Our aim was to examine the clinical and radiographic outcomes
in 257 consecutive Oxford unicompartmental knee arthroplasties (OUKAs)
(238 patients), five years post-operatively. A retrospective evaluation was undertaken of patients treated
between April 2008 and October 2010 in a regional centre by two
non-designing surgeons with no previous experience of UKAs. The
Oxford Knee Scores (OKSs) were recorded and fluoroscopically aligned
radiographs were assessed post-operatively at one and five years.Aims
Patients and Methods
Unicompartmental knee arthroplasty (UKA) is a demanding procedure, with tibial component subsidence or pain from high tibial strain being potential causes of revision. The optimal position in terms of load transfer has not been documented for lateral UKA. Our aim was to determine the effect of tibial component position on proximal tibial strain. A total of 16 composite tibias were implanted with an Oxford Domed Lateral Partial Knee implant using cutting guides to define tibial slope and resection depth. Four implant positions were assessed: standard (5° posterior slope); 10° posterior slope; 5° reverse tibial slope; and 4 mm increased tibial resection. Using an electrodynamic axial-torsional materials testing machine (Instron 5565), a compressive load of 1.5 kN was applied at 60 N/s on a meniscal bearing via a matching femoral component. Tibial strain beneath the implant was measured using a calibrated Digital Image Correlation system.Objectives
Methods
Mobile-bearing unicompartmental knee replacements
(UKRs) with a flat tibial plateau have not performed well in the
lateral compartment, owing to a high dislocation rate. This led
to the development of the Domed Lateral Oxford UKR (Domed OUKR)
with a biconcave bearing. The aim of this study was to assess the
survival and clinical outcomes of the Domed OUKR in a large patient
cohort in the medium term. We prospectively evaluated 265 consecutive knees with isolated
disease of the lateral compartment and a mean age at surgery of
64 years (32 to 90). At a mean follow-up of four years ( The Domed Lateral OUKR gives good clinical outcomes, low re-operation
and revision rates and a low dislocation rate in patients with isolated
lateral compartmental disease, in the hands of the designer surgeons. Cite this article:
Unicompartmental knee arthroplasty (UKA) is a potential treatment
for isolated bone on bone osteoarthritis when limited to a single
compartment. The risk for revision of UKA is three times higher
than for total knee arthroplasty (TKA). The aim of this review was
to discuss the different revision options after UKA failure. A search was performed for English language articles published
between 2006 and 2016. After reviewing titles and abstracts, 105
papers were selected for further analysis. Of these, 39 papers were
deemed to contain clinically relevant data to be included in this review.Objectives
Materials and Methods
The October 2014 Knee Roundup360 looks at: microfracture equivalent to OATS; examination better than MRI in predicting hamstrings re-injury; a second view on return to play with hamstrings injuries; dislocation risks in the Oxford Unicompartmental Knee; what about the tibia?; getting on top of lateral facet pain post TKR; readmission in TKR; patient-specific instrumentation; treating infrapatellar saphenous neuralgia; and arthroscopy in the middle-aged.
Approved by the Food and Drug Administration in 2004, the Phase
III Oxford Medial Partial Knee is used to treat anteromedial osteoarthritis
(AMOA) in patients with an intact anterior cruciate ligament. This
unicompartmental knee arthroplasty (UKA) is relatively new in the
United States, and therefore long-term American results are lacking. This is a single surgeon, retrospective study based on prospectively
collected data, analysing a consecutive series of primary UKAs using
the Phase III mobile-bearing Oxford Knee and Phase III instrumentation. Between July 2004 and December 2006, the senior author (RHE)
carried out a medial UKA in 173 patients (213 knees) for anteromedial
osteoarthritis or avascular necrosis (AVN). A total of 95 patients were men and 78 were women. Their mean
age at surgery was 67 years (38 to 89) and mean body mass index
29.87 kg/m2 (17 to 62). The mean follow-up was ten years (4 to 11).Aims
Patients and Methods
The interest in unicompartmental knee arthroplasty (UKA) for
medial osteoarthritis has increased rapidly but the long-term follow-up
of the Oxford UKAs has yet to be analysed in non-designer centres.
We have examined our ten- to 15-year clinical and radiological follow-up
data for the Oxford Phase III UKAs. Between January 1999 and January 2005 a total of 138 consecutive
Oxford Phase III arthroplasties were performed by a single surgeon
in 129 patients for medial compartment osteoarthritis (71 right
and 67 left knees, mean age 72.0 years (47 to 91), mean body mass
index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were
prospectively recorded and obtained at intervals. Of the 129 patients,
32 patients (32 knees) died, ten patients (12 knees) were not able
to take part in the final clinical and radiological assessment due
to physical and mental conditions, but via telephone interview it
was confirmed that none of these ten patients (12 knees) had a revision
of the knee arthroplasty. One patient (two knees) was lost to follow-up.Aims
Patients and Methods
The contraindications for unicompartmental knee replacement (UKR) remain controversial. The views of many surgeons are based on Kozinn and Scott’s 1989 publication which stated that patients who weighed more than 82 kg, were younger than 60 years, undertook heavy labour, had exposed bone in the patellofemoral joint or chondrocalcinosis, were not ideal candidates for UKR. Our aim was to determine whether these potential contraindications should apply to patients with a mobile-bearing UKR. In order to do this the outcome of patients with these potential contraindications was compared with that of patients without the contraindications in a prospective series of 1000 UKRs. The outcome was assessed using the Oxford knee score, the American Knee Society score, the Tegner activity score, revision rate and survival. The clinical outcome of patients with each of the potential contraindications was similar to or better than those without each contraindication. Overall, 678 UKRs (68%) were performed in patients who had at least one potential contraindication and only 322 (32%) in patients deemed to be ideal. The survival at ten years was 97.0% (95% confidence interval 93.4 to 100.0) for those with potential contraindications and 93.6% (95% confidence interval 87.2 to 100.0) in the ideal patients. We conclude that the thresholds proposed by Kozinn and Scott using weight, age, activity, the state of the patellofemoral joint and chondrocalcinosis should not be considered to be contraindications for the use of the Oxford UKR.
Medium-term survivorship of the Oxford phase
3 unicompartmental knee replacement (UKR) has not yet been established
in an Asian population. We prospectively evaluated the outcome of
400 phase 3 Oxford UKRs in 320 Korean patients with a mean age at
the time of operation of 69 years (48 to 82). The mean follow-up
was 5.2 years (1 to 10). Clinical and radiological assessment was
carried out pre- and post-operatively. At five years, the mean Knee
Society knee and functional scores had increased significantly from
56.2 (30 to 91) pre-operatively to 87.2 (59 to 98) (p = 0.034) and
from 59.2 (30 to 93) to 88.3 (50 to 100) (p = 0.021), respectively.
The Oxford knee score increased from a mean of 25.8 (12 to 39) pre-operatively
to 39.8 (25 to 58) at five years (p = 0.038). The ten-year survival
rate was 94% (95% confidence interval 90.1 to 98.0). A total of
14 UKRs (3.5%) required revision. The most common reason for revision
was dislocation of the bearing in 12 (3%). Conversion to a total
knee replacement was required in two patients who developed osteoarthritis
of the lateral compartment. This is the largest published series of UKR in Korean patients.
It shows that the mid-term results after a minimally invasive Oxford
phase 3 UKR can yield satisfactory clinical and functional results
in this group of patients.
This prospective study describes the outcome of the first 1000 phase 3 Oxford medial unicompartmental knee replacements (UKRs) implanted using a minimally invasive surgical approach for the recommended indications by two surgeons and followed up independently. The mean follow-up was 5.6 years (1 to 11) with 547 knees having a minimum follow-up of five years. At five years their mean Oxford knee score was 41.3 ( The incidence of implant-related re-operations was 2.9%; of these 29 re-operations two were revisions requiring revision knee replacement components with stems and wedges, 17 were conversions to a primary total knee replacement, six were open reductions for dislocation of the bearing, three were secondary lateral UKRs and one was revision of a tibial component. The most common reason for further surgical intervention was progression of arthritis in the lateral compartment (0.9%), followed by dislocation of the bearing (0.6%) and revision for unexplained pain (0.6%). If all implant-related re-operations are considered failures, the ten-year survival rate was 96% (95% confidence interval, 92.5 to 99.5). If only revisions requiring revision components are considered failures the ten-year survival rate is 99.8% (confidence interval 99 to 100). This is the largest published series of UKRs implanted through a minimally invasive surgical approach and with ten-year survival data. The survival rates are similar to those obtained with a standard open approach whereas the function is better. This demonstrates the effectiveness and safety of a minimally invasive surgical approach for implanting the Oxford UKR.
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article:
The most common reasons for revision of unicompartmental
knee arthroplasty (UKA) are loosening and pain. Cementless components
may reduce the revision rate. The aim of this study was to compare
the fixation and clinical outcome of cementless and cemented Oxford
UKAs. A total of 43 patients were randomised to receive either a cemented
or a cementless Oxford UKA and were followed for two years with
radiostereometric analysis (RSA), radiographs aligned with the bone–implant
interfaces and clinical scores. The femoral components migrated significantly during the first
year (mean 0.2 mm) but not during the second. There was no significant
difference in the extent of migration between cemented and cementless
femoral components in either the first or the second year. In the
first year the cementless tibial components subsided significantly
more than the cemented components (mean 0.28 mm ( As second-year migration is predictive of subsequent loosening,
and as radiolucency is suggestive of reduced implant–bone contact,
these data suggest that fixation of the cementless components is
at least as good as, if not better than, that of cemented devices. Cite this article:
Patellofemoral joint degeneration is often considered a contraindication to medial unicompartmental knee replacement. We examined the validity of this preconception using information gathered prospectively on the intra-operative status of the patellofemoral joint in 824 knees in 793 consecutive patients who underwent Oxford unicondylar knee replacement for anteromedial osteoarthritis. All operations were performed between January 1998 and September 2005. A five-point grading system classified degeneration of the patellofemoral joint from none to full-thickness cartilage loss. A subclassification of the presence or absence of any full-thickness cartilage loss was subsequently performed to test selected hypotheses. Outcome was evaluated independently by physiotherapists using the Oxford and the American Knee Society Scores with a minimum follow-up of one year. Full-thickness cartilage loss on the trochlear surface was observed in 100 of 785 knees (13%), on the medial facet of the patella in 69 of 782 knees (9%) and on the lateral facet in 29 of 784 knees (4%). Full-thickness cartilage loss at any location was seen in 128 knees (16%) and did not produce a significantly worse outcome than those with a normal or near-normal joint surface. The severity of the degeneration at any of the intra-articular locations also had no influence on outcome. We concluded that, provided there is not bone loss and grooving of the lateral facet, damage to the articular cartilage of the patellofemoral joint to the extent of full-thickness cartilage loss is not a contraindication to the Oxford mobile-bearing unicompartmental knee replacement.
Medium-term results of the Discovery elbow replacement
are presented. We reviewed 51 consecutive primary Discovery total
elbow replacements (TERs) implanted in 48 patients. The mean age
of the patients was 69.2 years (49 to 92), there were 19 males and
32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69).
A total of six patients were lost to follow-up. Statistically significant
improvements in range movement and Oxford Elbow Score were found
(p <
0.001). Radiolucent lines were much more common in, and
aseptic loosening was exclusive to, the humeral component. Kaplan–Meier
survivorship at five years was 92.2% (95% CI 74.5% to 96.4%) for
aseptic loosening. In four TERs, periprosthetic infection occurred
resulting in failure. A statistically significant association between
infection and increased BMI was found (p = 0.0268). Triceps failure
was more frequent after the Mayo surgical approach and TER performed
after previous trauma surgery. No failures of the implant were noted. Our comparison shows that the Discovery has early clinical results
that are similar to other semi-constrained TERs. We found continued
radiological surveillance with particular focus on humeral lucency
is warranted and has not previously been reported. Despite advances
in the design of total elbow replacement prostheses, rates of complication
remain high. Cite this article:
We present a comparison of the results of the Oxford unicompartmental knee arthroplasty in patients younger and older than 60 years of age. The ten-year all-cause survival of the <
60 years of age group (52) was 91% (95% confidence interval (CI) 12), while in the ≥ 60 years of age group (512), the figure was 96% (95% CI 3). For the younger group, the mean Hospital for Special Surgery score at ten-year follow-up (n = 21) was 94 of 100, compared with a mean of 86 of 100 for the older group (n = 135). The results show that the Oxford unicompartmental arthroplasty can achieve ten-year results that are comparable to total knee arthroplasty in patients <
60 years of age. We conclude that for patients aged over 50, age should not be considered a contraindication for this procedure.
This prospective study describes the complications and survival of the first 688 Phase 3 Oxford medial unicompartmental knee replacements implanted using a minimally-invasive technique by two surgeons and followed up independently. None was lost to follow-up. We had carried out 132 of the procedures more than five years ago. The clinical assessment of 101 of these which were available for review at five years is also presented. Nine of the 688 knees were revised: four for infection, three for dislocation of the bearing and two for unexplained pain. A further seven knees (1%) required other procedures: four had a manipulation under anaesthesia, two an arthroscopy and one a debridement for superficial infection. The survival rate at seven years was 97.3% (95% confidence interval 5.3). At five years, 96% of the patients had a good or excellent American Knee Society score, the mean Oxford knee score was 39 and the mean flexion was 133°. This study demonstrates that the minimally-invasive Oxford unicompartmental knee replacement is a reliable and effective procedure.
The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term.
The December 2012 Knee Roundup360 looks at: the demand for knee replacement; a Japanese knee outcome score; smoking and TKR; coronal alignment as a determinant of outcome in TKR; fixed flexion; MRI detected knee lesions; and lateral domed Oxford unicompartmental knee replacements.
The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients
with end-stage anteromedial osteoarthritis of the knee, with radiological
evidence of improved fixation compared with previous reports using
cemented fixation. Cite this article: