Abstract
Aims
The purpose of this study was to determine the sensitivity, specificity and predictive values of previously reported thresholds of proximal translation and sagittal rotation of cementless acetabular components used for revision total hip arthroplasty (THA) at various times during early follow-up.
Patients and Methods
Migration of cementless acetabular components was measured retrospectively in 84 patients (94 components) using Ein-Bild-Rontgen-Analyse (EBRA-Cup) in two groups of patients. In Group A, components were recorded as not being loose intra-operatively at re-revision THA (52 components/48 patients) and Group B components were recorded to be loose at re-revision (42 components/36 patients).
Results
The mean proximal translation and sagittal rotation were significantly higher in Group B than in Group A from three months onwards (p < 0.02). Proximal translation > 1.0 mm within 24 months had a positive predictive value (PPV) of 90% and a specificity of 94%, but a sensitivity of 64%. Proximal translation > 1.0 mm within the first 24 months correctly identified 76 of 94 (81%) of components to be either loose or not loose. However, ten components in Group B (24%) did not migrate proximally above 1.0 mm within the first 60 months.
Conclusion
The high PPV of EBRA-Cup measurements of proximal translation (90%) shows that this can be used in early follow-up to identify patients at risk of aseptic loosening. The absence of proximal translation within the first 60 months indicates a component is not likely to be loose at re-revision THA although it does not exclude late aseptic loosening as a cause of failure.
Cite this article: Bone Joint J 2017;99-B:465–74.