Objectives. Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis
Objectives. Because posterior cruciate ligament (PCL) resection makes flexion
gaps wider in total knee replacement (TKR), preserving or sacrificing
a PCL affects the gap equivalence; however, there are no criteria
for the PCL resection that consider gap situations of each knee.
This study aims to investigate gap characteristics of knees and
to consider the criteria for PCL resection. Methods. The extension and flexion gaps were measured, first with the
PCL preserved and subsequently with the PCL removed (in cases in
which posterior substitute
Arthrofibrosis is a relatively common complication after joint injuries and surgery, particularly in the knee. The present study used a previously described and validated rabbit model to assess the biomechanical, histopathological, and molecular effects of the mast cell stabilizer ketotifen on surgically induced knee joint contractures in female rabbits. A group of 12 skeletally mature rabbits were randomly divided into two groups. One group received subcutaneous (SQ) saline, and a second group received SQ ketotifen injections. Biomechanical data were collected at eight, ten, 16, and 24 weeks. At the time of necropsy, posterior capsule tissue was collected for histopathological and gene expression analyses (messenger RNA (mRNA) and protein).Aims
Methods
Objectives. Computed tomography (CT) plays an important role in evaluating wear and periacetabular osteolysis (PAO) in total hip replacements. One concern with CT is the high radiation exposure since standard pelvic CT provides approximately 3.5 millisieverts (mSv) of radiation exposure, whereas a planar radiographic examination with three projections totals approximately 0.5 mSv. The objective of this study was to evaluate the lowest acceptable radiation dose for dual-energy CT (DECT) images when measuring wear and periacetabular osteolysis in uncemented metal
Objectives. Wear debris released from bearing surfaces has been shown to
provoke negative immune responses in the recipient. Excessive wear
has been linked to early failure of prostheses. Analysis using coordinate
measuring machines (CMMs) can provide estimates of total volumetric
material loss of explanted prostheses and can help to understand
device failure. The accuracy of volumetric testing has been debated,
with some investigators stating that only protocols involving hundreds
of thousands of measurement points are sufficient. We looked to
examine this assumption and to apply the findings to the clinical
arena. . Methods. We examined the effects on the calculated material loss from
a ceramic femoral head when different CMM scanning parameters were
used. Calculated wear volumes were compared with gold standard gravimetric
tests in a blinded study. . Results. Various scanning parameters including point pitch, maximum point
to point distance, the number of scanning contours or the total
number of points had no clinically relevant effect on volumetric
wear calculations. Gravimetric testing showed that material loss
can be calculated to provide clinically relevant degrees of accuracy. . Conclusions. Prosthetic surfaces can be analysed accurately and rapidly with
currently available technologies. Given these results, we believe
that routine analysis of explanted hip
Wound complications are reported in up to 10% hip and knee arthroplasties and there is a proven association between wound complications and deep prosthetic infections. In this randomised controlled trial (RCT) we explore the potential benefits of a portable, single use, incisional negative pressure wound therapy dressing (iNPWTd) on wound exudate, length of stay (LOS), wound complications, dressing changes and cost-effectiveness following total hip and knee arthroplasties. A total of 220 patients undergoing elective primary total hip and knee arthroplasties were recruited into in a non-blinded RCT. For the final analysis there were 102 patients in the study group and 107 in the control group.Objectives
Methods
Hips with metal-on-metal total hip arthroplasty (MoM THA) have a high rate of adverse local tissue reactions (ALTR), often associated with hypersensitivity reactions. Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) measures tissue perfusion with the parameter Ktrans (volume transfer constant of contrast agent). Our purpose was 1) to evaluate the feasibility of DCE-MRI in patients with THA and 2) to compare DCE-MRI in patients with MoM bearings with metal-on-polyethylene (MoP) bearings, hypothesising that the perfusion index Ktrans in hips with MoM THA is higher than in hips with MoP THA. In this pilot study, 16 patients with primary THA were recruited (eight MoM, eight MoP). DCE-MRI of the hip was performed at 1.5 Tesla (T). For each patient, Ktrans was computed voxel-by-voxel in all tissue lateral to the bladder. The mean Ktrans for all voxels was then calculated. These values were compared with respect to implant type and gender, and further correlated with clinical parameters.Objectives
Methods
Effective analgesia after total knee arthroplasty (TKA) improves
patient satisfaction, mobility and expedites discharge. This study
assessed whether continuous femoral nerve infusion (CFNI) was superior
to a single-shot femoral nerve block in primary TKA surgery completed
under subarachnoid blockade including morphine. We performed an adequately powered, prospective, randomised,
placebo-controlled trial comparing CFNI of 0.125% bupivacaine Objectives
Methods