Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset femoral stems, and occurs within a central concentric
zone of the femoral head bore taper. Further investigation is required
to determine the effect of increased head size, and variations in
head–neck taper design. Cite this article:
This review summarises the evidence for the treatment
of displaced fractures of the femoral neck in elderly patients.
Results from randomised clinical trials and national register studies are
presented when available. The advantages of arthroplasty compared with internal fixation
are supported by several studies. A number of studies contribute
to the discussions of total hip arthroplasty (THA) The direct lateral approach reduces the risk of dislocation compared
with the posterior approach. Cemented implants lower the risk of
periprosthetic fracture and its subsequent morbidity and mortality.
As the risk of peri-operative death related to bone cement can be reduced
by adequate measures, cemented implants are recommended in fracture
cases. Take home message: There remains a great variation in the surgical
management of patients with a hip fracture, and an evidence-based
approach should improve the outcomes for this vulnerable patient
group. Cite this article:
The optimal management of intracapsular fractures of the femoral
neck in independently mobile patients remains open to debate. Successful
fixation obviates the limitations of arthroplasty for this group
of patients. However, with fixation failure rates as high as 30%,
the outcome of revision surgery to salvage total hip arthroplasty
(THA) must be considered. We carried out a systematic review to
compare the outcomes of salvage THA and primary THA for intracapsular
fractures of the femoral neck. We performed a Preferred Reporting Items for Systematic Reviews
and Meta-Analysis (PRISMA) compliant systematic review, using the
PubMed, EMBASE and Cochrane libraries databases. A meta-analysis
was performed where possible, and a narrative synthesis when a meta-analysis
was not possible.Aims
Patients and Methods
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
Bone allografts can store and release high levels of vancomycin. We present our results of a two-stage treatment for infected hip arthroplasty with acetabular and femoral impaction grafting using vancomycin-loaded allografts. We treated 29 patients (30 hips) by removal of the implants, meticulous debridement, parenteral antibiotic therapy and second-stage reconstruction using vancomycin-supplemented impacted bone allografts and a standard cemented Charnley femoral component. The mean follow-up was 32.4 months (24 to 60). Infection control was obtained in 29 cases (re-infection rate of 3.3%; 95% confidence interval 0.08 to 17) without evidence of progressive radiolucent lines, demarcation or graft resorption. One patient had a further infection ten months after revision caused by a different pathogen. Associated post-operative complications were one traumatic periprosthetic fracture at 14 months, a single dislocation in two hips and four displacements of the greater trochanter. Vancomycin-supplemented allografts restored bone stock and provided sound fixation with a low incidence of further infection.
Tapered, fluted, modular, titanium stems have
a long history in Europe and are increasing in popularity in North America.
We have reviewed the results at our institution looking at stem
survival and clinical outcomes. Radiological outcomes and quality
of life assessments have been performed and compared to cylindrical
non-modular cobalt chromium stems. Survival at five years was 94%.
This fell to 85% at ten years due to stem breakage with older designs.
Review of radiology showed maintenance or improvement of bone stock
in 87% of cases. Outcome scores were superior in tapered stems despite
worse pre-operative femoral deficiency. Tapered stems have proved
to be a useful alternative in revision total hip arthroplasty across
the spectrum of femoral bone deficiency.
Large femoral heads have been used with increasing
frequency over the last decade. The prime reason is likely the effect
of large heads on stability. The larger head neck ratio, combined
with the increased jump distance of larger heads result in a greater
arc of impingement free motion, and greater resistance to dislocation
in a provocative position. Multiple studies have demonstrated clear
clinical efficacy in diminishing dislocation rates with the use
of large femoral heads. With crosslinked polyethylene, wear has
been shown to be equivalent between larger and smaller heads. However,
the stability advantages of increasing diameter beyond 38 mm have
not been clearly demonstrated. More importantly, recent data implicates
large heads in the increasing prevalence of groin pain and psoas impingement.
There are clear benefits with larger femoral head diameters, but
the advantages of diameters beyond 38 mm have not yet been demonstrated
clinically.
We describe a new technique of reconstruction of the deficient acetabulum in cementless total hip arthroplasty. The outer iliac table just above the deficient acetabulum is osteotomised and slid downwards. We have termed this an iliac sliding graft. Between October 1997 and November 2001, cementless total hip arthroplasty with an iliac sliding graft was performed on 19 patients (19 hips) with acetabular dysplasia. The mean follow-up was 3.4 years (2 to 6). The mean pre-operative Harris hip score was 45.1 which improved significantly to 85.3 at the time of the final follow-up. No patient had post-operative abductor dysfunction. Incorporation of the graft was seen after two to three months in all patients. Resorption of the graft and radiolucencies were infrequent. This technique is a useful alternative to femoral head autografting when the patient’s own femoral head cannot be used.
Failure of total hip arthroplasty with acetabular deficiency occurred in 55 patients (60 hips) and was treated with acetabular revision using morsellised allograft and a cemented metal-backed component. A total of 50 patients (55 hips) were available for clinical and radiological evaluation at a mean follow-up of 5.8 years (3 to 9.5). No hip required further revision of the acetabular component because of aseptic loosening. All the hips except one had complete incorporation of the allograft demonstrated on the radiographs. A complete radiolucent line of >
1 mm was noted in two hips post-operatively. A good to excellent result occurred in 50 hips (91%). With radiological evidence of aseptic loosening of the acetabular component as the end-point, the survivorship at a mean of 5.8 years after surgery was 96.4%. The use of impacted allograft chips in combination with a cemented metal-backed acetabular component and screw fixation can achieve good medium-term results in patients with acetabular bone deficiency.
The purpose of this paper is to discuss the risk
factors, prevention strategies, classification, and treatment of
intra-operative femur fractures sustained during primary and revision
total hip arthroplasty.
We report the clinical and radiological outcome of 86 revisions of cemented hip arthroplasties using JRI-Furlong hydroxyapatite-ceramic-coated acetabular and femoral components. The acetabular component was revised in 62 hips and the femoral component in all hips. The mean follow-up was 12.6 years and no patient was lost to follow-up. The mean age of the patients was 71.2 years. The mean Harris hip and Oxford scores were 82 (59 to 96) and 23.4 (14 to 40), respectively. The mean Charnley modification of the Merle d’Aubigné and Postel score was 5 (3 to 6) for pain, 4.9 (3 to 6) for movement and 4.4 (3 to 6) for mobility. Migration of the acetabular component was seen in two hips and the mean acetabular inclination was 42.6°. The mean linear polyethylene wear was 0.05 mm/year. The mean subsidence of the femoral component was 1.9 mm and stress shielding was seen in 23 (28%) with bony ingrowth in 76 (94%). Heterotopic ossification was seen in 12 hips (15%). There were three re-revisions, two for deep sepsis and one for recurrent dislocation and there were no re-revisions for aseptic loosening. The mean EuroQol EQ-5D description scores and health thermometer scores were 0.69 (0.51 to 0.89) and 79 (54 to 95), respectively. With an end-point of definite or probable loosening, the probability of survival at 12 years was 93.9% and 95.6% for the acetabular and femoral components, respectively. Overall survival at 12 years, with removal or further revision of either component for any reason as the end-point, was 92.3%. Our study supports the continued use of this arthroplasty and documents the durability of hydroxyapatite-ceramic-coated components.
In this paper we make the case for the use of
single-stage revision for infected knee arthroplasty.
The in-cement technique for revision hip arthroplasty involves retaining the original cement-bone interface. This has been proven to be a biomechanically stronger method than recementing after complete removal of the original cement mantle. This study reviewed a series of 54 consecutive revision hip arthroplasty procedures, using the in-cement technique, between November 1999 and November 2003. Clinical and radiological follow-up included functional assessment. There were 54 procedures performed in 51 patients, whose mean age at surgery was 70.3 years (45 to 85). A total of 42 were available at a mean follow-up of 29.2 months (6 to 51). There was no radiological evidence of loosening. Functional assessments were available for 40 patients who had a mean Harris hip score of 85.2 (51.9 to 98.5), a mean Oxford hip score of 19.6 (12 to 41), a mean UCLA activity profile score of 5.9 (3 to 8) and a mean SF-36 score of 78.0 (31.6 to 100). The in-cement technique provides consistent, high functional outcomes and should be considered in appropriately selected cases.
Impacted morsellised allografts have been used successfully to address the problem of poor bone stock in revision surgery. However, there are concerns about the transmission of pathogens, the high cost and the shortage of supply of donor bone. Bone-graft extenders, such as tricalcium phosphate (TCP) and hydroxyapatite (HA), have been developed to minimise the use of donor bone. In a human cadaver model we have evaluated the surgical and mechanical feasibility of a TCP/HA bone-graft extender during impaction grafting revision surgery. A TCP/HA allograft mix increased the risk of producing a fissure in the femur during the impaction procedure, but provided a higher initial mechanical stability when compared with bone graft alone. The implications of the use of this type of graft extender in impaction grafting revision surgery are discussed.
There is no diagnostic, non-invasive method for the early detection of loosening after total hip arthroplasty. In a pilot study, we have analysed two serum markers of bone remodelling, procollagen I C-terminal extension peptide (PICP) and cross-linked N-terminal telopeptide (NTx), as well as the diagnostic performance of NTx for the assessment of osteolysis. We recruited 21 patients with loosening (group I), 18 with a well-fixed prosthesis (group II) and 17 at the time of primary arthroplasty for osteoarthritis (OA) (group III). Internal normal reference ranges were obtained from 30 healthy subjects (group IV). The serum PICP level was found to be significantly lower in patients with OA and those with loosening, when compared with those with stable implants, while the NTx level was significantly increased only in the group with loosening, suggesting that collagen degradation depended on the altered bone turnover induced by the implant. This hypothesis was reinforced by the finding that the values in the pre-surgery patients and stable subjects were comparable with the reference range of younger healthy subjects. A high specificity and positive predictive value for NTx provided good diagnostic evidence of agreement between the test and the clinical and radiological evaluations. The NTx level could be used to indicate stability of the implant. However, further prospective, larger studies are necessary.