We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
This paper describes the methodology, validation and reliability
of a new computer-assisted method which uses models of the patient’s
bones and the components to measure their migration and polyethylene
wear from radiographs after total hip arthroplasty (THA). Models of the patient’s acetabular and femoral component obtained
from the manufacturer and models of the patient’s pelvis and femur
built from a single computed tomography (CT) scan, are used by a
computer program to measure the migration of the components and
the penetration of the femoral head from anteroposterior and lateral radiographs
taken at follow-up visits. The program simulates the radiographic
setup and matches the position and orientation of the models to
outlines of the pelvis, the acetabular and femoral component, and
femur on radiographs. Changes in position and orientation reflect
the migration of the components and the penetration of the femoral
head. Validation was performed using radiographs of phantoms simulating
known migration and penetration, and the clinical feasibility of
measuring migration was assessed in two patients.Aims
Materials and Methods
The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison.Aims
Patients and Methods
Tantalum (Ta) trabecular metal components are increasingly used
to reconstruct major bone defects in revision arthroplasty surgery.
It is known that some metals such as silver have antibacterial properties.
Recent reports have raised the question regarding whether Ta components
are protective against infection in revision surgery. This laboratory
study aimed to establish whether Ta has intrinsic antibacterial
properties against planktonic bacteria, or the ability to inhibit
biofilm formation. Equal-sized pieces of Ta and titanium (Ti) acetabular components
were sterilised and incubated with a low dose inoculum of either Aims
Materials and Methods
Our aim was to evaluate the radiographic characteristics of patients
undergoing total hip arthroplasty (THA) for the potential of posterior
bony impingement using CT simulations. Virtual CT data from 112 patients who underwent THA were analysed.
There were 40 men and 72 women. Their mean age was 59.1 years (41
to 76). Associations between radiographic characteristics and posterior
bony impingement and the range of external rotation of the hip were
evaluated. In addition, we investigated the effects of pelvic tilt
and the neck/shaft angle and femoral offset on posterior bony impingement.Aims
Patients and Methods
The determination of the volumetric polyethylene wear on explanted material requires complicated equipment, which is not available in many research institutions. Our aim in this study was to present and validate a method that only requires a set of polyetheretherketone balls and a laboratory balance to determine wear. The insert to be measured was placed on a balance, and a ball of the appropriate diameter was inserted. The cavity remaining between the ball and insert caused by wear was filled with contrast medium and the weight of the contrast medium was recorded. The volume was calculated from the known density of the liquid. The precision, inter- and intraobserver reliability, were determined by four investigators on four days using nine inserts with specified wear (0.094 ml to 1.626 ml), and the intra-class correlation coefficient was calculated. The feasibility of using this method in routine clinical practice and the time required for measurement were tested on 84 explanted inserts by one investigator.Objectives
Methods
To determine whether there is any association between glomerular
filtration rate (GFR) and blood cobalt (Co) and chromium (Cr) levels
in patients with metal-on-metal (MoM) hip arthroplasty. We identified 179 patients with a unilateral 36 mm diameter head
as part of a stemmed Summit-Pinnacle MoM hip arthroplasty. GFR was
calculated using the Modification of Diet in Renal Disease formula.Aims
Patients and Methods
The National Hip Fracture Database (NHFD) publishes hospital-level risk-adjusted mortality rates following hip fracture surgery in England, Wales and Northern Ireland. The performance of the risk model used by the NHFD was compared with the widely-used Nottingham Hip Fracture Score. Data from 94 hospitals on patients aged 60 to 110 who had hip fracture surgery between May 2013 and July 2013 were analysed. Data were linked to the Office for National Statistics (ONS) death register to calculate the 30-day mortality rate. Risk of death was predicted for each patient using the NHFD and Nottingham models in a development dataset using logistic regression to define the models’ coefficients. This was followed by testing the performance of these refined models in a second validation dataset.Objectives
Methods
Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding ceramic fracture.
This study aims to report the risk of revision for fracture of modern
CoC bearings and identify factors that might influence this risk,
using data from the National Joint Registry (NJR) for England, Wales, Northern
Ireland and the Isle of Man. We analysed data on 223 362 bearings from 111 681 primary CoC
THAs and 182 linked revisions for bearing fracture recorded in the
NJR. We used implant codes to identify ceramic bearing composition
and generated Kaplan-Meier estimates for implant survivorship. Logistic
regression analyses were performed for implant size and patient specific
variables to determine any associated risks for revision.Aims
Patients and Methods
The objective of this study was to assess all evidence comparing the Thompson monoblock hemiarthroplasty with modular unipolar implants for patients requiring hemiarthroplasty of the hip with respect to mortality and complications. A literature search was performed to identify all relevant literature. The population consisted of patients undergoing hemiarthroplasty of the hip for fracture. The intervention was hemiarthroplasty of the hip with a comparison between Thompson and modular unipolar prostheses. Pubmed, Embase, CINAHL, Web of Science, PROSPERO and the Cochrane Central Register of Controlled Trials. The study designs included were randomised controlled trials (RCTs), well designed case control studies and retrospective or prospective cohort studies. Studies available in any language, published at any time until September 2015 were considered. Studies were included if they contained mortality or complications.Objectives
Methods
The aim of this study was to assess the reproducibility and validity
of cross table radiographs for measuring the anteversion of the
acetabular component after total hip arthroplasty (THA) and to compare
it with measurements using CT scans. A total of 29 patients who underwent THA between June 2010 and
January 2016 were included. There were 17 men and 12 women. Their
mean age was 43 years (26 to 65). Seven patients underwent a bilateral
procedure. Thus, 36 THAs were included in the study. Lateral radiographs
and CT scans were obtained post-operatively and radiographs repeated
three weeks later. The anteversion of the acetabular component was
measured using the method described by Woo and Morrey and the ischiolateral
method described by Pulos et al and these were compared with the
results obtained from CT scans.Aims
Patients and Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
The purpose of this study was to evaluate the existing literature from 2005 to 2016 reporting on the efficacy of surgical management of patients with femoroacetabular impingement (FAI) secondary to slipped capital femoral epiphysis (SCFE). The electronic databases MEDLINE, EMBASE, and PubMed were searched and screened in duplicate. Data such as patient demographics, surgical technique, surgical outcomes and complications were retrieved from eligible studies.Objectives
Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
The injured anterior cruciate ligament (ACL) is thought to exhibit an impaired healing response, and attempts at surgical repair have not been successful. Connective tissue growth factor (CTGF) is reported to be associated with wound healing, probably through transforming growth factor beta 1 (TGF-β1). A rabbit ACL injury model was used to study the effect of CTGF on ligament recovery. Quantitative real-time PCR (qRT-PCR) was performed for detection of changes in RNA levels of TGF-β1, type 1 collagen (COL1), type 2 collagen (COL2), SRY-related high mobility group-box gene9 (SOX9), tissue inhibitor of metalloproteinase-1 (TIMP-1) and matrix metallopeptidase 13 (MMP-13). Expression of related proteins was detected by Western blotting.Objectives
Methods
The aim of this study was to compare the incidence of aseptic
loosening after the use of a cemented acetabular component and a
Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw,
Indiana) at acetabular revision with bone impaction grafting. A total of 42 patients were included in the study. Patients were
randomised to receive an all- polyethylene cemented acetabular component
(n = 19) or a TM component (n = 23). Radiostereometric analysis
and conventional radiographic examinations were performed regularly
up to two years post-operatively or until further revision.Aims
Patients and Methods
A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS
ion studies performed more than one year after surgery. The mean
age of the patients at the time of surgery was 51.7 years (29 to
70), with 38 women (68%) and 18 men (32%).Aims
Patients and Methods
There is no consensus about the best method of achieving equal
leg lengths at total hip arthroplasty (THA) in patients with Crowe
type-IV developmental dysplasia of the hip (DDH). We reviewed our
experience of a consecutive series of patients who underwent THA
for this indication. We retrospectively reviewed 78 patients (86 THAs) with Crowe
type-IV DDH, including 64 women and 14 men, with a minimum follow-up
of two years. The mean age at the time of surgery was 52.2 years
(34 to 82). We subdivided Crowe type-IV DDH into two major types
according to the number of dislocated hips, and further categorised
them into three groups according to the occurrence of pelvic obliquity
or spinal curvature. Leg length discrepancy (LLD) and functional
scores were analysed.Aims
Patients and Methods
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
The aim of this study was to evaluate the long-term clinical
and radiographic outcomes of the Birmingham Interlocking Pelvic
Osteotomy (BIPO). In this prospective study, we report the mid- to long-term clinical
outcomes of the first 100 consecutive patients (116 hips; 88 in
women, 28 in men) undergoing BIPO, reflecting the surgeon’s learning
curve. Failure was defined as conversion to hip arthroplasty. The
mean age at operation was 31 years (7 to 57). Three patients (three
hips) were lost to follow-up.Aims
Patients and Methods
We conducted a prospective study of a delta ceramic total hip
arthroplasty (THA) to determine the rate of ceramic fracture, to
characterise post-operative noise, and to evaluate the mid-term
results and survivorship. Between March 2009 and March 2011, 274 patients (310 hips) underwent
cementless THA using a delta ceramic femoral head and liner. At
each follow-up, clinical and radiological outcomes were recorded.
A Kaplan-Meier analysis was undertaken to estimate survival.Aims
Patients and Methods
We used the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man (NJR) to investigate the risk of revision due to prosthetic joint infection (PJI) for patients undergoing primary and revision hip arthroplasty, the changes in risk over time, and the overall burden created by PJI. We analysed revision total hip arthroplasties (THAs) performed due to a diagnosis of PJI and the linked index procedures recorded in the NJR between 2003 and 2014. The cohort analysed consisted of 623 253 index primary hip arthroplasties, 63 222 index revision hip arthroplasties and 7585 revision THAs performed due to a diagnosis of PJI. The prevalence, cumulative incidence functions and the burden of PJI (total procedures) were calculated. Overall linear trends were investigated with log-linear regression.Objectives
Methods
The aims of this study were to describe the prevalence of previous
lumbar surgery in patients who undergo total hip arthroplasty (THA)
and to investigate their patient-reported outcomes (PROMs) one year
post-operatively. Data from the Swedish Hip Arthroplasty Register and the Swedish
Spine Register gathered from 2002 to 2013 were merged to identify
a group of patients who had undergone lumbar surgery before THA
(n = 997) and a carefully matched one-to-one control group. We investigated
differences in the one-year post-operative PROMs between the groups.
Linear regression analyses were used to explore the associations
between previous lumbar surgery and these PROMs following THA. The
prevalence of prior lumbar surgery was calculated as the ratio of
patients identified with previous lumbar surgery between 2002 and
2012, and divided by the total number of patients who underwent
a THA in 2012.Aims
Patients and Methods
Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. Aims
Materials and Methods
The most effective surgical approach for total hip arthroplasty
(THA) remains controversial. The direct anterior approach may be
associated with a reduced risk of dislocation, faster recovery,
reduced pain and fewer surgical complications. This systematic review
aims to evaluate the current evidence for the use of this approach
in THA. Following the Cochrane collaboration, an extensive literature
search of PubMed, Medline, Embase and OvidSP was conducted. Randomised
controlled trials, comparative studies, and cohort studies were
included. Outcomes included the length of the incision, blood loss,
operating time, length of stay, complications, and gait analysis.Aims
Materials and Methods
It may not be possible to undertake revision total hip arthroplasty
(THA) in the presence of massive loss of acetabular bone stock using
standard cementless hemispherical acetabular components and metal
augments, as satisfactory stability cannot always be achieved. We
aimed to study the outcome using a reconstruction cage and a porous
metal augment in these patients. A total of 22 acetabular revisions in 19 patients were performed
using a combination of a reconstruction cage and porous metal augments.
The augments were used in place of structural allografts. The mean
age of the patients at the time of surgery was 70 years (27 to 85)
and the mean follow-up was 39 months (27 to 58). The mean number
of previous THAs was 1.9 (1 to 3). All patients had segmental defects
involving more than 50% of the acetabulum and seven hips had an
associated pelvic discontinuity. Aims
Patients and Methods
The stability of cementless acetabular components is an important
factor for surgical planning in the treatment of patients with pelvic
osteolysis after total hip arthroplasty (THA). However, the methods
for determining the stability of the acetabular component from pre-operative
radiographs remain controversial. Our aim was to develop a scoring
system to help in the assessment of the stability of the acetabular
component under these circumstances. The new scoring system is based on the mechanism of failure of
these components and the location of the osteolytic lesion, according
to the DeLee and Charnley classification. Each zone is evaluated
and scored separately. The sum of the individual scores from the
three zones is reported as a total score with a maximum of 10 points.
The study involved 96 revision procedures which were undertaken
for wear or osteolysis in 91 patients between July 2002 and December
2012. Pre-operative anteroposterior pelvic radiographs and Judet
views were reviewed. The stability of the acetabular component was
confirmed intra-operatively.Aims
Patients and Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
Lumbar fusion is known to reduce the variation in pelvic tilt
between standing and sitting. A flexible lumbo-pelvic unit increases
the stability of total hip arthroplasty (THA) when seated by increasing
anterior clearance and acetabular anteversion, thereby preventing
impingement of the prosthesis. Lumbar fusion may eliminate this protective
pelvic movement. The effect of lumbar fusion on the stability of
total hip arthroplasty has not previously been investigated. The Medicare database was searched for patients who had undergone
THA and spinal fusion between 2005 and 2012. PearlDiver software
was used to query the database by the International Classification
of Diseases, 9th Revision, Clinical Modification (ICD-9-CM) procedural
code for primary THA and lumbar spinal fusion. Patients who had
undergone both lumbar fusion and THA were then divided into three
groups: 1 to 2 levels, 3 to 7 levels and 8+ levels of fusion. The
rate of dislocation in each group was established using ICD-9-CM codes.
Patients who underwent THA without spinal fusion were used as a
control group. Statistical significant difference between groups
was tested using the chi-squared test, and significance set at p
<
0.05.Aims
Patients and Methods
Advocates of debridement, antibiotics and implant retention (DAIR)
in hip periprosthetic joint infection (PJI) argue that a procedure
not disturbing a sound prosthesis-bone interface is likely to lead
to better survival and functional outcome compared with revision.
This case-control study aims were to compare outcome of DAIRs for
infected primary total hip arthroplasty (THA) with outcomes following
primary THA and two-stage revision of infected primary THAs. We retrospectively reviewed all DAIRs, performed for confirmed
infected primary hip arthropasty (n = 82) at out institution, between
1997 and 2013. Data recorded included full patient information and
type of surgery. Outcome measures included complications, mortality,
implant survivorship and functional outcome. Outcome was compared with
two control groups matched for gender and age; a cohort of primary
THAs (n = 120) and a cohort of two-stage revisions for infection
(n = 66).Aims
Patients and Methods
Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.Objectives
Methods
The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design. This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40 trunnions paired with either CoCr (n = 21) or ceramic (n = 27) heads. The taper junction of all hips was evaluated for fretting/corrosion damage and volumetric material loss using a roundness-measuring machine. We used linear regression analysis to investigate taper damage differences after adjusting for potential confounding variables.Objectives
Methods
Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix. Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.Objectives
Methods
The objective of this five-year prospective, blinded, randomised
controlled trial (RCT) was to compare femoral head penetration into
a vitamin E diffused highly cross-linked polyethylene (HXLPE) liner
with penetration into a medium cross-linked polyethylene control
liner using radiostereometric analysis. Patients scheduled for total hip arthroplasty (THA) were randomised
to receive either the study E1 (32 patients) or the control ArComXL
polyethylene (35 patients). The median age (range) of the overall
cohort was 66 years (40 to 76).Aims
Patients and Methods
Modular or custom-made femoral components have been preferred
for total hip arthroplasty (THA) in patients with a history of Perthes’
disease because of the distortion in the anatomy of the proximal
femur. However, it has not been established whether a monobloc cementless
stem will fit the distorted proximal femur or whether the results
of the procedure are satisfactory in this group of patients. We reviewed 68 consecutive patients who had undergone THA for
childhood Perthes’ disease between June 2003 and December 2008.
There were 35 men and 33 women with a mean age of 48 years (16 to
73) at the time of index arthroplasty. Their mean body mass index
was 24.4 (18.3 to 32.9). Of the 68 hips, 32 were classified as Stulberg
class III and 36 as class IV. The mean pre-operative shortening
of the affected leg was 17.2 mm (5 to 34). The minimum follow-up
was five years (mean 8.5 years; 5.2 to 10).Aims
Patients and Methods
This study aimed to determine the diagnostic performance of radiographic
criteria to detect aseptic acetabular loosening after revision total
hip arthroplasty (THA). Secondary aims were to determine the predictive
values of different thresholds of migration and to determine the
predictive values of radiolucency criteria. Acetabular component migration to re-revision was measured retrospectively
using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements
(Sutherland method) in two groups: Group A, 52 components (48 patients) found
not loose at re-revision and Group B, 42 components (36 patients)
found loose at re-revision between 1980 and 2015. The presence and
extent of radiolucent lines was also assessed.Aims
Patients and Methods
Recently, there has been considerable interest in quantifying
the associations between bony abnormalities around and in the hip
joint and osteoarthritis (OA). Our aim was to investigate the relationships
between acetabular undercoverage, acetabular overcoverage, and femoroacetabular
impingement (FAI) with OA of the hip, which currently remain controversial. A total of 545 cadaveric skeletons (1090 hips) from the Hamann-Todd
osteological collection were obtained. Femoral head volume (FHV),
acetabular volume (AV), the FHV/AV ratio, acetabular version, alpha
angle and anterior femoral neck offset (AFNO) were measured. A validated
grading system was used to quantify OA of the hip as minimal, moderate,
or severe. Multiple linear and multinomial logistic regression were
used to determine the factors that correlated independently with
the FHV, AV, and the FHV/AV ratio. Aims
Materials and Methods
The aim of this study was to investigate the outcomes of Vancouver
type B2 and B3 fractures by performing a systematic review of the
methods of surgical treatment which have been reported. A systematic search was performed in Ovid MEDLINE, Embase and
the Cochrane Central Register of Controlled Trials. For inclusion,
studies required a minimum of ten patients with a Vancouver type
B2 and/or ten patients with a Vancouver type B3 fracture, a minimum
mean follow-up of two years and outcomes which were matched to the type
of fracture. Studies were also required to report the rate of re-operation
as an outcome measure. The protocol was registered in the PROSPERO
database. Aims
Materials and Methods
The aim of this study was to examine the real time A total of 50 patients (83 hips) underwent 4D dynamic CT scanning
of the hip, producing real time osseous models of the pelvis and
femur being moved through flexion, adduction, and internal rotation.
The location and size of the cam deformity and its relationship
to the angle of flexion of the hip and pelvic tilt, and the position
of impingement were recorded.Aims
Patients and Methods
Our first aim was to determine whether there are significant
changes in the level of metal ions in the blood at mid-term follow-up,
in patients with an Articular Surface Replacement (ASR) arthroplasty.
Secondly, we sought to identify risk factors for any increases. The study involved 435 patients who underwent unilateral, metal-on-metal
(MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These
patients all had one measurement of the level of metal ions in the
blood before seven years had passed post-operatively (early evaluation)
and one after seven years had passed post-operatively (mid-term evaluation).
Changes in ion levels were tested using a Wilcoxon signed-rank test.
We identified subgroups at the highest risk of increase using a
multivariable linear logistic regression model.Aims
Patients and Methods
The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated tapered short stem femoral component and
53 patients with a standard conventional femoral component, at a
minimum follow-up of two years. The mean follow-up was 31.4 months
(24 to 39). Patients with poor proximal femoral bone quality were
excluded. The mean age of the patients was 66.6 years (59 to 77)
and the mean body mass index was 30.2 kg/m2 (24.1 to
41.0). Outcome was assessed using the Oxford Hip Score (OHS) and
the University College Hospital (UCH) hip score which is a validated
performance based instrument. HRQoL was assessed using the EuroQol
5D (EQ-5D).Aims
Patients and Methods
To evaluate the hypothesis that failed osteosynthesis of periprosthetic
Vancouver type B1 fractures can be treated successfully with stem
revision using a transfemoral approach and a cementless, modular,
tapered revision stem with reproducible rates of fracture healing,
stability of the revision stem, and clinically good results. A total of 14 patients (11 women, three men) with a mean age
of 72.4 years (65 to 90) undergoing revision hip arthroplasty after
failed osteosynthesis of periprosthetic fractures of Vancouver type
B1 were treated using a transfemoral approach to remove the well-fixed
stem before insertion of a modular, fluted titanium stem which obtained
distal fixation. These patients were clinically and radiologically
followed up for a mean 52.2 months (24 to 144).Aims
Patients and Methods
The Sliding Hip Screw (SHS) is commonly used to treat trochanteric hip fractures. Fixation failure is a devastating complication requiring complex revision surgery. One mode of fixation failure is lag screw cut-out which is greatest in unstable fracture patterns and when the tip-apex distance of the lag screw is > 25 mm. The X-Bolt Dynamic Hip Plating System (X-Bolt Orthopaedics, Dublin, Ireland) is a new device which aims to reduce this risk of cut-out. However, some surgeons have reported difficulty minimising the tip-apex distance with subsequent concerns that this may lead to an increased risk of cut-out. We measured the tip-apex distance from the intra-operative radiographs of 93 unstable trochanteric hip fractures enrolled in a randomised controlled trial (Warwick Hip Trauma Evaluation, WHiTE One trial). Participants were treated with either the sliding hip screw or the X-Bolt dynamic hip plating system. We also recorded the incidence of cut-out in both groups, at a median follow-up time of 17 months.Objectives
Patients and Methods
The World Health Organization (WHO) and the Centre
for Disease Control and Prevention (CDC) recently published guidelines
for the prevention of surgical site infection. The WHO guidelines,
if implemented worldwide, could have an immense impact on our practices
and those of the CDC have implications for healthcare policy in
the United States. Our aim was to review the strategies for prevention of periprosthetic
joint infection in light of these and other recent guidelines. Cite this article:
The purpose of this study was to evaluate whether an innovative
templating technique could predict the need for acetabular augmentation
during primary total hip arthroplasty for patients with dysplastic
hips. We developed a simple templating technique to estimate acetabular
component coverage at total hip arthroplasty, the True Cup: False
Cup (TC:FC) ratio. We reviewed all patients with dysplastic hips
who underwent primary total hip arthroplasty between 2005 and 2012.
Traditional radiological methods of assessing the degree of acetabular dysplasia
(Sharp’s angle, Tönnis angle, centre-edge angle) as well as the
TC:FC ratio were measured from the pre-operative radiographs. A
comparison of augmented and non-augmented hips was undertaken to
determine any difference in pre-operative radiological indices between
the two cohorts. The intra- and inter-observer reliability for all
radiological indices used in the study were also calculated.Aims
Patients and Methods
The aim of this study was to identify patient- and surgery-related
risk factors for sustaining an early periprosthetic fracture following
primary total hip arthroplasty (THA) performed using a double-tapered
cementless femoral component (Bi-Metric femoral stem; Biomet Inc.,
Warsaw, Indiana). A total of 1598 consecutive hips, in 1441 patients receiving
primary THA between January 2010 and June 2015, were retrospectively
identified. Level of pre-operative osteoarthritis, femoral Dorr
type and cortical index were recorded. Varus/valgus placement of
the stem and canal fill ratio were recorded post-operatively. Periprosthetic
fractures were identified and classified according to the Vancouver
classification. Regression analysis was performed to identify risk
factors for early periprosthetic fracture.Aims
Patients and Methods
The purpose of this study was to determine the sensitivity, specificity
and predictive values of previously reported thresholds of proximal
translation and sagittal rotation of cementless acetabular components
used for revision total hip arthroplasty (THA) at various times
during early follow-up. Migration of cementless acetabular components was measured retrospectively
in 84 patients (94 components) using Ein-Bild-Rontgen-Analyse (EBRA-Cup)
in two groups of patients. In Group A, components were recorded
as not being loose intra-operatively at re-revision THA (52 components/48
patients) and Group B components were recorded to be loose at re-revision
(42 components/36 patients).Aims
Patients and Methods
Compared with primary total hip arthroplasty (THA), revision
surgery can be challenging. The cement-in-cement femoral revision
technique involves removing a femoral component from a well-fixed
femoral cement mantle and cementing a new stem into the original
mantle. This technique is widely used and when carried out for the
correct indications, is fast, relatively inexpensive and carries
a reduced short-term risk for the patient compared with the alternative
of removing well-fixed cement. We report the outcomes of this procedure
when two commonly used femoral stems are used. We identified 1179 cement-in-cement stem revisions involving
an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty
Register (SHAR) between January 1999 and December 2015. Kaplan-Meier
survival analysis was performed.Aims
Patients and Methods
Post-operative migration of cemented acetabular components as
measured by radiostereometric analysis (RSA) has a strong predictive
power for late, aseptic loosening. Also, radiolucent lines predict
late loosening. Migration has been reduced by systemic bisphosphonate
treatment in randomised trials of hip and knee arthroplasty. Used
as a local treatment, a higher local dose of bisphosphonate can
be achieved without systemic exposure. We wished to see if this
principle could be applied usefully in total hip arthroplasty (THA). In this randomised placebo-controlled, double-blinded trial with
60 participants, we compressed gauze soaked in bisphosphonate solution
(ibandronate) or saline against the acetabular bone bed immediately
before cementing the acetabular component. RSA, classification of
radiolucent lines, the Harris Hip Score (HHS) and the Western Ontario McMasters
Universities Osteoarthritis Index (WOMAC) were carried out at three-,
six-, 12-, and 24-month follow-up.Aims
Patients and Methods
Joint-preserving surgery of the hip (JPSH) has evolved considerably
and now includes a number of procedures, including arthroscopy,
surgical dislocation, and redirectional osteotomies of the femur
and acetabulum. There are a number of different factors which lead
to failure of JPSH. Consequently, it is of interest to assess the
various modes of failure in order to continue to identify best practice
and the indications for these procedures. Using a retrospective observational study design, we reviewed
1013 patients who had undergone JPSH by a single surgeon between
2005 and 2015. There were 509 men and 504 women with a mean age
of 39 years (16 to 78). Of the 1013 operations, 783 were arthroscopies,
122 surgical dislocations, and 108 peri-acetabular osteotomies (PAO).
We analysed the overall failure rates and modes of failure. Re-operations
were categorised into four groups: Mode 1 was arthritis progression
or organ failure leading to total hip arthroplasty (THA); Mode 2
was an Incorrect diagnosis/procedure; Mode 3 resulted from malcorrection
of femur (type A), acetabulum (type B), or labrum (type C) and Mode 4
resulted from an unintended consequence of the initial surgical
intervention.Aims
Patients and Methods