The August 2013 Hip &
Pelvis Roundup360 looks at: are we getting it right first time?; tantalum augments in revision hip surgery; lower wear in dual mobility?; changing faces changes outcomes; synovial fluid aspiration in MOM hips; taper disease: the new epidemic of hip surgery; the super-obese and THR; and whether well fixed stems can remain in infected hips
We evaluated the outcome of 41 consecutive Charnley
low-friction arthroplasties (LFAs) performed by a single surgeon
in 28 patients aged ≤ 35 years at operation between 23 and 36 years
previously. There were 20 women and eight men with a mean age of
32 years (23 to 35) at surgery. Two patients (three hips) were lost
to follow-up at 12 and 17 years post-operatively, respectively,
and one patient (one hip) died at 13 years post-operatively. These patients
were excluded from the final evaluation. The survival rate of the
acetabular components was 92.7% (95% confidence interval (CI) 88.7
to 96.7) at ten years, 67.1% (95% CI 59.75 to 74.45) at 20 years
and 53.2% (95% CI 45.3 to 61.1) at 25 years. For the femoral component
the survival was 95.1% (95% CI 91.8 to 98.5) at ten years, 77.1%
(95% CI 73.9 to 80.3) at 20 years and 68.2% (95% CI 60.7 to 75.8)
at 25 years. The results indicate that the Charnley LFA remains
a reasonable choice in the treatment of young patients and can serve
for comparison with newer techniques and implants. Cite this article:
We report on the long-term results of 163 bicruciate-retaining
Hermes 2C total knee replacements in 130 patients at a mean follow-up
of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament
had a partially degenerative appearance it was preserved as long
as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The
description and surgical technique of this minimally constrained
prosthesis were published in 1983 and the ten-year clinical results
in 1999. A total of 12% of the knees (20 of 163) in this study were revised
because of wear of the polyethylene tibial insert. Excellent stability
was achieved and the incidence of aseptic component loosening was
4.3% (seven of 163). The survival rate using revision for any reason
as the endpoint was 82% (95% confidence interval 76.2 to 88.0). Although this series included a relatively small number of replacements,
it demonstrated that the anterior cruciate ligament, even when partially
degenerated at the time of TKR, remained functional and provided
adequate stability at a long-term follow-up. Cite this article:
A retrospective review was performed of patients
undergoing primary cementless total knee replacement (TKR) using
porous tantalum performed by a group of surgical trainees. Clinical
and radiological follow-up involved 79 females and 26 males encompassing
115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up
was 7 years (2 to 11). Tibial and patellar components were porous
tantalum monoblock implants, and femoral components were posterior
stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological
assessments were made for implant positioning, alignment, radiolucencies,
lysis, and loosening. There was 95.7% survival of implants. There
was no radiological evidence of loosening and no osteolysis found.
No revisions were performed for aseptic loosening. Average tibial
component alignment was 1.4° of varus (4°of valgus to 9° varus),
and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean
femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral
alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar
tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient
satisfaction with improvement in pain was 91%. Cementless TKR incorporating
porous tantalum yielded good clinical and radiological outcomes
at a mean of follow-up of seven-years. Cite this article:
The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age >
70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p <
0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age <
55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure. We attribute the success of the AGC implant to its relatively unconstrained articular geometry and the durability of a non-modular metal-backed tibial component with compression moulded polyethylene.
Stress shielding resulting in diminished bone
density following total knee replacement (TKR) may increase the
risk of migration and loosening of the prosthesis. This retrospective
study was designed to quantify the effects of the method of fixation
on peri-prosthetic tibial bone density beneath cemented and uncemented
tibial components of similar design and with similar long-term survival
rates. Standard radiographs taken between two months and 15 years
post-operatively were digitised from a matched group of TKRs using
cemented (n = 67) and uncemented (n = 67) AGC tibial prostheses.
Digital radiograph densitometry was used to quantify changes in
bone density over time. Age, length of follow-up, gender, body mass
index and alignment each significantly influenced the long-term
pattern of peri-prosthetic bone density. Similar long-term changes
in density irrespective of the method of fixation correlated well
with the high rate of survival of this TKR at 20 years, and suggest
that cemented and uncemented fixation are both equally viable. Cite this article:
In this retrospective study we evaluated the
proficiency of shelf autograft in the restoration of bone stock
as part of primary total hip replacement (THR) for hip dysplasia,
and in the results of revision arthroplasty after failure of the primary
arthroplasty. Of 146 dysplastic hips treated by THR and a shelf
graft, 43 were revised at an average of 156 months, 34 of which
were suitable for this study (seven hips were excluded because of
insufficient bone-stock data and two hips were excluded because
allograft was used in the primary THR). The acetabular bone stock
of the hips was assessed during revision surgery. The mean implant–bone
contact was 58% (50% to 70%) at primary THR and 78% (40% to 100%)
at the time of the revision, which was a significant improvement
(p <
0.001). At primary THR all hips had had a segmental acetabular
defect >
30%, whereas only five (15%) had significant segmental
bone defects requiring structural support at the time of revision.
In 15 hips (44%) no bone graft or metal augments were used during
revision. A total of 30 hips were eligible for the survival study. At a
mean follow-up of 103 months (27 to 228), two aseptic and two septic
failures had occurred. Kaplan-Meier survival analysis of the revision
procedures demonstrated a ten-year survival rate of 93.3% (95% confidence
interval (CI) 78 to 107) with clinical or radiological failure as
the endpoint. The mean Oxford hip score was 38.7 (26 to 46) for
non-revised cases at final follow-up. Our results indicate that the use of shelf autografts during
THR for dysplastic hips restores bone stock, contributing to the
favourable survival of the revision arthroplasty should the primary
procedure fail. Cite this article:
The Exeter femoral stem is a double-tapered highly
polished collarless cemented implant with good long-term clinical
results. In order to determine why the stem functions well we have
undertaken a long-term radiostereometric analysis (RSA) study. A total of 20 patients undergoing primary Exeter total hip replacement
for osteoarthritis using the Hardinge approach were recruited and
followed with RSA for ten years. The stems progressively subsided
and internally rotated with posterior head migration. The mean subsidence
was 0.7 mm (95% confidence interval (CI) 0.5 to 0.9) at two years
and 1.3 mm (95% CI 1.0 to 1.6) at ten years. The mean posterior
migration of the head was 0.7 mm (95% CI 0.5 to 0.9) at two years
and 1.2 mm (95% CI 1.0 to 1.4) at ten years. There was no significant
cement restrictor migration. The Exeter stem continues to subside slowly into the cement mantle
in the long term. This appears to compress the cement and the cement
bone interface, contributing to secure fixation in the long term. Cite this article:
We report the mid- to long-term (mean 20.3 years, 10 to 32.5) results of the Chiari pelvic osteotomy in patients with pre- to advanced stage osteoarthritis in dysplastic hips. We followed 163 Japanese patients (173 hips) with a mean age at surgery of 20 years (9 to 54). Overall, 124 hips (72%) had satisfactory results, with Harris hip scores ≥ 80. Satisfactory results were seen in 105 of 134 hips with pre- or early osteoarthritis (78%) and 19 of 39 hips with advanced osteoarthritis (49%). A total of 15 hips (9%) underwent a total hip replacement (THR) with a mean interval between osteotomy and THR of 16.4 years. With conversion to THR as the endpoint, the 30-year survival rate was 85.9% (95% confidence interval 82.3 to 89.5). It was 91.8% for patients with pre- or early osteoarthritis and 43.6% for those with advanced osteoarthritis (p < 0.001). We now perform the Chiari osteotomy for patients with dysplastic hips showing poor joint congruency and who prefer a joint-conserving procedure to THR.
With greater numbers of younger patients undergoing
total hip replacement (THR), the effect of patient age on the diameter
of the femoral canal may become more relevant. This study aimed
to investigate the relationship between the diameter of the diaphysis
of the femoral canal with increasing age in a large number of patients
who underwent THR. A total of 1685 patients scheduled for THR had
their femoral dimensions recorded from calibrated radiographs. There
were 736 males and 949 females with mean ages of 67.1 years (34
to 92) and 70.2 years (29 to 92), respectively. The mean diameter
of the femoral canal was 13.3 mm (8.0 to 23.0) for males and 12.7
mm (6.0 to 26.0) for females. There was a poor correlation between
age and the diameter of the canal in males (r = 0.071, p = 0.05)
but a stronger correlation in females (r = 0.31, p <
0.001). The diameter of the femoral canal diameter of a female patient
undergoing THR could be predicted to increase by 3.2 mm between
the ages of 40 and 80 years, in contrast a male would be expected
to experience only a 0.6 mm increase during the same period. This
increase in the diameter of the canal with age might affect the
long-term survival of the femoral component in female patients. Cite this article:
The combination of an irreparable tear of the rotator cuff and destructive arthritis of the shoulder joint may cause severe pain, disability and loss of independence in the aged. Standard anatomical shoulder replacements depend on a functioning rotator cuff, and hence may fail in the presence of tears in the cuff. Many designs of non-anatomical constrained or semi-constrained prostheses have been developed for cuff tear arthropathy, but have proved unsatisfactory and were abandoned. The DePuy Delta III reverse prosthesis, designed by Grammont, medialises and stabilises the centre of rotation of the shoulder joint and has shown early promise. This study evaluated the mid-term clinical and radiological results of this arthroplasty in a consecutive series of 50 shoulders in 43 patients with a painful pseudoparalysis due to an irreparable cuff tear and destructive arthritis, performed over a period of seven years by a single surgeon. A follow-up of 98% was achieved, with a mean duration of 39 months (8 to 81). The mean age of the patients at the time of surgery was 81 years (59 to 95). The female to male ratio was 5:1. During the seven years, six patients died of natural causes. The clinical outcome was assessed using the American Shoulder and Elbow score, the Oxford Shoulder Score and the Short-form 36 score. A radiological review was performed using the Sirveaux score for scapular notching. The mean American Shoulder and Elbow score was 19 (95% confidence interval (CI) 14 to 23) pre-operatively, and 65 (95% CI 48 to 82) (paired The mean maximum elevation improved from 55° pre-operatively to 105° at final follow-up. There were seven complications during the whole series, although only four patients required further surgery.
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
Mobile-bearing (MB) total knee replacement (TKR)
was introduced to reduce the risk of aseptic loosening and wear of
polyethylene inserts. However, no consistent clinical advantages
of mobile- over fixed-bearing (FB) TKR have been found. In this
study we evaluated whether mobile bearings have an advantage over
fixed bearings with regard to revision rates and clinical outcome
scores. Furthermore, we determined which modifying variables affected
the outcome. A systematic search of the literature was conducted to collect
clinical trials comparing MB and FB in primary TKR. The primary
outcomes were revision rates for any reason, aseptic loosening and
wear. Secondary outcomes included range of movement, Knee Society
score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score
and radiological parameters. Meta-regression techniques were used
to explore factors modifying the observed effect. Our search yielded 1827 publications, of which 41 studies met
our inclusion criteria, comprising over 6000 TKRs. Meta-analyses
showed no clinically relevant differences in terms of revision rates,
clinical outcome scores or patient-reported outcome measures between
MB and FB TKRs. It appears that theoretical assumptions of superiority
of MB over FB TKR are not borne out in clinical practice. Cite this article:
We evaluated the long-term outcome of patients with an osteosarcoma who had undergone prior manipulative therapy, a popular treatment in Asia, and investigated its effects on several prognostic factors. Of the 134 patients in this study, 70 (52%) patients had manipulative therapy and 64 (48%) did not. The age, location, and size of tumour were not significantly different between the groups. The five-year overall survival rate was 58% and 92% in the groups with and without manipulative therapy (p = 0.004). Both the primary and overall rates of lung metastasis were significantly higher in the manipulative group (primary: 32% This form of therapy may serve as a mechanism to accelerate the spread of tumour cells, and therefore must be avoided in order to improve the outcome for patients with an osteosarcoma.
A total of 157 hindquarter amputations were carried
out in our institution during the last 30 years. We have investigated
the reasons why this procedure is still required and the outcome.
This operation was used as treatment for 13% of all pelvic bone
sarcomas. It was curative in 140 and palliative in 17, usually to
relieve pain. There were 90 primary procedures (57%) with the remaining
67 following the failure of previous operations to control the disease
locally. The indication for amputation in primary disease was for
large tumours for which limb-salvage surgery was no longer feasible.
The peri-operative mortality was 1.3% (n = 2) and major complications
of wound healing or infection arose in 71 (45%) patients. The survival
at five years after hindquarter amputation with the intent to cure
was 45%, and at ten years 38%. Local recurrence occurred in 23 patients
(15%). Phantom pain was a significant problem, and only 20% used
their prosthesis regularly. Functional scores were a mean of 57%.
With careful patient selection the oncological results and functional outcomes
of hindquarter amputation justify its continued use. Cite this article:
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients
with end-stage anteromedial osteoarthritis of the knee, with radiological
evidence of improved fixation compared with previous reports using
cemented fixation. Cite this article:
The October 2012 Hip &
Pelvis Roundup360 looks at: diagnosing the infected hip replacement; whether tranexamic acid has a low complication rate; the relationship between poor cementing technique and early failure of resurfacing; debridement and retention for the infected replacement; triple-tapered stems and bone mineral density; how early discharge can be bad for your sleep; an updated QFracture algorithm to predict the risk of an osteoporotic fracture; and local infiltration analgesia and total hip replacement.
Rarely, the extent of a malignant bone tumour
may necessitate resection of the complete humerus to achieve adequate
oncological clearance. We present our experience with reconstruction
in such cases using a total humeral endoprosthesis (THER) in 20
patients (12 male and eight female) with a mean age of 22 years
(6 to 59). We assessed the complications, the oncological and functional
outcomes and implant survival. Surgery was performed between June
2001 and October 2009. The diagnosis included osteosarcoma in nine,
Ewing’s sarcoma in eight and chondrosarcoma in three. One patient
was lost to follow-up. The mean follow-up was 41 months (10 to 120)
for all patients and 56 months (25 to 120) in survivors. There were
five local recurrences (26.3%) and 11 patients were alive at time
of last follow-up, with overall survival for all patients being
52% (95% confidence interval (CI) 23.8 to 74) at five years. The
mean Musculoskeletal Tumor Society score for the survivors was 22
(73%; 16 to 23). The implant survival was 95% (95% CI 69.5 to 99.3)
at five years. The use of a THER in the treatment of malignant tumours of bone
is oncologically safe; it gives consistent and predictable results
with low rates of complication.
Between 1990 and 2000, 123 hips in 110 patients were reconstructed for aseptic loosening using impaction bone grafting with frozen, irradiated, morsellised femoral heads and cemented acetabular components. This series was reported previously at a mean follow-up of five years. We have extended this follow-up and now describe the outcome of 86 hips in 74 patients at a mean of ten years. There have been 19 revisions, comprising nine for infection, seven for aseptic loosening and three for dislocation. In surviving acetabular reconstructions, union of the graft had occurred in 64 of 67 hips (95.5%). Survival analysis for all indications at ten years was 83.3% (95% confidence interval (CI) 68 to 89) and 71.3% (95% CI 58 to 84) at 15 years. Acetabular reconstruction using irradiated allograft and a cemented acetabular component is an effective method of reconstruction, providing results in the medium- to long-term comparable with those of reported series where non-irradiated freshly-frozen bone was used.