The removal of all prosthetic material and a
two-stage revision procedure is the established standard management of
an infected total hip replacement (THR). However, the removal of
well-fixed femoral cement is time-consuming and can result in significant
loss of bone stock and femoral shaft perforation or fracture. We
report our results of two-stage revision THR for treating infection,
with retention of the original well-fixed femoral cement mantle
in 15 patients, who were treated between 1989 and 2002. Following
partial excision arthroplasty, patients received local and systemic
antibiotics and underwent reconstruction and re-implantation at
a second-stage procedure, when the infection had resolved. The mean follow-up of these 15 patients was 82 months (60 to
192). Two patients had positive microbiology at the second stage
and were treated with six weeks of appropriate antibiotics; one
of these developed recurrent infection requiring further revision.
Successful eradication of infection was achieved in the remaining
14 patients. We conclude that when two-stage revision is used for the treatment
of peri-prosthetic infection involving a THR, a well-fixed femoral
cement mantle can be safely left
Juvenile idiopathic arthritis (JIA) is a chronic disease of childhood; it causes joint damage which may require surgical intervention, often in the young adult. The aim of this study was to describe the long-term outcome and survival of hip replacement in a group of adult patients with JIA and to determine predictors of survival for the prosthesis. In this retrospective comparative study patients were identified from the database of a regional specialist adult JIA clinic. This documented a series of 47 hip replacements performed in 25 adult patients with JIA. Surgery was performed at a mean age of 27 years (11 to 47), with a mean follow-up of 19 years (2 to 36). The mean Western Ontario and McMaster Universities osteoarthritis index questionnaire (WOMAC) score at the last follow-up was 53 (19 to 96) and the mean Health Assessment Questionnaire score was 2.25 (0 to 3). The mean pain component of the WOMAC score (60 (20 to 100)) was significantly higher than the mean functional component score (46 (0 to 97)) (p = 0.02). Kaplan-Meier survival analysis revealed a survival probability of 46.6% (95% confidence interval 37.5 to 55.7) at 19 years, with a trend towards enhanced survival with the use of a cemented acetabular component and a cementless femoral component. This was not, however, statistically significant (acetabular component, p = 0.76, femoral component, p = 0.45). Cox’s proportional hazards regression analysis showed an implant survival rate of 54.9% at 19 years at the mean of covariates. Survival of the prosthesis was significantly poorer (p = 0.001) in patients who had been taking long-term corticosteroids and significantly better (p = 0.02) in patients on methotrexate.
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the BHR from an independent centre. In this cohort of 554 patients (646 BHRs) with a mean age of
51.9 years (16.5 to 81.5) followed for a mean of eight years (1
to 12), the survival and patient-reported functional outcome depended
on gender and the size of the implant. In female hips (n = 267)
the ten-year survival was 74% (95% confidence interval (CI) 83 to
91), the ten-year revision rate for pseudotumour was 7%, the mean
Oxford hip score (OHS) was 43 ( This study supports the ongoing use of resurfacing in young active
men, who are a subgroup of patients who tend to have problems with
conventional THR. In contrast, the results in women have been poor
and we do not recommend metal-on-metal resurfacing in women. Continuous
follow-up is recommended because of the increasing incidence of
pseudotumour with the passage of time.
Fracture of a pedicle is a rare complication of spinal instrumentation using pedicular screws, but it can lead to instability and pain and may necessitate extension of the fusion. Osteosynthesis of the fractured pedicle by cerclage-wire fixation and augmentation of the screw fixation by vertebroplasty or temporary elongation of the fixation, allows stabilisation without sacrifice of the adjacent healthy segment. We describe three patients who developed a fracture of the pedicle in the most caudal instrumented vertebra early after lumbar spinal fusion. During revision surgery the pedicles were reduced and secured by a soft cerclage wire bilaterally. Fusion was obtained at the site of the primary instrumentation and healing of the pedicles was achieved. Cerclage wiring of the fractured pedicle seems to be safe and avoids permanent extension of the fusion without the sacrifice of an otherwise healthy segment.
In order to compare the outcome from surgical repair and physiotherapy, 103 patients with symptomatic small and medium-sized tears of the rotator cuff were randomly allocated to one of the two approaches. The primary outcome measure was the Constant score, and secondary outcome measures included the self-report section of the American Shoulder and Elbow Surgeons score, the Short Form 36 Health Survey and subscores for shoulder movement, pain, strength and patient satisfaction. Scores were taken at baseline and after six and 12 months by a blinded assessor. Nine patients (18%) with insufficient benefit from physiotherapy after at least 15 treatment sessions underwent secondary surgical treatment. Analysis of between-group differences showed better results for the surgery group on the Constant scale (difference 13.0 points, p − 0.002), on the American Shoulder and Elbow surgeons scale (difference 16.1 points, p <
0.0005), for pain-free abduction (difference 28.8°, p = 0.003) and for reduction in pain (difference on a visual analogue scale −1.7 cm, p <
0.0005).
We report the long-term results of the management
of neglected chronically infected total knee replacements with a
two-stage re-implantation protocol. In 18 of 34 patients (53%) a
resistant organism was isolated. All cases were treated by the same
surgical team in a specialist centre and had a mean follow-up of
12.1 years (10 to 14). They were evaluated clinically and radiologically
using the Knee Society Score (KSS) and the American Knee Society Roentgenographic
scoring system, respectively. One patient died after eight years
from an unrelated cause and two were lost to follow-up. Three patients
(8.8%) developed a recurrent infection for which further surgery
was required. The infection was eradicated successfully in 31 patients
(91.1%). There was one case of aseptic loosening after 13 years.
We found a significant improvement in the KSS at final follow-up
(p <
0.001).
Blood metal ions have been widely used to investigate
metal-on-metal hip replacements, but their ability to discriminate
between well-functioning and failed hips is not known. The Medicines
and Healthcare products Regulatory Agency (MHRA) has suggested a
cut-off level of 7 parts per billion (ppb). We performed a pair-matched, case-control study to investigate
the sensitivity and specificity of blood metal ion levels for diagnosing
failure in 176 patients with a unilateral metal-on-metal hip replacement.
We recruited 88 cases with a pre-revision, unexplained failed hip
and an equal number of matching controls with a well-functioning
hip. We investigated the 7 ppb cut-off level for the maximum of
cobalt or chromium and determined optimal mathematical cut-off levels
from receiver-operating characteristic curves. The 7 ppb cut-off level for the maximum of cobalt or chromium
had a specificity of 89% and sensitivity 52% for detecting a pre-operative
unexplained failed metal on metal hip replacement. The optimal cut-off
level for the maximum of cobalt or chromium was 4.97 ppb and had
sensitivity 63% and specificity 86%. Blood metal ions had good discriminant ability to separate failed
from well-functioning hip replacements. The MHRA cut-off level of
7 ppb provides a specific test but has poor sensitivity.
We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p <
0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p <
0.01). Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.
We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips). Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant. This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p <
0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the
Arthroscopy of the native hip is an established diagnostic and therapeutic procedure. Its application in the symptomatic replaced hip is still being explored. We describe the use of arthroscopy of the hip in 24 symptomatic patients following total hip replacement, resurfacing arthroplasty of the hip and partial resurfacing (study group), and compared it with arthroscopy of the native hip in 24 patients (control group). A diagnosis was made or confirmed at arthroscopy in 23 of the study group and a therapeutic arthroscopic intervention resulted in relief of symptoms in ten of these. In a further seven patients it led to revision hip replacement. In contrast, arthroscopy in the control group was diagnostic in all 24 patients and the resulting arthroscopic therapeutic intervention provided symptomatic relief in 21. The mean operative time in the study group (59.7 minutes (35 to 93)) was less than in the control group (71 minutes (40 to 100), p = 0.04) but the arthroscopic approach was more difficult in the arthroplasty group. We suggest that arthroscopy has a role in the management of patients with a symptomatic arthroplasty when other investigations have failed to provide a diagnosis.
The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had radiolucent lines and two acetabular components migrated initially before stabilising. Acetabular reconstruction in both primary and revision surgery using a 1:1 mixture of frozen, ground, irriadiated bone and Apapore 60 appears to be a reliable method of managing acetabular defects. Longer follow-up will be required to establish whether this technique is as effective as using fresh-frozen allograft.
The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the early results of the use of this prosthesis for established isolated patellofemoral arthritis in 51 consecutive patellofemoral replacements in 35 patients. The mean follow-up was 25 months (5 to 60). The estimated survival rate at three years was 63% (95% confidence interval 47 to 80) with revision as the endpoint and 46% (95% confidence interval 30 to 63) with revision and ongoing moderate or severe pain as the endpoint. The early results of the use of the low contact stress patellofemoral replacement are disappointing with a high rate of revision. We cannot therefore recommend its use.
Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements. We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant. At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 ( The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.
In a systematic review, reports from national registers and clinical studies were identified and analysed with respect to revision rates after joint replacement, which were calculated as revisions per 100 observed component years. After primary hip replacement, a mean of 1.29 revisions per 100 observed component years was seen. The results after primary total knee replacement are 1.26 revisions per 100 observed component years, and 1.53 after medial unicompartmental replacement. After total ankle replacement a mean of 3.29 revisions per 100 observed component years was seen. The outcomes of total hip and knee replacement are almost identical. Revision rates of about 6% after five years and 12% after ten years are to be expected.
This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
We have evaluated the extent to which diabetes affects the revision rate following total hip replacement (THR). Through the Danish Hip Arthroplasty Registry we identified all patients undergoing a primary THR (n = 57 575) between 1 January 1996 and 31 December 2005, of whom 3278 had diabetes. The presence of diabetes among these patients was identified through the Danish National Registry of Patients and the Danish National Drug Prescription Database. We estimated the relative risk for revision and the 95% confidence intervals for patients with diabetes compared to those without, adjusting for the confounding factors. Diabetes is associated with an increased risk of revision due to deep infection (relative risk = 1.45 (95% confidence interval 1.00 to 2.09), particularly in those with type 2 diabetes (relative risk = 1.49 (95% confidence interval 1.02 to 2.18)), those with diabetes for less than five years prior to THR (relative risk = 1.69 (95% confidence interval 1.24 to 2.32)), those with complications due to diabetes (relative risk = 2.11 (95% confidence interval 1.41 to 3.17)), and those with cardiovascular comorbidities prior to surgery (relative risk = 2.35 (95% confidence interval 1.39 to 3.98)). Patients and surgeons should be aware of the relatively elevated risk of revision due to deep infection following THR in diabetes particularly in those with insufficient control of their glucose level.
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p <
0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with <
1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.
We evaluated the outcome in a series of patients with recurrent patellar dislocation who had either medial transfer of the tibial tuberosity and lateral release or an isolated lateral release as the primary treatment. The decision to use one or other procedure was based on a pre-operative distance between the tibial tuberosity to the trochlear groove (TTTG) of less than 10 mm to include the tibial tuberosity transfer in addition to the lateral release. Between April 2002 and December 2006, 49 patients (63 knees) underwent one of these procedures. A total of 35 patients (46 knees) was evaluated at a mean of 38 months (13 to 71) post-operatively. Medial transfer of the tibial tuberosity was performed in 33 knees and isolated lateral release in the remaining 13. Evaluation included the International Knee Documentation Committee (IKDC), the Kujala and the Short-form 36 scores. From the tibial tuberosity group 23 knees also underwent radiological examination at follow-up. There were further episodes of patellar dislocation in six of the 46 knees available for review. Further dislocation was noted in five of 33 knees (15.2%) in the tibial tuberosity transfer group and in one of 13 knees (7.7%) in the lateral release group. The mean subjective IKDC score was 80.4 (