Advertisement for orthosearch.org.uk
Results 21 - 40 of 120
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 793 - 801
1 Jun 2015
Liddle AD Pandit H Judge A Murray DW

Whether to use total or unicompartmental knee replacement (TKA/UKA) for end-stage knee osteoarthritis remains controversial. Although UKA results in a faster recovery, lower rates of morbidity and mortality and fewer complications, the long-term revision rate is substantially higher than that for TKA. The effect of each intervention on patient-reported outcome remains unclear. The aim of this study was to determine whether six-month patient-reported outcome measures (PROMs) are better in patients after TKA or UKA, using data from a large national joint registry (NJR).

We carried out a propensity score-matched cohort study which compared six-month PROMs after TKA and UKA in patients enrolled in the NJR for England and Wales, and the English national PROM collection programme. A total of 3519 UKA patients were matched to 10 557 TKAs.

The mean six-month PROMs favoured UKA: the Oxford Knee Score was 37.7 (95% confidence interval (CI) 37.4 to 38.0) for UKA and 36.1 (95% CI 35.9 to 36.3) for TKA; the mean EuroQol EQ-5D index was 0.772 (95% CI 0.764 to 0.780) for UKA and 0.751 (95% CI 0.747 to 0.756) for TKA. UKA patients were more likely to achieve excellent results (odds ratio (OR) 1.59, 95% CI 1.47 to 1.72, p < 0.001) and to be highly satisfied (OR 1.27, 95% CI 1.17 to 1.39, p <  0.001), and were less likely to report complications than those who had undergone TKA.

UKA gives better early patient-reported outcomes than TKA; these differences are most marked for the very best outcomes. Complications and readmission are more likely after TKA. Although the data presented reflect the short-term outcome, they suggest that the high revision rate for UKA may not be because of poorer clinical outcomes. These factors should inform decision-making in patients eligible for either procedure.

Cite this article: Bone Joint J 2015;97-B:793–801.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 164 - 172
1 Feb 2015
Grammatopoulos G Thomas GER Pandit H Beard DJ Gill HS Murray DW

We assessed the orientation of the acetabular component in 1070 primary total hip arthroplasties with hard-on-soft, small diameter bearings, aiming to determine the size and site of the target zone that optimises outcome. Outcome measures included complications, dislocations, revisions and ΔOHS (the difference between the Oxford Hip Scores pre-operatively and five years post-operatively). A wide scatter of orientation was observed (2sd 15°). Placing the component within Lewinnek’s zone was not associated withimproved outcome. Of the different zone sizes tested (± 5°, ± 10° and ± 15°), only ± 15° was associated with a decreased rate of dislocation. The dislocation rate with acetabular components inside an inclination/anteversion zone of 40°/15° ± 15° was four times lower than those outside. The only zone size associated with statistically significant and clinically important improvement in OHS was ± 5°. The best outcomes (ΔOHS > 26) were achieved with a 45°/25° ± 5° zone.

This study demonstrated that with traditional technology surgeons can only reliably achieve a target zone of ±15°. As the optimal zone to diminish the risk of dislocation is also ±15°, surgeons should be able to achieve this. This is the first study to demonstrate that optimal orientation of the acetabular component improves the functional outcome. However, the target zone is small (± 5°) and cannot, with current technology, be consistently achieved.

Cite this article: Bone Joint J 2015;97-B:164–72.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 185 - 191
1 Feb 2015
Kendrick BJL Kaptein BL Valstar ER Gill HS Jackson WFM Dodd CAF Price AJ Murray DW

The most common reasons for revision of unicompartmental knee arthroplasty (UKA) are loosening and pain. Cementless components may reduce the revision rate. The aim of this study was to compare the fixation and clinical outcome of cementless and cemented Oxford UKAs.

A total of 43 patients were randomised to receive either a cemented or a cementless Oxford UKA and were followed for two years with radiostereometric analysis (RSA), radiographs aligned with the bone–implant interfaces and clinical scores.

The femoral components migrated significantly during the first year (mean 0.2 mm) but not during the second. There was no significant difference in the extent of migration between cemented and cementless femoral components in either the first or the second year. In the first year the cementless tibial components subsided significantly more than the cemented components (mean 0.28 mm (sd 0.17) vs. 0.09 mm (sd 0.19 mm)). In the second year, although there was a small amount of subsidence (mean 0.05 mm) there was no significant difference (p = 0.92) between cemented and cementless tibial components. There were no femoral radiolucencies. Tibial radiolucencies were narrow (< 1 mm) and were significantly (p = 0.02) less common with cementless (6 of 21) than cemented (13 of 21) components at two years. There were no complete radiolucencies with cementless components, whereas five of 21 (24%) cemented components had complete radiolucencies. The clinical scores at two years were not significantly different (p = 0.20).

As second-year migration is predictive of subsequent loosening, and as radiolucency is suggestive of reduced implant–bone contact, these data suggest that fixation of the cementless components is at least as good as, if not better than, that of cemented devices.

Cite this article: Bone Joint J 2015; 97-B:185–91.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1623 - 1630
1 Dec 2014
Monk AP Choji K O’Connor JJ Goodfellow† JW Murray DW

We scanned 25 left knees in healthy human subjects using MRI. Multiplanar reconstruction software was used to take measurements of the inferior and posterior facets of the femoral condyles and the trochlea.

A ‘basic circle’ can be defined which, in the sagittal plane, fits the posterior and inferior facets of the lateral condyle, the posterior facet of the medial condyle and the floor of the groove of the trochlea. It also approximately fits both condyles in the coronal plane (inferior facets) and the axial plane (posterior facets). The circle fitting the inferior facet of the medial condyle in the sagittal plane was consistently 35% larger than the other circles and was termed the ‘medial inferior circle’. There were strong correlations between the radii of the circles, the relative positions of the centres of the condyles, the width of the condyles, the total knee width and skeletal measurements including height. There was poor correlation between the radii of the circles and the position of the trochlea relative to the condyles.

In summary, the condyles are approximately spherical except for the inferior facet medially, which has a larger radius in the sagittal plane. The size and position of the condyles are consistent and change with the size of the person. However, the position of the trochlea is variable even though its radius is similar to that of the condyles. This information has implications for understanding anterior knee pain and for the design of knee replacements.

Cite this article: Bone Joint J 2014;96-B:1623–30.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1290 - 1297
1 Oct 2014
Grammatopoulos G Pandit HG da Assunção R McLardy-Smith P De Smet KA Gill HS Murray DW

There is great variability in acetabular component orientation following hip replacement. The aims of this study were to compare the component orientation at impaction with the orientation measured on post-operative radiographs and identify factors that influence the difference between the two. A total of 67 hip replacements (52 total hip replacements and 15 hip resurfacings) were prospectively studied. Intra-operatively, the orientation of the acetabular component after impaction relative to the operating table was measured using a validated stereo-photogrammetry protocol. Post-operatively, the radiographic orientation was measured; the mean inclination/anteversion was 43° (sd 6°)/ 19° (sd 7°). A simulated radiographic orientation was calculated based on how the orientation would have appeared had an on-table radiograph been taken intra-operatively. The mean difference between radiographic and intra-operative inclination/anteversion was 5° (sd 5°)/ -8° (sd 8°). The mean difference between simulated radiographic and intra-operative inclination/anteversion, which quantifies the effect of the different way acetabular orientation is measured, was 3°/-6° (sd 2°). The mean difference between radiographic and simulated radiographic orientation inclination/anteversion, which is a manifestation of the change in pelvic position between component impaction and radiograph, was 1°/-2° (sd 7°).

This study demonstrated that in order to achieve a specific radiographic orientation target, surgeons should implant the acetabular component 5° less inclined and 8° more anteverted than their target. Great variability (2 sd about ± 15°) in the post-operative radiographic cup orientation was seen. The two equally contributing causes for this are variability in the orientation at which the cup is implanted, and the change in pelvic position between impaction and post-operative radiograph.

Cite this article: Bone Joint J 2014;96-B:1290–7


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 876 - 883
1 Jul 2014
Grammatopoulos G Pandit HG da Assunção R Taylor A McLardy-Smith P De Smet KA Murray DW Gill HS

The orientation of the acetabular component is influenced not only by the orientation at which the surgeon implants the component, but also the orientation of the pelvis at the time of implantation. Hence, the orientation of the pelvis at set-up and its movement during the operation, are important. During 67 hip replacements, using a validated photogrammetric technique, we measured how three surgeons orientated the patient’s pelvis, how much the pelvis moved during surgery, and what effect these had on the final orientation of the acetabular component. Pelvic orientation at set-up, varied widely (mean (± 2, standard deviation (sd))): tilt 8° (2sd ±32), obliquity –4° (2sd ±12), rotation –8° (2sd ±14). Significant differences in pelvic positioning were detected between surgeons (p < 0.001). The mean angular movement of the pelvis between set-up and component implantation was 9° (sd 6). Factors influencing pelvic movement included surgeon, approach (posterior >  lateral), procedure (hip resurfacing > total hip replacement) and type of support (p < 0.001). Although, on average, surgeons achieved their desired acetabular component orientation, there was considerable variability (2sd ±16) in component orientation. We conclude that inconsistency in positioning the patient at set-up and movement of the pelvis during the operation account for much of the variation in acetabular component orientation. Improved methods of positioning and holding the pelvis are required.

Cite this article: Bone Joint J 2014; 96-B:876–83.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 345 - 349
1 Mar 2014
Liddle AD Pandit HG Jenkins C Lobenhoffer P Jackson WFM Dodd CAF Murray DW

The cementless Oxford unicompartmental knee replacement has been demonstrated to have superior fixation on radiographs and a similar early complication rate compared with the cemented version. However, a small number of cases have come to our attention where, after an apparently successful procedure, the tibial component subsides into a valgus position with an increased posterior slope, before becoming well-fixed. We present the clinical and radiological findings of these six patients and describe their natural history and the likely causes. Two underwent revision in the early post-operative period, and in four the implant stabilised and became well-fixed radiologically with a good functional outcome.

This situation appears to be avoidable by minor modifications to the operative technique, and it appears that it can be treated conservatively in most patients.

Cite this article: Bone Joint J 2014;96-B:345–9.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 59 - 64
1 Jan 2014
Weston-Simons JS Pandit H Kendrick BJL Jenkins C Barker K Dodd CAF Murray DW

Mobile-bearing unicompartmental knee replacements (UKRs) with a flat tibial plateau have not performed well in the lateral compartment, owing to a high dislocation rate. This led to the development of the Domed Lateral Oxford UKR (Domed OUKR) with a biconcave bearing. The aim of this study was to assess the survival and clinical outcomes of the Domed OUKR in a large patient cohort in the medium term.

We prospectively evaluated 265 consecutive knees with isolated disease of the lateral compartment and a mean age at surgery of 64 years (32 to 90). At a mean follow-up of four years (sd 2.2, (0.5 to 8.3)) the mean Oxford knee score was 40 out of 48 (sd 7.4). A total of 12 knees (4.5%) had re-operations, of which four (1.5%) were for dislocation. All dislocations occurred in the first two years. Two (0.8%) were secondary to significant trauma that resulted in ruptured ligaments, and two (0.8%) were spontaneous. In four patients (1.5%) the UKR was converted to a primary TKR. Survival at eight years, with failure defined as any revision, was 92.1% (95% confidence interval 81.3 to 100).

The Domed Lateral OUKR gives good clinical outcomes, low re-operation and revision rates and a low dislocation rate in patients with isolated lateral compartmental disease, in the hands of the designer surgeons.

Cite this article: Bone Joint J 2014;96-B:59–64.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 51 - 52
1 Nov 2013
Murray DW

There is a North Atlantic divide, with cementless femoral stems being used more frequently in the USA and cemented stems being used more frequently in many countries in Europe. This is primarily because different cemented stems have been used on different sides of the Atlantic and the results of the cemented stems in the US have often been poor, whereas the results of the stems used in Europe have been good. In the National registers in Europe, cemented stems have tended to achieve better results than cementless.

Cite this article: Bone Joint J 2013;95-B, Supple A:51–2.


The Bone & Joint Journal
Vol. 95-B, Issue 6 | Pages 793 - 797
1 Jun 2013
Williams DP Pandit HG Athanasou NA Murray DW Gibbons CLMH

The aim of this study was to review the early outcome of the Femoro-Patella Vialla (FPV) joint replacement. A total of 48 consecutive FPVs were implanted between December 2007 and June 2011. Case-note analysis was performed to evaluate the indications, operative histology, operative findings, post-operative complications and reasons for revision. The mean age of the patients was 63.3 years (48.2 to 81.0) and the mean follow-up was 25.0 months (6.1 to 48.9). Revision was performed in seven (14.6%) at a mean of 21.7 months, and there was one re-revision. Persistent pain was observed in three further patients who remain unrevised. The reasons for revision were pain due to progressive tibiofemoral disease in five, inflammatory arthritis in one, and patellar fracture following trauma in one. No failures were related to the implant or the technique. Trochlear dysplasia was associated with a significantly lower rate of revision (5.9% vs 35.7%, p = 0.017) and a lower incidence of revision or persistent pain (11.8% vs 42.9%, p = 0.045).

Focal patellofemoral osteoarthritis secondary to trochlear dysplasia should be considered the best indication for patellofemoral replacement. Standardised radiological imaging, with MRI to exclude overt tibiofemoral disease should be part of the pre-operative assessment, especially for the non-dysplastic knee.

Cite this article: Bone Joint J 2013;95-B:793–7.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 605 - 608
1 May 2013
Murray DW Gulati A Gill HS

The Exeter femoral stem is a double-tapered highly polished collarless cemented implant with good long-term clinical results. In order to determine why the stem functions well we have undertaken a long-term radiostereometric analysis (RSA) study.

A total of 20 patients undergoing primary Exeter total hip replacement for osteoarthritis using the Hardinge approach were recruited and followed with RSA for ten years. The stems progressively subsided and internally rotated with posterior head migration. The mean subsidence was 0.7 mm (95% confidence interval (CI) 0.5 to 0.9) at two years and 1.3 mm (95% CI 1.0 to 1.6) at ten years. The mean posterior migration of the head was 0.7 mm (95% CI 0.5 to 0.9) at two years and 1.2 mm (95% CI 1.0 to 1.4) at ten years. There was no significant cement restrictor migration.

The Exeter stem continues to subside slowly into the cement mantle in the long term. This appears to compress the cement and the cement bone interface, contributing to secure fixation in the long term.

Cite this article: Bone Joint J 2013;95-B:605–8.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 486 - 492
1 Apr 2013
Breeman S Campbell MK Dakin H Fiddian N Fitzpatrick R Grant A Gray A Johnston L MacLennan GS Morris RW Murray DW

There is conflicting evidence about the merits of mobile bearings in total knee replacement, partly because most randomised controlled trials (RCTs) have not been adequately powered. We report the results of a multicentre RCT of mobile versus fixed bearings. This was part of the knee arthroplasty trial (KAT), where 539 patients were randomly allocated to mobile or fixed bearings and analysed on an intention-to-treat basis. The primary outcome measure was the Oxford Knee Score (OKS) plus secondary measures including Short Form-12, EuroQol EQ-5D, costs, cost-effectiveness and need for further surgery.

There was no significant difference between the groups pre-operatively: mean OKS was 17.18 (sd 7.60) in the mobile-bearing group and 16.49 (sd 7.40) in the fixed-bearing group. At five years mean OKS was 33.19 (sd 16.68) and 33.65 (sd 9.68), respectively. There was no significant difference between trial groups in OKS at five years (-1.12 (95% confidence interval -2.77 to 0.52) or any of the other outcome measures. Furthermore, there was no significant difference in the proportion of patients with knee-related re-operations or in total costs.

In this appropriately powered RCT, over the first five years after total knee replacement functional outcomes, re-operation rates and healthcare costs appear to be the same irrespective of whether a mobile or fixed bearing is used.

Cite this article: Bone Joint J 2013;95-B:486–92.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 181 - 187
1 Feb 2013
Liddle AD Pandit H O’Brien S Doran E Penny ID Hooper GJ Burn PJ Dodd CAF Beverland DE Maxwell AR Murray DW

The Cementless Oxford Unicompartmental Knee Replacement (OUKR) was developed to address problems related to cementation, and has been demonstrated in a randomised study to have similar clinical outcomes with fewer radiolucencies than observed with the cemented device. However, before its widespread use it is necessary to clarify contraindications and assess the complications. This requires a larger study than any previously published.

We present a prospective multicentre series of 1000 cementless OUKRs in 881 patients at a minimum follow-up of one year. All patients had radiological assessment aligned to the bone–implant interfaces and clinical scores. Analysis was performed at a mean of 38.2 months (19 to 88) following surgery. A total of 17 patients died (comprising 19 knees (1.9%)), none as a result of surgery; there were no tibial or femoral loosenings. A total of 19 knees (1.9%) had significant implant-related complications or required revision. Implant survival at six years was 97.2%, and there was a partial radiolucency at the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies. There was no significant increase in complication rate compared with cemented fixation (p = 0.87), and no specific contraindications to cementless fixation were identified.

Cementless OUKR appears to be safe and reproducible in patients with end-stage anteromedial osteoarthritis of the knee, with radiological evidence of improved fixation compared with previous reports using cemented fixation.

Cite this article: Bone Joint J 2013;95-B:181–7.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 45 - 51
1 Jan 2013
Williams DP Blakey CM Hadfield SG Murray DW Price AJ Field RE

The Oxford knee score (OKS) is a validated and widely accepted disease-specific patient-reported outcome measure, but there is limited evidence regarding any long-term trends in the score. We reviewed 5600 individual OKS questionnaires (1547 patients) from a prospectively-collected knee replacement database, to determine the trends in OKS over a ten-year period following total knee replacement. The mean OKS pre-operatively was 19.5 (95% confidence interval (CI) 18.8 to 20.2). The maximum post-operative OKS was observed at two years (mean score 34.4 (95% CI 33.7 to 35.2)), following which a gradual but significant decline was observed through to the ten-year assessment (mean score 30.1 (95% CI 29.1 to 31.1)) (p < 0.001). A similar trend was observed for most of the individual OKS components (p < 0.001). Kneeling ability initially improved in the first year but was then followed by rapid deterioration (p < 0.001). Pain severity exhibited the greatest improvement, although residual pain was reported in over two-thirds of patients post-operatively, and peak improvement in the night pain component did not occur until year four. Post-operative OKS was lower for women (p < 0.001), those aged < 60 years (p < 0.003) and those with a body mass index > 35 kg/m2 (p < 0.014), although similar changes in scores were observed. This information may assist surgeons in advising patients of their expected outcomes, as well as providing a comparative benchmark for evaluating longer-term outcomes following knee replacement.

Cite this article: Bone Joint J 2013;95-B:45–51.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 38 - 44
1 Jan 2013
Williams DP Price AJ Beard DJ Hadfield SG Arden NK Murray DW Field RE

We present a comparison of patient-reported outcomes (PROMs) in relation to patient age, in patients who had received a total (TKR) or unicompartmental knee replacement (UKR). The outcome was evaluated using the Oxford knee score (OKS), EuroQol (EQ-5D) and satisfaction scores. Patients aged 65 to 84 years demonstrated better pre-operative function scores than those aged < 65 years (OKS, p = 0.03; EQ-5D, p = 0.048) and those aged ≥ 85 years (OKS, p = 0.03). Post-operative scores were comparable across age groups, but a linear trend for greater post-operative improvement in OKS and EQ-5D was seen with decreasing age (p < 0.033). The overall mean satisfaction score at six months was 84.9, but those aged <  55 years exhibited a lower mean level of satisfaction (78.3) compared with all other age groups (all p < 0.031). The cumulative overall two-year revision rate was 1.3%. This study demonstrates that good early outcomes, as measured by the OKS and EQ-5D, can be anticipated following knee replacement regardless of the patient’s age, although younger patients gain greater improvement. However, the lower satisfaction in those aged < 55 years is a concern, and suggests that outcome is not fully encapsulated by the OKS and EQ-5D evaluation, and raises the question whether the OKS alone is an appropriate measure of pain and function in younger, more active individuals.

Cite this article: Bone Joint J 2013;95-B:38–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1216 - 1220
1 Sep 2012
Weston-Simons JS Pandit H Jenkins C Jackson WFM Price AJ Gill HS Dodd CAF Murray DW

The Oxford unicompartmental knee replacement (UKR) is an established treatment option in the management of symptomatic end-stage medial compartmental osteoarthritis (MCOA), which works well in the young and active patient. However, previous studies have shown that it is reliable only in the presence of a functionally intact anterior cruciate ligament (ACL). This review reports the outcomes, at a mean of five years and a maximum of ten years, of 52 consecutive patients with a mean age of 51 years (36 to 57) who underwent staged or simultaneous ACL reconstruction and Oxford UKR. At the last follow-up (with one patient lost to follow-up), the mean Oxford knee score was 41 (sd 6.3; 17 to 48). Two patients required conversion to TKR: one for progression of lateral compartment osteoarthritis and one for infection. Implant survival at five years was 93% (95% CI 83 to 100). All but one patient reported being satisfied with the procedure. The outcome was not significantly influenced by age, gender, femoral or tibial tunnel placement, or whether the procedure was undertaken at one- or two-stages.

In summary, ACL reconstruction and Oxford UKR gives good results in patients with end-stage MCOA secondary to ACL deficiency.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1180 - 1186
1 Sep 2012
Murray DW Grammatopoulos G Pandit H Gundle R Gill HS McLardy-Smith P

Recent events have highlighted the importance of implant design for survival and wear-related complications following metal-on-metal hip resurfacing arthroplasty. The mid-term survival of the most widely used implant, the Birmingham Hip Resurfacing (BHR), has been described by its designers. The aim of this study was to report the ten-year survival and patient-reported functional outcome of the BHR from an independent centre.

In this cohort of 554 patients (646 BHRs) with a mean age of 51.9 years (16.5 to 81.5) followed for a mean of eight years (1 to 12), the survival and patient-reported functional outcome depended on gender and the size of the implant. In female hips (n = 267) the ten-year survival was 74% (95% confidence interval (CI) 83 to 91), the ten-year revision rate for pseudotumour was 7%, the mean Oxford hip score (OHS) was 43 (sd 8) and the mean UCLA activity score was 6.4 (sd 2). In male hips (n = 379) the ten-year survival was 95% (95% CI 92.0 to 97.4), the ten-year revision rate for pseudotumour was 1.7%, the mean OHS was 45 (sd 6) and the mean UCLA score was 7.6 (sd 2). In the most demanding subgroup, comprising male patients aged < 50 years treated for primary osteoarthritis, the survival was 99% (95% CI 97 to 100).

This study supports the ongoing use of resurfacing in young active men, who are a subgroup of patients who tend to have problems with conventional THR. In contrast, the results in women have been poor and we do not recommend metal-on-metal resurfacing in women. Continuous follow-up is recommended because of the increasing incidence of pseudotumour with the passage of time.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 42 - 49
1 Apr 2012
Kwon Y Mellon SJ Monk P Murray DW Gill HS

Objectives

Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities.

Methods

The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient’s specific hip joint centre, based on CT scans.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 477 - 482
1 Apr 2012
Merle C Waldstein W Pegg E Streit MR Gotterbarm T Aldinger PR Murray DW Gill HS

The aim of this retrospective cohort study was to identify any difference in femoral offset as measured on pre-operative anteroposterior (AP) radiographs of the pelvis, AP radiographs of the hip and corresponding CT scans in a consecutive series of 100 patients with primary end-stage osteoarthritis of the hip (43 men and 57 women with a mean age of 61 years (45 to 74) and a mean body mass index of 28 kg/m2 (20 to 45)).

Patients were positioned according to a standardised protocol to achieve reproducible projection and all images were calibrated. Inter- and intra-observer reliability was evaluated and agreement between methods was assessed using Bland-Altman plots.

In the entire cohort, the mean femoral offset was 39.0 mm (95% confidence interval (CI) 37.4 to 40.6) on radiographs of the pelvis, 44.0 mm (95% CI 42.4 to 45.6) on radiographs of the hip and 44.7 mm (95% CI 43.5 to 45.9) on CT scans. AP radiographs of the pelvis underestimated femoral offset by 13% when compared with CT (p < 0.001). No difference in mean femoral offset was seen between AP radiographs of the hip and CT (p = 0.191).

Our results suggest that femoral offset is significantly underestimated on AP radiographs of the pelvis but can be reliably and accurately assessed on AP radiographs of the hip in patients with primary end-stage hip osteoarthritis.

We, therefore, recommend that additional AP radiographs of the hip are obtained routinely for the pre-operative assessment of femoral offset when templating before total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1610 - 1616
1 Dec 2011
Pegg E Pandit H Gill HS Keys GW Svard UGC O’Connor JJ Murray DW

Since the Oxford knee was first used unicompartmentally in 1982, a small number of bearings have fractured. Of 14 retrieved bearings, we examined ten samples with known durations in situ (four Phase 1, four Phase 2 and two Phase 3). Evidence of impingement and associated abnormally high wear (> 0.05 mm per year) as well as oxidation was observed in all bearings. In four samples the fracture was associated with the posterior radio-opaque wire. Fracture surfaces indicated fatigue failure, and scanning electron microscopy suggested that the crack initiated in the thinnest region. The estimated incidence of fracture was 3.20% for Phase 1, 0.74% for Phase 2, 0.35% for Phase 3, and 0% for Phase 3 without the posterior marker wire. The important aetiological factors for bearing fracture are impingement leading to high wear, oxidation, and the posterior marker wire. With improved surgical technique, impingement and high wear should be prevented and modern polyethylene may reduce the oxidation risk. A posterior marker wire is no longer used in the polyethylene meniscus. Therefore, the rate of fracture, which is now very low, should be reduced to a negligible level.