Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Bone & Joint Research

Research
Dates
Year From

Year To
Bone & Joint Research
Vol. 4, Issue 2 | Pages 23 - 28
1 Feb 2015
Auston DA Werner FW Simpson RB

Objectives. This study tests the biomechanical properties of adjacent locked plate constructs in a femur model using Sawbones. Previous studies have described biomechanical behaviour related to inter-device distances. We hypothesise that a smaller lateral inter-plate distance will result in a biomechanically stronger construct, and that addition of an anterior plate will increase the overall strength of the construct. Methods. Sawbones were plated laterally with two large-fragment locking compression plates with inter-plate distances of 10 mm or 1 mm. Small-fragment locking compression plates of 7-hole, 9-hole, and 11-hole sizes were placed anteriorly to span the inter-plate distance. Four-point bend loading was applied, and the moment required to displace the constructs by 10 mm was recorded. Results . We found that a 1 mm inter-plate distance supported greater moments than a 10 mm distance in constructs with only lateral plates. Moments supported after the addition of a 9- or 11-hole anterior plate were greater for both 10 mm and 1 mm inter-plate distance, with the 11-hole anterior plate supporting a greater moment than a 9-hole plate. Femurs with a 7-hole anterior plate fractured regardless of lateral inter-plate distance size. Conclusion. This suggests that the optimal plate configuration is to minimise lateral inter-plate distance and protect it with an anterior plate longer than seven holes. Cite this article: Bone Joint Res 2015;4:23–8