Vulcano E Myerson MS

The last decade has seen a considerable increase in the use of in total ankle arthroplasty (TAA) to treat patients with end-stage arthritis of the ankle. However, the longevity of the implants is still far from that of total knee and hip arthroplasties.

The aim of this review is to outline a diagnostic and treatment algorithm for the painful TAA to be used when considering revision surgery.

Cite this article: Bone Joint J 2017;99-B:5–11.


Bone & Joint 360
Vol. 7, Issue 1 | Pages 15 - 17
1 Feb 2018


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 190 - 196
1 Feb 2018
Chraim M Krenn S Alrabai HM Trnka H Bock P

Aims

Hindfoot arthrodesis with retrograde intramedullary nailing has been described as a surgical strategy to reconstruct deformities of the ankle and hindfoot in patients with Charcot arthropathy. This study presents case series of Charcot arthropathy patients treated with two different retrograde intramedullary straight compression nails in order to reconstruct the hindfoot and assess the results over a mid-term follow-up.

Patients and Methods

We performed a retrospective analysis of 18 consecutive patients and 19 operated feet with Charcot arthropathy who underwent a hindfoot arthrodesis using a retrograde intramedullary compression nail. Patients were ten men and eight women with a mean age of 63.43 years (38.5 to 79.8). We report the rate of limb salvage, complications requiring additional surgery, and fusion rate in both groups. The mean duration of follow-up was 46.36 months (37 to 70).


Bone & Joint Research
Vol. 7, Issue 1 | Pages 20 - 27
1 Jan 2018
Kang K Son J Suh D Kwon SK Kwon O Koh Y

Objectives

Patient-specific (PS) implantation surgical technology has been introduced in recent years and a gradual increase in the associated number of surgical cases has been observed. PS technology uses a patient’s own geometry in designing a medical device to provide minimal bone resection with improvement in the prosthetic bone coverage. However, whether PS unicompartmental knee arthroplasty (UKA) provides a better biomechanical effect than standard off-the-shelf prostheses for UKA has not yet been determined, and still remains controversial in both biomechanical and clinical fields. Therefore, the aim of this study was to compare the biomechanical effect between PS and standard off-the-shelf prostheses for UKA.

Methods

The contact stresses on the polyethylene (PE) insert, articular cartilage and lateral meniscus were evaluated in PS and standard off-the-shelf prostheses for UKA using a validated finite element model. Gait cycle loading was applied to evaluate the biomechanical effect in the PS and standard UKAs.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 199 - 203
1 Feb 2017
Sandiford NA Jameson SS Wilson MJ Hubble MJW Timperley AJ Howell JR

Aims

We present the clinical and radiological results at a minimum follow-up of five years for patients who have undergone multiple cement-in-cement revisions of their femoral component at revision total hip arthroplasty (THA).

Patients and Methods

We reviewed the outcome on a consecutive series of 24 patients (10 men, 14 women) (51 procedures) who underwent more than one cement-in-cement revision of the same femoral component. The mean age of the patients was 67.5 years (36 to 92) at final follow-up.

Function was assessed using the original Harris hip score (HHS), Oxford Hip Score (OHS) and the Merle D’Aubigné Postel score (MDP).


Bone & Joint 360
Vol. 6, Issue 5 | Pages 24 - 27
1 Oct 2017


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1031 - 1037
1 Aug 2015
da Assunção RE Pollard TCB Hrycaiczuk A Curry J Glyn-Jones S Taylor A

Periprosthetic femoral fracture (PFF) is a potentially devastating complication after total hip arthroplasty, with historically high rates of complication and failure because of the technical challenges of surgery, as well as the prevalence of advanced age and comorbidity in the patients at risk.

This study describes the short-term outcome after revision arthroplasty using a modular, titanium, tapered, conical stem for PFF in a series of 38 fractures in 37 patients.

The mean age of the cohort was 77 years (47 to 96). A total of 27 patients had an American Society of Anesthesiologists grade of at least 3. At a mean follow-up of 35 months (4 to 66) the mean Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly associated with a poorer OHS. All fractures united and no stem needed to be revised. Three hips in three patients required further surgery for infection, recurrent PFF and recurrent dislocation and three other patients required closed manipulation for a single dislocation. One stem subsided more than 5 mm but then stabilised and required no further intervention.

In this series, a modular, tapered, conical stem provided a versatile reconstruction solution with a low rate of complications.

Cite this article: Bone Joint J 2015;97-B:1031–7.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 312 - 317
1 Mar 2015
Amanatullah DF Howard JL Siman H Trousdale RT Mabry TM Berry DJ

Revision total hip arthroplasty (THA) is challenging when there is severe loss of bone in the proximal femur. The purpose of this study was to evaluate the clinical and radiographic outcomes of revision THA in patients with severe proximal femoral bone loss treated with a fluted, tapered, modular femoral component. Between January 1998 and December 2004, 92 revision THAs were performed in 92 patients using a single fluted, tapered, modular femoral stem design. Pre-operative diagnoses included aseptic loosening, infection and peri-prosthetic fracture. Bone loss was categorised pre-operatively as Paprosky types III-IV, or Vancouver B3 in patients with a peri-prosthetic fracture. The mean clinical follow-up was 6.4 years (2 to 12). A total of 47 patients had peri-operative complications, 27 of whom required further surgery. However, most of these further operations involved retention of a well-fixed femoral stem, and 88/92 femoral components (97%) remained in situ. Of the four components requiring revision, three were revised for infection and were well fixed at the time of revision; only one (1%) was revised for aseptic loosening. The most common complications were post-operative instability (17 hips, 19%) and intra-operative femoral fracture during insertion of the stem (11 hips, 12%). Diaphyseal stress shielding was noted in 20 hips (22%). There were no fractures of the femoral component. At the final follow-up 78% of patients had minimal or no pain.

Revision THA in patients with extensive proximal femoral bone loss using the Link MP fluted, tapered, modular stem led to a high rate of osseointegration of the stem at mid-term follow-up.

Cite this article: Bone Joint J 2015; 97-B:312–17.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (sd 1.71) vs 0.32 mm (sd 0.45), p = 0.66; rotation 0.96° (sd 1.49) vs 1.41° (sd 2.95), p = 0.88; and total migration 1.09 mm (sd 1.74) vs 0.73 mm (sd 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem.

Cite this article: Bone Joint J 2015; 97-B:595–602.


Aims

To investigate the longevity of uncemented fixation of a femoral component in total hip arthroplasty (THA) in patients with Dorr type C proximal femoral morphology.

Patients and Methods

A total of 350 consecutive uncemented THA in 320 patients were performed between 1983 and 1987, by a single surgeon using the Taperloc femoral component. The 63 patients (68 hips) with Dorr type C proximal femoral morphology were the focus of this review. The mean age of the patients was 69 years (24 to 88) and mean follow-up was 16.6 years (ten to 29). Survival analysis included eight patients (eight hips) who died without undergoing revision surgery prior to obtaining ten years follow-up. All 55 surviving patients (60 hips) were available for clinical assessment and radiographic review. As a comparator group, the survival and implant fixation in the remaining 282 THAs (257 patients) with Dorr type A and B morphology were evaluated. The mean age of these patients was 52 years (20 to 82).


Bone & Joint Research
Vol. 6, Issue 1 | Pages 22 - 30
1 Jan 2017
Scott CEH Eaton MJ Nutton RW Wade FA Evans SL Pankaj P

Objectives

Up to 40% of unicompartmental knee arthroplasty (UKA) revisions are performed for unexplained pain which may be caused by elevated proximal tibial bone strain. This study investigates the effect of tibial component metal backing and polyethylene thickness on bone strain in a cemented fixed-bearing medial UKA using a finite element model (FEM) validated experimentally by digital image correlation (DIC) and acoustic emission (AE).

Materials and Methods

A total of ten composite tibias implanted with all-polyethylene (AP) and metal-backed (MB) tibial components were loaded to 2500 N. Cortical strain was measured using DIC and cancellous microdamage using AE. FEMs were created and validated and polyethylene thickness varied from 6 mm to 10 mm. The volume of cancellous bone exposed to < -3000 µε (pathological loading) and < -7000 µε (yield point) minimum principal (compressive) microstrain and > 3000 µε and > 7000 µε maximum principal (tensile) microstrain was computed.


Aims

The purpose of this study was to compare the clinical and radiographic outcomes of total ankle arthroplasty (TAA) in patients with pre-operatively moderate and severe arthritic varus ankles to those achieved for patients with neutral ankles.

Patients and Methods

A total of 105 patients (105 ankles), matched for age, gender, body mass index, and follow-up duration, were divided into three groups by pre-operative coronal plane tibiotalar angle; neutral (< 5°), moderate (5° to 15°) and severe (> 15°) varus deformity. American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot score, a visual analogue scale (VAS), and Short Form (SF)-36 score were used to compare the clinical outcomes after a mean follow-up period of 51 months (24 to 147).


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 56 - 59
1 Nov 2014
Konan S Garbuz DS Masri BA Duncan CP

Non-modular tapered fluted, titanium stems are available for use in femoral revision. The combination of taper and flutes on the stem provides axial and rotational stability, respectively. The material and surface properties of the stem promotes bone on-growth. If the surgeon is confident and reasonably experienced in the surgical use of this sort of design and the case is relatively straightforward, a non-modular design is effective. It also potentially reduces implant inventory, and circumvents the potential problems of taper junction corrosion and fatigue fracture. There are reports of excellent survival, good clinical and functional results and evidence of subsequent increase in proximal bone stock.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):56–9.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1319 - 1328
1 Oct 2017
Shelton TJ Nedopil AJ Howell SM Hull ML

Aims

The aims of this study were to determine the proportion of patients with outlier varus or valgus alignment in kinematically aligned total knee arthroplasty (TKA), whether those with outlier varus or valgus alignment have higher forces in the medial or lateral compartments of the knee than those with in-range alignment and whether measurements of the alignment of the limb, knee and components predict compartment forces.

Patients and Methods

The intra-operative forces in the medial and lateral compartments were measured with an instrumented tibial insert in 67 patients who underwent a kinematically aligned TKA during passive movement. The mean of the forces at full extension, 45° and 90° of flexion determined the force in the medial and lateral compartments. Measurements of the alignment of the limb and the components included the hip-knee-ankle (HKA) angle, proximal medial tibial angle (PMTA), and distal lateral femoral angle (DLFA). Measurements of the alignment of the knee and the components included the tibiofemoral angle (TFA), tibial component angle (TCA) and femoral component angle (FCA). Alignment was measured on post-operative, non-weight-bearing anteroposterior (AP) scanograms and categorised as varus or valgus outlier or in-range in relation to mechanically aligned criteria.


Bone & Joint Research
Vol. 6, Issue 7 | Pages 433 - 438
1 Jul 2017
Pan M Chai L Xue F Ding L Tang G Lv B

Objectives

The aim of this study was to compare the biomechanical stability and clinical outcome of external fixator combined with limited internal fixation (EFLIF) and open reduction and internal fixation (ORIF) in treating Sanders type 2 calcaneal fractures.

Methods

Two types of fixation systems were selected for finite element analysis and a dual cohort study. Two fixation systems were simulated to fix the fracture in a finite element model. The relative displacement and stress distribution were analysed and compared. A total of 71 consecutive patients with closed Sanders type 2 calcaneal fractures were enrolled and divided into two groups according to the treatment to which they chose: the EFLIF group and the ORIF group. The radiological and clinical outcomes were evaluated and compared.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 65 - 74
1 Jan 2016
Phadnis J Huang T Watts A Krishnan J Bain GI

Aims

To date, there is insufficient evidence available to compare the outcome of cemented and uncemented fixation of the humeral stem in reverse shoulder arthroplasty (RSA).

Methods

A systemic review comprising 41 clinical studies was performed to compare the functional outcome and rate of complications of cemented and uncemented stems in RSA. These included 1455 cemented and 329 uncemented shoulders. The clinical characteristics of the two groups were similar.

Variables were compared using pooled frequency-weighted means and relative risk ratios (RR).


Bone & Joint 360
Vol. 6, Issue 2 | Pages 10 - 13
1 Apr 2017


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 623 - 631
1 May 2017
Blaney J Harty H Doran E O’Brien S Hill J Dobie I Beverland D

Aims

Our aim was to examine the clinical and radiographic outcomes in 257 consecutive Oxford unicompartmental knee arthroplasties (OUKAs) (238 patients), five years post-operatively.

Patients and Methods

A retrospective evaluation was undertaken of patients treated between April 2008 and October 2010 in a regional centre by two non-designing surgeons with no previous experience of UKAs. The Oxford Knee Scores (OKSs) were recorded and fluoroscopically aligned radiographs were assessed post-operatively at one and five years.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 14 - 17
1 Jan 2017
Carlson SW Liu SS Callaghan JJ

Aims

The aim of this study was to compare the survivorship and radiographic outcomes at ten-year follow-up of three prospective consecutive series of patients each of which received a different design of cementless femoral components for total hip arthroplasty (THA).

Patients and Methods

In Cohort 1, 91 consecutive patients (100 hips) underwent THA with a cementless porous-coated anatomic femoral stem (PCA) between October 1983 and January 1986. In Cohort 2, 86 consecutive patients (100 hips) underwent THA with an extensively porous-coated cementless femoral stem (Prodigy) between June 1994 and October 1997. In Cohort 3, 88 consecutive patients (100 hips) underwent THA with a proximally porous-coated triple-tapered cementless stem (Summit) between April 2002 and October 2003. All three groups underwent prospective clinical and radiographic evaluation.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 65 - 69
1 Jan 2017
Thienpont E

Objectives

Unicompartmental knee arthroplasty (UKA) is a potential treatment for isolated bone on bone osteoarthritis when limited to a single compartment. The risk for revision of UKA is three times higher than for total knee arthroplasty (TKA). The aim of this review was to discuss the different revision options after UKA failure.

Materials and Methods

A search was performed for English language articles published between 2006 and 2016. After reviewing titles and abstracts, 105 papers were selected for further analysis. Of these, 39 papers were deemed to contain clinically relevant data to be included in this review.