Advertisement for orthosearch.org.uk
Results 1 - 20 of 206
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1011 - 1016
1 Aug 2011
Langton DJ Jameson SS Joyce TJ Gandhi JN Sidaginamale R Mereddy P Lord J Nargol AVF

There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent National Joint Registry data have shown clear differences in the rates of failure of different designs of hip resurfacing. Our aim was to update the failure rates related to metal debris for the Articular Surface Replacement (ASR). A total of 505 of these were implanted. Kaplan-Meier analysis showed a failure rate of 25% at six years for the ASR resurfacing and of 48.8% for the ASR total hip replacement (THR). Of 257 patients with a minimum follow-up of two years, 67 (26.1%) had a serum cobalt concentration which was greater than 7 μg/l. Co-ordinate measuring machine analysis of revised components showed that all patients suffering adverse tissue reactions in the resurfacing group had abnormal wear of the bearing surfaces. Six THR patients had relatively low rates of articular wear, but were found to have considerable damage at the trunion-taper interface. Our results suggest that wear at the modular junction is an important factor in the development of adverse tissue reactions after implantation of a large-diameter MoM THR


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 4 | Pages 504 - 509
1 May 2004
Lie SA Havelin LI Furnes ON Engesæter LB Vollset SE

We present the results for 4762 revision total hip arthroplasties with no previous infection in the hip, which were reported to the Norwegian Arthroplasty Register between 1987 and 2003. The ten-year failure rate for revised prostheses was 26% (95% CI 25 to 26). Cox regression analyses were undertaken separately for acetabular and femoral revision components. Cemented revision components without allograft was the reference category. For acetabular components, we found a significantly reduced risk of failure for uncemented revisions both with (relative risk (RR) = 0.66; 95% CI 0.43 to 0.99) and without (RR = 0.37; 95% CI 0.22 to 0.61) allograft. For femoral components, we found a significantly reduced risk of failure for uncemented revisions, both with (RR = 0.27; 95% CI 0.16 to 0.46) and without (RR = 0.22; 95% CI 0.11 to 0.46) unimpacted allograft. This reduced risk of failure also applied to cemented revision components with allograft (RR = 0.53; 95% CI 0.33 to 0.84) and with impaction bone grafting (RR = 0.34; 95% CI 0.19 to 0.62). Revision prostheses have generally inferior results when compared with primary prostheses. Recementation without allograft, and uncemented revision with bone impaction, were associated with worse results than the other revision techniques which we studied


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 608 - 615
1 May 2011
Bolland BJRF Culliford DJ Langton DJ Millington JPS Arden NK Latham JM

This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


Bone & Joint Open
Vol. 4, Issue 5 | Pages 329 - 337
8 May 2023
Khan AQ Chowdhry M Sherwani MKA McPherson EJ

Aims. Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). Methods. In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years. Results. Mean follow-up was 4.6 years (4.1 to 5.0) in the CCS group and 5.5 years (5.25 to 5.75) in the VOOF group. The mean Harris Hip Score at two-year follow-up was 83.85 in the CCS group versus 88.00 in the VOOF group (p < 0.001). At the latest follow-up, all-cause failure rate was 29.1% in the CCS group and 11.7% in the VOOF group (p = 0.003). The total cost of the VOOF technique was 7.2% of a THA, and total cost of the CCS technique was 6.3% of a THA. Conclusion. The VOOF technique decreased all-cause failure rate compared to CCS. The total cost of VOOF was 13.5% greater than CCS, but 92.8% less than a THA. Increased cost of VOOF was considered acceptable to all patients in this series. VOOF technique provides a reasonable alternative to THA in patients who cannot afford a THA procedure. Cite this article: Bone Jt Open 2023;4(5):329–337


Bone & Joint Open
Vol. 3, Issue 6 | Pages 485 - 494
13 Jun 2022
Jaubert M Le Baron M Jacquet C Couvreur A Fabre-Aubrespy M Flecher X Ollivier M Argenson J

Aims. Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. Methods. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test. Results. Our staged exchange practice shows poor results, which means a 42% mechanical spacer failure rate, and a 20% recurrent infection rate over the two years average follow-up period. Moreover, we found a high rate of spacer dislocation (23%) and a low rate of spacer fracture (8%) compared to the previous literature. Our findings stress that the majority of spacer complications and failures is reflecting a population with high comorbid burden, highlighted by the American Society of Anesthesiology grade, Charlson Comorbidity Index, and Lee score associations, as well as the cardiac, pulmonary, kidney, or hepatic chronic conditions. Conclusion. Our experience of a two-stage hip exchange revision noted important complication rates associated with high failure rates of polymethylmethacrylate spacers. These findings must be interpreted in the light of the patient’s comorbidity profiles, as the elective population for staged exchange has an increasing comorbid burden leading to poor results. In order to provide better results for this specific population, our conclusion suggests that comparative strategy studies are required to improve our therapeutic indication. Cite this article: Bone Jt Open 2022;3(6):485–494


Bone & Joint Open
Vol. 4, Issue 3 | Pages 182 - 187
14 Mar 2023
Sheridan GA Hanlon M Welch-Phillips A Spratt K Hagan R O'Byrne JM Kenny PJ Kurmis AP Masri BA Garbuz DS Hurson CJ

Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings. Methods. A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size. Results. Using a previously-reported ALVAL classification, the majority of pseudotumour-positive caseswere found to have intermediate-grade group 2 (n = 10; 63%) or group 3 (n = 4; 25%) histological findings. Two further patients (13%) had high-grade group 4 lesions. HLA-DQB1*05:03:01 (p = 0.0676) and HLA-DRB1*14:54:01 (p = 0.0676) alleles were significantly associated with a higher risk of pseudotumour formation, while HLA-DQA1*03:01:01 (p = 0.0240), HLA-DRB1*04:04:01 (p = 0.0453), HLA-C*01:02:01 (p = 0.0453), and HLA-B*27:05:02 (p = 0.0855) were noted to confer risk reduction. Conclusion. These findings confirm the association between specific HLA genotypes and the risk of pseudotumour development around MoM hip resurfacings. Specifically, the two ‘at risk’ alleles (DQB1*05:03:01 and DRB1*14:54:01) may hold clinical value in preoperative screening and prospective surgical decision-making. Cite this article: Bone Jt Open 2023;4(3):182–187


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 859 - 866
1 Jul 2022
Innocenti M Smulders K Willems JH Goosen JHM van Hellemondt G

Aims. The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). Methods. We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups. Results. The indication for revision influenced PROMs improvement over time. This finding mainly reflected preoperative differences between the groups, but diminished between the first and second postoperative years. Preoperatively, patients revised due to infection and aseptic loosening had a lower mOHS than patients with other indications for revision. Pain scores at baseline were highest in patients being revised for dislocation. Infection and aseptic loosening groups showed marked changes over time in both mOHS and EQ-5D-3L. Overall complications and re-revision rates were 35.4% and 9.7% respectively, with no differences between the groups (p = 0.351 and p = 0.470, respectively). Conclusion. Good outcomes were generally obtained regardless of the reason for revision, with patients having the poorest preoperative scores exhibiting the greatest improvement in PROMs. Furthermore, overall complication and reoperation rates were in line with previous reports and did not differ between different indications for rTHA. Cite this article: Bone Joint J 2022;104-B(7):859–866


Bone & Joint Open
Vol. 3, Issue 3 | Pages 182 - 188
1 Mar 2022
Boktor J Badurudeen A Rijab Agha M Lewis PM Roberts G Hills R Johansen A White S

Aims. In UK there are around 76,000 hip fractures occur each year 10% to 15% of which are undisplaced intracapsular. There is considerable debate whether internal fixation is the most appropriate treatment for undisplaced fractures in older patients. This study describes cannulated hip screws survivorship analysis for patients aged ≥ 60 years with undisplaced intra-capsular fractures. Methods. This was a retrospective cohort study of consecutive patients aged ≥ 60 years who had cannulated screws fixation for Garden I and II fractures in a teaching hospital between March 2013 and March 2016. The primary outcome was further same-side hip surgery. Descriptive statistics were used and Kaplan-Meier estimates calculated for implant survival. Results. A total of 114 operations were performed on 112 patients with a mean age of 80.2 years (SD 8.9). The 30-day and one-year mortality were 1% (n = 1) and 13% (n = 15), respectively. Median follow-up was 6.6 years (interquartile range 6.0 to 7.3). Kaplan-Meier estimates showed a survivorship of 95% at one year and 90% at five years (95% confidence interval 84% to 95%) for cannulated screws. Nine patients underwent further hip surgery: four revision to total hip arthroplasty, one revision to hemiarthroplasty, three removals of screws, and one haematoma washout. Posterior tilt was assessable in 106 patients; subsequent surgery was required in two of the six patients identified with a posterior angle > 20° (p = 0.035 vs angle < 20°). Of the 100 patients with angle < 20°, five-year survivorship was 91%, with seven patients requiring further surgery. Conclusion. This study of cannulated hip screw fixation for undisplaced fractures in patients aged ≥ 60 years reveals a construct survivorship without further operation of 90% at five years. Cannulated screws can be considered a safe reliable treatment option for Garden I and II fractures. Caution should be taken if posterior tilt angle on lateral view exceeds 20°, due to a higher failure rate and reoperation, and considered for similar management to Garden III and IV injuries. Cite this article: Bone Jt Open 2022;3(3):182–188


Bone & Joint Open
Vol. 3, Issue 1 | Pages 61 - 67
18 Jan 2022
van Lingen CP Ettema HB Bosker BH Verheyen CCPM

Aims. Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. Methods. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up. Results. Mean follow-up was 10.9 years (10 to 12), with a cumulative survival rate of 82.4%. Reason for revision was predominantly pseudotumour formation (68%), apart from loosening, pain, infection, and osteolysis. The prevalence of pseudotumour formation around the prostheses was 41%, while our previous report of this cohort (with a mean follow-up of 3.6 years) revealed a 39% prevalence. The ten-year revision-free survival with pseudotumour was 66.7% and without pseudotumour 92.4% (p < 0.05). There was poor discriminatory ability for cobalt for pseudotumour formation. Conclusion. This prospective study reports a minimum ten-year follow-up of large-head MoM THA. Revision rates are high, with the main reason being the sequelae of pseudotumour formation, which were rarely observed after five years of implantation. Blood ion measurements show limited discriminatory capacity in diagnosing pseudotumour formation. Our results evidence that an early comprehensive follow-up strategy is essential for MoM THA to promptly identify and manage early complications and revise on time. After ten years follow-up, we do not recommend continuing routine CT scanning or whole cobalt blood measurements, but instead enrolling these patients in routine follow-up protocols for THA. Cite this article: Bone Jt Open 2022;3(1):61–67


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1247 - 1253
1 Jul 2021
Slullitel PA Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims. There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA. Methods. We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality. Results. The cumulative incidence of septic failure was 8% (95% confidence interval (CI) 3.5 to 15) at one year, 13.8% (95% CI 7.6 to 22) at two years, and 19.7% (95% CI 12 to 28.6) at five and ten years of follow-up. A femoral bone defect worse than Paprosky IIIA (hazard ratio (HR) 13.58 (95% CI 4.86 to 37.93); p < 0.001) and obesity (BMI ≥ 30 kg/m. 2. ; HR 3.88 (95% CI 1.49 to 10.09); p = 0.005) were significantly associated with septic failure. Instability and periprosthetic fracture were the most common reasons for mechanical failure (5.7% and 4.5%, respectively). The cumulative incidence of aseptic failure was 2% (95% CI 0.4 to 7) at two years, 9% (95% CI 4 to 17) at five years, and 12% (95% CI 5 to 22) at ten years. A previous revision to treat PJI was significantly associated with non-septic failure (HR 9.93 (95% CI 1.77 to 55.46); p = 0.009). At the five-year timepoint, 93% of the patients were alive (95% CI 84% to 96%), which fell to 86% (95% CI 75% to 92%) at ten-year follow-up. Conclusion. Massive femoral bone loss was associated with greater chances of developing a further septic failure. All septic failures occurred within the first five years following the one-stage exchange. Surgeons should be aware of instability and periprosthetic fracture being potential causes of further aseptic revision surgery. Cite this article: Bone Joint J 2021;103-B(7):1247–1253


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 592 - 600
1 May 2017
Matharu GS Nandra RS Berryman F Judge A Pynsent PB Dunlop DJ

Aims. To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure. Patients and Methods. We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression. Results. Failure occurred in 97 hips (17.0%). The ten-year cumulative failure rate was 27.1% (95% confidence interval (CI) 21.6 to 33.7). Primary implantation from 2006 onwards (hazard ratio (HR) 4.30; 95% CI 1.82 to 10.1; p = 0.001) and bilateral MoM hip arthroplasty (HR 1.59; 95% CI 1.03 to 2.46; p = 0.037) predicted failure. The effect of implantation year on failure varied over time. From four years onwards following surgery, hips implanted since 2006 had significantly higher failure rates (eight years 28.3%; 95% CI 23.1 to 34.5) compared with hips implanted before 2006 (eight years 6.3%; 95% CI 2.4 to 15.8) (HR 15.2; 95% CI 2.11 to 110.4; p = 0.007). Conclusion. We observed that 36 mm MoM Pinnacle THAs have an unacceptably high ten-year failure rate, especially if implanted from 2006 onwards or in bilateral MoM hip patients. Our findings regarding implantation year and failure support recent concerns about the device manufacturing process. We recommend all patients undergoing implantation since 2006 and those with bilateral MoM hips undergo regular investigation, regardless of symptoms. Cite this article: Bone Joint J 2017;99-B:592–600


Bone & Joint Open
Vol. 4, Issue 2 | Pages 53 - 61
1 Feb 2023
Faraj S de Windt TS van Hooff ML van Hellemondt GG Spruit M

Aims

The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA).

Methods

Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 801 - 807
23 Oct 2023
Walter N Szymski D Kurtz SM Lowenberg DW Alt V Lau EC Rupp M

Aims

This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures, intertrochanteric fractures, and subtrochanteric fractures in the elderly USA population? and 2) Which factors influence adverse outcomes?

Methods

Proximal femoral fractures occurred between 1 January 2009 and 31 December 2019 were identified from the Medicare Physician Service Records Data Base. The Kaplan-Meier method with Fine and Gray sub-distribution adaptation was used to determine rates for nonunion, infection, and mechanical complications. Semiparametric Cox regression model was applied incorporating 23 measures as covariates to identify risk factors.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 775 - 782
1 Jul 2023
Koper MC Spek RWA Reijman M van Es EM Baart SJ Verhaar JAN Bos PK

Aims

The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate.

Methods

A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.


Bone & Joint Open
Vol. 5, Issue 11 | Pages 971 - 976
5 Nov 2024
Baker G Hill J O'Neill F McChesney J Stevenson M Beverland D

Aims

In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes.

Methods

A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 946 - 952
1 Sep 2023
Dhawan R Young DA Van Eemeren A Shimmin A

Aims

The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit.

Methods

Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.


Bone & Joint Open
Vol. 5, Issue 1 | Pages 28 - 36
18 Jan 2024
Selmene MA Moreau PE Zaraa M Upex P Jouffroy P Riouallon G

Aims

Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications.

Methods

Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims

Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up.

Methods

This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 511 - 517
1 May 2023
Petrie MJ Panchani S Al-Einzy M Partridge D Harrison TP Stockley I

Aims

The duration of systemic antibiotic treatment following first-stage revision surgery for periprosthetic joint infection (PJI) after total hip arthroplasty (THA) is contentious. Our philosophy is to perform an aggressive debridement, and to use a high local concentration of targeted antibiotics in cement beads and systemic prophylactic antibiotics alone. The aim of this study was to assess the success of this philosophy in the management of PJI of the hip using our two-stage protocol.

Methods

The study involved a retrospective review of our prospectively collected database from which we identified all patients who underwent an intended two-stage revision for PJI of the hip. All patients had a diagnosis of PJI according to the major criteria of the Musculoskeletal Infection Society (MSIS) 2013, a minimum five-year follow-up, and were assessed using the MSIS working group outcome-reporting tool. The outcomes were grouped into ‘successful’ or ‘unsuccessful’.