The primary aim of this study was to identify independent predictors associated with nonunion and delayed union of tibial diaphyseal fractures treated with intramedullary nailing. The secondary aim was to assess the Radiological Union Scale for Tibial fractures (RUST) score as an early predictor of tibial fracture nonunion. A consecutive series of 647 patients who underwent intramedullary nailing for tibial diaphyseal fractures were identified from a trauma database. Demographic data, comorbidities, smoking status, alcohol consumption, use of non-steroidal anti-inflammatory drugs (NSAIDs), and steroid use were documented. Details regarding mechanism of injury, fracture classification, complications, and further surgery were recorded. Nonunion was defined as the requirement for revision surgery to achieve union. Delayed union was defined as a RUST score < 10 at six months postoperatively.Aims
Methods
Osteopetrosis (OP) is a rare hereditary disease that causes reduced bone resorption and increased bone density as a result of osteoclastic function defect. Our aim is to review the difficulties, mid-term follow-up results, and literature encountered during the treatment of OP. This is a retrospective and observational study containing data from nine patients with a mean age of 14.1 years (9 to 25; three female, six male) with OP who were treated in our hospital between April 2008 and October 2018 with 20 surgical procedures due to 17 different fractures. Patient data included age, sex, operating time, length of stay, genetic type of the disease, previous surgery, fractures, complications, and comorbidity.Aims
Methods
Our aim was to compare the one-year post-operative
outcomes following retention or removal of syndesmotic screws in
adult patients with a fracture of the ankle that was treated surgically.
A total of 51 patients (35 males, 16 females), with a mean age of
33.5 years (16 to 62), undergoing fibular osteosynthesis and syndesmotic
screw fixation, were randomly allocated to retention of the syndesmotic
screw or removal at three months post-operatively. The two groups
were comparable at baseline. One year post-operatively, there was no significant difference
in the mean Olerud–Molander ankle score (82.4 retention We conclude that removal of a syndesmotic screw produces no significant
functional, clinical or radiological benefit in adult patients who
are treated surgically for a fracture of the ankle. Cite this article:
Small animal models of fracture repair primarily investigate
indirect fracture healing via external callus formation. We present
the first described rat model of direct fracture healing. A rat tibial osteotomy was created and fixed with compression
plating similar to that used in patients. The procedure was evaluated
in 15 cadaver rats and then Objectives
Methods
To investigate the differences of open reduction and internal
fixation (ORIF) of complex AO Type C distal radius fractures between
two different models of a single implant type. A total of 136 patients who received either a 2.4 mm (n = 61)
or 3.5 mm (n = 75) distal radius locking compression plate (LCP
DR) using a volar approach were followed over two years. The main
outcome measurements included motion, grip strength, pain, and the
scores of Gartland and Werley, the Short-Form 36 (SF-36) and the
Disabilities of the Arm, Shoulder, and Hand (DASH). Differences
between the treatment groups were evaluated using regression analysis
and the likelihood ratio test with significance based on the Bonferroni
corrected p-value of <
0.003.Objectives
Methods
Filling the empty holes in peri-articular locking
plates may improve the fatigue strength of the fixation. The purpose of
this A locking/compression plate was applied to 33 synthetic femurs
and then a 6 cm metaphyseal defect was created (AO Type 33-A3).
The specimens were then divided into three groups: unplugged, plugged
with locking screw only and fully plugged holes. They were then
tested using a stepwise or run-out fatigue protocol, each applying
cyclic physiological multiaxial loads. All specimens in the stepwise group failed at the 770 N load
level. The mean number of cycles to failure for the stepwise specimen
was 25 500 cycles ( In conclusion, filling the empty combination locking/compression
holes in peri-articular distal femur locking plates at the level
of supracondylar comminution does not increase the fatigue life
of the fixation in a comminuted supracondylar femoral fracture model
(AO 33-A3) with a 6 cm gap.
We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm3) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations. In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.
The aim of this study was to review the number of patients operated on for traumatic disruption of the pubic symphysis who developed radiological signs of movement of the anterior pelvic metalwork during the first post-operative year, and to determine whether this had clinical implications. A consecutive series of 49 patients undergoing internal fixation of a traumatic diastasis of the pubic symphysis were studied. All underwent anterior fixation of the diastasis, which was frequently combined with posterior pelvic fixation. The fractures were divided into groups using the Young and Burgess classification for pelvic ring fractures. The different combinations of anterior and posterior fixation adopted to stabilise the fractures and the type of movement of the metalwork which was observed were analysed and related to functional outcome during the first post-operative year. In 15 patients the radiographs showed movement of the anterior metalwork, with broken or mobile screws or plates, and in six there were signs of a recurrent diastasis. In this group, four patients required revision surgery; three with anterior fixation and one with removal of anterior pelvic metalwork; the remaining 11 functioned as well as the rest of the study group. We conclude that radiological signs of movement in the anterior pelvic metalwork, albeit common, are not in themselves an indication for revision surgery.
The purpose of this study was to compare the clinical and radiological outcome of patients with intact, broken and removed syndesmosis screws after Weber B or C ankle fracture with an associated injury to the syndesmosis. We hypothesised that there would be no difference. Of a possible 142 patients who fulfilled our inclusion criteria, 52 returned for clinical and radiological assessment at least one year after surgery. Of these, 27 had intact syndesmosis screws, ten had broken screws, and 15 had undergone elective removal of the screw. The mean American Orthopaedic Foot and Ankle Society ankle/hindfoot score was 83.07 ( There was no difference in clinical outcome of patients with intact or removed syndesmotic screws. Paradoxically, patients with a broken syndesmosis screw had the best clinical outcome. Our data do not support the removal of intact or broken syndesmosis screws, and we caution against attributing post-operative ankle pain to breakage of the syndesmosis screw.
Secure fixation of displaced proximal fractures of the humerus is a challenging problem. A total of 32 patients with acutely displaced three- or four-part proximal fractures of the humerus were treated by open reduction and internal fixation using the proximal humeral internal locking system (PHILOS) plate. There were 23 women and nine men with a mean age of 59.9 years (18 to 87). Data were collected prospectively and the outcomes were assessed using the Constant score. The mean follow-up was for 11 months (3 to 24). In 31 patients (97%) the fracture united clinically and radiologically at a mean of 10 weeks (8 to 24). The mean Constant score at final review was 66.5 (30 to 92). There was no significant difference in outcome when comparing patients aged more than 60 years (18 patients) with those aged less than 60 years (14 patients) ( This plate provides an alternative method of fixation for fractures of the proximal humerus. It provides a stable fixation in young patients with good-quality bone sufficient to permit early mobilisation. Failure of the screws to maintain fixation in the elderly remains a problem.