We assessed the orientation of the acetabular
component in 1070 primary total hip arthroplasties with hard-on-soft, small
diameter bearings, aiming to determine the size and site of the
target zone that optimises outcome. Outcome measures included complications,
dislocations, revisions and ΔOHS (the difference between the Oxford
Hip Scores pre-operatively and five years post-operatively). A wide
scatter of orientation was observed (2 This study demonstrated that with traditional technology surgeons
can only reliably achieve a target zone of ±15°. As the optimal
zone to diminish the risk of dislocation is also ±15°, surgeons
should be able to achieve this. This is the first study to demonstrate
that optimal orientation of the acetabular component improves the
functional outcome. However, the target zone is small (± 5°) and
cannot, with current technology, be consistently achieved. Cite this article:
Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements. We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant. At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 ( The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.
A prospective, multi-centre study was carried out on 1421 total hip replacements between January 1999 and July 2007 to examine if obesity has an effect on clinical outcomes. The patients were categorised into three groups: non-obese (body mass index (BMI) <
30 kg/m2), obese (BMI 30 to 40 kg/m2) and morbidly obese (BMI >
40 kg/m2). The primary outcome measure was the change in Oxford hip score at five years. Secondary outcome measures included dislocation and revision rates, increased haemorrhage, deep infection, deep-vein thrombosis and pulmonary embolism, mean operating time and length of hospital stay. Radiological analysis assessing heterotopic ossification, femoral osteolysis and femoral stem positioning was performed. Data were incomplete for 362 hips (25.5%) There was no difference in the change in the Oxford hip score, complication rates or radiological changes at five years between the groups. The morbidly obese group was significantly younger and required a significantly longer operating time. Obese and morbidly obese patients have as much to gain from total hip replacement as non-obese patients.
Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series. A total of 610 Birmingham Hip Resurfacing arthroplasties were performed in 532 patients with a mean age of 51.8 years (16.5 to 81.6). They were followed for between two and eight years; 107 patients (120 hips) had been followed up for more than five years. Two patients were lost to follow-up. At a minimum of five years’ follow-up, 79 of 85 hips (93%) had an excellent or good outcome according to the Harris hip score. The mean Oxford hip score was 16.1 points ( Considering that these patients are young and active these results are good, and support the use of resurfacing. Further study is needed to address the early failures, particularly those related to fracture and metal debris.
We performed a clinical and radiological study to determine the rate of failure of the Charnley Elite-Plus femoral component. Our aim was to confirm or refute the predictions of a previous roentgen stereophotogrammetric analysis study in which 20% of the Charnley Elite-Plus stems had shown rapid posterior head migration. It was predicted that this device would have a high early rate of failure. We examined 118 patients at a mean of nine years after hip replacement, including the 19 patients from the original roentgen stereophotogrammetric study. The number of revision procedures was recorded and clinical and radiological examinations were performed. The rate of survival of the femoral stems at ten years was 83% when revision alone was considered to be a failure. It decreased to 59% when a radiologically loose stem was also considered to be a failure. All the patients previously shown in the roentgen stereophotogrammetric study to have high posterior head migration went on to failure. There was a highly significant difference (p = 0.002) in posterior head migration measured at two years after operation between failed and non-failed femoral stems, but there was no significant difference in subsidence between these two groups. Our study has shown that the Charnley Elite-Plus femoral component has an unacceptably high rate of failure. It confirms that early evaluation of new components is important and that roentgen stereophotogrammetric is a good tool for this. Our findings have also shown that rapid posterior head migration is predictive of premature loosening and a better predictor than subsidence.
We inserted an electrode up the femoral neck into the femoral head of ten patients undergoing a metal-on-metal hip resurfacing arthroplasty through a posterior surgical approach and measured the oxygen concentration during the operation. In every patient the blood flow was compromised during surgery, but the extent varied. In three patients, the oxygen concentration was zero at the end of the procedure. The surgical approach caused a mean 60% drop (p <
0.005) in oxygen concentration while component insertion led to a further 20% drop (p <
0.04). The oxygen concentration did not improve significantly on wound closure. This study demonstrates that during hip resurfacing arthroplasty, patients experience some compromise to their femoral head blood supply and some have complete disruption.
Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability. Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°,