We have compared the changes in the pattern of the principal strains in the proximal femur after
We examined macroscopically and microscopically 55 cadaver rotator-cuff tendons attached to their humeral heads to determine the distance between the edge of the articular cartilage and the tendon
The cortical strains on the femoral neck and proximal femur were measured before and after implantation of a resurfacing femoral component in 13 femurs from human cadavers. These were loaded into a hip simulator for single-leg stance and stair-climbing. After resurfacing, the mean tensile strain increased by 15% (95% confidence interval (CI) 6 to 24, p = 0.003) on the lateral femoral neck and the mean compressive strain increased by 11% (95% CI 5 to 17, p = 0.002) on the medial femoral neck during stimulation of single-leg stance. On the proximal femur the deformation pattern remained similar to that of the unoperated femurs. The small increase of strains in the neck area alone would probably not be sufficient to cause fracture of the neck However, with patient-related and surgical factors these strain changes may contribute to the risk of early periprosthetic fracture.
Hydroxyapatite-coated standard anatomical and customised femoral stems are designed to transmit load to the metaphyseal part of the proximal femur in order to avoid stress shielding and to reduce resorption of bone. In a randomised in vitro study, we compared the changes in the pattern of cortical strain after the
The aim of this biomechanical study was to investigate the role of the dorsal vertebral cortex in transpedicular screw fixation. Moss transpedicular screws were introduced into both pedicles of each vertebra in 25 human cadaver vertebrae. The dorsal vertebral cortex and subcortical bone corresponding to the entrance site of the screw were removed on one side and preserved on the other. Biomechanical testing showed that the mean peak pull-out strength for the inserted screws, following removal of the dorsal cortex, was 956.16 N. If the dorsal cortex was preserved, the mean peak pullout strength was 1295.64 N. The mean increase was 339.48 N (26.13%; p = 0.033). The bone mineral density correlated positively with peak pull-out strength. Preservation of the dorsal vertebral cortex at the site of
We dissected 20 cadaver hips in order to investigate the anatomy and excursion of the trochanteric muscles in relation to the posterior approach for total hip replacement. String models of each muscle were created and their excursion measured while the femur was moved between its anatomical position and the dislocated position. The position of the hip was determined by computer navigation. In contrast to previous studies which showed a separate
Interfacial defects between the cement mantle and a hip implant may arise from constrained shrinkage of the cement or from air introduced during
We have investigated the role of the penetration of saline on the shear strength of the cement-stem interface for stems inserted at room temperature and those preheated to 37°C using a variety of commercial bone cements. Immersion in saline for two weeks at 37°C reduced interfacial strength by 56% to 88% after
Three palmar wrist ligaments from fresh human cadavers were dissected from the proximal to the distal
Six pairs of human cadaver femora were divided equally into two groups one of which received a non-cemented reference implant and the other a very short non-dependent experimental implant. Thirteen strain-gauge rosettes were attached to the external surface of each specimen and, during application of combined axial and torsional loads to the femoral head, the strains in both groups were measured. After the
We measured the
We investigated several factors which affect the stability of cortical screws in osteoporotic bone using 18 femora from cadavers of women aged between 45 and 96 years (mean 76). We performed bone densitometry to measure the bone mineral density of the cortical and cancellous bone of the shaft and head of the femur, respectively. The thickness and overall bone mass of the cortical layer of the shaft of the femur were measured using a microCT scanner. The force required to pull-out a 3.5 mm titanium cortical bone screw was determined after standardised
The aim of this study was to validate the use of three models of fracture fixation in the assessment of technical skills. We recruited 21 subjects (six experts, seven intermediates, and eight novices) to perform three procedures: application of a dynamic compression plate on a cadaver porcine model,
In a study on ten fresh human cadavers we examined the change in the height of the intervertebral disc space, the angle of lordosis and the geometry of the facet joints after
Little is known about the tissue reactions to various implant materials which coincide with an inflammatory reaction. We used the avridine arthritis rat model to evaluate the tissue response in the synovial, interstitial and subcutaneous tissues after implant
An innovative Kirschner (K-) wire point was developed and compared in fresh pig femora in terms of drilling efficiency and temperature elevation with the trochar and diamond points currently used in clinical practice. The tips of thermal couples were machined to the defined geometry and the temperature measured during drilling. Using the same drill speed (rev/min) and feed rate, the new K-wire point produced the lowest thrust force and torque as measured by a Kistler dynamometer. Drill point temperatures were highest with the trochar geometry (129 ± 6°C), followed by the diamond (98 ± 7°C). The lowest temperatures were recorded with the Medin K-wire (66 ± 2°C). On repeated drilling it could be used for up to 30 holes before reaching the less satisfactory drill performance of the diamond tip. The new K-wire provides a better alternative as it requires less effort for
We aimed to evaluate the temperature around the nerve root during drilling of the lamina and to
determine whether irrigation during drilling can reduce the chance of nerve root injury. Lumbar nerve roots were exposed to frictional heat by high-speed drilling of the lamina in a live
rabbit model, with saline (room temperature (RT) or chilled saline) or without saline (control)
irrigation. We measured temperatures surrounding the nerve root and made histological
evaluations.Aims
Materials and Methods
Malpositioning of the trochanteric entry point
during the introduction of an intramedullary nail may cause iatrogenic
fracture or malreduction. Although the optimal point of insertion
in the coronal plane has been well described, positioning in the
sagittal plane is poorly defined. The paired femora from 374 cadavers were placed both in the anatomical
position and in internal rotation to neutralise femoral anteversion.
A marker was placed at the apparent apex of the greater trochanter,
and the lateral and anterior offsets from the axis of the femoral
shaft were measured on anteroposterior and lateral photographs. Greater
trochanteric morphology and trochanteric overhang were graded. The mean anterior offset of the apex of the trochanter relative
to the axis of the femoral shaft was 5.1 mm ( Placement of the entry position at the apex of the greater trochanter
in the anteroposterior view does not reliably centre an intramedullary
nail in the sagittal plane. Based on our findings, the site of insertion
should be about 5 mm posterior to the apex of the trochanter to
allow for its anterior offset. Cite this article:
Reported rates of dislocation in hip hemiarthroplasty
(HA) for the treatment of intra-capsular fractures of the hip, range
between 1% and 10%. HA is frequently performed through a direct
lateral surgical approach. The aim of this study is to determine
the contribution of the anterior capsule to the stability of a cemented
HA through a direct lateral approach. A total of five whole-body cadavers were thawed at room temperature,
providing ten hip joints for investigation. A Thompson HA was cemented
in place via a direct lateral approach. The cadavers were then positioned
supine, both knee joints were disarticulated and a digital torque
wrench was attached to the femur using a circular frame with three
half pins. The wrench applied an external rotation force with the
hip in extension to allow the hip to dislocate anteriorly. Each
hip was dislocated twice; once with a capsular repair and once without
repairing the capsule. Stratified sampling ensured the order in
which this was performed was alternated for the paired hips on each
cadaver. Comparing peak torque force in hips with the capsule repaired
and peak torque force in hips without repair of the capsule, revealed
a significant difference between the ‘capsule repaired’ (mean 22.96
Nm, standard deviation ( Cite this article:
High-intensity narrow-spectrum (HINS) light is
a novel violet-blue light inactivation technology which kills bacteria through
a photodynamic process, and has been shown to have bactericidal
activity against a wide range of species. Specimens from patients
with infected hip and knee arthroplasties were collected over a
one-year period (1 May 2009 to 30 April 2010). A range of these
microbial isolates were tested for sensitivity to HINS-light. During
testing, suspensions of the pathogens were exposed to increasing
doses of HINS-light (of 123mW/cm2 irradiance). Non-light exposed
control samples were also used. The samples were then plated onto
agar plates and incubated at 37°C for 24 hours before enumeration.
Complete inactivation (greater than 4-log10 reduction)
was achieved for all of the isolates. The typical inactivation curve
showed a slow initial reaction followed by a rapid period of inactivation.
The doses of HINS-light required ranged between 118 and 2214 J/cm2.
Gram-positive bacteria were generally found to be more susceptible
than Gram-negative. As HINS-light uses visible wavelengths, it can be safely used
in the presence of patients and staff. This unique feature could
lead to its possible use in the prevention of infection during surgery
and post-operative dressing changes. Cite this article: