The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1118 - 1125
4 Oct 2022
Suda Y Hiranaka T Kamenaga T Koide M Fujishiro T Okamoto K Matsumoto T

Aims

A fracture of the medial tibial plateau is a serious complication of Oxford mobile-bearing unicompartmental knee arthroplasty (OUKA). The risk of these fractures is reportedly lower when using components with a longer keel-cortex distance (KCDs). The aim of this study was to examine how slight varus placement of the tibial component might affect the KCDs, and the rate of tibial plateau fracture, in a clinical setting.

Methods

This retrospective study included 255 patients who underwent 305 OUKAs with cementless tibial components. There were 52 males and 203 females. Their mean age was 73.1 years (47 to 91), and the mean follow-up was 1.9 years (1.0 to 2.0). In 217 knees in 187 patients in the conventional group, tibial cuts were made orthogonally to the tibial axis. The varus group included 88 knees in 68 patients, and tibial cuts were made slightly varus using a new osteotomy guide. Anterior and posterior KCDs and the origins of fracture lines were assessed using 3D CT scans one week postoperatively. The KCDs and rate of fracture were compared between the two groups.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1637 - 1640
1 Dec 2012
Clark DA Upadhyay N Gillespie G Wakeley C Eldridge JD

Ensuring correct rotation of the femoral component is a challenging aspect of patellofemoral replacement surgery. Rotation equal to the epicondylar axis or marginally more external rotation is acceptable. Internal rotation is associated with poor outcomes. This paper comprises two studies evaluating the use of the medial malleolus as a landmark to guide rotation. We used 100 lower-leg anteroposterior radiographs to evaluate the reliability of the medial malleolus as a landmark. Assessment was made of the angle between the tibial shaft and a line from the intramedullary rod entry site to the medial malleolus. The femoral cut was made in ten cadaver knees using the inferior tip of the medial malleolus as a landmark for rotation. Rotation of the cut relative to the anatomical epicondylar axis was assessed using CT. The study of radiographs found the position of the medial malleolus relative to the tibial axis is consistent. Using the inferior tip of the medial malleolus in the cadaver study produced a mean external rotation of 1.6° (0.1° to 3.7°) from the anatomical epicondylar axis. Using the inferior tip of the medial malleolus to guide the femoral cutting jig avoids internal rotation and introduces an acceptable amount of external rotation of the femoral component


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1686 - 1694
1 Nov 2021
Yang H Kwak W Kang SJ Song E Seon J

Aims

To determine the relationship between articular cartilage status and clinical outcomes after medial opening-wedge high tibial osteotomy (MOHTO) for medial compartmental knee osteoarthritis at intermediate follow-up.

Methods

We reviewed 155 patients (155 knees) who underwent MOHTO from January 2008 to December 2016 followed by second-look arthroscopy with a mean 5.3-year follow-up (2.0 to 11.7). Arthroscopic findings were assessed according to the International Cartilage Repair Society (ICRS) Cartilage Repair Assessment (CRA) grading system. Patients were divided into two groups based on the presence of normal or nearly normal quality cartilage in the medial femoral condyle: good (second-look arthroscopic) status (ICRS grade I or II; n = 70), and poor (second-look arthroscopic) status (ICRS grade III or IV; n = 85) groups at the time of second-look arthroscopy. Clinical outcomes were assessed using the International Knee Documentation Committee (IKDC) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 36-Item Short Form survey.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 74 - 80
1 Jun 2021
Deckey DG Rosenow CS Verhey JT Brinkman JC Mayfield CK Clarke HD Bingham JS

Aims

Robotic-assisted total knee arthroplasty (RA-TKA) is theoretically more accurate for component positioning than TKA performed with mechanical instruments (M-TKA). Furthermore, the ability to incorporate soft-tissue laxity data into the plan prior to bone resection should reduce variability between the planned polyethylene thickness and the final implanted polyethylene. The purpose of this study was to compare accuracy to plan for component positioning and precision, as demonstrated by deviation from plan for polyethylene insert thickness in measured-resection RA-TKA versus M-TKA.

Methods

A total of 220 consecutive primary TKAs between May 2016 and November 2018, performed by a single surgeon, were reviewed. Planned coronal plane component alignment and overall limb alignment were all 0° to the mechanical axis; tibial posterior slope was 2°; and polyethylene thickness was 9 mm. For RA-TKA, individual component position was adjusted to assist gap-balancing but planned coronal plane alignment for the femoral and tibial components and overall limb alignment remained 0 ± 3°; planned tibial posterior slope was 1.5°. Mean deviations from plan for each parameter were compared between groups for positioning and size and outliers were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 137 - 144
1 Jun 2021
Lachiewicz PF Steele JR Wellman SS

Aims

To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period.

Methods

This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 36 - 42
1 Jun 2020
Nishitani K Kuriyama S Nakamura S Umatani N Ito H Matsuda S

Aims

This study aimed to evaluate the association between the sagittal alignment of the femoral component in total knee arthroplasty (TKA) and new Knee Society Score (2011KSS), under the hypothesis that outliers such as the excessive extended or flexed femoral component were related to worse clinical outcomes.

Methods

A group of 156 knees (134 F:22 M) in 133 patients with a mean age 75.8 years (SD 6.4) who underwent TKA with the cruciate-substituting Bi-Surface Knee prosthesis were retrospectively enrolled. On lateral radiographs, γ angle (the angle between the distal femoral axis and the line perpendicular to the distal rear surface of the femoral component) was measured, and the patients were divided into four groups according to the γ angle. The 2011KSSs among groups were compared using the Kruskal-Wallis test. A secondary regression analysis was used to investigate the association between the 2011KSS and γ angle.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 366 - 371
1 Apr 2004
Nabeyama R Matsuda S Miura H Mawatari T Kawano T Iwamoto Y

Our study evaluated the accuracy of an image-guided total knee replacement system based on CT with regard to preparation of the femoral and tibial bone using nine limbs from five cadavers. The accuracy was assessed by direct measurement using an extramedullary alignment rod without radiographs. The mean angular errors of the femur and tibia, which represent angular gaps from the real mechanical axis in the coronal plane, were 0.3° and 1.1°, respectively. The CT-based system, provided almost perfect alignment of the femoral component with less than 1° of error and excellent alignment with less than 3° of error for the tibial component. Our results suggest that standardisation of knee replacement by the use of this system will lead to improved long-term survival of total knee arthroplasty


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 10 - 18
1 Jun 2020
Ueyama H Kanemoto N Minoda Y Taniguchi Y Nakamura H

Aims

The aim of this study was to assess the effectiveness of perioperative essential amino acid (EAA) supplementation to prevent rectus femoris muscle atrophy and facilitate early recovery of function after total knee arthroplasty (TKA).

Methods

The study involved 60 patients who underwent unilateral TKA for primary knee osteo-arthritis (OA). This was a double-blind, placebo-controlled, randomized control trial with patients randomly allocated to two groups, 30 patients each: the essential amino acid supplementation (9 g daily) and placebo (lactose powder, 9 g daily) groups. Supplementation and placebo were provided from one week before to two weeks after surgery. The area of the rectus femoris muscle were measured by ultrasound imaging one month before surgery and one, two, three, and four weeks postoperatively. The serum albumin level, a visual analogue knee pain score, and mobility were also measured at each time point. The time to recovery of activities of daily living (ADLs) was recorded. Postoperative nutrition and physiotherapy were identical in both groups.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 115 - 120
1 Jul 2019
Hooper J Schwarzkopf R Fernandez E Buckland A Werner J Einhorn T Walker PS

Aims

This aim of this study was to assess the feasibility of designing and introducing generic 3D-printed instrumentation for routine use in total knee arthroplasty.

Materials and Methods

Instruments were designed to take advantage of 3D-printing technology, particularly ensuring that all parts were pre-assembled, to theoretically reduce the time and skill required during surgery. Concerning functionality, ranges of resection angle and distance were restricted within a safe zone, while accommodating either mechanical or anatomical alignment goals. To identify the most suitable biocompatible materials, typical instrument shapes and mating parts, such as dovetails and screws, were designed and produced.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 79 - 84
1 Jun 2020
Abdelfadeel W Houston N Star A Saxena A Hozack WJ

Aims

The aim of this study was to analyze the true costs associated with preoperative CT scans performed for robotic-assisted total knee arthroplasty (RATKA) planning and to determine the value of a formal radiologist’s report of these studies.

Methods

We reviewed 194 CT reports of 176 sequential patients who underwent primary RATKA by a single surgeon at a suburban teaching hospital. CT radiology reports were reviewed for the presence of incidental findings that might change the management of the patient. Payments for the scans, including the technical and professional components, for 330 patients at two hospitals were also recorded and compared.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 53 - 58
1 Jan 2019
Billi F Kavanaugh A Schmalzried H Schmalzried TP

Aims

Loosening of the tibial component after total knee arthroplasty (TKA) is a common indication for revision. Increasing the strength of the initial tibial implant/cement interface is desirable. There is little information about the surgical techniques that lead to the highest strength. We investigated the effects of eight variables on the strength of the initial tibial baseplate/cement interface.

Materials and Methods

A total of 48 tibial trays were cemented into acrylic holders using cement from two manufacturers, at three different times (early, normal, and late) using two techniques: cementing the tibial plateau or the plateau and the keel; and involving two conditions of contamination with marrow fat (at the metal/cement and cement/cement interfaces). Push-out tests were performed with load continuously recorded.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1087 - 1092
1 Sep 2019
Garceau S Warschawski Y Dahduli O Alshaygy I Wolfstadt J Backstein D

Aims

The aim of this study was to assess the effects of transferring patients to a specialized arthroplasty centre between the first and second stages (interstage) of prosthetic joint infection (PJI) of the knee.

Patients and Methods

A search of our institutional database was performed to identify patients having undergone two-stage revision total knee arthroplasty (TKA) for PJI. Two cohorts were created: continuous care (CC) and transferred care (TC). Baseline characteristics and outcomes were collected and compared between cohorts.


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1047 - 1052
1 Aug 2017
Ikawa T Takemura S Kim M Takaoka K Minoda Y Kadoya Y

Aims

The aim of this study was to evaluate the effects of using a portable, accelerometer-based surgical navigation system (KneeAlign2) in total knee arthroplasty (TKA) on the alignment of the femoral component, and blood loss.

Patients and Methods

A total of 241 consecutive patients with primary osteoarthritis of the knee were enrolled in this prospective, randomised controlled study. There were 207 women and 34 men. The mean age of the patients was 74.0 years (57 to 89). The KneeAlign2 system was used for distal femoral resection in 121 patients (KA2 group) and a conventional intramedullary femoral guide was used in 120 patients (IM group).


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1319 - 1328
1 Oct 2017
Shelton TJ Nedopil AJ Howell SM Hull ML

Aims

The aims of this study were to determine the proportion of patients with outlier varus or valgus alignment in kinematically aligned total knee arthroplasty (TKA), whether those with outlier varus or valgus alignment have higher forces in the medial or lateral compartments of the knee than those with in-range alignment and whether measurements of the alignment of the limb, knee and components predict compartment forces.

Patients and Methods

The intra-operative forces in the medial and lateral compartments were measured with an instrumented tibial insert in 67 patients who underwent a kinematically aligned TKA during passive movement. The mean of the forces at full extension, 45° and 90° of flexion determined the force in the medial and lateral compartments. Measurements of the alignment of the limb and the components included the hip-knee-ankle (HKA) angle, proximal medial tibial angle (PMTA), and distal lateral femoral angle (DLFA). Measurements of the alignment of the knee and the components included the tibiofemoral angle (TFA), tibial component angle (TCA) and femoral component angle (FCA). Alignment was measured on post-operative, non-weight-bearing anteroposterior (AP) scanograms and categorised as varus or valgus outlier or in-range in relation to mechanically aligned criteria.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1499 - 1505
1 Nov 2012
Kim J Kasat NS Bae J Kim S Oh S Lim H

The purpose of this study was to measure the radiological parameters of femoral component alignment of the Oxford Phase 3 unicompartmental knee replacement (UKR), and evaluate their effect on clinical outcome. Multiple regression analysis was used to examine the relative contributions of the radiological assessment of femoral component alignment in 189 consecutive UKRs performed by a single surgeon. The American Knee Society scores were compared between groups, defined as being within or outside recommended tolerances of the position of the femoral component. For the flexion/extension position 21 UKRs (11.1%) lay outside the recommended limits, and for posterior overhang of the femoral component nine (4.8%) lay outside the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest canal distance from the distal femoral entry point of the alignment jig and coronal entry-point position had significant effects on the flexion/extension position. Pre-operative HKA angle had a significant influence on posterior overhang of the femoral component. However, there was no significant difference in American Knee Society scores relative to the position of the femoral component.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1628 - 1633
1 Dec 2015
Elmadag M Uzer G Yildiz F Erden T Bilsel K Büyükpinarbasili N Üsümez A Bozdag E Sen C

This animal study compares different methods of performing an osteotomy, including using an Erbium-doped Yttrium Aluminum Garnet laser, histologically, radiologically and biomechanically. A total of 24 New Zealand rabbits were divided into four groups (Group I: multihole-drilling; Group II: Gigli saw; Group III: electrical saw blade and Group IV: laser). A proximal transverse diaphyseal osteotomy was performed on the right tibias of the rabbits after the application of a circular external fixator. The rabbits were killed six weeks after the procedure, the operated tibias were resected and radiographs taken.

The specimens were tested biomechanically using three-point bending forces, and four tibias from each group were examined histologically. Outcome parameters were the biomechanical stability of the tibias as assessed by the failure to load and radiographic and histological examination of the osteotomy site.

The osteotomies healed in all specimens both radiographically and histologically. The differences in the mean radiographic (p = 0.568) and histological (p = 0.71) scores, and in the mean failure loads (p = 0.180) were not statistically significant between the groups.

Different methods of performing an osteotomy give similar quality of union. The laser osteotomy, which is not widely used in orthopaedics is an alternative to the current methods.

Cite this article: Bone Joint J 2015;97-B:1628–33.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 864 - 869
1 Jul 2008
Amis AA Oguz C Bull AMJ Senavongse W Dejour D

Objective patellar instability has been correlated with dysplasia of the femoral trochlea. This in vitro study tested the hypothesis that trochleoplasty would increase patellar stability and normalise the kinematics of a knee with a dysplastic trochlea. Six fresh-frozen knees were loaded via the heads of the quadriceps. The patella was displaced 10 mm laterally and the displacing force was measured from 0° to 90° of flexion. Patellar tracking was measured from 0° to 130° of knee flexion using magnetic sensors. These tests were repeated after raising the central anterior trochlea to simulate dysplasia, and repeated again after performing a trochleoplasty on each specimen. The simulated dysplasia significantly reduced stability from that of the normal knee (p < 0.001). Trochleoplasty significantly increased the stability (p < 0.001), so that it did not then differ significantly from the normal knee (p = 0.244). There were small but statistically significant changes in patellar tracking (p< 0.001).

This study has provided objective biomechanical data to support the use of trochleoplasty in the treatment of patellar instability associated with femoral trochlear dysplasia.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 198 - 202
1 Feb 2006
Kalairajah Y Cossey AJ Verrall GM Ludbrook G Spriggins AJ

We undertook a prospective, randomised study using a non-invasive transcranial Doppler device to evaluate cranial embolisation in computer-assisted navigated total knee arthroplasty (n = 14) and compared this with a standard conventional surgical technique using intramedullary alignment guides (n = 10). All patients were selected randomly without the knowledge of the patient, anaesthetists (before the onset of the procedure) and ward staff. The operations were performed by a single surgeon at one hospital using a uniform surgical approach, instrumentation, technique and release sequence. The only variable in the two groups of patients was the use of single tracker pins of the imageless navigation system in the tibia and femur of the navigated group and intramedullary femoral and tibial alignment jigs in the non-navigated group.

Acetabular Doppler signals were obtained in 14 patients in the computer-assisted group and nine (90%) in the conventional group, in whom high-intensity signals were detected in seven computer-assisted patients (50%) and in all of the non-navigated patients.

In the computer-assisted group no patient had more than two detectable emboli, with a mean of 0.64 (SD 0.74). In the non-navigated group the number of emboli ranged from one to 43 and six patients had more than two detectable emboli, with a mean of 10.7 (sd 13.5). The difference between the two groups was highly significant using the Wilcoxon non-parametric test (p = 0.0003).

Our findings show that computer-assisted total knee arthroplasty, when compared with conventional jig-based surgery, significantly reduces systemic emboli as detected by transcranial Doppler ultrasonography.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 354 - 359
1 Mar 2013
Chareancholvanich K Narkbunnam R Pornrattanamaneewong C

Patient-specific cutting guides (PSCGs) are designed to improve the accuracy of alignment of total knee replacement (TKR). We compared the accuracy of limb alignment and component positioning after TKR performed using PSCGs or conventional instrumentation. A total of 80 patients were randomised to undergo TKR with either of the different forms of instrumentation, and radiological outcomes and peri-operative factors such as operating time were assessed. No significant difference was observed between the groups in terms of tibiofemoral angle or femoral component alignment. Although the tibial component in the PSCGs group was measurably closer to neutral alignment than in the conventional group, the size of the difference was very small (89.8° (sd 1.2) vs 90.5° (sd 1.6); p = 0.030). This new technology slightly shortened the bone-cutting time by a mean of 3.6 minutes (p < 0.001) and the operating time by a mean 5.1 minutes (p = 0.019), without tangible differences in post-operative blood loss (p = 0.528) or need for blood transfusion (p = 0.789). This study demonstrated that both PSCGs and conventional instrumentation restore limb alignment and place the components with the similar accuracy. The minimal advantages of PSCGs in terms of consistency of alignment or operative time are unlikely to be clinically relevant.

Cite this article: Bone Joint J 2013;95-B:354–9.