Aims. Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. Patients and Methods. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared. Results. In 20 cases (15 patients) the Precice lengthening nail was used
and in 13 cases (seven patients) the LRS external fixator system.
The desired length was achieved in all cases in the Precice group
and in 12 of 13 cases in the LRS group. The mean healing index was
31.3 days/cm in the Precice and 47.1 days/cm in the LRS group (p
<
0.001). This was associated with an earlier ability to bear
full weight without aids in the Precice group. There were more complications
with LRS lengthening, including pin site infections and regenerate
deformity. Implant tolerance and the patients’ perception of the
cosmetic result were better with the Precice treatment. Conclusion. Femoral lengthening with the Precice femoral nail achieved excellent
functional results with fewer complications and greater patient
satisfaction when compared with the LRS system in our patients. Cite this article: Bone
Aims. This study describes the Osseointegration Group of Australia’s
Accelerated Protocol two-stage strategy (OGAAP-1) for the osseointegrated
reconstruction of amputated limbs. Patients and Methods. We report clinical outcomes in 50 unilateral trans-femoral amputees
with a mean age of 49.4 years (24 to 73), with a minimum one-year
follow-up. Outcome measures included the Questionnaire for persons
with a Trans-Femoral Amputation, the health assessment questionnaire
Short-Form-36 Health Survey, the Amputation Mobility Predictor scores
presented as K-levels, 6 Minute Walk Test and timed up and go tests.
Adverse events included soft-tissue problems, infection, fractures
and failure of the implant. Results. Our results demonstrated statistically significant improvements
in all five outcome measures. A total of 27 patients experienced
adverse events but at the conclusion of the study, all 50 were walking
on osseointegrated prostheses. Conclusion. These results demonstrate that osseointegrated prostheses are
a suitable alternative to socket-fit devices for amputees experiencing
socket-related discomfort and that our strategy offers more rapid
progress to walking than other similar protocols. Cite this article: Bone
Objectives. Induced membrane technique is a relatively new technique in the reconstruction of large bone defects. It involves the implantation of polymethylmethacrylate (PMMA) cement in the bone defects to induce the formation of membranes after radical debridement and reconstruction of bone defects using an autologous cancellous bone graft in a span of four to eight weeks. The purpose of this study was to explore the clinical outcomes of the induced membrane technique for the treatment of post-traumatic osteomyelitis in 32 patients. Methods. A total of 32 cases of post-traumatic osteomyelitis were admitted to our department between August 2011 and October 2012. This retrospective study included 22 men and ten women, with a mean age of 40 years (19 to 70). Within this group there were 20 tibias and 12 femurs with a mean defect of 5 cm (1.5 to 12.5). Antibiotic-loaded PMMA cement was inserted into the defects after radical debridement. After approximately eight weeks, the defects were implanted with bone graft. Results. The patients were followed for 27.5 months (24 to 32). Radiographic bone union occurred at six months for 26 cases (81%) and clinical healing occurred in 29 cases (90%) at ten months. A total of six cases had a second debridement before bone grafting because of recurrence of infection and one patient required a third debridement. No cases of osteomyelitis had recurred at the time of the last follow-up visit. Conclusion. The induced membrane technique for the treatment of post-traumatic osteomyelitis is a simple, reliable method, with good early results. However, there are many challenges in determining the scope of the debridement, type of limb fixation and source of bone graft to be used. Cite this article: Dr Z. Xie. Induced membrane technique for the treatment of bone defects due to post-traumatic osteomyelitis. Bone
Infected nonunion of a long bone continues to
present difficulties in management. In addition to treating the infection,
it is necessary to establish bony stability, encourage fracture
union and reconstruct the soft-tissue envelope. We present a series of 67 infected nonunions of a long bone in
66 patients treated in a multidisciplinary unit. The operative treatment
of patients suitable for limb salvage was performed as a single
procedure. Antibiotic regimes were determined by the results of
microbiological culture. At a mean follow-up of 52 months (22 to 97), 59 patients (88%)
had an infection-free united fracture in a functioning limb. Seven
others required amputation (three as primary treatment, three after
late failure of limb salvage and one for recalcitrant pain after
union). The initial operation achieved union in 54 (84%) of the salvaged
limbs at a mean of nine months (three to 26), with recurrence of
infection in 9%. Further surgery in those limbs that remained ununited
increased the union rate to 62 (97%) of the 64 limbs treated by
limb salvage at final follow-up. The use of internal fixation was
associated with a higher risk of recurrent infection than external
fixation. Cite this article: Bone
Deep vein thrombosis is a common complication
of immobilising the lower limb after surgery. We hypothesised that
intermittent pneumatic compression (IPC) therapy in outpatients
who had undergone surgical repair of acute ruptures of the Achilles
tendon could reduce the incidence of this problem. A total of 150 patients who had undergone surgical repair of
the Achilles tendon were randomised to either treatment with IPC
for six hours per day (n = 74) under an orthosis or treatment as
usual (n = 74) in a plaster cast without IPC. At two weeks post-operatively,
the incidence of deep vein thrombosis was assessed using blinded, double-reported
compression duplex ultrasound. At this point, IPC was discontinued
and all patients were immobilised in an orthosis for a further four
weeks. At six weeks post-operatively, a second compression duplex ultrasound
scan was performed. At two weeks, the incidence of deep vein thrombosis was 21% in
the treated group and 37% in the control group (p = 0.042). Age
over 39 years was found to be a strong risk factor for deep vein
thrombosis (odds ratio (OR) = 4.84, 95% confidence interval (CI)
2.14 to 10.96). Treatment with IPC, corrected for age differences
between groups, reduced the risk of deep vein thrombosis at the
two-week point (OR = 2.60; 95% CI 1.15 to 5.91; p =0.022). At six weeks,
the incidence of deep vein thrombosis was 52% in the treated group
and 48% in the control group (OR 0.94, 95% CI 0.49 to 1.83). IPC
appears to be an effective method of reducing the risk of deep vein
thrombosis in the early stages of post-operative immobilisation
of outpatients. Further research is necessary to elucidate whether
it can confer similar benefits over longer periods of immobilisation
and in a more heterogeneous group of patients. Cite this article: Bone
Although patients with a history of venous thromboembolism
(VTE) who undergo lower limb joint replacement are thought to be
at high risk of further VTE, the actual rate of recurrence has not
been reported. The purpose of this study was to identify the recurrence rate
of VTE in patients who had undergone lower limb joint replacement,
and to compare it with that of patients who had undergone a joint
replacement without a history of VTE. . From a pool of 6646 arthroplasty procedures (3344 TKR, 2907 THR,
243 revision THR, 152 revision TKR) in 5967 patients (68% female,
mean age 67.7; 21 to 96) carried out between 2009 and 2011, we retrospectively
identified 118 consecutive treatment episodes in 106 patients (65%
female, mean age 70; 51 to 88,) who had suffered a previous VTE.
Despite mechanical prophylaxis and anticoagulation with warfarin,
we had four recurrences by three months (3.4% of 118) and six by
one year (5.1% of 118). In comparison, in all our other
The treatment of infected nonunions is difficult.
Antibiotic cement-coated (ACC) rods provide stability as well as delivering
antibiotics. We conducted a review of 110 infected nonunions treated
with ACC rods. Patients were divided into two groups: group A (67
patients) with an infected arthrodesis, and group B (43 patients)
with an infected nonunion in a long bone. In group A, infected arthrodesis,
the success rate after the first procedure was 38/67 (57%), 29/67
(43%) required further surgery for either control of infection or
non-union. At last follow-up, five patients required amputation,
representing a limb salvage rate of 62/67 (93%) overall. In all,
29/67 (43%) presented with a bone defect with a mean size of 6.78
cm (2 to 25). Of those with a bone defect, 13/29 (45%) required
further surgery and had a mean size of defect of 7.2 cm (3.5 to
25). The cultures were negative in 17/67 (26%) and the most common
organism cultured was methicillin-resistant staphylococcus
aureus (MRSA) (23/67, (35%)). In group B, long bones nonunion,
the success rate after the first procedure was 26/43 (60%), 17/43
(40%) required further surgery for either control of infection or
nonunion. The limb salvage rate at last follow-up was 43/43 (100%).
A total of 22/43 (51%) had bone defect with a mean size of 4.7 cm
(1.5 to 11.5). Of those patients with a bone defect, 93% required further
surgery with a mean size of defect of 5.4 cm (3 to 8.5). The cultures
were negative in 10/43 (24%) and the most common organism cultured
was MRSA, 15/43 (35%). ACC rods are an effective form of treatment
for an infected nonunion, with an acceptable rate of complications. Cite this article: Bone
Intramedullary infection in long bones represents
a complex clinical challenge, with an increasing incidence due to the
increasing use of intramedullary fixation. We report a prospective
case series using an intramedullary reaming device, the Reamer–Irrigator–Aspirator
(RIA) system, in association with antibiotic cement rods for the
treatment of lower limb long bone infections. A total of 24 such
patients, 16 men and eight women, with a mean age of 44.5 years
(17 to 75), 14 with femoral and 10 with tibial infection, were treated
in a staged manner over a period of 2.5 years in a single referral
centre. Of these, 21 patients had had previous surgery, usually
for fixation of a fracture (seven had sustained an open fracture
originally and one had undergone fasciotomies). According to the
Cierny–Mader classification system, 18 patients were classified
as type 1A, four as 3A (discharging sinus tract), one as type 4A
and one as type 1B. Staphylococcus species were
isolated in 20 patients (83.3%). Local antibiotic delivery was used
in the form of impregnated cement rods in 23 patients. These were
removed at a mean of 2.6 months (1 to 5). Pathogen-specific antibiotics
were administered systemically for a mean of six weeks (3 to 18).
At a mean follow-up of 21 months (8 to 36), 23 patients (96%) had
no evidence of recurrent infection. One underwent a planned trans-tibial
amputation two weeks post-operatively due to peripheral vascular
disease and chronic recalcitrant osteomyelitis of the tibia and
foot. The combination of RIA reaming, the administration of systemic
pathogen-specific antibiotics and local delivery using impregnated
cement rods proved to be a safe and efficient form of treatment
in these patients. Cite this article: Bone
Although gradual bone transport may permit the
restoration of large-diameter bones, complications are common owing
to the long duration of external fixation. In order to reduce such
complications, a new technique of bone transport involving the use
of an external fixator and a locking plate was devised for segmental
tibial bone defects. A total of ten patients (nine men, one woman) with a mean age
at operation of 40.4 years (16 to 64) underwent distraction osteogenesis
with a locking plate to treat previously infected post-traumatic
segmental tibial defects. The locking plate was fixed percutaneously
to bridge proximal and distal segments, and was followed by external fixation.
After docking, percutaneous screws were fixed at the transported
segment through plate holes. At the same time, bone grafting was
performed at the docking site with the external fixator removed. The mean defect size was 5.9 cm (3.8 to 9.3) and mean external
fixation index was
13.4 days/cm (11.8 to 19.5). In all cases, primary union of the
docking site and distraction callus was achieved, with an excellent
bony result. There was no recurrence of deep infection or osteomyelitis,
and with the exception of one patient with a pre-existing peroneal
nerve injury, all achieved an excellent or good functional result. With short external fixation times and low complication rates,
bone transport with a locking plate could be recommended for patients
with segmental tibial defects. Cite this article: Bone
We carried out a systematic review of the literature
to evaluate the evidence regarding the clinical results of the Ilizarov
method in the treatment of long bone defects of the lower limbs. Only 37 reports (three non-randomised comparative studies, one
prospective study and 33 case-series) met our inclusion criteria.
Although several studies were unsatisfactory in terms of statistical
heterogeneity, our analysis appears to show that the Ilizarov method
of distraction osteogenesis significantly reduced the risk of deep
infection in infected osseous lesions (risk ratio 0.14 (95% confidence
interval (CI) 0.10 to 0.20), p <
0.001). However, there was a
rate of re-fracture of 5% (95% CI 3 to 7), with a rate of neurovascular
complications of 2.2% (95% CI 0.3 to 4) and an amputation rate of
2.9% (95% CI 1.4 to 4.4).The data was generally not statistically
heterogeneous. Where tibial defects were >
8 cm, the risk of re-fracture
increased (odds ratio 3.7 (95% CI 1.1 to 12.5), p = 0.036). . The technique is demanding for patients, illustrated by the voluntary
amputation rate of 1.6% (95% CI 0 to 3.1), which underlines the
need for careful patient selection. Cite this article: Bone
Objectives. To conduct a pilot randomised controlled trial to evaluate the
feasibility of conducting a larger trial to evaluate the difference
in Victorian Institute of Sports Assessment-Achilles (VISA-A) scores
at six months between patients with Achilles tendinopathy treated
with a platelet-rich plasma (PRP) injection compared with an eccentric
loading programme. Methods. Two groups of patients with mid-substance Achilles tendinopathy
were randomised to receive a PRP injection or an eccentric loading
programme. A total of 20 patients were randomised, with a mean age
of 49 years (35 to 66). All outcome measures were recorded at baseline,
six weeks, three months and six months. Results. The mean VISA-A score for the injection group at the primary
endpoint of six months was 76.0 (95% confidence interval (CI) 58.3
to 93.7) and for the exercise group was 57.4 (95% CI 38.1 to 76.7).
There was no statistically significant difference between these
scores (p = 0.171), which was expected from such a pilot study. Conclusions. This pilot study has been key to providing data to inform a larger
study and shows that the methodology is feasible. Cite this article: Bone
We hypothesised that adjuvant intermittent pneumatic
compression (IPC) beneath a plaster cast would reduce the risk of
deep-vein thrombosis (DVT) during post-operative immobilisation
of the lower limb. Of 87 patients with acute tendo Achillis (TA)
rupture, 26 were prospectively randomised post-operatively after
open TA repair. The treatment group (n = 14) received two weeks of
IPC of the foot for at least six hours daily under a plaster cast.
The control group (n = 12) had no additional treatment. At two weeks
post-operatively all patients received an orthosis until follow-up
at six weeks. At two and six weeks the incidence of DVT was assessed
using colour duplex sonography by two ultrasonographers blinded
to the treatment. Two patients withdrew from the study due to inability
to tolerate IPC treatment. An interim analysis demonstrated a high incidence of DVT in both
the IPC group (9 of 12, 75%) and the controls (6 of 12, 50%) (p
= 0.18). No significant differences in incidence were detected at
two (p = 0.33) or six weeks (p = 0.08) post-operatively. Malfunction
of the IPC leading to a second plaster cast was found to correlate
with an increased DVT risk at two weeks (φ = 0.71; p = 0.019), leading
to a premature abandonment of the study. We cannot recommend adjuvant treatment with foot IPC under a
plaster cast for outpatient DVT prevention during post-operative
immobilisation, owing to a high incidence of DVT related to malfunctioning
of this type of IPC application. Cite this article: Bone
We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m2 (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus knees. The alignment seen or measured intra-operatively during TKR is
not necessarily the same as the deformity seen on a standing long-leg
radiograph either pre- or post-operatively. Further investigation
into the effect of weight-bearing and surgical exposure of the joint
on the mechanical femorotibial angle is required to enable the most appropriate
intra-operative alignment to be selected.
Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along
the anatomical axis of the femur results in a lateral shift of the
mechanical axis by approximately 1 mm for each 1 cm of lengthening.
The use of autograft bone is the best option
when undertaking a procedure that requires bone graft because it
is osteogenic, osteoconductive and osseo-inductive. Pain, morbidity
and complications associated with harvesting iliac or non-iliac
sites occur in between 6% and 30% of cases. An alternative source
of graft with possibly a lower morbidity is the intramedullary canal.
In this study, 28 patients undergoing 30 arthrodesis procedures
on the hindfoot had a mean of 48 cm3 (43 to 50) of bone
harvested locally from the hindfoot or the tibial shaft by antegrade or
retrograde reaming. No patient sustained a fracture of the calcaneum,
talus or tibia. There was no morbidity except for one complication
when the reamer breached the medial tibial cortex. This healed uneventfully. This method of using the reamer–irrigator–aspirator system is
an extension of the standard technique of intramedullary reaming
of the lower limb: it produces good-quality bone graft with viable
growth factors consistent with that of the iliac crest, and donor
site morbidity is low. This is an efficient method of obtaining
autologous bone for use in arthrodesis of the ankle or hindfoot.
The purpose of this study was to investigate the effects of right leg restriction at the knee, ankle or both, on a driver’s braking times. Previous studies have not investigated the effects of knee restriction on braking performance. A total of 23 healthy drivers performed a series of emergency braking tests in a driving simulator in either an above-knee plaster cast, a below-knee cast, or in a knee brace with an increasing range of restriction. The study showed that total braking reaction time was significantly longer when wearing an above-knee plaster cast, a below-knee plaster cast or a knee brace fixed at 0°, compared with braking normally (p <
0.001). Increases in the time taken to move the foot from the accelerator to the brake accounted for some of the increase in the total braking reaction time. Unexpectedly, thinking time also increased with the level of restriction (p <
0.001). The increase in braking time with an above-knee plaster cast in this study would increase the stopping distance at 30 miles per hour by almost 3 m. These results suggest that all patients wearing any lower-limb plaster cast or knee brace are significantly impaired in their ability to perform an emergency stop. We suggest changes to the legislation to prevent patients from driving with lower-limb plaster casts or knee braces.
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).
We compared the complications and outcome of tibial lengthening using the Ilizarov method with and without the use of a supplementary intramedullary nail. In a retrospective case-matched series assembled from 176 patients with tibial lengthening, we matched 52 patients (26 pairs, group A with nail and group B without) according to the following criteria in order of importance: 1) difference in amount of lengthening (± 2 cm); 2) percentage difference in lengthening (± 5%); 3) difference in patient’s age (± seven years); 4) aetiology of the shortening, and 5) level of difficulty in obtaining the correction. The outcome was evaluated using the external fixator index, the healing index and an outcome score according to the criteria of Paley. It was found that some complications were specific to group A or B respectively, but others were common to both groups. The outcome was generally better in lengthenings with a nail, although there was a higher incidence of rectifiable equinus deformity in these patients.
We present the results of the surgical correction of lower-limb deformities caused by metabolic bone disease. Our series consisted of 17 patients with a diagnosis of hypophosphataemic rickets and two with renal osteodystrophy; their mean age was 25.6 years (14 to 57). In all, 43 lower-limb segments (27 femora and 16 tibiae) were osteotomised and the deformity corrected using a monolateral external fixator. The segment was then stabilised with locked intramedullary nailing. In addition, six femora in three patients were subsequently lengthened by distraction osteogenesis. The mean follow-up was 60 months (18 to 120). The frontal alignment parameters (the mechanical axis deviation, the lateral distal femoral angle and the medial proximal tibial angle) and the sagittal alignment parameters (the posterior distal femoral angle and the posterior proximal tibial angle) improved post-operatively. The external fixator was removed either at the end of surgery or at the end of the lengthening period, allowing for early mobilisation and weight-bearing. We encountered five problems and four obstacles in the programme of treatment. The use of intramedullary nails prevented recurrence of deformity and refracture.
The results of primary total knee replacement performed on a group of haemophiliac patients in a single institution by the same surgeon using the same surgical technique and prosthesis are reported. A total of 35 primary replacements in 30 patients were carried out between 1996 and 2005 and were reviewed retrospectively. The mean age of the patients was 31 years (24 to 42) and the mean follow-up was for 7.5 years (1 to 10). There were 25 patients with haemophilia A and five with haemophilia B. The HIV status and CD4 count were recorded, and Knee Society scores determined. Two patients had inhibitors to the deficient coagulation factor. There were no early wound infections and only one late deep infection which required a two-stage revision arthroplasty, with a good final result. The incidence of infection in HIV-positive and negative patients was thus similar. One knee in a patient with inhibitor had excessive bleeding due to a pseudoaneurysm which required embolisation. The results were excellent in 27 knees (77%), good in six (17%) and fair in two (6%). The survival rate at 7.5 years taking removal of the prosthesis for loosening or infection as the end-point was 97%. The mechanical survival of total knee replacements in haemophiliacs is very good. Our results confirm that this is a reproducible procedure in haemophilia, even in HIV-positive patients with a CD4 count >
200 mm3 and those with inhibitors. Our rate of infection was lower than previously reported. This could be due to better control of the HIV status with highly active anti-retroviral therapy and the use of antibiotic-loaded cement.