Advertisement for orthosearch.org.uk
Results 1 - 100 of 204
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 540 - 547
1 Jun 2024
Nandra RS Elnahal WA Mayne A Brash L McBryde CW Treacy RBC

Aims

The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years.

Methods

The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 555 - 564
1 Jun 2024
Leal J Holland CT Cochrane NH Seyler TM Jiranek WA Wellman SS Bolognesi MP Ryan SP

Aims

This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection.

Methods

A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 118 - 124
1 May 2024
Macheras GA Argyrou C Tzefronis D Milaras C Tsivelekas K Tsiamtsouris KG Kateros K Papadakis SA

Aims

Accurate diagnosis of chronic periprosthetic joint infection (PJI) presents a significant challenge for hip surgeons. Preoperative diagnosis is not always easy to establish, making the intraoperative decision-making process crucial in deciding between one- and two-stage revision total hip arthroplasty (THA). Calprotectin is a promising point-of-care novel biomarker that has displayed high accuracy in detecting PJI. We aimed to evaluate the utility of intraoperative calprotectin lateral flow immunoassay (LFI) in THA patients with suspected chronic PJI.

Methods

The study included 48 THAs in 48 patients with a clinical suspicion of PJI, but who did not meet European Bone and Joint Infection Society (EBJIS) PJI criteria preoperatively, out of 105 patients undergoing revision THA at our institution for possible PJI between November 2020 and December 2022. Intraoperatively, synovial fluid calprotectin was measured with LFI. Cases with calprotectin levels ≥ 50 mg/l were considered infected and treated with two-stage revision THA; in negative cases, one-stage revision was performed. At least five tissue cultures were obtained; the implants removed were sent for sonication.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 482 - 491
1 May 2024
Davies A Sabharwal S Liddle AD Zamora Talaya MB Rangan A Reilly P

Aims

Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey.

Methods

NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.


Bone & Joint Research
Vol. 13, Issue 4 | Pages 149 - 156
4 Apr 2024
Rajamäki A Lehtovirta L Niemeläinen M Reito A Parkkinen J Peräniemi S Vepsäläinen J Eskelinen A

Aims

Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium.

Methods

Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 17 - 23
1 Mar 2024
LaValva SM LeBrun DG Canoles HG Ren R Padgett DE Su EP

Aims

Professional dancers represent a unique patient population in the setting of hip arthroplasty, given the high degree of hip strength and mobility required by their profession. We sought to determine the clinical outcomes and ability to return to professional dance after total hip arthroplasty (THA) or hip resurfacing arthroplasty (HRA).

Methods

Active professional dancers who underwent primary THA or HRA at a single institution with minimum one-year follow-up were included in the study. Primary outcomes included the rate of return to professional dance, three patient-reported outcome measures (PROMs) (modified Harris Hip Score (mHHS), Hip disability and Osteoarthritis Outcome Score for Joint Replacement (HOOS-JR), and Lower Extremity Activity Scale (LEAS)), and postoperative complications.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 38 - 43
1 Mar 2024
Buckner BC Urban ND Cahoy KM Lyden ER Deans CF Garvin KL

Aims

Oxidized zirconium (OxZi) and highly cross-linked polyethylene (HXLPE) were developed to minimize wear and risk of osteolysis in total hip arthroplasty (THA). However, retrieval studies have shown that scratched femoral heads may lead to runaway wear, and few reports of long-term results have been published. The purpose of this investigation is to report minimum ten-year wear rates and clinical outcomes of THA with OxZi femoral heads on HXLPE, and to compare them with a retrospective control group of cobalt chrome (CoCr) or ceramic heads on HXLPE.

Methods

From 2003 to 2006, 108 THAs were performed on 96 patients using an OxZi head with a HXLPE liner with minimum ten-year follow-up. Harris Hip Scores (HHS) were collected preoperatively and at the most recent follow-up (mean 13.3 years). Linear and volumetric liner wear was measured on radiographs of 85 hips with a minimum ten-year follow-up (mean 14.5 years). This was compared to a retrospective control group of 45 THAs using ceramic or CoCr heads from October 1999 to February 2005, with a minimum of ten years’ follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 137 - 142
1 Mar 2024
van Veghel MHW van der Koelen RE Hannink G Schreurs BW Rijnen WHC

Aims

The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA).

Methods

We included all primary 394 THAs with a cemented short Exeter femoral component (≤ 125 mm) used in our tertiary referral centre between October 1993 and December 2021. A total of 83 patients (21%) were male. The median age of the patients at the time of surgery was 42 years (interquartile range (IQR) 30 to 55). The main indication for THA was a childhood hip disease (202; 51%). The median follow-up was 6.7 years (IQR 3.1 to 11.0). Kaplan-Meier survival analyses were performed to determine the rates of survival with femoral revision for any indication, for septic loosening, for fracture of the femoral component and for aseptic loosening as endpoints. The indications for revision were evaluated. Fractures of the femoral component were described in detail.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 97 - 103
1 Mar 2024
Baujard A Martinot P Demondion X Dartus J Faure PA Girard J Migaud H

Aims

Mechanical impingement of the iliopsoas (IP) tendon accounts for 2% to 6% of persistent postoperative pain after total hip arthroplasty (THA). The most common initiator is anterior acetabular component protrusion, where the anterior margin is not covered by anterior acetabular wall. A CT scan can be used to identify and measure this overhang; however, no threshold exists for determining symptomatic anterior IP impingement due to overhang. A case-control study was conducted in which CT scan measurements were used to define a threshold that differentiates patients with IP impingement from asymptomatic patients after THA.

Methods

We analyzed the CT scans of 622 patients (758 THAs) between May 2011 and May 2020. From this population, we identified 136 patients with symptoms suggestive of IP impingement. Among them, six were subsequently excluded: three because the diagnosis was refuted intraoperatively, and three because they had another obvious cause of impingement, leaving 130 hips (130 patients) in the study (impingement) group. They were matched to a control group of 138 asymptomatic hips (138 patients) after THA. The anterior acetabular component overhang was measured on an axial CT slice based on anatomical landmarks (orthogonal to the pelvic axis).


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 89 - 96
1 Mar 2024
Heckmann ND Chung BC Liu KC Chen XT Lovro LR Kistler NM White E Christ AB Longjohn DB Oakes DA Lieberman JR

Aims

Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations.

Methods

A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 51 - 58
1 Mar 2024
Jenkinson MRJ Meek DRM Tate R Brady A MacMillan S Grant H Currie S

Aims

Elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties are a suggested risk factor for developing cardiovascular complications including cardiomyopathy. Clinical studies assessing patients with MoM hips using left ventricular ejection fraction (LVEF) have found conflicting evidence of cobalt-induced cardiomyopathy. Global longitudinal strain (GLS) is an echocardiography measurement known to be more sensitive than LVEF when diagnosing early cardiomyopathies. The extent of cardiovascular injury, as measured by GLS, in patients with elevated blood cobalt levels has not previously been examined.

Methods

A total of 16 patients with documented blood cobalt ion levels above 13 µg/l (13 ppb, 221 nmol/l) were identified from a regional arthroplasty database. They were matched with eight patients awaiting hip arthroplasty. All patients underwent echocardiography, including GLS, investigating potential signs of cardiomyopathy.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 31 - 37
1 Mar 2024
Bunyoz KI Tsikandylakis G Mortensen K Gromov K Mohaddes M Malchau H Troelsen A

Aims. In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients. Methods. A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively. Results. At five years, the median levels of chromium, cobalt, and titanium were 0.5 μg/l (interquartile range (IQR) 0.50 to 0.62), 0.24 μg/l (IQR 0.18 to 0.30), and 1.16 μg/l (IQR 1.0 to 1.68) for the 32 mm group, and 0.5 μg/l (IQR 0.5 to 0.54), 0.23 μg/l (IQR 0.17 to 0.39), and 1.30 μg/l (IQR 1 to 2.05) for the 36 mm to 44 mm group, with no significant difference between the groups (p = 0.825, p = 1.000, p = 0.558). There were increased levels of metal ions at two years postoperatively in seven patients in the 32 mm group, compared with four in the 36 mm to 44 mm group, and at five years postoperatively in six patients in the 32 mm group, compared with seven in the 36 mm to 44 mm group. There was no significant difference in either the OHS (p = 0.665) or UCLA (p = 0.831) scores between patients with or without an increased level of metal ions. Conclusion. In patients who underwent MoP THA, we found no differences in the levels of metal ions five years postoperatively between those with a femoral head of 32 mm and those with a femoral head of between 36 mm and 44 mm, and no corrosion-related revisions. As taper corrosion can start after five years, there remains a need for longer-term studies investigating the relationship between the size of the femoral head size and corrosion in patients undergoing MoP THA. Cite this article: Bone Joint J 2024;106-B(3 Supple A):31–37


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 467 - 470
1 May 2023
McBryde CW Prakash R Haddad FS


Bone & Joint 360
Vol. 12, Issue 1 | Pages 26 - 29
1 Feb 2023

The February 2023 Wrist & Hand Roundup360 looks at: ‘Self-care’ protocol for minimally displaced distal radius fractures; Treatment strategies for acute Seymour fractures in children and adolescents: including crushed open fractures; Routinely collected outcomes of proximal row carpectomy; Moving minor hand surgeries in the office-based procedure room: a population-based trend analysis; A comparison between robotic-assisted scaphoid screw fixation and a freehand technique for acute scaphoid fracture: a randomized, controlled trial; Factors associated with conversion to surgical release after a steroid injection in patients with a trigger finger; Two modern total wrist arthroplasties: a randomized comparison; Triangular fibrocartilage complex suture repair reliable even in ulnar styloid nonunion.


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1132 - 1141
1 Oct 2022
Holm-Glad T Røkkum M Röhrl SM Roness S Godang K Reigstad O

Aims. To analyze the short-term outcome of two types of total wrist arthroplasty (TWA) in terms of wrist function, migration, and periprosthetic bone behaviour. Methods. A total of 40 patients suffering from non-rheumatoid wrist arthritis were enrolled in a randomized controlled trial comparing the ReMotion and Motec TWAs. Patient-rated and functional outcomes, radiological changes, blood metal ion levels, migration measured by model-based radiostereometric analysis (RSA), bone mineral density (BMD) measured by dual-energy X-ray absorptiometry (DXA), complications, loosening, and revision rates at two years were compared. Results. Patient-Rated Wrist and Hand Evaluation (PRWHE) scores, abbreviated version of the Disabilities of the Arm, Shoulder and Hand questionnaire (QuickDASH) scores, and pain improved similarly and significantly in both groups. Wrist motion improved significantly in the Motec group only, and forearm rotation in the ReMotion group only. Cobalt (Co) and chromium (Cr) blood ion levels were significantly higher in the metal-on-metal (MoM) Motec group than in the metal-on-polyethylene (MoP) ReMotion group. Mean total translation was 0.65 mm (95% confidence interval (CI) 0.26 to 1.12) and 0.27 mm (95% CI 0.14 to 0.47) for the ReMotion carpal and radial components, and 0.32 mm (95% CI 0.22 to 0.45) and 0.26 mm (95% CI 0.20 to 0.34) for the Motec metacarpal and radial components, respectively. Apart from dorsal and volar tilts, which were significantly higher for the radial ReMotion than for the Motec component, no significant differences in absolute migration occurred. BMD around the radial components never returned to baseline. Almost one-third of patients required reoperation due to complications. Two ReMotion implants were revised to Motec TWAs due to carpal component loosening, and three Motec MoM articulations were revised to metal-on-polyether ether ketone due to painful synovitis. Conclusion. Both implants provided matched function and were stable at short-term follow-up, but with a high complication rate. This procedure should be restricted to specialist centres undertaking prospective analysis until its role is clarified. Cite this article: Bone Joint J 2022;104-B(10):1132–1141


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 852 - 858
1 Jul 2022
Grothe T Günther K Hartmann A Blum S Haselhoff R Goronzy J

Aims

Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors.

Methods

Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.


Bone & Joint 360
Vol. 11, Issue 2 | Pages 52 - 54
1 Apr 2022
Evans JT Evans JP Whitehouse MR


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 376 - 385
1 Mar 2022
Gramlich Y Hofmann L Kress S Ruckes C Kemmerer M Klug A Hoffmann R Kremer M

Aims

This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant.

Methods

A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 359 - 367
1 Mar 2022
Deere K Matharu GS Ben-Shlomo Y Wilkinson JM Blom AW Sayers A Whitehouse MR

Aims

A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database.

Methods

Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 341 - 351
1 Mar 2022
Fowler TJ Aquilina AL Reed MR Blom AW Sayers A Whitehouse MR

Aims

Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA.

Methods

We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 200 - 205
1 Feb 2022
Orita K Goto K Kuroda Y Kawai T Okuzu Y Matsuda S

Aims

The aim of this study was to evaluate the performance of first-generation annealed highly cross-linked polyethylene (HXLPE) in cementless total hip arthroplasty (THA).

Methods

We retrospectively evaluated 29 patients (35 hips) who underwent THA between December 2000 and February 2002. The survival rate was estimated using the Kaplan-Meier method. Hip joint function was evaluated using the Japanese Orthopaedic Association (JOA) score. Two-dimensional polyethylene wear was estimated using Martell’s Hip Analysis Suite. We calculated the wear rates between years 1 and 5, 5 and 10, 10 and 15, and 15 and final follow-up.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 206 - 211
1 Feb 2022
Bloch BV White JJE Matar HE Berber R Manktelow ARJ

Aims

Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost.

Methods

In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.


Aims

To establish the survivorship, function, and metal ion levels in an unselected series of metal-on-metal hip resurfacing arthroplasties (HRAs) performed by a non-designer surgeon.

Methods

We reviewed 105 consecutive HRAs in 83 patients, performed by a single surgeon, at a mean follow-up of 14.9 years (9.3 to 19.1). The cohort included 45 male and 38 female patients, with a mean age of 49.5 years (SD 12.5)


Bone & Joint Open
Vol. 2, Issue 12 | Pages 1067 - 1074
1 Dec 2021
El-Bakoury A Khedr W Williams M Eid Y Hammad AS

Aims

After failed acetabular fractures, total hip arthroplasty (THA) is a challenging procedure and considered the gold standard treatment. The complexity of the procedure depends on the fracture pattern and the initial fracture management. This study’s primary aim was to evaluate patient-reported outcome measures (PROMs) for patients who underwent delayed uncemented acetabular THA after acetabular fractures. The secondary aims were to assess the radiological outcome and the incidence of the associated complications in those patients.

Methods

A total of 40 patients underwent cementless acetabular THA following failed treatment of acetabular fractures. The postoperative clinical and radiological outcomes were evaluated for all the cohort.


Bone & Joint Open
Vol. 2, Issue 11 | Pages 1004 - 1016
26 Nov 2021
Wight CM Whyne CM Bogoch ER Zdero R Chapman RM van Citters DW Walsh WR Schemitsch E

Aims

This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads.

Methods

In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1633 - 1640
1 Oct 2021
Lex JR Evans S Parry MC Jeys L Stevenson JD

Aims

Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs.

Methods

This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims

Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues.

Methods

A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims

Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA.

Methods

This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m2, and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported.


Bone & Joint Research
Vol. 10, Issue 6 | Pages 340 - 347
1 Jun 2021
Jenkinson MRJ Meek RMD Tate R MacMillan S Grant MH Currie S

Elevated levels of circulating cobalt ions have been linked with a wide range of systemic complications including neurological, endocrine, and cardiovascular symptoms. Case reports of patients with elevated blood cobalt ions have described significant cardiovascular complications including cardiomyopathy. However, correlation between the actual level of circulating cobalt and extent of cardiovascular injury has not previously been performed. This review examines evidence from the literature for a link between elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties and cardiomyopathy. Correlation between low, moderate, and high blood cobalt with cardiovascular complications has been considered. Elevated blood cobalt at levels over 250 µg/l have been shown to be a risk factor for developing systemic complications and published case reports document cardiomyopathy, cardiac transplantation, and death in patients with severely elevated blood cobalt ions. However, it is not clear that there is a hard cut-off value and cardiac dysfunction may occur at lower levels. Clinical and laboratory research has found conflicting evidence of cobalt-induced cardiomyopathy in patients with MoM hips. Further work needs to be done to clarify the link between severely elevated blood cobalt ions and cardiomyopathy.

Cite this article: Bone Joint Res 2021;10(6):340–347.


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1009 - 1020
1 Jun 2021
Ng N Gaston P Simpson PM Macpherson GJ Patton JT Clement ND

Aims

The aims of this systematic review were to assess the learning curve of semi-active robotic arm-assisted total hip arthroplasty (rTHA), and to compare the accuracy, patient-reported functional outcomes, complications, and survivorship between rTHA and manual total hip arthroplasty (mTHA).

Methods

Searches of PubMed, Medline, and Google Scholar were performed in April 2020 in line with the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “hip”, and “arthroplasty”. The criteria for inclusion were published clinical research articles reporting the learning curve for rTHA (robotic arm-assisted only) and those comparing the implantation accuracy, functional outcomes, survivorship, or complications with mTHA.


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1070 - 1077
1 Jun 2021
Hipfl C Mooij W Perka C Hardt S Wassilew GI

Aims

The purpose of this study was to evaluate unexpected positive cultures in total hip arthroplasty (THA) revisions for presumed aseptic loosening, to assess the prevalence of low-grade infection using two definition criteria, and to analyze its impact on implant survival after revision.

Methods

A total of 274 THA revisions performed for presumed aseptic loosening from 2012 to 2016 were reviewed. In addition to obtaining intraoperative tissue cultures from all patients, synovial and sonication fluid samples of the removed implant were obtained in 215 cases (79%) and 101 cases (37%), respectively. Histopathological analysis was performed in 250 cases (91%). Patients were classified as having low-grade infections according to institutional criteria and Musculoskeletal Infection Society (MSIS) International Consensus Meeting (ICM) 2013 criteria. Low-grade infections according to institutional criteria were treated with targeted antibiotics for six weeks postoperatively. Implant failure was defined as the need for re-revision resulting from periprosthetic joint infection (PJI) and aseptic reasons. The mean follow-up was 68 months (26 to 95).


Bone & Joint 360
Vol. 9, Issue 6 | Pages 5 - 11
1 Dec 2020
Sharma V Turmezei T Wain J McNamara I


Bone & Joint Open
Vol. 1, Issue 12 | Pages 743 - 748
1 Dec 2020
Mahon J McCarthy CJ Sheridan GA Cashman JP O'Byrne JM Kenny P

Aims

The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre.

Methods

All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery.


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1505 - 1510
2 Nov 2020
Klemt C Limmahakhun S Bounajem G Xiong L Yeo I Kwon Y

Aims

The complex relationship between acetabular component position and spinopelvic mobility in patients following total hip arthroplasty (THA) renders it difficult to optimize acetabular component positioning. Mobility of the normal lumbar spine during postural changes results in alterations in pelvic tilt (PT) to maintain the sagittal balance in each posture and, as a consequence, markedly changes the functional component anteversion (FCA). This study aimed to investigate the in vivo association of lumbar degenerative disc disease (DDD) with the PT angle and with FCA during postural changes in THA patients.

Methods

A total of 50 patients with unilateral THA underwent CT imaging for radiological evaluation of presence and severity of lumbar DDD. In all, 18 patients with lumbar DDD were compared to 32 patients without lumbar DDD. In vivo PT and FCA, and the magnitudes of changes (ΔPT; ΔFCA) during supine, standing, swing-phase, and stance-phase positions were measured using a validated dual fluoroscopic imaging system.


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1491 - 1496
1 Nov 2020
Buddhdev PK Vanhegan IS Khan T Hashemi-Nejad A

Aims

Despite advances in the treatment of paediatric hip disease, adolescent and young adult patients can develop early onset end-stage osteoarthritis. The aims of this study were to address the indications and medium-term outcomes for total hip arthroplasty (THA) with ceramic bearings for teenage patients.

Methods

Surgery was performed by a single surgeon working in the paediatric orthopaedic unit of a tertiary referral hospital. Databases were interrogated from 2003 to 2017 for all teenage patients undergoing THA with a minimum 2.3 year follow-up. Data capture included patient demographics, the underlying hip pathology, number of previous surgeries, and THA prostheses used. Institutional ethical approval was granted to contact patients for prospective clinical outcomes and obtain up-to-date radiographs. In total, 60 primary hips were implanted in 51 patients (35 female, 16 male) with nine bilateral cases. The mean age was 16.7 years (12 to 19) and mean follow-up was 9.3 years (2.3 to 16.8).


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1303 - 1310
3 Oct 2020
Kjærgaard K Ding M Jensen C Bragdon C Malchau H Andreasen CM Ovesen O Hofbauer C Overgaard S

Aims

The most frequent indication for revision surgery in total hip arthroplasty (THA) is aseptic loosening. Aseptic loosening is associated with polyethylene liner wear, and wear may be reduced by using vitamin E-doped liners. The primary objective of this study was to compare proximal femoral head penetration into the liner between a) two cross-linked polyethylene (XLPE) liners (vitamin E-doped (vE-PE)) versus standard XLPE liners, and b) two modular femoral head diameters (32 mm and 36 mm).

Methods

Patients scheduled for a THA were randomized to receive a vE-PE or XLPE liner with a 32 mm or 36 mm metal head (four intervention groups in a 2 × 2 factorial design). Head penetration and acetabular component migration were measured using radiostereometric analysis at baseline, three, 12, 24, and 60 months postoperatively. The Harris Hip Score, University of California, Los Angeles (UCLA) Activity Score, EuroQol five-dimension questionnaire (EQ-5D), and 36-Item Short-Form Health Survey questionnaire (SF-36) were assessed at baseline, three, 12, 36, and 60 months.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1003 - 1009
1 Aug 2020
Mononen H Sund R Halme J Kröger H Sirola J

Aims

There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision.

Methods

Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims. We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Methods. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time. Results. Metal ion levels were repeated in 33 patients. When comparing the results of serum metal ion levels over time, regardless of head size, there was a significant increase in both cobalt and chromium levels (p < 0.001). Two patients with larger head sizes had undergone revision arthroplasty with evidence of trunnion damage at surgery. Two patients within the 40 mm subgroup had metal ion levels above the MHRA (Medicines and Healthcare Products Regulatory Agency) threshold for detailed investigation. The increase in cobalt and chromium, when comparing the 36 mm and 40 mm groups with those of the 28 mm group, was not significant (36 mm vs 28 mm; p = 0.092/p = 0.191; 40 mm vs 28 mm; p = 0.200/p = 0.091, respectively). There was no difference, between femoral head sizes, when comparing outcome as measured by the Oxford Hip Score. Conclusion. This study shows an increase in cobalt and chromium levels over time for all modular femoral head sizes in patients with metal-on-polyethylene bearings, with two patients demonstrating ion levels above the MHRA threshold for failure, and a further two patients requiring revision surgery. These results may have clinical implications regarding longer term follow-up of patients and future implant choice, particularly among younger patients. Cite this article: Bone Joint J 2020;102-B(7):832–837


Bone & Joint 360
Vol. 9, Issue 1 | Pages 4 - 9
1 Feb 2020
Logishetty K Muirhead-Allwood SK Cobb JP


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 90 - 101
1 Jan 2020
Davis ET Pagkalos J Kopjar B

Aims

The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA).

Methods

Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad).


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1199 - 1208
1 Oct 2019
Lamb JN Matharu GS Redmond A Judge A West RM Pandit HG

Aims

We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls.

Patients and Methods

This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression.


Bone & Joint Research
Vol. 8, Issue 10 | Pages 443 - 450
1 Oct 2019
Treacy RBC Holland JP Daniel J Ziaee H McMinn DJW

Objectives

Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated.

Methods

This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (sd 11; 21 to 82), and 73% of implantations were in female patients. All patients were individually counselled, and a detailed informed consent was obtained prior to operation. Primary resurfacing was carried out in 85 hips, and three cases involved revision of previous MoM HRA. Clinical, radiological, and Oxford Hip Score (OHS) assessments were studied, along with implant survival.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 57 - 61
1 Jun 2019
Chalmers BP Mangold DG Hanssen AD Pagnano MW Trousdale RT Abdel MP

Aims

Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR.

Patients and Methods

Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 540 - 546
1 May 2019
Juneau D Grammatopoulos G Alzahrani A Thornhill R Inacio JR Dick A Vogel KI Dobransky J Beaulé PE Dwivedi G

Aims

Cardiac magnetic resonance (CMR) was used to assess whether cardiac function or tissue composition was affected in patients with well-functioning metal-on-metal hip resurfacing arthroplasties (MoMHRA) when compared with a group of controls, and to assess if metal ion levels correlated with any of the functional or structural parameters studied.

Patients and Methods

In all, 30 participants with no significant cardiac history were enrolled: 20 patients with well-functioning MoMHRA at mean follow-up of 8.3 years post-procedure (ten unilateral, ten bilateral; 17 men, three women) and a case-matched control group of ten non-MoM total hip arthroplasty patients (six men, four women). The mean age of the whole cohort (study group and controls) at the time of surgery was 50.6 years (41.0 to 64.0). Serum levels of cobalt and chromium were measured, and all patients underwent CMR imaging, including cine, T2* measurements, T1 and T2 mapping, late gadolinium enhancement, and strain measurements.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 378 - 385
1 Apr 2019
García-Rey E Carbonell-Escobar R Cordero-Ampuero J García-Cimbrelo E

Aims

We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up.

Patients and Methods

Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 386 - 389
1 Apr 2019
Hampton C Weitzler L Baral E Wright TM Bostrom MPG

Aims

The aim of this study was to evaluate fretting and corrosion in retrieved oxidized zirconium (OxZr; OXINIUM, Smith & Nephew, Memphis, Tennessee) femoral heads and compare the results with those from a matched cohort of cobalt-chromium (CoCr) femoral heads.

Patients and Methods

A total of 28 OxZr femoral heads were retrieved during revision total hip arthroplasty (THA) and matched to 28 retrieved CoCr heads according to patient demographics. The mean age at index was 56 years (46 to 83) in the OxZr group and 70 years (46 to 92) in the CoCr group. Fretting and corrosion scores of the female taper of the heads were measured according to the modified Goldberg scoring method.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 365 - 371
1 Apr 2019
Nam D Salih R Nahhas CR Barrack RL Nunley RM

Aims

Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design.

Patients and Methods

This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m2 and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (sd 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis.


Bone & Joint Research
Vol. 8, Issue 3 | Pages 146 - 155
1 Mar 2019
Langton DJ Natu S Harrington CF Bowsher JG Nargol AVF

Objectives

We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr.

Methods

For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood.


Bone & Joint Research
Vol. 8, Issue 2 | Pages 65 - 72
1 Feb 2019
Cowie RM Aiken SS Cooper JJ Jennings LM

Objectives

Bone void fillers are increasingly being used for dead space management in arthroplasty revision surgery. The aim of this study was to investigate the influence of calcium sulphate bone void filler (CS-BVF) on the damage and wear of total knee arthroplasty using experimental wear simulation.

Methods

A total of 18 fixed-bearing U2 total knee arthroplasty system implants (United Orthopedic Corp., Hsinchu, Taiwan) were used. Implants challenged with CS-BVF were compared with new implants (negative controls) and those intentionally scratched with a diamond stylus (positive controls) representative of severe surface damage (n = 6 for each experimental group). Three million cycles (MC) of experimental simulation were carried out to simulate a walking gait cycle. Wear of the ultra-high-molecular-weight polyethylene (UHMWPE) tibial inserts was measured gravimetrically, and damage to articulating surfaces was assessed using profilometry.


Bone & Joint 360
Vol. 7, Issue 6 | Pages 39 - 40
1 Dec 2018


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1592 - 1599
1 Dec 2018
Galea VP Connelly JW Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen MB Muratoglu O Bragdon C Malchau H

Aims

The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure.

Patients and Methods

A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery.


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1434 - 1441
1 Nov 2018
Blakeney WG Beaulieu Y Puliero B Lavigne M Roy A Massé V Vendittoli P

Aims

This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing.

Patients and Methods

Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1310 - 1319
1 Oct 2018
Langton DJ Wells SR Joyce TJ Bowsher JG Deehan D Green S Nargol AVF Holland JP

Aims. There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. Patients and Methods. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling. Results. A total of 95 Exeter MoP and 249 MoM THAs were examined. The median volumetric loss from the MoM cohort was over four times larger than that from the MoP cohort (1.01 mm. 3. vs 0.23 mm. 3. , p < 0.001), despite a significantly shorter median period in vivo for the MoM group (48 months vs 90 months, p < 0.001). Multiple regression modelling indicated that the dominant variables leading to greater female taper material loss were bearing diameter (p < 0.001), larger female taper angles (p < 0.001), and male titanium stem tapers (p < 0.001). Conclusion. Consistent with the long-term clinical success of the device, the volumetric material loss from Exeter femoral head tapers was, in general, small compared with that from larger-diameter MoM head tapers. Cite this article: Bone Joint J 2018;100-B:1310–9


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1320 - 1329
1 Oct 2018
Metcalfe D Peterson N Wilkinson JM Perry DC

Aims. The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. Patients and Methods. A descriptive observational study was undertaken using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man between April 2003 and March 2017. All patients aged ≤ 20 years at the time of THA were included and the primary outcome was revision surgery. Descriptive statistics were used to summarize the data and Kaplan–Meier estimates calculated for the cumulative implant survival. Results. A total of 769 THAs were performed in 703 patients. The median follow-up was 5.1 years (interquartile range (IQR) 2.6 to 7.8). Eight patients died and 35 THAs were revised. The use of metal-on-metal (MoM) bearings and resurfacing procedures declined after 2008. The most frequently recorded indications for revision were loosening (20%) and infection (20%), although the absolute risk of these events occurring was low (0.9%). Factors associated with lower implant survival were MoM and metal-on-polyethylene (MoP) bearings and resurfacing arthroplasty (vs ceramic-on-polyethylene (CoP) and ceramic-on-ceramic (CoC) bearings, p = 0.002), and operations performed by surgeons who undertook few THAs in this age group as recorded in the NJR (vs those with five or more recorded operations, p = 0.030). Kaplan–Meier estimates showed 96% (95% confidence interval (CI) 94% to 98%) survivorship of implants at five years. Conclusion. Within the NJR, the overall survival for very young patients undergoing THA exceeded 96% during the first five postoperative years. In the absence of studies that can better account for differences in the characteristics of the patients, surgeons should consider the association between early revision and the type of implant, the number of THAs performed in these patients, and the bearing surface when performing THA in very young patients. Cite this article: Bone Joint J 2018;100-B:1320–9


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 991 - 1001
1 Aug 2018
Findlay C Ayis S Demetriades AK

Aims

The aim of this study was to determine how the short- and medium- to long-term outcome measures after total disc replacement (TDR) compare with those of anterior cervical discectomy and fusion (ACDF), using a systematic review and meta-analysis.

Patients and Methods

Databases including Medline, Embase, and Scopus were searched. Inclusion criteria involved prospective randomized control trials (RCTs) reporting the surgical treatment of patients with symptomatic degenerative cervical disc disease. Two independent investigators extracted the data. The strength of evidence was assessed using the Grading of Recommendations, Assessment, Development and Evaluation (GRADE) criteria. The primary outcome measures were overall and neurological success, and these were included in the meta-analysis. Standardized patient-reported outcomes, including the incidence of further surgery and adjacent segment disease, were summarized and discussed.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 898 - 902
1 Jul 2018
Lachiewicz PF O’Dell JA

Aims. To report our experience with trunnion corrosion following metal-on-polyethylene total hip arthroplasty, in particular to report the spectrum of presentation and determine the mean time to presentation. Patients and Methods. We report the presenting symptoms and signs, intraoperative findings, and early results and complications of operative treatment in nine patients with a mean age of 74 years (60 to 86). The onset of symptoms was at a mean of seven years (3 to 18) after index surgery. Results. Patients presented with a variety of symptoms including pain, limp and rash. The preoperative mean serum cobalt level was 7.1 ppb (2.2 to 12.8) and mean serum chromium level was 2.2 ppb (0.5 to 5.2). Metal artifact reduction sequence (MARS) MRI showed fluid collection and possible pseudotumour formation in five hips, fluid collection in two hips, and synovitis/debris in one hip, with no MRI in one patient. Acetabular revision was performed in three patients, six patients underwent liner and head exchange only. The postoperative metal levels decreased in all patients: mean cobalt 0.5 ppb (0 to 1.8) and mean chromium 0.9 ppb (0 to 2.6) at a mean of five months (3 to 8) postoperatively. Seven patients had good pain relief and no complications at one year. There were two major complications requiring reoperation: acute infection at six weeks, for which the patient required two-stage reimplantation; and recurrent dislocation, for which the patient was revised to a dual mobility component. Conclusion. Trunnion corrosion in metal-on-polyethylene THA has a range of presenting symptoms, and may present later than previously described. A high index of suspicion is warranted, and serum cobalt and chromium levels are recommended for diagnosis. Patients should be counselled about possible postoperative complications. Cite this article: Bone Joint J 2018;100-B:898–902


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 867 - 874
1 Jul 2018
Makarewich CA Anderson MB Gililland JM Pelt CE Peters CL

Aims

For this retrospective cohort study, patients aged ≤ 30 years (very young) who underwent total hip arthroplasty (THA) were compared with patients aged ≥ 60 years (elderly) to evaluate the rate of revision arthroplasty, implant survival, the indications for revision, the complications, and the patient-reported outcomes.

Patients and Methods

We retrospectively reviewed all patients who underwent primary THA between January 2000 and May 2015 from our institutional database. A total of 145 very young and 1359 elderly patients were reviewed. The mean follow-up was 5.3 years (1 to 18). Logistic generalized estimating equations were used to compare characteristics and the revision rate. Survival was evaluated using Kaplan–Meier curves and hazard rates were created using Cox regression.


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 720 - 724
1 Jun 2018
Waterson HB Whitehouse MR Greidanus NV Garbuz DS Masri BA Duncan CP

Aims. Fretting and corrosion at the modular head/neck junction, known as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse reaction to metal debris (ARMD). We describe the outcome of revision of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis with emphasis on the risk of major complications. Patients and Methods. A total of 36 patients with a MoP THA who underwent revision for ARMD due to trunnionosis were identified. Three were excluded as their revision had been to another metal head. The remaining 33 were revised to a ceramic head with a titanium sleeve. We describe the presentation, revision findings, and risk of complications in these patients. Results. The patients presented with pain, swelling, stiffness, or instability and an inflammatory mass was confirmed radiologically. Macroscopic material deposition on the trunnion was seen in all patients, associated with ARMD. Following revision, six (18.2%) dislocated, requiring further revision in four. Three (9.1%) developed a deep infection and six (18.2%) had significant persistent pain without an obvious cause. One developed a femoral artery thrombosis after excision of an iliofemoral pseudotumor, requiring a thrombectomy. Conclusion. The risk of serious complications following revision MoP THA for ARMD associated with trunnionosis is high. In the presence of extensive tissue damage, a constrained liner or dual mobility construct is recommended in these patients. Cite this article: Bone Joint J 2018;100-B:720–4


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 134 - 142
1 Feb 2018
Hexter AT Hislop SM Blunn GW Liddle AD

Aims. Periprosthetic joint infection (PJI) is a serious complication of total hip arthroplasty (THA). Different bearing surface materials have different surface properties and it has been suggested that the choice of bearing surface may influence the risk of PJI after THA. The objective of this meta-analysis was to compare the rate of PJI between metal-on-polyethylene (MoP), ceramic-on-polyethylene (CoP), and ceramic-on-ceramic (CoC) bearings. Patients and Methods. Electronic databases (Medline, Embase, Cochrane library, Web of Science, and Cumulative Index of Nursing and Allied Health Literature) were searched for comparative randomized and observational studies that reported the incidence of PJI for different bearing surfaces. Two investigators independently reviewed studies for eligibility, evaluated risk of bias, and performed data extraction. Meta-analysis was performed using the Mantel–Haenzel method and random-effects model in accordance with methods of the Cochrane group. Results. Our search strategy revealed 2272 studies, of which 17 met the inclusion criteria and were analyzed. These comprised 11 randomized controlled trials and six observational studies. The overall quality of included studies was high but the observational studies were at high risk of bias due to inadequate adjustment for confounding factors. The overall cumulative incidence of PJI across all studies was 0.78% (1514/193 378). For each bearing combination, the overall incidence was as follows: MoP 0.85% (1353/158 430); CoP 0.38% (67/17 489); and CoC 0.53% (94/17 459). The meta-analysis showed no significant difference between the three bearing combinations in terms of risk of PJI. Conclusion. On the basis of the clinical studies available, there is no evidence that bearing choice influences the risk of PJI. Future research, including basic science studies and large, adequately controlled registry studies, may be helpful in determining whether implant materials play a role in determining the risk of PJI following arthroplasty surgery. Cite this article: Bone Joint J 2018;100-B:134–42


Bone & Joint 360
Vol. 7, Issue 1 | Pages 12 - 14
1 Feb 2018


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 44 - 49
1 Jan 2018
Berstock JR Whitehouse MR Duncan CP

Aims

To present a surgically relevant update of trunnionosis.

Materials and Methods

Systematic review performed April 2017.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 28 - 32
1 Jan 2018
Goodnough LH Bala A Huddleston III JI Goodman SB Maloney WJ Amanatullah DF

Aims

Many case reports and small studies have suggested that cobalt ions are a potential cause of cardiac complications, specifically cardiomyopathy, after metal-on-metal (MoM) total hip arthroplasty (THA). The impact of metal ions on the incidence of cardiac disease after MoM THA has not been evaluated in large studies. The aim of this study was to compare the rate of onset of new cardiac symptoms in patients who have undergone MoM THA with those who have undergone metal-on-polyethylene (MoP) THA.

Patients and Methods

Data were extracted from the Standard Analytics Files database for patients who underwent MoM THA between 2005 and 2012. Bearing surface was selected using International Classification of Diseases ninth revision codes. Patients with a minimum five-year follow-up were selected. An age and gender-matched cohort of patients who underwent MoP THA served as a comparison group. New diagnoses of cardiac disease were collected during the follow-up period. Comorbidities and demographics were identified and routine descriptive statistics were used.


Bone & Joint 360
Vol. 6, Issue 6 | Pages 14 - 16
1 Dec 2017


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1435 - 1441
1 Nov 2017
Buttaro MA Oñativia JI Slullitel PA Andreoli M Comba F Zanotti G Piccaluga F

Aims

The Corail stem has good long-term results. After four years of using this stem, we have detected a small group of patients who have presented with symptomatic metaphyseal debonding. The aim of this study was to quantify the incidence of this complication, to delineate the characteristics of patients presenting with this complication and to compare these patients with asymptomatic controls to determine any important predisposing factors.

Patients and Methods

Of 855 Corail collarless cementless stems implanted for osteoarthritis, 18 presented with symptomatic metaphyseal debonding. A control group of 74 randomly selected patients was assembled. Clinical and radiological parameters were measured and a logistic regression model was created to evaluate factors associated with metaphyseal debonding.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1399 - 1408
1 Oct 2017
Scott CEH MacDonald D Moran M White TO Patton JT Keating JF

Aims

To evaluate the outcomes of cemented total hip arthroplasty (THA) following a fracture of the acetabulum, with evaluation of risk factors and comparison with a patient group with no history of fracture.

Patients and Methods

Between 1992 and 2016, 49 patients (33 male) with mean age of 57 years (25 to 87) underwent cemented THA at a mean of 6.5 years (0.1 to 25) following acetabular fracture. A total of 38 had undergone surgical fixation and 11 had been treated non-operatively; 13 patients died at a mean of 10.2 years after THA (0.6 to 19). Patients were assessed pre-operatively, at one year and at final follow-up (mean 9.1 years, 0.5 to 23) using the Oxford Hip Score (OHS). Implant survivorship was assessed. An age and gender-matched cohort of THAs performed for non-traumatic osteoarthritis (OA) or avascular necrosis (AVN) (n = 98) were used to compare complications and patient-reported outcome measures (PROMs).


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1298 - 1303
1 Oct 2017
Schouten R Malone AA Frampton CM Tiffen C Hooper G

Aims

The primary aim of this independent prospective randomised trial was to compare serum metal ion levels for ceramic-on-metal (CoM) and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty (THA). Our one-year results demonstrated elevation in metal ion levels above baseline with no significant difference between the CoM and MoM groups. This paper reviews the five-year data.

Patients and Methods

The implants used in each patient differed only in respect to the type of femoral head (ceramic or metal). At five-year follow-up of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available for comparison.


The Bone & Joint Journal
Vol. 99-B, Issue 9 | Pages 1147 - 1152
1 Sep 2017
Lainiala O Reito A Jämsä P Eskelinen A

Aims

To determine whether there is any association between glomerular filtration rate (GFR) and blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasty.

Patients and Methods

We identified 179 patients with a unilateral 36 mm diameter head as part of a stemmed Summit-Pinnacle MoM hip arthroplasty. GFR was calculated using the Modification of Diet in Renal Disease formula.


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 993 - 995
1 Aug 2017
Skinner JA Haddad FS


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1020 - 1027
1 Aug 2017
Matharu GS Judge A Pandit HG Murray DW

Aims

To determine the outcomes following revision surgery of metal-on-metal hip arthroplasties (MoMHA) performed for adverse reactions to metal debris (ARMD), and to identify factors predictive of re-revision.

Patients and Methods

We performed a retrospective observational study using National Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery for ARMD between 2008 and 2014. The outcomes studied following revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using competing-risk regression modelling.


Bone & Joint Research
Vol. 6, Issue 7 | Pages 405 - 413
1 Jul 2017
Matharu GS Judge A Murray DW Pandit HG

Objectives

Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision.

Methods

We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.


Bone & Joint Research
Vol. 6, Issue 5 | Pages 345 - 350
1 May 2017
Di Laura A Hothi H Henckel J Swiatkowska I Liow MHL Kwon Y Skinner JA Hart AJ

Objectives

The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design.

Methods

This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40 trunnions paired with either CoCr (n = 21) or ceramic (n = 27) heads. The taper junction of all hips was evaluated for fretting/corrosion damage and volumetric material loss using a roundness-measuring machine. We used linear regression analysis to investigate taper damage differences after adjusting for potential confounding variables.


Bone & Joint 360
Vol. 6, Issue 2 | Pages 10 - 13
1 Apr 2017


Bone & Joint 360
Vol. 6, Issue 1 | Pages 10 - 13
1 Feb 2017


Bone & Joint Research
Vol. 6, Issue 1 | Pages 52 - 56
1 Jan 2017
Hothi HS Kendoff D Lausmann C Henckel J Gehrke T Skinner J Hart A

Objectives. Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. Methods. In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons. Results. Evidence of corrosion was observed in 55% of hips. The median Goldberg taper corrosion score was 2 (1 to 4) and the annual rate of material loss at the taper was 0.084 mm. 3. /year (0 to 0.239). The median trunnion corrosion score was 1 (1 to 3). Conclusions. We have reported a level of trunnionosis for MOP hips with large-diameter heads that were revised for reasons other than trunnionosis, and therefore may be clinically insignificant. Cite this article: H. S. Hothi, D. Kendoff, C. Lausmann, J. Henckel, T. Gehrke, J. Skinner, A. Hart. Clinically insignificant trunnionosis in large-diameter metal-on-polyethylene total hip arthroplasty. Bone Joint Res 2017;6:52–56. DOI: 10.1302/2046-3758.61.BJR-2016-0150.R2


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 100 - 106
1 Jan 2017
Aujla RS Sheikh N Divall P Bhowal B Dias JJ

Aims

We performed a systematic review of the current literature regarding the outcomes of unconstrained metacarpophalangeal joint (MCPJ) arthroplasty.

Materials and Methods

We initially identified 1305 studies, and 406 were found to be duplicates. After exclusion criteria were applied, seven studies were included. Outcomes extracted included pre- and post-operative pain visual analogue scores, range of movement (ROM), strength of pinch and grip, satisfaction and patient reported outcome measures (PROMs). Clinical and radiological complications were recorded. The results are presented in three groups based on the design of the arthroplasty and the aetiology (pyrocarbon-osteoarthritis (pyro-OA), pyrocarbon-inflammatory arthritis (pyro-IA), metal-on-polyethylene (MoP)).


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 44 - 50
1 Jan 2017
Salo PP Honkanen PB Ivanova I Reito A Pajamäki J Eskelinen A

Aims

We evaluated the short-term functional outcome and prevalence of bearing-specific generation of audible noise in 301 patients (336 hips) operated on with fourth generation (Delta) medium diameter head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs).

Patients and Methods

There were 191 female (63%) and 110 male patients (37%) with a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years (1.3 to 3.4). Patients completed three questionnaires: Oxford Hip Score (OHS), Research and Development 36-item health survey (RAND-36) and a noise-specific symptom questionnaire. Plain radiographs were also analysed. A total of three hips (0.9%) were revised.


Bone & Joint Research
Vol. 5, Issue 10 | Pages 461 - 469
1 Oct 2016
Liu YK Deng XX Yang H

Objectives

The cytotoxicity induced by cobalt ions (Co2+) and cobalt nanoparticles (Co-NPs) which released following the insertion of a total hip prosthesis, has been reported. However, little is known about the underlying mechanisms. In this study, we investigate the toxic effect of Co2+ and Co-NPs on liver cells, and explain further the potential mechanisms.

Methods

Co-NPs were characterised for size, shape, elemental analysis, and hydrodynamic diameter, and were assessed by Transmission Electron Microscope, Scanning Electron Microscope, Energy Dispersive X-ray Spectroscopy and Dynamic Light Scattering. BRL-3A cells were used in this study. Cytotoxicity was evaluated by MTT and lactate dehydrogenase release assay. In order to clarify the potential mechanisms, reactive oxygen species, Bax/Bcl-2 mRNA expression, IL-8 mRNA expression and DNA damage were assessed on BRL-3A cells after Co2+ or Co-NPs treatment.


Bone & Joint Research
Vol. 5, Issue 9 | Pages 379 - 386
1 Sep 2016
Pahuta M Smolders JM van Susante JL Peck J Kim PR Beaule PE

Objectives

Alarm over the reported high failure rates for metal-on-metal (MoM) hip implants as well as their potential for locally aggressive Adverse Reactions to Metal Debris (ARMDs) has prompted government agencies, internationally, to recommend the monitoring of patients with MoM hip implants. Some have advised that a blood ion level >7 µg/L indicates potential for ARMDs. We report a systematic review and meta-analysis of the performance of metal ion testing for ARMDs.

Methods

We searched MEDLINE and EMBASE to identify articles from which it was possible to reconstruct a 2 × 2 table. Two readers independently reviewed all articles and extracted data using explicit criteria. We computed a summary receiver operating curve using a Bayesian random-effects hierarchical model.


Bone & Joint Research
Vol. 5, Issue 8 | Pages 338 - 346
1 Aug 2016
MacLeod AR Sullivan NPT Whitehouse MR Gill HS

Objectives

Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection.

Materials and Methods

Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.


Bone & Joint 360
Vol. 5, Issue 3 | Pages 10 - 12
1 Jun 2016


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 579 - 584
1 May 2016
Osman K Panagiotidou AP Khan M Blunn G Haddad FS

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface.

We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice.

Cite this article: Bone Joint J 2016;98-B:579–84.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 162 - 168
1 May 2016
Athanasou NA

Pathological assessment of periprosthetic tissues is important, not only for diagnosis, but also for understanding the pathobiology of implant failure. The host response to wear particle deposition in periprosthetic tissues is characterised by cell and tissue injury, and a reparative and inflammatory response in which there is an innate and adaptive immune response to the material components of implant wear. Physical and chemical characteristics of implant wear influence the nature of the response in periprosthetic tissues and account for the development of particular complications that lead to implant failure, such as osteolysis which leads to aseptic loosening, and soft-tissue necrosis/inflammation, which can result in pseudotumour formation. The innate response involves phagocytosis of implant-derived wear particles by macrophages; this is determined by pattern recognition receptors and results in expression of cytokines, chemokines and growth factors promoting inflammation and osteoclastogenesis; phagocytosed particles can also be cytotoxic and cause cell and tissue necrosis. The adaptive immune response to wear debris is characterised by the presence of lymphoid cells and most likely occurs as a result of a cell-mediated hypersensitivity reaction to cell and tissue components altered by interaction with the material components of particulate wear, particularly metal ions released from cobalt-chrome wear particles.

Cite this article: Professor N. A. Athanasou. The pathobiology and pathology of aseptic implant failure. Bone Joint Res 2016;5:162–168. DOI: 10.1302/2046-3758.55.BJR-2016-0086.


Bone & Joint Research
Vol. 5, Issue 3 | Pages 73 - 79
1 Mar 2016
Anwander H Cron GO Rakhra K Beaule PE

Objectives. Hips with metal-on-metal total hip arthroplasty (MoM THA) have a high rate of adverse local tissue reactions (ALTR), often associated with hypersensitivity reactions. Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) measures tissue perfusion with the parameter Ktrans (volume transfer constant of contrast agent). Our purpose was 1) to evaluate the feasibility of DCE-MRI in patients with THA and 2) to compare DCE-MRI in patients with MoM bearings with metal-on-polyethylene (MoP) bearings, hypothesising that the perfusion index Ktrans in hips with MoM THA is higher than in hips with MoP THA. Methods. In this pilot study, 16 patients with primary THA were recruited (eight MoM, eight MoP). DCE-MRI of the hip was performed at 1.5 Tesla (T). For each patient, Ktrans was computed voxel-by-voxel in all tissue lateral to the bladder. The mean Ktrans for all voxels was then calculated. These values were compared with respect to implant type and gender, and further correlated with clinical parameters. Results. There was no significant difference between the two bearing types with both genders combined. However, dividing patients by THA bearing and gender, women with MoM bearings had the highest Ktrans values, exceeding those of women with MoP bearings (0.067 min. −1. versus 0.053 min. −1. ; p-value < 0.05) and men with MoM bearings (0.067 min. −1. versus 0.034 min. −1. ; p-value < 0.001). Considering only the men, patients with MoM bearings had lower Ktrans than those with MoP bearings (0.034 min. −1. versus 0.046 min. −1. ; p < 0.05). Conclusion. DCE-MRI is feasible to perform in tissues surrounding THA. Females with MoM THA show high Ktrans values in DCE-MRI, suggesting altered tissue perfusion kinematics which may reflect relatively greater inflammation. Cite this article: Dr P. E. Beaule. Perfusion MRI in hips with metal-on-metal and metal-on-polyethylene total hip arthroplasty: A pilot stud. Bone Joint Res 2016;5:73–79. DOI: 10.1302/2046-3758.53.2000572


The Bone & Joint Journal
Vol. 98-B, Issue 2 | Pages 187 - 193
1 Feb 2016
Lash NJ Whitehouse MR Greidanus NV Garbuz DS Masri BA Duncan CP

Aims. We present a case series of ten metal-on-polyethylene total hip arthroplasties (MoP THAs) with delayed dislocation associated with unrecognised adverse local tissue reaction due to corrosion at the trunnion and pseudotumour formation. . Methods. The diagnosis was not suspected in nine of the ten patients (six female/four male; mean age 66 years), despite treatment in a specialist unit (mean time from index surgery to revision was 58 months, 36 to 84). It was identified at revision surgery and subsequently confirmed by histological examination of resected tissue. Pre-operative assessment and culture results ruled out infection. A variety of treatment strategies were used, including resection of the pseudotumour and efforts to avoid recurrent dislocation. . Results. The rate of complications was high and included three deep infections, two patients with recurrent dislocation, and one recurrent pseudotumour. . Conclusion. This series (mean follow-up of 76 months following index procedure and 19 months following revision THA) demonstrates that pseudotumour is an infrequent but important contributor to delayed instability following MoP THA. It is easy to overlook in the differential diagnosis, especially if the alignment of the components is less than optimal, leading to an assumption that malalignment is the cause of the dislocation. The instability is likely to be multifactorial and the revision surgery is complex. Take home message: Due to the high complication rate associated with revision in this cohort, the diagnosis should be borne in mind when counselling patients regarding the risks of revision surgery. . Cite this article: Bone Joint J 2016;98-B:187–93


Bone & Joint Research
Vol. 5, Issue 2 | Pages 52 - 60
1 Feb 2016
Revell PA Matharu GS Mittal S Pynsent PB Buckley CD Revell MP

Objectives. T-cells are considered to play an important role in the inflammatory response causing arthroplasty failure. The study objectives were to investigate the composition and distribution of CD4+ T-cell phenotypes in the peripheral blood (PB) and synovial fluid (SF) of patients undergoing revision surgery for failed metal-on-metal (MoM) and metal-on-polyethylene (MoP) hip arthroplasties, and in patients awaiting total hip arthroplasty. Methods. In this prospective case-control study, PB and SF were obtained from 22 patients (23 hips) undergoing revision of MoM (n = 14) and MoP (n = 9) hip arthroplasties, with eight controls provided from primary hip osteoarthritis cases awaiting arthroplasty. Lymphocyte subtypes in samples were analysed using flow cytometry. Results. The percentages of CD4+ T-cell subtypes in PB were not different between groups. The CD4+ T-cells in the SF of MoM hips showed a completely different distribution of phenotypes compared with that found in the PB in the same patients, including significantly decreased CD4+ T-central memory cells (p < 0.05) and increased T-effector memory cells (p < 0.0001) in the SF. Inducible co-stimulator (ICOS) was the only co-stimulatory molecule with different expression on CD4+ CD28+ cells between groups. In PB, ICOS expression was increased in MoM (p < 0.001) and MoP (p < 0.05) cases compared with the controls. In SF, ICOS expression was increased in MoM hips compared with MoP hips (p < 0.05). Conclusions. Increased expression of ICOS on CD4+ T-cells in PB and SF of patients with failed arthroplasties suggests that these cells are activated and involved in generating immune responses. Variations in ICOS expression between MoM and MoP hips may indicate different modes of arthroplasty failure. Cite this article: Professor P. A. Revell. Increased expression of inducible co-stimulator on CD4+ T-cells in the peripheral blood and synovial fluid of patients with failed hip arthroplasties. Bone Joint Res 2016;5:52–60. doi: 10.1302/2046-3758.52.2000574


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 28 - 32
1 Jan 2016
Hanna SA Somerville L McCalden RW Naudie DD MacDonald SJ

Aims

The purpose of this study was to compare the long-term results of primary total hip arthroplasty (THA) in young patients using either a conventional (CPE) or a highly cross-linked (HXLPE) polyethylene liner in terms of functional outcome, incidence of osteolysis, radiological wear and rate of revision.

Methods

We included all patients between the ages of 45 and 65 years who, between January 2000 and December 2001, had undergone a primary THA for osteoarthritis at our hospital using a CPE or HXLPE acetabular liner and a 28 mm cobalt-chrome femoral head.

From a total of 160 patients, 158 (177 hips) were available for review (CPE 89; XLPE 88). The mean age, body mass index (BMI) and follow-up in each group were: CPE: 56.8 years (46 to 65); 30.7 kg/m2 (19 to 58); 13.2 years (2.1 to 14.7) and HXLPE: 55.6 years (45 to 65); BMI: 30 kg/m2 (18 to 51); 13.1 years (5.7 to 14.4).


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 6 - 13
1 Jan 2016
Cheung AC Banerjee S Cherian JJ Wong F Butany J Gilbert C Overgaard C Syed K Zywiel MG Jacobs JJ Mont MA

Recently, the use of metal-on-metal articulations in total hip arthroplasty (THA) has led to an increase in adverse events owing to local soft-tissue reactions from metal ions and wear debris. While the majority of these implants perform well, it has been increasingly recognised that a small proportion of patients may develop complications secondary to systemic cobalt toxicity when these implants fail. However, distinguishing true toxicity from benign elevations in cobalt ion levels can be challenging.

The purpose of this two part series is to review the use of cobalt alloys in THA and to highlight the following related topics of interest: mechanisms of cobalt ion release and their measurement, definitions of pathological cobalt ion levels, and the pathophysiology, risk factors and treatment of cobalt toxicity. Historically, these metal-on-metal arthroplasties are composed of a chromium-cobalt articulation.

The release of cobalt is due to the mechanical and oxidative stresses placed on the prosthetic joint. It exerts its pathological effects through direct cellular toxicity.

This manuscript will highlight the pathophysiology of cobalt toxicity in patients with metal-on-metal hip arthroplasties.

Take home message: Patients with new or evolving hip symptoms with a prior history of THA warrant orthopaedic surgical evaluation. Increased awareness of the range of systemic symptoms associated with cobalt toxicity, coupled with prompt orthopaedic intervention, may forestall the development of further complications.

Cite this article: Bone Joint J 2016;98-B:6–13.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1481 - 1487
1 Nov 2015
van der Veen HC Reininga IHF Zijlstra WP Boomsma MF Bulstra SK van Raay JJAM

We compared the incidence of pseudotumours after large head metal-on-metal (MoM) total hip arthroplasty (THA) with that after conventional metal-on-polyethylene (MoP) THA and assessed the predisposing factors to pseudotumour formation. . From a previous randomised controlled trial which compared large head (38 mm to 60 mm) cementless MoM THA with conventional head (28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males, 20 females, mean age of 64 years, standard deviation (. sd. ) 4) and 55 MoP (25 males, 30 females, mean age of 65 years, . sd. 5) were recruited after a mean follow-up of 50 months (36 to 64). . The incidence of pseudotumours, measured using a standardised CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP THA. Women with a MoM THA were more likely to develop a pseudotumour than those with a MoP THA (15 vs 7, odds ratio (OR) = 13.4, p < 0.001). There was a similar incidence of pseudotumours in men with MoM THAs and those with MoP THAs (7 vs 5, OR = 2.1, p = 0.30). Elevated cobalt levels (≥ 5 microgram/L) were only associated with pseudotumours in women with a MoM THA. There was no difference in mean Oxford and Harris hip scores between patients with a pseudotumour and those without. . Contrary to popular belief, pseudotumours occur frequently around MoP THAs. Women with a MoM THA and an elevated cobalt level are at greatest risk. In this study, pseudotumours had no effect on the functional outcome after either large head MoM or conventional MoP THA. Cite this article: Bone Joint J 2015;97-B:1481–7


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1175 - 1182
1 Sep 2015
Bisseling P de Wit BWK Hol AM van Gorp MJ van Kampen A van Susante JLC

Patients from a randomised trial on resurfacing hip arthroplasty (RHA) (n = 36, 19 males; median age 57 years, 24 to 65) comparing a conventional 28 mm metal-on-metal total hip arthroplasty (MoM THA) (n = 28, 17 males; median age 59 years, 37 to 65) and a matched control group of asymptomatic patients with a 32 mm ceramic-on-polyethylene (CoP) THA (n = 33, 18 males; median age 63 years, 38 to 71) were cross-sectionally screened with metal artefact reducing sequence-MRI (MARS-MRI) for pseudotumour formation at a median of 55 months (23 to 72) post-operatively. MRIs were scored by consensus according to three different classification systems for pseudotumour formation.

Clinical scores were available for all patients and metal ion levels for MoM bearing patients.

Periprosthetic lesions with a median volume of 16 mL (1.5 to 35.9) were diagnosed in six patients in the RHA group (17%), one in the MoM THA group (4%) and six in the CoP group (18%). The classification systems revealed no clear differences between the groups. Solid lesions (n = 3) were exclusively encountered in the RHA group. Two patients in the RHA group and one in the MoM THA group underwent a revision for pseudotumour formation. There was no statistically significant relationship between clinical scoring, metal ion levels and periprosthetic lesions in any of the groups.

Periprosthetic fluid collections are seen on MARS-MRI after conventional CoP THA and RHA and may reflect a soft-tissue collection or effusion.

Currently available MRI classification systems seem to score these collections as pseudotumours, causing an-overestimatation of the incidence of pseudotumours.

Cite this article: Bone Joint J 2015;97-B:1175–82.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1183 - 1191
1 Sep 2015
Briggs TWR Hanna SA Kayani B Tai S Pollock RC Cannon SR Blunn GW Carrington RWJ

The long term biological effects of wear products following total hip arthroplasty (THA) are unclear. However, the indications for THA are expanding, with increasingly younger patients undergoing the procedure. This prospective, randomised study compared two groups of patients undergoing THA after being randomised to receive one of two different bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal (MoM) n = 23. We investigated the relationship between three variables: bearing surface (MoP vs MoM), whole blood levels of chromium (Cr) and cobalt (Co) and chromosomal aberrations in peripheral lymphocyte pre-operatively and at one, two and five years post-surgery. Our results demonstrated significantly higher mean cobalt and chromium (Co and Cr) blood levels in the MoM group at all follow-up points following surgery (p < 0.01), but there were no significant differences in the chromosomal aberration indices between MoM and MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p = 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34) for translocations, breaks, loss and gain of chromosomes respectively. Regression analysis showed a strong linear relationship between Cr levels and the total chromosomal aberration indices in the MoM group (R. 2.  = 0.90016), but this was not as strong for Co (R. 2. = 0.68991). In the MoP group, the analysis revealed a poor relationship between Cr levels and the total chromosomal aberration indices (R. 2. = 0.23908) but a slightly stronger relationship for Co (R. 2. = 0.64292). Across both groups, Spearman’s correlation detected no overall association between Co and Cr levels and each of the studied chromosomal aberrations. There remains no clear indication which THA bearing couple is the most biocompatible, especially in young active patients. While THA continues to be very successful at alleviating pain and restoring function, the long-term biological implications of the procedure still require further scrutiny. Cite this article: Bone Joint J 2015;97-B:1183–91


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 917 - 923
1 Jul 2015
Singh G Nuechtern JV Meyer H Fiedler GM Awiszus F Junk-Jantsch S Bruegel M Pflueger G Lohmann CH

The peri-prosthetic tissue response to wear debris is complex and influenced by various factors including the size, area and number of particles. We hypothesised that the ‘biologically active area’ of all metal wear particles may predict the type of peri-prosthetic tissue response.

Peri-prosthetic tissue was sampled from 21 patients undergoing revision of a small diameter metal-on-metal (MoM) total hip arthroplasty (THA) for aseptic loosening. An enzymatic protocol was used for tissue digestion and scanning electron microscope was used to characterise particles. Equivalent circle diameters and particle areas were calculated. Histomorphometric analyses were performed on all tissue specimens. Aspirates of synovial fluid were collected for analysis of the cytokine profile analysis, and compared with a control group of patients undergoing primary THA (n = 11) and revision of a failed ceramic-on-polyethylene arthroplasty (n = 6).

The overall distribution of the size and area of the particles in both lymphocyte and non-lymphocyte-dominated responses were similar; however, the subgroup with lymphocyte-dominated peri-prosthetic tissue responses had a significantly larger total number of particles.

14 cytokines (interleukin (IL)-1ß, IL-2, IL-4, IL-5, IL-6, IL-10, IL-13, IL-17, interferon (IFN)-γ, and IFN-gamma-inducible protein 10), chemokines (macrophage inflammatory protein (MIP)-1α and MIP-1ß), and growth factors (granulocyte macrophage colony stimulating factor (GM-CSF) and platelet derived growth factor) were detected at significantly higher levels in patients with metal wear debris compared with the control group.

Significantly higher levels for IL-1ß, IL-5, IL-10 and GM-CSF were found in the subgroup of tissues from failed MoM THAs with a lymphocyte-dominated peri-prosthetic response compared with those without this response.

These results suggest that the ‘biologically active area’ predicts the type of peri-prosthetic tissue response. The cytokines IL-1ß, IL-5, IL-10, and GM-CSF are associated with lymphocyte-dominated tissue responses from failed small-diameter MoM THA.

Cite this article: Bone Joint J 2015;97-B:917–23.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 911 - 916
1 Jul 2015
Del Balso C Teeter MG Tan SC Lanting BA Howard JL

Tribocorrosion at the head–neck taper interface – so-called ‘taperosis’ – may be a source of metal ions and particulate debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and corrosion in retrieved head–neck tapers in vivo for a minimum of two years (mean 8.7 years; 2.6 to 15.9). A total of 56 femoral heads ranging from 28 mm to 3 mm to 28 mm + 8 mm, and 17 femoral stems featuring a single taper design were included in the study. Fretting and corrosion were scored in three horizontally oriented concentric zones of each taper by stereomicroscopy. Head length was observed to affect fretting (p = 0.03), with 28 mm + 8 mm femoral heads showing greater total fretting scores than all other head lengths. The central zone of the femoral head bore taper was subject to increased fretting damage (p = 0.01), regardless of head length or stem offset. High-offset femoral stems were associated with greater total fretting of the bore taper (p = 0.04). Increased fretting damage is seen with longer head lengths and high-offset femoral stems, and occurs within a central concentric zone of the femoral head bore taper. Further investigation is required to determine the effect of increased head size, and variations in head–neck taper design. Cite this article: Bone Joint J 2015; 97-B:911–16


Bone & Joint 360
Vol. 4, Issue 3 | Pages 10 - 12
1 Jun 2015

The June 2015 Hip & Pelvis Roundup360 looks at: neuraxial anaesthesia and large joint arthroplasty; revision total hip arthoplasty: factors associated with re-revision surgery; acetabular version and clinical outcomes in impingement surgery; hip precautions may be ineffective; implant selection and cost effectiveness; femoroacetabular impingement in the older age group; multiple revision in hip arthroplasty


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 449 - 457
1 Apr 2015
Pulikottil-Jacob R Connock M Kandala N Mistry H Grove A Freeman K Costa M Sutcliffe P Clarke A

Many different designs of total hip arthroplasty (THA) with varying performance and cost are available. The identification of those which are the most cost-effective could allow significant cost-savings. We used an established Markov model to examine the cost effectiveness of five frequently used categories of THA which differed according to bearing surface and mode of fixation, using data from the National Joint Registry for England and Wales. Kaplan–Meier analyses of rates of revision for men and women were modelled with parametric distributions. Costs of devices were provided by the NHS Supply Chain and associated costs were taken from existing studies. Lifetime costs, lifetime quality-adjusted-life-years (QALYs) and the probability of a device being cost effective at a willingness to pay £20 000/QALY were included in the models.

The differences in QALYs between different categories of implant were extremely small (<  0.0039 QALYs for men or women over the patient’s lifetime) and differences in cost were also marginal (£2500 to £3000 in the same time period). As a result, the probability of any particular device being the most cost effective was very sensitive to small, plausible changes in quality of life estimates and cost.

Our results suggest that available evidence does not support recommending a particular device on cost effectiveness grounds alone. We would recommend that the choice of prosthesis should be determined by the rate of revision, local costs and the preferences of the surgeon and patient.

Cite this article: Bone Joint J 2015;97-B:449–57.


Bone & Joint 360
Vol. 4, Issue 1 | Pages 2 - 5
1 Feb 2015
Wright GM Porteous MJ


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 150 - 153
1 Feb 2015
Rogers BA Alolabi B Carrothers AD Kreder HJ Jenkinson RJ

In this study we evaluated whether pre-operative Western Ontario and McMaster Universities (WOMAC) osteoarthritis scores can predict satisfaction following total hip arthroplasty (THA). Prospective data for a cohort of patients undergoing THA from two large academic centres were collected, and pre-operative and one-year post-operative WOMAC scores and a 25-point satisfaction questionnaire were obtained for 446 patients. Satisfaction scores were dichotomised into either improvement or deterioration. Scatter plots and Spearman’s rank correlation coefficient were used to describe the association between pre-operative WOMAC and one-year post-operative WOMAC scores and patient satisfaction. Satisfaction was compared using receiver operating characteristic (ROC) analysis against pre-operative, post-operative and δ WOMAC scores.

We found no relationship between pre-operative WOMAC scores and one-year post-operative WOMAC or satisfaction scores, with Spearman’s rank correlation coefficients of 0.16 and –0.05, respectively. The ROC analysis showed areas under the curve (AUC) of 0.54 (pre-operative WOMAC), 0.67 (post-operative WOMAC) and 0.43 (δ WOMAC), respectively, for an improvement in satisfaction.

We conclude that the pre-operative WOMAC score does not predict the post-operative WOMAC score or patient satisfaction after THA, and that WOMAC scores can therefore not be used to prioritise patient care.

Cite this article: Bone Joint J 2015;97-B:150–3.


Bone & Joint 360
Vol. 4, Issue 1 | Pages 6 - 11
1 Feb 2015
Manktelow A Bloch B

This review examines the future of total hip arthroplasty, aiming to avoid past mistakes


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1600 - 1609
1 Dec 2014
Matharu GS Pynsent PB Sumathi VP Mittal S Buckley CD Dunlop DJ Revell PA Revell MP

We undertook a retrospective cohort study to determine clinical outcomes following the revision of metal-on-metal (MoM) hip replacements for adverse reaction to metal debris (ARMD), and to identify predictors of time to revision and outcomes following revision. Between 1998 and 2012 a total of 64 MoM hips (mean age at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings and 18 (28%) total hip replacements) were revised for ARMD at one specialist centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision for ARMD there were 13 hips (20.3%) with post-operative complications and eight (12.5%) requiring re-revision.

The Kaplan–Meier five-year survival rate for ARMD revision was 87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding re-revisions, the median absolute Oxford hip score (OHS) following ARMD revision using the percentage method (0% best outcome and 100% worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%), which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using the modified OHS. Histopathological response did not affect time to revision for ARMD (p = 0.334) or the subsequent risk of re-revision (p = 0.879). Similarly, the presence or absence of a contralateral MoM hip bearing did not affect time to revision for ARMD (p = 0.066) or the subsequent risk of re-revision (p = 0.178).

Patients revised to MoM bearings had higher rates of re-revision (five of 16 MoM hips re-revised; p = 0.046), but those not requiring re-revision had good functional results (median absolute OHS 14.6% or 41.0/48). Short-term morbidity following revision for ARMD was comparable with previous reports. Caution should be exercised when choosing bearing surfaces for ARMD revisions.

Cite this article: Bone Joint J 2014;96-B:1600–9.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1610 - 1617
1 Dec 2014
Lainiala O Eskelinen A Elo P Puolakka T Korhonen J Moilanen T

We conducted a retrospective study to assess the prevalence of adverse reactions to metal debris (ARMD) in patients operated on at our institution with metal-on-metal (MoM) total hip replacements with 36 mm heads using a Pinnacle acetabular shell. A total of 326 patients (150 males, 175 hips; 176 females, 203 hips) with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5 years (0.1 to 10.8) participating in our in-depth modern MoM follow-up programme were included in the study, which involved recording whole blood cobalt and chromium ion measurements, Oxford hip scores (OHS) and plain radiographs of the hip and targeted cross-sectional imaging. Elevated blood metal ion levels (> 5 parts per billion) were seen in 32 (16.1%) of the 199 patients who underwent unilateral replacement. At 23 months after the start of our modern MoM follow-up programme, 29 new cases of ARMD had been revealed. Hence, the nine-year survival of this cohort declined from 96% (95% CI 95 to 98) with the old surveillance routine to 86% (95% CI 82 to 90) following the new protocol. Although ARMD may not be as common in 36 mm MoM THRs as in those with larger heads, these results support the Medicines and Healthcare Products Regulatory Agency guidelines on regular reviews and further investigations, and emphasise the need for specific a follow-up programme for patients with MoM THRs.

Cite this article: Bone Joint J 2014; 96-B:1610–17.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 23 - 26
1 Nov 2014
Cooper HJ Della Valle CJ

Dislocation remains among the most common complications of, and reasons for, revision of both primary and revision total hip replacements (THR). Hence, there is great interest in maximising stability to prevent this complication. Head size has been recognised to have a strong influence on the risk of dislocation post-operatively. As femoral head size increases, stability is augmented, secondary to an increase in impingement-free range of movement. Larger head sizes also greatly increase the ‘jump distance’ required for the head to dislocate in an appropriately positioned cup. Level-one studies support the use of larger diameter heads as they decrease the risk of dislocation following primary and revision THR. Highly cross-linked polyethylene has allowed us to increase femoral head size, without a marked increase in wear. However, the thin polyethylene liners necessary to accommodate larger heads may increase the risk of liner fracture and larger heads have also been implicated in causing soft-tissue impingement resulting in groin pain. Larger diameter heads also impart larger forces on the femoral trunnion, which may contribute to corrosion, metal release, and adverse local tissue reactions. Alternative large bearings including large ceramic heads and dual mobility bearings may mitigate some of these risks, and several of these devices have been used with clinical success.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):23–6.