We divided 309 patients with an inflammatory arthritis who had undergone primary
We present the results of 525 primary Souter
The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function. No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).
Between 1978 and 1986, 59 patients received a GSB III elbow prosthesis, six of them in both elbows. Rheumatoid arthritis (RA) was the underlying cause in 51 of the patients and post-traumatic osteoarthritis (PTOA) in eight. Of these, 24 patients (28 prostheses) have since died; two, both operated on bilaterally, had had their implants for more than ten years and had already been assessed for inclusion in the long-term follow-up. Two patients, each with one elbow prosthesis, have been lost to follow-up and three males who are still living (two with PTOA, one with juvenile RA) had their prosthesis removed before ten years had elapsed. The remaining 32 patients (28 RA, 4 PTOA) with 36 GSB III elbows were examined clinically and radiologically after a mean period of 13.5 years. Pain was considerably reduced in 91.6%. Mobility was increased by 37° in those with RA and by 67° in those with PTOA. There were three cases of aseptic loosening and three of deep infection. The main complication was disassembly of the prosthetic component in nine elbows (13.8%). This last group included two patients with postoperative fractures unrelated to the operative technique and one with neuropathic arthritis. Ulnar neuritis occurred in two patients. Since 87.7% of all the GSB III prostheses implanted in this period remained in situ, our results are comparable with those for hip and knee arthroplasty.
From 1984 to 1995, 68 ankylosed elbows and 11 which were unstable after trauma were replaced in India by Baksi sloppy hinge prostheses. The mean age of the patients was 28.6 years (17 to 70) and the mean follow-up 9.6 years (2 to 13.5). Of the 68 ankylosed elbows, 59 (87%) regained a mean arc of painless movement of 88.5° (27 to 115). The mean improvement of supination was 24° and of pronation 16.5°. There were 54 good results (80%), eight fair and three poor. There were two complete failures due to infection, and one due to a broken humeral stem. Of the 11 unstable elbows, the nine with good results had a mean arc of 125° (15 to 140) of painless stable movement, with a mean improvement in supination of 26° and of pronation of 19.5°. There was one fair result and one failure due to loosening with subsequent late infection. There were significant complications in 14 cases with infection in seven and aseptic loosening in four. Patients with loosening or late removal of the prosthesis often retained reasonably stable elbow movement because periprosthetic fibrosis had connected the approximated bone ends, and muscle balance had been restored.
There have been few reports in the literature of total elbow arthroplasty extending beyond 10 to 15 years. We reviewed 40 patients (41 elbows) with a mean age of 56 years (19 to 83) who had undergone a Coonrad/Coonrad-Morrey
Ulnar neuropathy presents as a complication in 5% to 10% of total elbow replacements, but subsequent ulnar neurolysis is rarely performed. Little information is available on the surgical management of persistent ulnar neuropathy after
We studied retrospectively the results of revision arthroplasty of the elbow using a linked Coonrad-Morrey implant in 23 patients (24 elbows) after a mean follow-up period of 55 months. According to the Mayo Elbow Performance Score, 19 elbows were satisfactory, nine were excellent and ten good. The median total score had improved from 35 points (20 to 75) before the primary arthroplasty to 85 points (40 to 100) at the latest follow-up. There was a marked relief of pain, but the range of movement showed no overall improvement. Two patients had a second revision because of infection and two for aseptic loosening. The estimated five-year survival rate of the prosthesis was 83.1% (95% confidence interval 61.1 to 93.3). Revision
We examined the effects of previous resection of the radial head and synovectomy on the outcome of subsequent total elbow arthroplasty in patients with rheumatoid arthritis. Fifteen elbows with a history of resection and synovectomy were compared with a control group of patients who had
We present six patients with chronic dislocation of the elbow who were treated by primary semiconstrained total elbow arthroplasty. All were women with a mean age of 65 years (51 to 76), the mean interval between dislocation and surgery was 17 weeks (5 to 52) and the mean follow-up 58 months (24 to 123). The most dramatic improvement was in function. The mean American Shoulder and Elbow Surgeon score was 5.2 times better (p <
0.001) and the mean total range of movement increased from 33° to 121° (p <
0.001) after operation. Three patients developed wear of polyethylene. One required revision for a periprosthetic fracture, and another required a bushing exchange. Primary semiconstrained
Seven patients with chronic insufficiency of the triceps were treated by either a rotation flap using anconeus (4) or an allograft of tendo Achillis (3). The latter procedure was selected for patients with a large defect in whom the anconeus muscle had been devitalised. Five disruptions were in patients who had previously undergone an
Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde
We report the results of a series of 59 unconstrained total arthroplasties of the elbow after a mean follow-up of 6.5 years (4 to 10). All the patients had rheumatoid arthritis. The indication for surgery was pain in all but one. Outcome was assessed by the Mayo Elbow Performance Index (MEPI). Of the 59 arthroplasties, two were lost to follow-up. Ten patients died, but as two of their arthroplasties were failures we included them in the results. The outcome in the remaining 49 was excellent in 26 (53%), good in 15 (31%), fair in one (2%) and poor or a failure in seven (14%). There was an improvement in the pain score (p <
0.001), movement (p <
0.001) and function (p <
0.001). Two patients developed instability, but neither required further surgery. There was a mean increase of 21° in flexion and of 7° in extension. The overall rate of complications was 33.9%. Lesions of the ulnar nerve, one of which did not recover, occurred in four patients, deep infection in two and stiffness in five. The rates of complications were similar to those in recent reports of other
This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection. Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was
We investigated the incidence of and risk factors
for venous thromboembolism (VTE) following surgery of the shoulder
and elbow and assessed the role of thromboprophylaxis in upper limb
surgery. All papers describing VTE after shoulder and elbow surgery
published in the English language literature before 31 March 2012
were reviewed. A total of 14 papers were available for analysis,
most of which were retrospective studies and case series. The incidence
of VTE was 0.038% from 92 440 shoulder arthroscopic procedures,
0.52% from 42 261 shoulder replacements, and 0.64% from 4833 procedures
for fractures of the proximal humerus (open reduction and internal fixation
or hemiarthroplasty). The incidence following replacement of the
elbow was 0.26% from 2701 procedures. Diabetes mellitus, rheumatoid
arthritis and ischaemic heart disease were identified as the major
risk factors. The evidence that exists on thromboprophylaxis is based on level
III and IV studies, and we therefore cannot make any recommendations
on prophylaxis based on the current evidence. It seems reasonable
to adopt a multimodal approach that involves all patients receiving
mechanical prophylaxis, with chemical prophylaxis reserved for those
who are at high risk for VTE. Cite this article:
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
We present the outcome of 47 Souter-Strathclyde replacements of the elbow with a mean follow-up of 82 months (12 to 129). The clinical results were assessed using a condition-specific outcome measure. The mean total score (maximum 100) before the operation was 47.21 and improved to 79.92 (p <
0.001). The mean pain score (maximum 50) improved from 21.41 to 46.70 (p <
0.001) and the mean functional component of the score (maximum 30) from 11.19 to 18.65 (p <
0.001). There was negligible change in the score for the range of movement although a significant improvement in mean flexion from 124° to 136° was noted (p <
0.001). Revision surgery was required in four patients, for dislocation, wound dehiscence and early infection in one, late infection in two and aseptic loosening in one. The cumulative survival was 75% at nine years for all causes of failure and 97% at ten years for aseptic loosening alone. Our study demonstrates the value of the Souter-Strathclyde total elbow arthroplasty in providing relief from pain and functional improvement in rheumatoid patients.
Arthrolysis and dynamic splinting have been used in the treatment of elbow contractures, but there is no standardised protocol for treatment of severe contractures with an arc of flexion <
30°. We present our results of radical arthrolysis with twin incisions with the use of a monolateral hinged fixator to treat very severe extra-articular contracture of the elbow. This retrospective study included 26 patients (15 males and 11 females) with a mean age of 30 years (12 to 60). The mean duration of stiffness was 9.1 months (5.4 to 18) with mean follow-up of 5.2 years (3.5 to 9.4). The mean pre-operative arc of movement was 15.6° (0° to 30°), with mean pre-operative flexion of 64.1° (30° to 120°) and mean pre-operative extension of 52.1° (10° to 90°). Post-operatively the mean arc improved to 102.4° (60° to 135°), the mean flexion improved to 119.1° (90° to 140°) and mean extension improved to 16.8° (0° to 30°) (p <
0.001). The Mayo elbow score improved from a mean of 45 (30 to 65) to 89 (75 to 100) points, and 13 had excellent, nine had good, three had fair and one had a poor result. We had one case of severe instability and one wound dehiscence which responded well to treatment. One case had deep infection with poor results which responded well to treatment. Our findings indicate that this method is very effective in the treatment of severe elbow contracture; however, a randomised controlled study is necessary for further evaluation.
We reviewed 20 patients who had undergone a Coonrad-Morrey total elbow arthroplasty after resection of a primary or metastatic tumour from the elbow or distal humerus between 1980 and 2002. Eighteen patients underwent reconstruction for palliative treatment with restoration of function after intralesional surgery and two after excision of a primary bone tumour. The mean follow-up was 30 months (1 to 192). Five patients (25%) were alive at the final follow-up; 14 (70%) had died of their disease and one of unrelated causes. Local control was achieved in 15 patients (75%). The mean Mayo Elbow Performance Score improved from 22 (5 to 45) to 75 points (55 to 95). Four reconstructions (20%) failed and required revision. Seven patients (35%) had early complications, the most frequent being nerve injury (25%). There were no infections or wound complications although 18 patients (90%) had radiotherapy, chemotherapy or both. The Coonrad-Morrey total elbow arthroplasty provides good relief from pain and a good functional outcome after resection of tumours of the elbow. The rates of complications involving local recurrence of tumour (25%) and nerve injury (25%) are of concern.
There is little information about the management
of peri-prosthetic fracture of the humerus after total shoulder replacement
(TSR). This is a retrospective review of 22 patients who underwent
a revision of their original shoulder replacement for peri-prosthetic
fracture of the humerus with bone loss and/or loose components.
There were 20 women and two men with a mean age of 75 years (61
to 90) and a mean follow-up 42 months (12 to 91): 16 of these had
undergone a previous revision TSR. Of the 22 patients, 12 were treated
with a long-stemmed humeral component that bypassed the fracture.
All their fractures united after a mean of 27 weeks (13 to 94).
Eight patients underwent resection of the proximal humerus with
endoprosthetic replacement to the level of the fracture. Two patients
were managed with a clam-shell prosthesis that retained the original
components. The mean Oxford shoulder score (OSS) of the original
TSRs before peri-prosthetic fracture was 33 (14 to 48). The mean
OSS after revision for fracture was 25 (9 to 31). Kaplan-Meier survival
using re-intervention for any reason as the endpoint was 91% (95%
confidence interval (CI) 68 to 98) and 60% (95% CI 30 to 80) at
one and five years, respectively. There were two revisions for dislocation of the humeral head,
one open reduction for modular humeral component dissociation, one
internal fixation for nonunion, one trimming of a prominent screw
and one re-cementation for aseptic loosening complicated by infection,
ultimately requiring excision arthroplasty. Two patients sustained
nerve palsies. Revision TSR after a peri-prosthetic humeral fracture associated
with bone loss and/or loose components is a salvage procedure that
can provide a stable platform for elbow and hand function. Good
rates of union can be achieved using a stem that bypasses the fracture.
There is a high rate of complications and function is not as good as
with the original replacement.