Aims.
Aims. The purpose of the present study was to compare
We conducted a meta-analysis, including randomised
controlled trials (RCTs) and cohort studies, to examine the effect
of patient-specific instruments (PSI) on radiological outcomes after
total knee replacement (TKR) including: mechanical axis alignment
and malalignment of the femoral and tibial components in the coronal,
sagittal and axial planes, at a threshold of >
3º from neutral.
Relative risks (RR) for malalignment were determined for all studies
and for RCTs and cohort studies separately. Of 325 studies initially identified, 16 met the eligibility criteria,
including eight RCTs and eight cohort studies. There was no significant
difference in the likelihood of mechanical axis malalignment with
PSI We conclude that PSI does not improve the accuracy of alignment
of the components in TKR compared with conventional instrumentation. Cite this article:
Accurate placement of the acetabular component during total hip
arthroplasty (THA) is an important factor in the success of the
procedure. However, the reported accuracy varies greatly and is
dependent upon whether free hand or navigated techniques are used.
The aim of this study was to assess the accuracy of an instrument
system that incorporates 3D printed, patient-specific guides designed
to optimise the placement of the acetabular component. A total of 100 consecutive patients were prospectively enrolled
and the accuracy of placement of the acetabular component was measured
using post-operative CT scans.Aims
Patients and Methods
The February 2024 Oncology Roundup. 360. looks at: Does primary tumour resection improve survival for patients with sarcomas of the pelvis with metastasis at diagnosis?; Proximal femur replacements for an oncologic indication offer a durable endoprosthetic reconstruction option: a 40-year experience; The importance of awaiting biopsy results in solitary pathological proximal femoral fractures: do we need to biopsy solitary pathological fractures?; Effect of radiotherapy on local recurrence, distant metastasis, and overall survival in 1,200 extremity soft-tissue sarcoma patients; What to choose in bone tumour resections?
Improvements in the surgical technique of total
knee replacement (TKR) are continually being sought. There has recently
been interest in three-dimensional (3D) pre-operative planning using
magnetic resonance imaging (MRI) and CT. The 3D images are increasingly
used for the production of patient-specific models, surgical guides
and custom-made implants for TKR. The users of
In this study we randomised 140 patients who
were due to undergo primary total knee arthroplasty (TKA) to have the
procedure performed using either patient-specific cutting guides
(PSCG) or conventional instrumentation (CI). The primary outcome measure was the mechanical axis, as measured
at three months on a standing long-leg radiograph by the hip–knee–ankle
(HKA) angle. This was undertaken by an independent observer who
was blinded to the instrumentation. Secondary outcome measures were
component positioning, operating time, Knee Society and Oxford knee
scores, blood loss and length of hospital stay. A total of 126 patients (67 in the CI group and 59 in the PSCG
group) had complete clinical and radiological data. There were 88
females and 52 males with a mean age of 69.3 years (47 to 84) and
a mean BMI of 28.6 kg/m2 (20.2 to 40.8). The mean HKA
angle was 178.9° (172.5 to 183.4) in the CI group and 178.2° (172.4
to 183.4) in the PSCG group (p = 0.34). Outliers were identified
in 22 of 67 knees (32.8%) in the CI group and 19 of 59 knees (32.2%)
in the PSCG group (p = 0.99). There was no significant difference
in the clinical results (p = 0.95 and 0.59, respectively). Operating time,
blood loss and length of hospital stay were not significantly reduced
(p = 0.09, 0.58 and 0.50, respectively) when using PSCG. The use of PSCG in primary TKA did not reduce the proportion
of outliers as measured by post-operative coronal alignment. Cite this article:
We conducted a randomised controlled trial to assess the accuracy
of positioning and alignment of the components in total knee arthroplasty
(TKA), comparing those undertaken using standard intramedullary
cutting jigs and those with patient-specific instruments (PSI). There were 64 TKAs in the standard group and 69 in the PSI group. The post-operative hip-knee-ankle (HKA) angle and positioning
was investigated using CT scans. Deviation of >
3° from the planned
position was regarded as an outlier. The operating time, Oxford
Knee Scores (OKS) and Short Form-12 (SF-12) scores were recorded.Aims
Patients and Methods
Aims. This is a multicentre, non-inventor, prospective observational study of 503 INFINITY fixed bearing total ankle arthroplasties (TAAs). We report our early experience, complications, and radiological and functional outcomes. Methods. Patients were recruited from 11 specialist centres between June 2016 and November 2019. Demographic, radiological, and functional outcome data (Ankle Osteoarthritis Scale, Manchester Oxford Questionnaire, and EuroQol five-dimension five-level score) were collected preoperatively, at six months, one year, and two years. The Canadian Orthopaedic Foot and Ankle Society (COFAS) grading system was used to stratify deformity. Early and late complications and reoperations were recorded as adverse events. Radiographs were assessed for lucencies, cysts, and/or subsidence. Results. In all, 500 patients reached six-month follow-up, 420 reached one-year follow-up, and 188 reached two-year follow-up. The mean age was 67.8 years (23.9 to 88.5). A total of 38 patients (7.5%) presented with inflammatory arthritis. A total of 101 (20.0%) of implantations used
Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
Advanced 3D imaging and CT-based navigation have emerged as valuable tools to use in total knee arthroplasty (TKA), for both preoperative planning and the intraoperative execution of different philosophies of alignment. Preoperative planning using CT-based 3D imaging enables more accurate prediction of the size of components, enhancing surgical workflow and optimizing the precision of the positioning of components. Surgeons can assess alignment, osteophytes, and arthritic changes better. These scans provide improved insights into the patellofemoral joint and facilitate tibial sizing and the evaluation of implant-bone contact area in cementless TKA. Preoperative CT imaging is also required for the development of
Aims.
The February 2024 Foot & Ankle Roundup. 360. looks at: Survival of revision ankle arthroplasty; Tibiotalocalcaneal nail for the management of open ankle fractures in the elderly patient; Accuracy of a
Substantial healthcare resources have been devoted
to computer navigation and
The October 2014 Knee Roundup. 360 . looks at: microfracture equivalent to OATS; examination better than MRI in predicting hamstrings re-injury; a second view on return to play with hamstrings injuries; dislocation risks in the Oxford Unicompartmental Knee; what about the tibia?; getting on top of lateral facet pain post TKR; readmission in TKR;
L. Spencer-Gardner, J. Pierrepont, M. Topham,
J. Baré, S. McMahon, A. J. Shimmin
Total hip and knee arthroplasty (THA, TKA) are largely successful procedures; however, both have variable outcomes, resulting in some patients being dissatisfied with the outcome. Surgeons are turning to technologies such as robotic-assisted surgery in an attempt to improve outcomes. Robust studies are needed to find out if these innovations are really benefitting patients. The Robotic Arthroplasty Clinical and Cost Effectiveness Randomised Controlled Trials (RACER) trials are multicentre, patient-blinded randomized controlled trials. The patients have primary osteoarthritis of the hip or knee. The operation is Mako-assisted THA or TKA and the control groups have operations using conventional instruments. The primary clinical outcome is the Forgotten Joint Score at 12 months, and there is a built-in analysis of cost-effectiveness. Secondary outcomes include early pain, the alignment of the components, and medium- to long-term outcomes. This annotation outlines the need to assess these technologies and discusses the design and challenges when conducting such trials, including surgical workflows, isolating the effect of the operation, blinding, and assessing the learning curve. Finally, the future of robotic surgery is discussed, including the need to contemporaneously introduce and evaluate such technologies. Cite this article:
Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA. A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.Aims
Methods
This study aimed to assess the carbon footprint associated with total hip arthroplasty (THA) in a UK hospital setting, considering various components within the operating theatre. The primary objective was to identify actionable areas for reducing carbon emissions and promoting sustainable orthopaedic practices. Using a life-cycle assessment approach, we conducted a prospective study on ten cemented and ten hybrid THA cases, evaluating carbon emissions from anaesthetic room to recovery. Scope 1 and scope 2 emissions were considered, focusing on direct emissions and energy consumption. Data included detailed assessments of consumables, waste generation, and energy use during surgeries.Aims
Methods