Objectives. Because of the contradictory body of evidence related to the
potential benefits of helical blades in trochanteric fracture fixation,
we studied the effect of bone compaction resulting from the insertion
of a proximal femoral nail anti-rotation (PFNA). . Methods. We developed a subject-specific computational model of a trochanteric
fracture (31-A2 in the AO classification) with lack of medial support
and varied the bone density to account for variability in bone properties
among hip fracture patients. Results. We show that for a bone density corresponding to 100% of the
bone density of the cadaveric femur, there does not seem to be any
advantage in using a PFNA with respect to the risk of blade cut-out.
On the other hand, in a more osteoporotic femoral head characterised
by a density corresponding to 75% of the initial bone density, local
bone compaction around the helical blade provides additional bone
purchase, thereby decreasing the risk of cut-out, as quantified
by the volume of bone susceptible to yielding. Conclusions. Our findings indicate benefits of using a PFNA over an intramedullary
nail with a conventional
The use of two implants to manage concomitant ipsilateral femoral
shaft and proximal femoral fractures has been indicated, but no
studies address the relationship of dynamic hip screw (DHS) side
plate screws and the intramedullary nail where failure might occur
after union. This study compares different implant configurations
in order to investigate bridging the gap between the distal DHS
and tip of the intramedullary nail. A total of 29 left synthetic femora were tested in three groups:
1) gapped short nail (GSN); 2) unicortical short nail (USN), differing
from GSN by the use of two unicortical bridging screws; and 3) bicortical
long nail (BLN), with two angled bicortical and one unicortical bridging
screws. With these findings, five matched-pairs of cadaveric femora
were tested in two groups: 1) unicortical long nail (ULN), with
a longer nail than USN and three bridging unicortical screws; and
2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally
rotated 90°/sec until failure.Objectives
Methods
The monitoring of fracture healing is a complex process. Typically, successive radiographs are performed and an emerging calcification of the fracture area is evaluated. The aim of this study was to investigate whether different bone healing patterns can be distinguished using a telemetric instrumented femoral internal plate fixator. An electronic telemetric system was developed to assess bone healing mechanically. The system consists of a telemetry module which is applied to an internal locking plate fixator, an external reader device, a sensor for measuring externally applied load and a laptop computer with processing software. By correlation between externally applied load and load measured in the implant, the elasticity of the osteosynthesis is calculated. The elasticity decreases with ongoing consolidation of a fracture or nonunion and is an appropriate parameter for the course of bone healing. At our centre, clinical application has been performed in 56 patients suffering nonunion or fracture of the femur.Objectives
Materials and Methods
Fractures of the proximal femur are one of the
greatest challenges facing the medical community, constituting a
heavy socioeconomic burden worldwide. Controversy exists regarding
the optimal treatment for patients with unstable trochanteric proximal
femoral fractures. The recognised treatment alternatives are extramedullary
fixation usually with a sliding hip screw and intramedullary fixation
with a cephalomedullary nail. Current evidence suggests that best
results and lowest complication rates occur using a sliding hip screw.
Complications in these difficult fractures are relatively common
regardless of type of treatment. We believe that a novel device,
the X-Bolt dynamic plating system, may offer superior fixation over
a sliding hip screw with lower reoperation risk and better function.
We therefore propose to investigate the clinical effectiveness of
the X-bolt dynamic plating system compared with standard sliding
hip screw fixation within the framework of a the larger WHiTE (Warwick
Hip Trauma Evaluation) Comprehensive Cohort Study. Cite this article:
The objective of this study was to determine if a synthetic bone
substitute would provide results similar to bone from osteoporotic
femoral heads during Pushout studies were performed with the dynamic hip screw (DHS)
and the DHS Blade in both cadaveric femoral heads and artificial
bone substitutes in the form of polyurethane foam blocks of different
density. The pushout studies were performed as a means of comparing
the force displacement curves produced by each implant within each
material.Introduction
Methods