We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231 knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196 patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1 mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the
Locking after total knee replacement is uncommon and is generally caused by the formation of fibrous tissue around the patella. We report an unusual cause of locking resulting from intermittent occlusion of the popliteal artery, which was tethered to cement at the posterior aspect of the
We randomised 62 knees to receive either cemented or cementless versions of the Oxford unicompartmental knee replacement. The implants used in both arms of the study were similar, except that the cementless components were coated with porous titanium and hydroxyapatite. The tibial interfaces were studied with fluoroscopically-aligned radiographs. At one year there was no difference in clinical outcome between the two groups. Narrow radiolucent lines were seen at the bone-implant interfaces in 75% of cemented
After obtaining informed consent, 80 patients were randomised to undergo a navigated or conventional total knee replacement. All received a cemented, unconstrained, cruciate-retaining implant with a rotating platform. Full-length standing and lateral radiographs and CT scans of the hip, knee and ankle joint were carried out five to seven days after operation. No notable differences were found between computer-assisted navigation and conventional implantation techniques as regards the rotational alignment of the femoral or
The purpose of this prospective, randomised study
was to evaluate the clinical and radiological results comparing the
identical cemented or cementless NexGen total knee prostheses implanted
bilaterally in the same patient. Sequential simultaneous bilateral
total knee replacements were performed in 50 patients (100 knees).
There were 39 women and 11 men with a mean age of 58.4 years (51
to 67) who received a cemented prosthesis in one knee and a cementless
prosthesis in the other. The mean follow-up was 13.6 years (13 to
14). At final review, the mean Knee Society scores (96.2 (82 to
100) versus 97.7 (90 to 100)), the mean Western
Ontario and McMaster Universities osteoarthritis index (34.5 (4
to 59) versus 35.6 (5 to 51)), the mean ranges
of knee movement (124° (100° to 140°) versus 128°
(110° to 140°)), mean patient satisfaction (8.1 (. sd. 1.9) versus 8.3
(. sd. 1.7)), and radiological results were similar in both
groups. The rate of survival of the femoral components was 100%
in both groups at 14 years. The rate of survival of the cemented
tibial component was 100% and 98% in the cementless
The relationship between post-operative bone
density and subsequent failure of total knee replacement (TKR) is
not known. This retrospective study aimed to determine the relationship
between bone density and failure, both overall and according to
failure mechanism. All 54 aseptic failures occurring in 50 patients
from 7760 consecutive primary cemented TKRs between 1983 and 2004
were matched with non-failing TKRs, and 47 failures in 44 patients
involved tibial failures with the matching characteristics of age
(65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis
(93.6% OA), date of operation, bilaterality, pre-operative alignment
(0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively).
In each case, the density of bone beneath the
Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating TKA patients with metal hypersensitivity/allergy. All studies reporting on diagnosing or managing metal hypersensitivity in TKA were included. Data were extracted and summarized based on study design, study population, interventions and outcomes. A practical guide is then formulated based on the available evidence.Aims
Methods
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB
This was a retrospective analysis of the medium-
to long-term results of 46 TC3 Sigma revision total knee replacements
using long uncemented stems in press-fit mode. Clinical and radiological analysis took place pre-operatively,
at two years post-operatively, and at a mean follow-up of 8.5 years
(4 to 12). The mean pre-operative International Knee Society (IKS)
clinical score was 42 points (0 to 74), improving to 83.7 (52 to
100) by the final follow-up. The mean IKS score for function improved
from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up.
At the final follow-up 30 knees (65.2%) had an excellent result, seven
(15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor
result. There were two failures, one with anteroposterior instability
and one with aseptic loosening. The TC3 revision knee system, when used with press-fit for long
intramedullary stems and cemented femoral and
Experiments were carried out to determine the optimum conformity between the femoral and tibial condyles in condylar replacement knee prostheses. Wear tests and observations from removed prostheses indicated that both high and low conformity produced characteristic abrasion and fatigue. Partly conforming condyles provided stability under load-bearing but allowed laxity to occur. Fixation to resist the various forces on the
We describe the medium-term results of a prospective study of 200 total ankle replacements at a single-centre using the Scandinavian Total Ankle Replacement. A total of 24 ankles (12%) have been revised, 20 by fusion and four by further replacement and 27 patients (33 ankles) have died. All the surviving patients were seen at a minimum of five years after operation. The five-year survival was 93.3% (95% confidence interval (CI) 89.8 to 96.8) and the ten-year survival 80.3% (95% CI 71.0 to 89.6). Anterior subluxation of the talus, often seen on the lateral radiograph in osteoarthritic ankles, was corrected and, in most instances, the anatomical alignment was restored by total ankle replacement. The orientation of the
The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit. A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.Aims
Methods
We studied the effect of a layer of cement placed under the
Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and
We measured the contact areas and contact stresses at the post-cam mechanism of a posterior-stabilised total knee arthroplasty when a posterior force of 500 N was applied to the Kirschner Performance, Scorpio Superflex, NexGen LPS Flex Fixed, and NexGen LPS Flex Mobile knee systems. Measurements were made at 90°, 120°, and 150° of flexion both in neutral rotation and 10° of internal rotation of the
We performed a prospective study of 54 patients (76 knees) who underwent Osteonics series 3000 cruciate-retaining cementless total knee arthroplasty between December 1990 and June 1993. Five patients (seven knees) were lost to follow-up (90.7% completion). One patient required revision at 10.5 years after operation. The rate of survival was 100% at ten years and 96.7% at 13 years. The mean ten-year knee and function scores were 79 and 59 respectively. Both were significantly better than the pre-operative scores. The range of movement also improved. Although a radiolucent line around the
Osteoarthritis of the medial compartment of the knee often shows a specific pattern of anterior wear. Review of our revisions from a series of medial metal-backed Brigham unicondylar knee replacements performed between 1983 and 1989 showed that this wear pattern was common on the tibial polyethylene surface. We reviewed these cases retrospectively to compare the pattern of preoperative erosion with the wear of the prosthesis. In all 14 knees with severe anterior wear in a unicompartmental replacement, the prearthroplasty radiographs showed similar patterns, suggesting that the implanted
We have studied 27 tibial prostheses retrieved from knee replacements after 1 to 9 years. In 22 the femoral components were of cobalt-chrome, in five polyacetal. The design of the components gave a nominal contact area of 320 mm2 on each condyle. The
It is unknown whether gap laxities measured in robotic arm-assisted total knee arthroplasty (TKA) correlate to load sensor measurements. The aim of this study was to determine whether symmetry of the maximum medial and lateral gaps in extension and flexion was predictive of knee balance in extension and flexion respectively using different maximum thresholds of intercompartmental load difference (ICLD) to define balance. A prospective cohort study of 165 patients undergoing functionally-aligned TKA was performed (176 TKAs). With trial components in situ, medial and lateral extension and flexion gaps were measured using robotic navigation while applying valgus and varus forces. The ICLD between medial and lateral compartments was measured in extension and flexion with the load sensor. The null hypothesis was that stressed gap symmetry would not correlate directly with sensor-defined soft tissue balance.Aims
Methods