Advertisement for orthosearch.org.uk
Results 1 - 20 of 21
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 1000 - 1007
1 Sep 2024
Gong T Lu M Sheng H Li Z Zhou Y Luo Y Min L Tu C

Aims

Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes.

Methods

This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 504 - 509
1 Apr 2022
Kennedy JW Farhan-Alanie OM Young D Kelly MP Young PS

Aims

The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases.

Methods

This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1410 - 1416
1 Oct 2013
Gebert C Wessling M Gosheger G Aach M Streitbürger A Henrichs MP Dirksen U Hardes J

To date, all surgical techniques used for reconstruction of the pelvic ring following supra-acetabular tumour resection produce high complication rates. We evaluated the clinical, oncological and functional outcomes of a cohort of 35 patients (15 men and 20 women), including 21 Ewing’s sarcomas, six chondrosarcomas, three sarcomas not otherwise specified, one osteosarcoma, two osseous malignant fibrous histiocytomas, one synovial cell sarcoma and one metastasis. The mean age of the patients was 31 years (8 to 79) and the latest follow-up was carried out at a mean of 46 months (1.9 to 139.5) post-operatively. We undertook a functional reconstruction of the pelvic ring using polyaxial screws and titanium rods. In 31 patients (89%) the construct was encased in antibiotic-impregnated polymethylmethacrylate. Preservation of the extremities was possible for all patients. The survival rate at three years was 93.9% (95% confidence interval (CI) 77.9 to 98.4), at five years it was 82.4% (95% CI 57.6 to 93.4). For the 21 patients with Ewing’s sarcoma it was 95.2% (95% CI 70.7 to 99.3) and 81.5% (95% CI 52.0 to 93.8), respectively. Wound healing problems were observed in eight patients, deep infection in five and clinically asymptomatic breakage of the screws in six. The five-year implant survival was 93.3% (95% CI 57.8 to 95.7). Patients were mobilised at a mean of 3.5 weeks (1 to 7) post-operatively. A post-operative neurological defect occurred in 12 patients. The mean Musculoskeletal Tumor Society score at last available follow-up was 21.2 (10 to 27). This reconstruction technique is characterised by simple and oncologically appropriate applicability, achieving high primary stability that allows early mobilisation, good functional results and relatively low complication rates. Cite this article: Bone Joint J 2013;95-B:1410–16


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 836 - 841
1 Jun 2012
Frisoni T Cevolani L Giorgini A Dozza B Donati DM

We retrospectively reviewed 101 consecutive patients with 114 femoral tumours treated by massive bone allograft at our institution between 1986 and 2005. There were 49 females and 52 males with a mean age of 20 years (4 to 74). At a median follow-up of 9.3 years (2 to 19.8), 36 reconstructions (31.5%) had failed. The allograft itself failed in 27 reconstructions (24%). Mechanical complications such as delayed union, fracture and failure of fixation were studied. The most adverse factor on the outcome was the use of intramedullary nails, followed by post-operative chemotherapy, resection length > 17 cm and age > 18 years at the time of intervention. The simultaneous use of a vascularised fibular graft to protect the allograft from mechanical complications improved the outcome, but the use of intramedullary cementing was not as successful. In order to improve the strength of the reconstruction and to advance the biology of host–graft integration, we suggest avoiding the use of intramedullary nails and titanium plates, but instead using stainless steel plates, as these gave better results. The use of a supplementary vascularised fibular graft should be strongly considered in adult patients with resection > 17 cm and in those who require post-operative chemotherapy


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 398 - 404
1 Feb 2021
Christ AB Fujiwara T Yakoub MA Healey JH

Aims

We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur.

Methods

A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory status, and Musculoskeletal Tumor Society (MSTS) score were recorded. The percentage of proximal femur remaining was calculated from follow-up radiographs.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 2 | Pages 245 - 248
1 Mar 2002
Tsuchiya H Abdel-Wanis ME Uehara K Tomita K Takagi Y Yasutake H

We describe a consecutive series of 26 patients with simple bone cysts who were treated by curettage, multiple drilling and continuous decompression by the insertion of either a cannulated screw or a pin. In the first 15 patients we used titanium cannulated screws (group 1) and in the next 11 a cannulated hydroxyapatite pin (group 2). Satisfactory healing was achieved in 12 patients in group 1 (80%) and in all in group 2. This technique seems to be a promising option for the treatment of simple bone cysts. The cannulated hydroxyapatite pin is recommended because of its higher success rate and the fact that it does not need to be removed


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 267 - 275
1 Feb 2017
Liang H Ji T Zhang Y Wang Y Guo W

Aims

The aims of this retrospective study were to report the feasibility of using 3D-printing technology for patients with a pelvic tumour who underwent reconstruction.

Patients and Methods

A total of 35 patients underwent resection of a pelvic tumour and reconstruction using 3D-printed endoprostheses between September 2013 and December 2015. According to Enneking’s classification of bone defects, there were three Type I lesions, 12 Type II+III lesions, five Type I+II lesions, two Type I+II+III lesions, ten type I+II+IV lesions and three type I+II+III+IV lesions. A total of three patients underwent reconstruction using an iliac prosthesis, 12 using a standard hemipelvic prosthesis and 20 using a screw-rod connected hemipelvic prosthesis.


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1689 - 1695
1 Dec 2017
Stevenson JD Wigley C Burton H Ghezelayagh S Morris G Evans S Parry M Jeys L

Aims

Following the resection of an extensive amount of bone in the treatment of a tumour, the residual segment may be insufficient to accept a standard length intramedullary cemented stem. Short-stemmed endoprostheses conceivably have an increased risk of aseptic loosening. Extra-cortical plates have been added to minimise this risk by supplementing fixation. The aim of this study was to investigate the survivorship of short-stemmed endoprostheses and extra-cortical plates.

Patients and Methods

The study involved 37 patients who underwent limb salvage surgery for a primary neoplasm of bone between 1998 and 2013. Endoprosthetic replacement involved the proximal humerus in nine, the proximal femur in nine, the distal femur in 13 and the proximal tibia in six patients. There were 12 primary (32%) and 25 revision procedures (68%). Implant survivorship was compared with matched controls. The amount of bone that was resected was > 70% of its length and statistically greater than the standard control group at each anatomical site.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 276 - 282
1 Feb 2017
Mumith A Coathup M Chimutengwende-Gordon M Aston W Briggs T Blunn G

Aims

Massive endoprostheses rely on extra-cortical bone bridging (ECBB) to enhance fixation. The aim of this study was to investigate the role of selective laser sintered (SLS) porous collars in augmenting the osseointegration of these prostheses.

Materials and Methods

The two novel designs of porous SLS collars, one with small pores (Ø700 μm, SP) and one with large pores (Ø1500 μm, LP), were compared in an ovine tibial diaphyseal model. Osseointegration of these collars was compared with that of a clinically used solid, grooved design (G). At six months post-operatively, the ovine tibias were retrieved and underwent radiological and histological analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 841 - 848
1 Jun 2017
Hipfl C Stihsen C Puchner SE Kaider A Dominkus M Funovics PT Windhager R

Aims

Pelvic reconstruction after the resection of a tumour around the acetabulum is a challenging procedure due to the complex anatomy and biomechanics. Several pelvic endoprostheses have been introduced, but the rates of complication remain high. Our aim was to review the use of a stemmed acetabular pedestal cup in the management of these patients.

Patients and Methods

The study involved 48 patients who underwent periacetabular reconstruction using a stemmed pedestal cup (Schoellner cup; Zimmer Biomet Inc., Warsaw, Indiana) between 2000 and 2013. The indications for treatment included a primary bone tumour in 27 patients and metastatic disease in 21 patients. The mean age of the patients at the time of surgery was 52 years (16 to 83).


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 252 - 257
1 Feb 2015
Wafa H Grimer RJ Reddy K Jeys L Abudu A Carter SR Tillman RM

We conducted a case-control study to examine the merit of silver-coated tumour prostheses. We reviewed 85 patients with Agluna-treated (silver-coated) tumour implants treated between 2006 and 2011 and matched them with 85 control patients treated between 2001 and 2011 with identical, but uncoated, tumour prostheses.

In all, 106 men and 64 women with a mean age of 42.2 years (18.4 to 90.4) were included in the study. There were 50 primary reconstructions (29.4%); 79 one-stage revisions (46.5%) and 41 two-stage revisions for infection (24.1%).

The overall post-operative infection rate of the silver-coated group was 11.8% compared with 22.4% for the control group (p = 0.033, chi-square test). A total of seven of the ten infected prostheses in the silver-coated group were treated successfully with debridement, antibiotics, and implant retention compared with only six of the 19 patients (31.6%) in the control group (p = 0.048, chi-square test). Three patients in the silver-coated group (3.5%) and 13 controls (15.3%) had chronic periprosthetic infection (p = 0.009, chi-square test).

The overall success rates in controlling infection by two-stage revision in the silver-coated group was 85% (17/20) compared with 57.1% (12/21) in the control group (p = 0.05, chi-square test). The Agluna-treated endoprostheses were associated with a lower rate of early periprosthetic infection. These silver-treated implants were particularly useful in two-stage revisions for infection and in those patients with incidental positive cultures at the time of implantation of the prosthesis.

Debridement with antibiotic treatment and retention of the implant appeared to be more successful with silver-coated implants.

Cite this article: Bone Joint J 2015;97-B:252–7.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1706 - 1712
1 Dec 2014
Bus MPA Boerhout EJ Bramer JAM Dijkstra PDS

Peri-acetabular tumour resections and their subsequent reconstruction are among the most challenging procedures in orthopaedic oncology. Despite the fact that a number of different pelvic endoprostheses have been introduced, rates of complication remain high and long-term results are mostly lacking.

In this retrospective study, we aimed to evaluate the outcome of reconstructing a peri-acetabular defect with a pedestal cup endoprosthesis after a type 2 or type 2/3 internal hemipelvectomy.

A total of 19 patients (11M:8F) with a mean age of 48 years (14 to 72) were included, most of whom had been treated for a primary bone tumour (n = 16) between 2003 and 2009. After a mean follow-up of 39 months (28 days to 8.7 years) seven patients had died. After a mean follow-up of 7.9 years (4.3 to 10.5), 12 patients were alive, of whom 11 were disease-free. Complications occurred in 15 patients. Three had recurrent dislocations and three experienced aseptic loosening. There were no mechanical failures. Infection occurred in nine patients, six of whom required removal of the prosthesis. Two patients underwent hindquarter amputation for local recurrence.

The implant survival rate at five years was 50% for all reasons, and 61% for non-oncological reasons. The mean Musculoskeletal Tumor Society score at final follow-up was 49% (13 to 87).

Based on these poor results, we advise caution if using the pedestal cup for reconstruction of a peri-acetabular tumour resection.

Cite this article: Bone Joint J 2014;96-B:1706–12.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1685 - 1689
1 Dec 2010
Gokaraju K Miles J Parratt MTR Blunn GW Pollock RC Skinner JA Cannon SR Briggs TWR

We have reviewed five adult patients treated with endoprosthetic reconstruction of the proximal radius following resection of non-traumatic lesions. The patients had a mean age of 33.4 years (20 to 60) at the time of surgery and the mean follow-up was 7.6 years (0.8 to 16).

Following surgery, all elbows were clinically stable and there was 100% survivorship of the prosthesis. Evaluation of function was assessed clinically and by the Mayo Elbow Performance Score, achieving a mean of 86% (70 to 100).

Results at medium-term follow-up are encouraging with regards to elbow stability, implant survivorship and functional outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1104 - 1110
1 Aug 2011
Ji T Guo W Yang R Tang S Sun X

We set out to determine the impact of surgery on quality of life and function in patients who had undergone surgery for symptomatic peri-acetabular metastases. From a prospective database we retrospectively reviewed 46 consecutive patients who had been treated operatively between June 2003 and June 2009. The mean age of the patients was 56.4 years (20 to 73) and the mean post-operative follow-up was 19.2 months (4 to 70). Functional evaluation and quality-of-life assessments were performed. At the most recent follow-up, 26 patients (56.5%) were alive. Their median survival time was 25.0 months. Ten major postoperative complications had occurred in eight patients (17.4%). The mean post-operative Musculoskeletal Tumor Society score (MSTS 93) was 56.3% (6.7% to 90.0%). Improvement in the Eastern Cooperative Oncology Group (ECOG) performance status was seen in 32 patients (69.6%). On the European Organisation for Research and Treatment of Cancer core quality-of-life questionnaire (QLQ-C30) measure of global health status there was a statistically significant improvement from the patients’ pre-operative status (42.8 (sd 13.7)) to that found at the latest follow-up (58.0 (sd 12.5)) (p = 0.001). The only statistically significant change in the nine symptom domains of the QLQ-C30 was a reduction in the mean level of pain (from 59.1 to 29.5 (out of 100)) (p < 0.001).

Surgery for patients with peri-acetabular metastases reduces pain and improves their quality of life, and has a low rate of surgical complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 425 - 430
1 Mar 2012
Picardo NE Blunn GW Shekkeris AS Meswania J Aston WJ Pollock RC Skinner JA Cannon SR Briggs TW

In skeletally immature patients, resection of bone tumours and reconstruction of the lower limb often results in leg-length discrepancy. The Stanmore non-invasive extendible endoprosthesis, which uses electromagnetic induction, allows post-operative lengthening without anaesthesia. Between 2002 and 2009, 55 children with a mean age of 11.4 years (5 to 16) underwent reconstruction with this prosthesis; ten patients (18.2%) died of disseminated disease and one child underwent amputation due to infection. We reviewed 44 patients after a mean follow-up of 41.2 months (22 to 104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to 30) and the Toronto Extremity Salvage score was 92.3% (55.2% to 99.0%). There was no local recurrence of tumour. Complications developed in 16 patients (29.1%) and ten (18.2%) underwent revision.

The mean length gained per patient was 38.6 mm (3.5 to 161.5), requiring a mean of 11.3 extensions (1 to 40), and ten component exchanges were performed in nine patients (16.4%) after attaining the maximum lengthening capacity of the implant. There were 11 patients (20%) who were skeletally mature at follow-up, ten of whom had equal leg lengths and nine had a full range of movement of the hip and knee.

This is the largest reported series using non-invasive extendible endoprostheses after excision of primary bone tumours in skeletally immature patients. The technique produces a good functional outcome, with prevention of limb-length discrepancy at skeletal maturity.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 399 - 405
1 Mar 2014
Zang J Guo W Yang Y Xie L

We report our early experience with the use of a new prosthesis, the Modular Hemipelvic Prosthesis II, for reconstruction of the hemipelvis after resection of a primary malignant peri-acetabular tumour involving the sacroiliac joint.

We retrospectively reviewed the outcome of 17 patients who had undergone resection of a pelvic tumour and reconstruction with this prosthesis between July 2002 and July 2010.

One patient had a type I+II+III+IV resection (ilium + peri-acetabulum + pubis/ischium + sacrum) and 16 had a type I+II+IV resection (ilium + acetabulum + sacrum). The outcome was assessed at a mean follow-up of 33 months (15 to 59). One patient was alive with disease, 11 were alive without disease and five had died of disease. The overall five-year survival rate was 62.4%. Six patients had a local recurrence. The mean Musculoskeletal Tumour Society score was 58% (33 to 77). Deep infection occurred in two patients, problems with wound healing in five and dislocation in one.

For patients with a primary malignant peri-acetabular sarcoma involving the sacroiliac joint, we believe that this new prosthesis is a viable option for reconstruction of the bony defect left following resection of the tumour. It results in a satisfactory functional outcome with an acceptable rate of complications.

Cite this article: Bone Joint J 2014;96-B:399–405.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1098 - 1103
1 Aug 2011
Ruggieri P Mavrogenis AF Guerra G Mercuri M

We retrospectively studied 14 patients with proximal and diaphyseal tumours and disappearing bone (Gorham’s) disease of the humerus treated with wide resection and reconstruction using an allograft-resurfacing composite (ARC). There were ten women and four men, with a mean age of 35 years (8 to 69). At a mean follow-up of 25 months (10 to 89), two patients had a fracture of the allograft. In one of these it was revised with a similar ARC and in the other with an intercalary prosthesis. A further patient had an infection and a fracture of the allograft that was revised with a megaprosthesis. In all patients with an ARC, healing of the ARC-host bone interface was observed. One patient had failure of the locking mechanism of the total elbow replacement. The mean post-operative Musculoskeletal Tumor Society score for the upper extremity was 77% (46.7% to 86.7%), which represents good and excellent results; one patient had a poor result (46.7%).

In the short term ARC effectively relieves pain and restores shoulder function in patients with wide resection of the proximal humerus. Fracture and infection remain significant complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 265 - 269
1 Feb 2012
Hwang N Grimer RJ Carter SR Tillman RM Abudu A Jeys LM

We reviewed our initial seven-year experience with a non-invasive extendible prosthesis in 34 children with primary bone tumours. The distal femur was replaced in 25 cases, total femur in five, proximal femur in one and proximal tibia in three. The mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain alive. The prostheses were lengthened by an electromagnetic induction mechanism in an outpatient setting and a mean extension of 32 mm (4 to 80) was achieved without anaesthesia. There were lengthening complications in two children: failed lengthening in one and the formation of scar tissue in the other. Deep infection developed in six patients (18%) and local recurrence in three. A total of 11 patients required further surgery to the leg. Amputation was necessary in five patients (20%) and a two-stage revision in another. There were no cases of loosening, but two patients had implant breakage and required revision. The mean Musculoskeletal Tumor Society functional score was 85% (60% to 100%) at last known follow-up. These early results demonstrate that the non-invasive extendible prosthesis allows successful lengthening without surgical intervention, but the high incidence of infection is a cause for concern.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1367 - 1372
1 Oct 2006
Gupta A Pollock R Cannon SR Briggs TWR Skinner J Blunn G

We used a knee-sparing distal femoral endoprosthesis in young patients with malignant bone tumours of the distal femur in whom it was possible to resect the tumour and to preserve the distal femoral condyles. The proximal shaft of the endoprosthesis had a coated hydroxyapatite collar, while the distal end had hydroxyapatite-coated extracortical plates to secure it to the small residual femoral condylar fragment. We reviewed the preliminary results of this endoprosthesis in eight patients with primary bone tumours of the distal femur. Their mean age at surgery was 17.years (14 to 21). The mean follow-up was 24 months (20 to 31). At final follow-up the mean flexion at the knee was 102° (20° to 120°) and the mean Musculoskeletal Tumour Society score was 80% (57% to 96.7%).

There was excellent osteointegration at the prosthesis-proximal bone interface with formation of new bone around the hydroxyapatite collar. The prosthesis allowed preservation of the knee and achieved a good functional result. Formation of new bone and remodelling at the interface make the implant more secure. Further follow-up is required to determine the long-term structural integrity of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1487 - 1491
1 Nov 2006
Ahlmann ER Menendez LR

Custom-made intercalary endoprostheses may be used for the reconstruction of diaphyseal defects following the resection of bone tumours. The aim of this study was to determine the survival of intercalary endoprostheses with a lap joint design, and to evaluate the clinical results, complications and functional outcome. We retrospectively reviewed six consecutive patients, three of whom underwent limb salvage with intercalary endoprostheses of the tibia, two of the femur, and one of the humerus. Their mean age was 42 years (28 to 64). The mean follow-up was 21.6 months (9 to 58). The humeral prosthesis required revision at 14 months owing to aseptic loosening. There were no implant-related failures. Musculoskeletal Tumour Society functional outcome scores indicated that patients achieved 90% of premorbid function.

Custom intercalary endoprostheses result in reconstructions comparable with, if not better than, those of allografts. Using this design of implant reduces the incidence of early complications and difficulties experienced with previous versions.