We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless,
The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach. A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.Aims
Methods
Cemented femoral stems with force closed fixation
designs have shown good clinical results despite high early subsidence.
A new triple-tapered stem in this category (C-stem AMT) was introduced
in 2005. This study compares this new stem with an established stem
of similar design (Exeter) in terms of migration (as measured using radiostereometric
analysis), peri-prosthetic bone remodelling (measured using dual
energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years
(53 to 78) were followed for two years. Owing to missing data of
miscellaneous reasons, the final analysis represents data from 51
(RSA) and 65 (DXA) patients. Both stems showed a typical pattern
of migration: Subsidence and retroversion that primarily occurred
during the first three months. C-stem AMT subsided less during the
first three months (p = 0.01), before stabilising at a subsidence rate
similar to the Exeter stem from years one to two. The rate of migration
into retroversion was slightly higher for C-stem AMT during the
second year (p = 0.03). Whilst there were slight differences in
movement patterns between the stems, the C-stem AMT exhibits good
early clinical outcomes and displays a pattern of migration and
bone remodelling that predicts good clinical performance. Cite this article:
Aims. The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck. Patients and Methods. From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan–Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis. Results. The one-year survival was 96.0% (95% confidence interval (CI) 95.6 to 96.4) for the Exeter stem, 97.0% (95% CI 96.4 to 97.6) for the Lubinus SP2 stem, 97.6% (95% CI 97.0 to 98.2) for the Charnley stem, 98.1% (95% CI 97.3 to 98.9) for the Spectron EF stem, and 96.4% (95% CI 95.6 to 97.2) for the Charnley Modular stem, respectively. The hazard ratio for reoperation after one year was lower for Lubinus SP2 (HR 0.77, 95% CI 0.60 to 0.97), Charnley (HR 0.64, 95% CI 0.48 to 0.86), and Spectron EF stems (HR 0.44, 95% CI 0.29 to 0.67) compared with the Exeter stem. Reoperation for periprosthetic fracture occurred almost exclusively after the use of
Aims. The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other
Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty. Patients recorded by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) between 1 September 1999 and 31 December 2020 who underwent hemiarthroplasty for the treatment of a hip fracture with the following implants were included: a cemented monoblock Exeter Trauma Stem (ETS), cemented Exeter V40 with a bipolar head, a monoblock Thompsons prosthesis (Cobalt/Chromium or Titanium), and an Exeter V40 with a Unitrax head. Overall and age-defined cumulative revision rates were compared over the ten years following surgery.Aims
Methods
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique. We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.Aims
Methods
Despite few good-quality studies on the subject, total hip arthroplasty (THA) is increasingly being performed for displaced intracapsular fractures of the neck of femur. We compared outcomes of all patients with displacement of these fractures treated surgically over a ten-year period in one institution. A total of 2,721 patients with intracapsular fractures of the femoral neck treated with either a cemented hemiarthroplasty or a THA at a single centre were retrospectively reviewed. The primary outcomes analyzed were readmission for any reason and revision surgery. We secondarily looked at mortality rates.Aims
Methods
There is a North Atlantic divide, with cementless
femoral stems being used more frequently in the USA and cemented
stems being used more frequently in many countries in Europe. This
is primarily because different cemented stems have been used on
different sides of the Atlantic and the results of the cemented
stems in the US have often been poor, whereas the results of the
stems used in Europe have been good. In the National registers in
Europe, cemented stems have tended to achieve better results than
cementless. Cite this article:
Vancouver type B periprosthetic femoral fractures (PFF) are challenging complications after total hip arthroplasty (THA), and some treatment controversies remain. The objectives of this study were: to evaluate the short-to-mid-term clinical outcomes after treatment of Vancouver type B PFF and to compare postoperative outcome in subgroups according to classifications and treatments; to report the clinical outcomes after conservative treatment; and to identify risk factors for postoperative complications in Vancouver type B PFF. A total of 97 consecutive PPFs (49 males and 48 females) were included with a mean age of 66 years (standard deviation (SD) 14.9). Of these, 86 patients were treated with surgery and 11 were treated conservatively. All living patients had a minimum two-year follow-up. Patient demographics details, fracture healing, functional scores, and complications were assessed. Clinical outcomes between internal fixation and revisions in patients with or without a stable femoral component were compared. Conservatively treated PPFs were evaluated in terms of mortality and healing status. A logistic regression analysis was performed to identify risk factors for complications.Aims
Methods
We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p <
0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p <
0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.
Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability. Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°,
The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.
This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless,
We determined the midterm survival, incidence
of peri-prosthetic fracture and the enhancement of the width of
the femur when combining struts and impacted bone allografts in
24 patients (25 hips) with severe femoral bone loss who underwent
revision hip surgery. The pre-operative diagnosis was aseptic loosening
in 16 hips, second-stage reconstruction in seven, peri-prosthetic
fracture in one and stem fracture in one hip. A total of 14 hips
presented with an Endoklinik grade 4 defect and 11 hips a grade
3 defect. The mean pre-operative Merle D’Aubigné and Postel score
was 5.5 points (1 to 8). The survivorship was 96% (95% confidence interval 72 to 98) at
a mean of 54.5 months (36 to 109). The mean functional score was
17.3 points (16 to 18). One patient in which the strut did not completely
bypass the femoral defect was further revised using a long cemented
stem due to peri-prosthetic fracture at six months post-operatively.
The mean subsidence of the stem was 1.6 mm (1 to 3). There was no
evidence of osteolysis, resorption or radiolucencies during follow-up
in any hip. Femoral width was enhanced by a mean of 41% (19% to
82%). A total of 24 hips had partial or complete bridging of the
strut allografts. This combined biological method was associated with a favourable
survivorship, a low incidence of peri-prosthetic fracture and enhancement
of the width of the femur in revision total hip replacement in patients
with severe proximal femoral bone loss.
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods