The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 473 - 477
1 Apr 2015
Bone MC Dold P Flohr M Preuss R Joyce TJ Aspden RM Holland J Deehan D

Concerns have been raised that deformation of acetabular shells may disrupt the assembly process of modular prostheses. In this study we aimed to examine the effect that the strength of bone has on the amount of deformation of the acetabular shell. The hypothesis was that stronger bone would result in greater deformation. A total of 17 acetabular shells were inserted into the acetabula of eight cadavers, and deformation was measured using an optical measuring system. Cores of bone from the femoral head were taken from each cadaver and compressed using a materials testing machine. The highest peak modulus and yield stress for each cadaver were used to represent the strength of the bone and compared with the values for the deformation and the surgeon’s subjective assessment of the hardness of the bone. The mean deformation of the shell was 129 µm (3 to 340). No correlation was found between deformation and either the maximum peak modulus (r² = 0.011, t = 0.426, p = 0.676) or the yield stress (r² = 0.024, t = 0.614, p = 0.549) of the bone. Although no correlation was found between the strength of the bone and deformation, the values for the deformation observed could be sufficient to disrupt the assembly process of modular acetabular components. Cite this article: Bone Joint J 2015; 97-B:473–7


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 261 - 268
1 Mar 2023
Ruhr M Huber G Niki Y Lohner L Ondruschka B Morlock MM

Aims. The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure. Methods. Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans. Results. Overimpaction reduced primary acetabular component stability (p = 0.038) but did not significantly increase strain release after implantation (p = 0.117) or plastic deformations (p = 0.193). Higher press-fits were associated with larger polar gaps for the 1 Hz reference impaction (p = 0.002, R. 2. = 0.77), with a similar trend for overimpaction (p = 0.082, R. 2. = 0.31). High-frequency impaction did not significantly increase primary stability (p = 0.170) at lower impaction forces (p = 0.001); it was associated with smaller plastic deformations (p = 0.035, R. 2. = 0.34) and a trend for increased acetabular component relaxation between strokes (p = 0.112). Higher press-fit was not related to larger polar gaps for the 6 Hz impaction (p = 0.346). Conclusion. Overimpaction of press-fit acetabular components should be prevented since additional strokes can be associated with increased bone damage and reduced primary stability as shown in this study. High-frequency impaction at 6 Hz was shown to be beneficial compared with 1 Hz impaction. This benefit has to be confirmed in clinical studies. Cite this article: Bone Joint J 2023;105-B(3):261–268


The Bone & Joint Journal
Vol. 101-B, Issue 12 | Pages 1585 - 1592
1 Dec 2019
Logishetty K Rudran B Cobb JP

Aims. Arthroplasty skills need to be acquired safely during training, yet operative experience is increasingly hard to acquire by trainees. Virtual reality (VR) training using headsets and motion-tracked controllers can simulate complex open procedures in a fully immersive operating theatre. The present study aimed to determine if trainees trained using VR perform better than those using conventional preparation for performing total hip arthroplasty (THA). Patients and Methods. A total of 24 surgical trainees (seven female, 17 male; mean age 29 years (28 to 31)) volunteered to participate in this observer-blinded 1:1 randomized controlled trial. They had no prior experience of anterior approach THA. Of these 24 trainees, 12 completed a six-week VR training programme in a simulation laboratory, while the other 12 received only conventional preparatory materials for learning THA. All trainees then performed a cadaveric THA, assessed independently by two hip surgeons. The primary outcome was technical and non-technical surgical performance measured by a THA-specific procedure-based assessment (PBA). Secondary outcomes were step completion measured by a task-specific checklist, error in acetabular component orientation, and procedure duration. Results. VR-trained surgeons performed at a higher level than controls, with a median PBA of Level 3a (procedure performed with minimal guidance or intervention) versus Level 2a (guidance required for most/all of the procedure or part performed). VR-trained surgeons completed 33% more key steps than controls (mean 22 (. sd. 3) vs 12 (. sd. 3)), were 12° more accurate in component orientation (mean error 4° (. sd. 6°) vs 16° (. sd. 17°)), and were 18% faster (mean 42 minutes (. sd. 7) vs 51 minutes (. sd. 9)). Conclusion. Procedural knowledge and psychomotor skills for THA learned in VR were transferred to cadaveric performance. Basic preparatory materials had limited value for trainees learning a new technique. VR training advanced trainees further up the learning curve, enabling highly precise component orientation and more efficient surgery. VR could augment traditional surgical training to improve how surgeons learn complex open procedures. Cite this article: Bone Joint J 2019;101-B:1585–1592


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 712 - 719
1 Jun 2018
Batailler C Weidner J Wyatt M Dalmay F Beck M

Aims. The primary aim of this study was to define and quantify three new measurements to indicate the position of the greater trochanter. Secondary aims were to define ‘functional antetorsion’ as it relates to abductor function in populations both with and without torsional abnormality. Patients and Methods. Three new measurements, functional antetorsion, posterior tilt, and posterior translation of the greater trochanter, were assessed from 61 CT scans of cadaveric femurs, and their reliability determined. These measurements and their relationships were also evaluated in three groups of patients: a control group (n = 22), a ‘high-antetorsion’ group (n = 22) and a ‘low-antetorsion’ group (n = 10). Results. In the cadaver group, the mean anatomical antetorsion was 14.7° (. sd. 8.5; 0 to 36.5) and the functional antetorsion 21.5° (. sd. 8.1; 3.6 to 44.3): the posterior tilt was 73.3° (. sd. 10.8; 46.9 to 88.7) and the posterior translation 0.59 (. sd. 0.2; 0.2 to 0.9). These measurements had excellent intra and interobserver agreement with a range from 0.93 to 0.99. When the anatomical antetorsion decreased, the greater trochanter was more tilted and translated posteriorly in relation to the axis of the femoral neck, and the difference between functional and anatomical antetorsion increased. The results the three patient groups were similar to those of the cadaver group. Conclusion. The position of the greater trochanter and functional antetorsion varied with anatomical antetorsion. In the surgical management of femoral retrotorsion, subtrochanteric osteotomy can result in an excessively posterior position of the greater trochanter and an increase in functional antetorsion. Cite this article: Bone Joint J 2018;100-B:712–19


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 898 - 906
1 Sep 2024
Kayani B Wazir MUK Mancino F Plastow R Haddad FS

Aims

The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system.

Methods

This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 426 - 434
1 Apr 2019
Logishetty K van Arkel RJ Ng KCG Muirhead-Allwood SK Cobb JP Jeffers JRT

Aims. The hip’s capsular ligaments passively restrain extreme range of movement (ROM) by wrapping around the native femoral head/neck. We determined the effect of hip resurfacing arthroplasty (HRA), dual-mobility total hip arthroplasty (DM-THA), conventional THA, and surgical approach on ligament function. Materials and Methods. Eight paired cadaveric hip joints were skeletonized but retained the hip capsule. Capsular ROM restraint during controlled internal rotation (IR) and external rotation (ER) was measured before and after HRA, DM-THA, and conventional THA, with a posterior (right hips) and anterior capsulotomy (left hips). Results. Hip resurfacing provided a near-native ROM with between 5° to 17° increase in IR/ER ROM compared with the native hip for the different positions tested, which was a 9% to 33% increase. DM-THA generated a 9° to 61° (18% to 121%) increase in ROM. Conventional THA generated a 52° to 100° (94% to 199%) increase in ROM. Thus, for conventional THA, the capsule function that exerts a limit on ROM is lost. It is restored to some extent by DM-THA, and almost fully restored by hip resurfacing. In positions of low flexion/extension, the posterior capsulotomy provided more normal function than the anterior, possibly because the capsule was shortened during posterior repair. However, in deep flexion positions, the anterior capsulotomy functioned better. Conclusion. Native head-size and capsular repair preserves capsular function after arthroplasty. The anterior and posterior approach differentially affect postoperative biomechanical function of the capsular ligaments. Cite this article: Bone Joint J 2019;101-B:426–434


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1252 - 1258
1 Dec 2023
Tanabe H Baba T Ozaki Y Yanagisawa N Homma Y Nagao M Kaneko K Ishijima M

Aims

Lateral femoral cutaneous nerve (LFCN) injury is a potential complication after the direct anterior approach for total hip arthroplasty (DAA-THA). The aim of this study was to determine how the location of the fasciotomy in DAA-THA affects LFCN injury.

Methods

In this trial, 134 patients were randomized into a lateral fasciotomy (n = 67) or a conventional fasciotomy (n = 67) group. This study was a dual-centre, double-blind, prospective randomized controlled two-arm trial with parallel group design and a 1:1 allocation ratio. The primary endpoint was the presence of LFCN injury, which was determined by the presence of numbness, decreased sensation, tingling, jolt-like sensation, or pain over the lateral aspect of the thigh, excluding the surgical scar, using a patient-based questionnaire. The secondary endpoints were patient-reported outcome measures (PROMs) using the Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ), and the Forgotten Joint Score-12 (FJS-12). Assessments were obtained three months after surgery.


The Bone & Joint Journal
Vol. 104-B, Issue 11 | Pages 1196 - 1201
1 Nov 2022
Anderson CG Brilliant ZR Jang SJ Sokrab R Mayman DJ Vigdorchik JM Sculco PK Jerabek SA

Aims

Although CT is considered the benchmark to measure femoral version, 3D biplanar radiography (hipEOS) has recently emerged as a possible alternative with reduced exposure to ionizing radiation and shorter examination time. The aim of our study was to evaluate femoral stem version in postoperative total hip arthroplasty (THA) patients and compare the accuracy of hipEOS to CT. We hypothesize that there will be no significant difference in calculated femoral stem version measurements between the two imaging methods.

Methods

In this study, 45 patients who underwent THA between February 2016 and February 2020 and had both a postoperative CT and EOS scan were included for evaluation. A fellowship-trained musculoskeletal radiologist and radiological technician measured femoral version for CT and 3D EOS, respectively. Comparison of values for each imaging modality were assessed for statistical significance.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 432 - 439
1 Apr 2017
Weinberg DS Williamson DFK Millis MB Liu RW

Aims. Recently, there has been considerable interest in quantifying the associations between bony abnormalities around and in the hip joint and osteoarthritis (OA). Our aim was to investigate the relationships between acetabular undercoverage, acetabular overcoverage, and femoroacetabular impingement (FAI) with OA of the hip, which currently remain controversial. . Materials and Methods. A total of 545 cadaveric skeletons (1090 hips) from the Hamann-Todd osteological collection were obtained. Femoral head volume (FHV), acetabular volume (AV), the FHV/AV ratio, acetabular version, alpha angle and anterior femoral neck offset (AFNO) were measured. A validated grading system was used to quantify OA of the hip as minimal, moderate, or severe. Multiple linear and multinomial logistic regression were used to determine the factors that correlated independently with the FHV, AV, and the FHV/AV ratio. . Results. Female cadavers had smaller FHVs (standardised beta -0.382, p < 0.001), and AVs (standardised beta -0.351, p < 0.001), compared with male patients, although the FHV/AV ratio was unchanged. Every 1° increase in alpha angle increased the probability of having moderate OA of the hip compared with minimal OA by 7.1%. Every 1 mm decrease in AFNO increased the probability of having severe or moderate OA of the hip, compared with minimal OA, by 11% and 9%, respectively. The relative risk ratios of having severe OA of the hip compared with minimal OA were 7.2 and 3.3 times greater for acetabular undercoverage and overcoverage, respectively, relative to normal acetabular cover. . Conclusion . Acetabular undercoverage and overcoverage were independent predictors of increased OA of the hip. The alpha angle and AFNO had modest effects, supporting the hypothesis that bony abnormalities both in acetabular dysplasia and FAI are associated with severe OA. Cite this article: Bone Joint J 2017;99-B:432–9


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 336 - 343
1 Apr 2024
Haertlé M Becker N Windhagen H Ahmad SS

Aims

Periacetabular osteotomy (PAO) is widely recognized as a demanding surgical procedure for acetabular reorientation. Reports about the learning curve have primarily focused on complication rates during the initial learning phase. Therefore, our aim was to assess the PAO learning curve from an analytical perspective by determining the number of PAOs required for the duration of surgery to plateau and the accuracy to improve.

Methods

The study included 118 consecutive PAOs in 106 patients. Of these, 28 were male (23.7%) and 90 were female (76.3%). The primary endpoint was surgical time. Secondary outcome measures included radiological parameters. Cumulative summation analysis was used to determine changes in surgical duration. A multivariate linear regression model was used to identify independent factors influencing surgical time.


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1056 - 1062
1 Aug 2015
Kanawati AJ Narulla RS Lorentzos P Facchetti G Smith A Stewart F

The aim of this cadaver study was to identify the change in position of the sciatic nerve during arthroplasty using the posterior surgical approach to the hip. We investigated the position of the nerve during this procedure by dissecting 11 formalin-treated cadavers (22 hips: 12 male, ten female). The distance between the sciatic nerve and the femoral neck was measured before and after dislocation of the hip, and in positions used during the preparation of the femur. The nerve moves closer to the femoral neck when the hip is flexed to > 30° and internally rotated to 90° (90° IR). The mean distance between the nerve and femoral neck was 43.1 mm (standard deviation (. sd. ) 8.7) with the hip at 0° of flexion and 90° IR; this significantly decreased to a mean of 36.1 mm (. sd. 9.5), 28.8 mm (. sd. 9.8) and 19.1 mm (. sd. 9.7) at 30°, 60° and 90° of hip flexion respectively (p < 0.001). In two hips the nerve was in contact with the femoral neck when the hip was flexed to 90°. . This study demonstrates that the sciatic nerve becomes closer to the operative field during hip arthroplasty using the posterior approach with progressive flexion of the hip. Cite this article: Bone Joint J 2015;97-B:1056–62


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 779 - 786
1 Jul 2019
Lamb JN Baetz J Messer-Hannemann P Adekanmbi I van Duren BH Redmond A West RM Morlock MM Pandit HG

Aims. The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model. Materials and Methods. A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an in vitro model using paired fresh frozen cadaveric femora. Results. The prevalence of early revision for periprosthetic fracture was 0.34% (1180/337 647) and 44.0% (520/1180) occurred within 90 days of surgery. Implant risk factors included: collarless stem, non-grit-blasted finish, and triple-tapered design. In the in vitro model, a medial calcar collar consistently improved the stability and resistance to fracture. Conclusion. Analysis of features of stem design in registry data is a useful method of identifying implant characteristics that affect the risk of early periprosthetic fracture around a cementless femoral stem. A collar on the calcar reduced the risk of an early periprosthetic fracture and this was confirmed by biomechanical testing. This approach may be useful in the analysis of other uncommon modes of failure after THA. Cite this article: Bone Joint J 2019;101-B:779–786


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1281 - 1286
1 Oct 2009
Olsen M Davis ET Chiu M Gamble P Tumia N Boyle RA Schemitsch EH

The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1036 - 1042
1 Aug 2016
Amanatullah DF Masini MA Roger DJ Pagnano MW

Aims. We wished to quantify the extent of soft-tissue damage sustained during minimally invasive total hip arthroplasty through the direct anterior (DA) and direct superior (DS) approaches. Materials and Methods. In eight cadavers, the DA approach was performed on one side, and the DS approach on the other, a single brand of uncemented hip prosthesis was implanted by two surgeons, considered expert in their surgical approaches. Subsequent reflection of the gluteus maximus allowed the extent of muscle and tendon damage to be measured and the percentage damage to each anatomical structure to be calculated. Results. The DA approach caused substantially greater damage to the gluteus minimus muscle and tendon when compared with the DS approach (t-test, p = 0.049 and 0.003, respectively). The tensor fascia lata and rectus femoris muscles were damaged only in the DA approach. There was no difference in the amount of damage to the gluteus medius muscle and tendon, piriformis tendon, obturator internus tendon, obturator externus tendon or quadratus femoris muscle between approaches. The posterior soft-tissue releases of the DA approach damaged the gluteus minimus muscle and tendon, piriformis tendon and obturator internus tendon. Conclusion. The DS approach caused less soft-tissue damage than the DA approach. However the clinical relevance is unknown. Further clinical outcome studies, radiographic evaluation of component position, gait analyses and serum biomarker levels are necessary to evaluate and corroborate the safety and efficacy of the DS approach. Cite this article: Bone Joint J 2016;98-B1036–42


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1582 - 1588
1 Dec 2016
Dewar DC Lazaro LE Klinger CE Sculco PK Dyke JP Ni AY Helfet DL Lorich DG

Aims. We aimed to quantify the relative contributions of the medial femoral circumflex artery (MFCA) and lateral femoral circumflex artery (LFCA) to the arterial supply of the head and neck of the femur. Materials and Methods. We acquired ten cadaveric pelvises. In each of these, one hip was randomly assigned as experimental and the other as a matched control. The MFCA and LFCA were cannulated bilaterally. The hips were designated LFCA-experimental or MFCA-experimental and underwent quantitative MRI using a 2 mm slice thickness before and after injection of MRI-contrast diluted 3:1 with saline (15 ml Gd-DTPA) into either the LFCA or MFCA. The contralateral control hips had 15 ml of contrast solution injected into the root of each artery. Next, the MFCA and LFCA were injected with a mixture of polyurethane and barium sulfate (33%) and their extra-and intra-arterial course identified by CT imaging and dissection. Results. The MFCA made a greater contribution than the LFCA to the vascularity of the femoral head (MFCA 82%, LFCA 18%) and neck (MFCA 67%, LFCA 33%). However, the LFCA supplied 48% of the anteroinferior femoral neck overall. Conclusion. This study clearly shows that the MFCA is the major arterial supply to the femoral head and neck. Despite this, the LFCA supplies almost half the anteroinferior aspect of the femoral neck. Cite this article: Bone Joint J 2016;98-B:1582–8


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 168 - 172
1 Feb 2006
Mayr E de la Barrera JM Eller G Bach C Nogler M

In navigated total hip arthroplasty, the pelvis and the femur are tracked by means of rigid bodies fixed directly to the bones. Exact tracking throughout the procedure requires that the connection between the marker and bone remains stable in terms of translation and rotation. We carried out a cadaver study to compare the intra-operative stability of markers consisting of an anchoring screw with a rotational stabiliser and of pairs of pins and wires of different diameters connected with clamps. These devices were tested at different locations in the femur. Three human cadavers were placed supine on an operating table, with a reference marker positioned in the area of the greater trochanter. K-wires (3.2 mm), Steinman pins (3 and 4 mm), Apex pins (3 and 4 mm), and a standard screw were used as fixation devices. They were positioned medially in the proximal third of the femur, ventrally in the middle third and laterally in the distal portion. In six different positions of the leg, the spatial positions were recorded with a navigation system. Compared with the standard single screw, with the exception of the 3 mm Apex pins, the two-pin systems were associated with less movement of the marker and could be inserted less invasively. With the knee flexed to 90° and the dislocated hip rotated externally until the lower leg was parallel to the table (figure-four position), all the anchoring devices showed substantial deflection of 1.5° to 2.5°. The most secure area for anchoring markers was the lateral aspect of the femur


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 193 - 199
1 Feb 2022
Wang Q Wang H A G Xiao T Kang P

Aims

This study aimed to use intraoperative free electromyography to examine how the placement of a retractor at different positions along the anterior acetabular wall may affect the femoral nerve during total hip arthroplasty (THA) when undertaken using the direct anterior approach (THA-DAA).

Methods

Intraoperative free electromyography was performed during primary THA-DAA in 82 patients (94 hips). The highest position of the anterior acetabular wall was defined as the “12 o’clock” position (middle position) when the patient was in supine position. After exposure of the acetabulum, a retractor was sequentially placed at the ten, 11, 12, one, and two o’clock positions (right hip; from superior to inferior positions). Action potentials in the femoral nerve were monitored with each placement, and the incidence of positive reactions (defined as explosive, frequent, or continuous action potentials, indicating that the nerve was being compressed) were recorded as the primary outcome. Secondary outcomes included the incidence of positive reactions caused by removing the femoral head, and by placing a retractor during femoral exposure; and the incidence of femoral nerve palsy, as detected using manual testing of the strength of the quadriceps muscle.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1204 - 1213
1 Sep 2015
Lazaro LE Klinger CE Sculco PK Helfet DL Lorich DG

This study investigates and defines the topographic anatomy of the medial femoral circumflex artery (MFCA) terminal branches supplying the femoral head (FH). Gross dissection of 14 fresh–frozen cadaveric hips was undertaken to determine the extra and intracapsular course of the MFCA’s terminal branches. A constant branch arising from the transverse MFCA (inferior retinacular artery; IRA) penetrates the capsule at the level of the anteroinferior neck, then courses obliquely within the fibrous prolongation of the capsule wall (inferior retinacula of Weitbrecht), elevated from the neck, to the posteroinferior femoral head–neck junction. This vessel has a mean of five (three to nine) terminal branches, of which the majority penetrate posteriorly. Branches from the ascending MFCA entered the femoral capsular attachment posteriorly, running deep to the synovium, through the neck, and terminating in two branches. The deep MFCA penetrates the posterosuperior femoral capsular. Once intracapsular, it divides into a mean of six (four to nine) terminal branches running deep to the synovium, within the superior retinacula of Weitbrecht of which 80% are posterior. Our study defines the exact anatomical location of the vessels, arising from the MFCA and supplying the FH. The IRA is in an elevated position from the femoral neck and may be protected from injury during fracture of the femoral neck. We present vascular ‘danger zones’ that may help avoid iatrogenic vascular injury during surgical interventions about the hip. Cite this article: Bone Joint J 2015;97-B:1204–13


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 484 - 491
1 Apr 2015
van Arkel RJ Amis AA Cobb JP Jeffers JRT

In this in vitro study of the hip joint we examined which soft tissues act as primary and secondary passive rotational restraints when the hip joint is functionally loaded. A total of nine cadaveric left hips were mounted in a testing rig that allowed the application of forces, torques and rotations in all six degrees of freedom. The hip was rotated throughout a complete range of movement (ROM) and the contributions of the iliofemoral (medial and lateral arms), pubofemoral and ischiofemoral ligaments and the ligamentum teres to rotational restraint was determined by resecting a ligament and measuring the reduced torque required to achieve the same angular position as before resection. The contribution from the acetabular labrum was also measured. Each of the capsular ligaments acted as the primary hip rotation restraint somewhere within the complete ROM, and the ligamentum teres acted as a secondary restraint in high flexion, adduction and external rotation. The iliofemoral lateral arm and the ischiofemoral ligaments were primary restraints in two-thirds of the positions tested. Appreciation of the importance of these structures in preventing excessive hip rotation and subsequent impingement/instability may be relevant for surgeons undertaking both hip joint preserving surgery and hip arthroplasty. Cite this article: Bone Joint J 2015; 97-B:484–91


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 616 - 621
1 May 2011
O’Neill F Condon F McGloughlin T Lenehan B Coffey JC Walsh M

We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study