of unexpected positive culture using sonication and to evaluate clinical characteristics of these patients. All patients who underwent a revision surgery after instrumented spinal surgery at our institution between July 2014 and August 2016 with spinal implants submitted for sonication were retrospectively analyzed. Only revisions presumed as aseptic are included in the study. During the study period, 204 spinal revisions were performed for diagnoses other than infection. In 38 cases, sonication cultures were not obtained, leaving a study cohort of 166 cases. The mean age of the cohort was 61.5 years (Aims
Patients and Methods