The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling. Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.Aims
Methods
Custom triflange acetabular components (CTACs) play an important role in reconstructive orthopaedic surgery, particularly in revision total hip arthroplasty (rTHA) and pelvic tumour resection procedures. Accurate CTAC positioning is essential to successful surgical outcomes. While prior studies have explored CTAC positioning in rTHA, research focusing on tumour cases and implant flange positioning precision remains limited. Additionally, the impact of intraoperative navigation on positioning accuracy warrants further investigation. This study assesses CTAC positioning accuracy in tumour resection and rTHA cases, focusing on the differences between preoperative planning and postoperative implant positions. A multicentre observational cohort study in Australia between February 2017 and March 2021 included consecutive patients undergoing acetabular reconstruction with CTACs in rTHA (Paprosky 3A/3B defects) or tumour resection (including Enneking P2 peri-acetabular area). Of 103 eligible patients (104 hips), 34 patients (35 hips) were analyzed.Aims
Methods
This study aimed to assess the impact of using the metal-augmented glenoid baseplate (AGB) on improving clinical and radiological outcomes, as well as reducing complications, in patients with superior glenoid wear undergoing reverse shoulder arthroplasty (RSA). From January 2016 to June 2021, out of 235 patients who underwent primary RSA, 24 received a superior-AGB after off-axis reaming (Group A). Subsequently, we conducted propensity score matching in a 1:3 ratio, considering sex, age, follow-up duration, and glenoid wear (superior-inclination and retroversion), and selected 72 well-balanced matched patients who received a standard glenoid baseplate (STB) after eccentric reaming (Group B). Superior-inclination, retroversion, and lateral humeral offset (LHO) were measured to assess preoperative glenoid wear and postoperative correction, as well as to identify any complications. Clinical outcomes were measured at each outpatient visit before and after surgery.Aims
Methods
Temporary hemiepiphysiodesis (HED) is applied to children and adolescents to correct angular deformities (ADs) in long bones through guided growth. Traditional Blount staples or two-hole plates are mainly used for this indication. Despite precise surgical techniques and attentive postoperative follow-up, implant-associated complications are frequently described. To address these pitfalls, a flexible staple was developed to combine the advantages of the established implants. This study provides the first results of guided growth using the new implant and compares these with the established two-hole plates and Blount staples. Between January 2013 and December 2016, 138 patients (22 children, 116 adolescents) with genu valgum or genu varum were treated with 285 flexible staples. The minimum follow-up was 24 months. These results were compared with 98 patients treated with 205 two-hole plates and 92 patients treated with 535 Blount staples. In long-standing anteroposterior radiographs, mechanical axis deviations (MADs) were measured before and during treatment to analyze treatment efficiency. The evaluation of the new flexible staple was performed according to the idea, development, evaluation, assessment, long-term (IDEAL) study framework (Stage 2a).Aims
Methods
The development of spinal deformity in children with underlying neurodisability can affect their ability to function and impact on their quality of life, as well as compromise provision of nursing care. Patients with neuromuscular spinal deformity are among the most challenging due to the number and complexity of medical comorbidities that increase the risk for severe intraoperative or postoperative complications. A multidisciplinary approach is mandatory at every stage to ensure that all nonoperative measures have been applied, and that the treatment goals have been clearly defined and agreed with the family. This will involve input from multiple specialities, including allied healthcare professionals, such as physiotherapists and wheelchair services. Surgery should be considered when there is significant impact on the patients’ quality of life, which is usually due to poor sitting balance, back or costo-pelvic pain, respiratory complications, or problems with self-care and feeding. Meticulous preoperative assessment is required, along with careful consideration of the nature of the deformity and the problems that it is causing. Surgery can achieve good curve correction and results in high levels of satisfaction from the patients and their caregivers. Modern modular posterior instrumentation systems allow an effective deformity correction. However, the risks of surgery remain high, and involvement of the family at all stages of decision-making is required in order to balance the risks and anticipated gains of the procedure, and to select those patients who can mostly benefit from spinal correction.
Debate continues regarding the optimum management of periprosthetic distal femoral fractures (PDFFs). This study aims to determine which operative treatment is associated with the lowest perioperative morbidity and mortality when treating low (Su type II and III) PDFFs comparing lateral locking plate fixation (LLP-ORIF) or distal femoral arthroplasty (DFA). This was a retrospective cohort study of 60 consecutive unilateral (PDFFs) of Su types II (40/60) and III (20/60) in patients aged ≥ 60 years: 33 underwent LLP-ORIF (mean age 81.3 years (SD 10.5), BMI 26.7 (SD 5.5); 29/33 female); and 27 underwent DFA (mean age 78.8 years (SD 8.3); BMI 26.7 (SD 6.6); 19/27 female). The primary outcome measure was reoperation. Secondary outcomes included perioperative complications, calculated blood loss, transfusion requirements, functional mobility status, length of acute hospital stay, discharge destination and mortality. Kaplan-Meier survival analysis was performed. Cox multivariate regression analysis was performed to identify risk factors for reoperation after LLP-ORIF.Aims
Methods
To devise a method to quantify and optimize tightness when inserting cortical screws, based on bone characterization and screw geometry. Cortical human cadaveric diaphyseal tibiae screw holes (n = 20) underwent destructive testing to firstly establish the relationship between cortical thickness and experimental stripping torque (Tstr), and secondly to calibrate an equation to predict Tstr. Using the equation’s predictions, 3.5 mm screws were inserted (n = 66) to targeted torques representing 40% to 100% of Tstr, with recording of compression generated during tightening. Once the target torque had been achieved, immediate pullout testing was performed.Aims
Methods
Arthroplasty for end-stage hallux rigidus (HR) is controversial. Arthrodesis remains the gold standard for surgical treatment, although is not without its complications, with rates of up to 10% for nonunion, 14% for reoperation and 10% for metatarsalgia. The aim of this study was to analyze the outcome of a double-stemmed silastic implant (Wright-Medical, Memphis, Tennessee, USA) for patients with end-stage HR. We conducted a retrospective review of 108 consecutive implants in 76 patients, between January 2005 and December 2016, with a minimum follow-up of two years. The mean age of the patients at the time of surgery was 61.6 years (42 to 84). There were 104 females and four males. Clinical, radiological, patient reported outcome measures (PROMS) data, a visual analogue score (VAS) for pain, and satisfaction scores were collected.Aims
Methods
Aims. Custom flange acetabular components (CFACs) are a patient-specific option for addressing large acetabular defects at revision total hip arthroplasty (THA), but patient and implant characteristics that affect survivorship remain unknown. This study aimed to identify patient and design factors related to survivorship. Patients and Methods. A retrospective review of 91 patients who underwent revision THA using 96 CFACs was undertaken, comparing features between radiologically failed and successful cases. Patient characteristics (demographic, clinical, and radiological) and implant features (design characteristics and intraoperative features) were collected. There were 74 women and 22 men; their mean age was 62 years (31 to 85). The mean follow-up was 24.9 months (. sd. 27.6; 0 to 116). Two sets of statistical analyses were performed: 1) univariate analyses (Pearson’s chi-squared and independent-samples Student’s t-tests) for each feature; and 2) bivariable logistic regressions using features identified from a random forest analysis. Results. Radiological failure and revision rates were 23% and 12.5%, respectively. Revisions were undertaken at a mean of 25.1 months (. sd. 26.4) postoperatively. Patients with radiological failure were younger at the time of the initial procedure, were less likely to have a diagnosis of primary osteoarthritis (OA), were more likely to have had ischial screws in previous surgery, had fewer ischial screw holes in their CFAC design, and had more proximal ischial fixation. Random forest analysis identified the age of the patient and the number of locking and
The aim of this study was to report the outcomes of different treatment options for glenoid loosening following reverse shoulder arthroplasty (RSA) at a minimum follow-up of two years. We retrospectively studied the records of 79 patients (19 men, 60 women; 84 shoulders) aged 70.4 years (21 to 87) treated for aseptic loosening of the glenosphere following RSA. Clinical evaluation included pre- and post-treatment active anterior elevation (AAE), external rotation, and Constant score.Aims
Patients and Methods
Intercalary allografts following resection of a primary diaphyseal
tumour have high rates of complications and failures. At our institution
intercalary allografts are augmented with intramedullary cement
and fixed using compression plating. Our aim was to evaluate their
long-term outcomes. A total of 46 patients underwent reconstruction with an intercalary
allograft between 1989 and 2014. The patients had a mean age of
32.8 years (14 to 77). The most common diagnoses were osteosarcoma
(n = 16) and chondrosarcoma (n = 9). The location of the tumours
was in the femur in 21, the tibia in 16 and the humerus in nine. Function
was assessed using the Musculoskeletal Tumor Society (MSTS) scoring
system and the Toronto Extremity Salvage Score (TESS). The survival
of the graft and the overall survival were assessed using the Kaplan-Meier method.Aims
Patients and Methods
We describe the technique and results of medial
submuscular plating of the femur in paediatric patients and discuss its
indications and limitations. Specifically, the technique is used
as part of a plate-after-lengthening strategy, where the period
of external fixation is reduced and the plate introduced by avoiding
direct contact with the lateral entry wounds of the external fixator
pins. The technique emphasises that vastus medialis is interposed
between the plate and the vascular structures. A total of 16 patients (11 male and five female, mean age 9.6
years (5 to 17)), had medial submuscular plating of the femur. All
underwent distraction osteogenesis of the femur with a mean lengthening
of 4.99 cm (3.2 to 12) prior to plating. All patients achieved consolidation
of the regenerate without deformity. The mean follow-up was 10.5 months
(7 to 15) after plating for those with plates still Placing the plate on the medial side is advantageous when the
external fixator is present on the lateral side, and is biomechanically
optimal in the presence of a femoral defect. We conclude that medial
femoral submuscular plating is a useful technique for specific indications
and can be performed safely with a prior understanding of the regional
anatomy. Cite this article:
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
The August 2013 Trauma Roundup360 looks at: reverse oblique fractures do better with a cephalomedullary device; locking screws confer no advantage in tibial plateau fractures; it’s all about the radius of curvature; radius of curvature revisited; radial head replacement in complex elbow reconstruction; stem cells in early fracture haematoma; heterotrophic ossification in forearms; and Boston in perspective.
Malunion is the most common complication of the
distal radius with many modalities of treatment available for such
a problem. The use of bone grafting after an osteotomy is still
recommended by most authors. We hypothesised that bone grafting
is not required; fixing the corrected construct with a volar locked
plate helps maintain the alignment, while metaphyseal defect fills
by itself. Prospectively, we performed the procedure on 30 malunited
dorsally-angulated radii using fixed angle volar locked plates without
bone grafting. At the final follow-up, 22 wrists were available.
Radiological evidence of union, correction of the deformity, clinical
and functional improvement was achieved in all cases. Without the
use of bone grafting, corrective open wedge osteotomy fixed by a
volar locked plate provides a high rate of union and satisfactory
functional outcomes.
Objectives. We aimed to further evaluate the biomechanical characteristics
of two locking screws versus three standard bicortical
screws in synthetic models of normal and osteoporotic bone. Methods. Synthetic tubular bone models representing normal bone density
and osteoporotic bone density were used. Artificial fracture gaps
of 1 cm were created in each specimen before fixation with one of
two constructs: 1) two locking screws using a five-hole locking
compression plate (LCP) plate; or 2) three
The ideal form of fixation for displaced, extra-articular
fractures of the distal tibia remains controversial. In the UK, open
reduction and internal fixation with locking-plates and intramedullary
nailing are the two most common forms of treatment. Both techniques
provide reliable fixation but both are associated with specific
complications. There is little information regarding the functional
recovery following either procedure. We performed a randomised pilot trial to determine the functional
outcome of 24 adult patients treated with either a locking-plate
(n = 12) or an intramedullary nailing (n = 12). At six months, there
was an adjusted difference of 13 points in the Disability Rating
Index in favour of the intramedullary nail. However, this was not
statistically significant in this pilot trial (p = 0.498). A total
of seven patients required further surgery in the locking-plate
group and one in the intramedullary nail group. This study suggests that there may be clinically relevant, functional
differences in patients treated with nail
Compression and absolute stability are important in the management of intra-articular fractures. We compared tension band wiring with plate fixation for the treatment of fractures of the olecranon by measuring compression within the fracture. Identical transverse fractures were created in models of the ulna. Tension band wires were applied to ten fractures and ten were fixed with Acumed plates. Compression was measured using a Tekscan force transducer within the fracture gap. Dynamic testing was carried out by reproducing cyclical contraction of the triceps of 20 N and of the brachialis of 10 N. Both methods were tested on each sample. Paired The mean compression for plating was 819 N ( During simulated movements, the mean compression was reduced in both groups, with tension band wiring at −14 N ( Pre-contoured plates provide significantly greater compression than tension bands in the treatment of transverse fractures of the olecranon, both over the whole fracture and specifically at the articular side of the fracture. In tension band wiring the overall compression was reduced and articular compression remained negligible during simulated contraction of the triceps, challenging the tension band principle.
Bone loss secondary to primary or metastatic lesions of the proximal humerus remains a challenging surgical problem. Options include preservation of the joint with stabilisation using internal fixation or resection of the tumour with prosthetic replacement. Resection of the proximal humerus often includes the greater tuberosity and adjacent diaphysis, which may result in poor function secondary to loss of the rotator cuff and/or deltoid function. Preservation of the joint with internal fixation may reduce the time in hospital and peri-operative morbidity compared with joint replacement, and result in a better functional outcome. We included 32 patients with pathological fractures of the proximal humerus in this study. Functional and radiological assessments were performed. At a mean follow-up of 17.6 months (8 to 61) there was no radiological evidence of failure of fixation. The mean revised musculoskeletal Tumour Society functional score was 94.6% (86% to 99%). There was recurrent tumour requiring further surgery in four patients (12.5%). Of the 22 patients who were employed prior to presentation all returned to work without restrictions. The use of a locking plate combined with augmentation with cement extends the indications for salvage of the proximal humerus with good function in patients with pathological and impending pathological fractures.