Modifications in the design of knee replacements have been proposed in order to maximise flexion. We performed a prospective double-blind randomised controlled trial to compare the functional outcome, including maximum knee flexion, in patients receiving either a standard or a
This prospective randomised controlled double-blind
trial compared two types of PFC Sigma total knee replacement (TKR),
differing in three design features aimed at improving flexion. The
outcome of a standard fixed-bearing posterior cruciate ligament-preserving
design (FB-S) was compared with that of a high-flexion rotating-platform
posterior-stabilised design (RP-F) at one year after TKR. The study group of 77 patients with osteoarthritis of the knee
comprised 37 men and 40 women, with a mean age of 69 years (44.9
to 84.9). The patients were randomly allocated either to the FB-S
or the RP-F group and assessed pre-operatively and at one year post-operatively.
The mean post-operative non-weight-bearing flexion was 107° (95%
confidence interval (CI) 104° to 110°)) for the FB-S group and 113°
(95% CI 109° to 117°) for the RP-F group, and this difference was
statistically significant (p = 0.032). However, weight-bearing range
of movement during both level walking and ascending a slope as measured
during flexible electrogoniometry was a mean of 4° lower in the RP-F
group than in the FB-S group, with 58° (95% CI 56° to 60°) Although the RP-F group achieved higher non-weight-bearing knee
flexion, patients in this group did not use this during activities
of daily living and reported more pain one year after surgery
Aims. Modern total knee arthroplasty (TKA) prostheses are designed to restore near normal kinematics including
The outcome of total knee replacement (TKR) using
components designed to increase the range of flexion is not fully
understood. The short- to mid-term risk of aseptic revision in high
flexion TKR was evaluated. The endpoint of the study was aseptic
revision and the following variables were investigated: implant
design (high flexion vs non-high flexion), the
thickness of the tibial insert (≤ 14 mm vs >
14
mm), cruciate ligament (posterior stabilised (PS) vs cruciate
retaining), mobility (fixed vs rotating), and the
manufacturer (Zimmer, Smith &
Nephew and DePuy). Covariates
included patient, implant, surgeon and hospital factors. Marginal
Cox proportional hazard models were used. In a cohort of 64 000 TKRs,
Aim. There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several national joint registry studies documenting high UKA revision rates when compared to total knee arthroplasty (TKA). With the recent introduction of robotically assisted UKA procedures, it is hoped that outcomes improve. This study examines the cumulative revision rate of UKA procedures implanted with a newly introduced robotic system and compares the results to one of the best performing non-robotically assisted UKA prostheses, as well as all other non-robotically assisted UKA procedures. Methods. Data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR) for all UKA procedures performed for osteoarthritis (OA) between 2015 and 2018 were analyzed. Procedures using the Restoris MCK UKA prosthesis implanted using the Mako Robotic-Arm Assisted System were compared to non-robotically assisted Zimmer Unicompartmental
We aimed to document the pre-operative expectations in Korean patients undergoing total knee replacement using an established survey form and to determine whether expectations were influenced by sociodemographic factors or pre-operative functional status. Expectations regarding 17 items in the Knee Replacement Expectation Survey form were investigated in 454 patients scheduled for total knee replacement. The levels and distribution patterns of summated expectation and of five expectation categories (relief from pain, baseline activity,
Aims. In Asia and the Middle-East, people often flex their knees deeply
in order to perform activities of daily living. The purpose of this
study was to investigate the 3D kinematics of normal knees during
high-flexion activities. Our hypothesis was that the femorotibial
rotation, varus-valgus angle, translations, and kinematic pathway
of normal knees during high-flexion activities, varied according
to activity. Materials and Methods. We investigated the in vivo kinematics of eight
normal knees in four male volunteers (mean age 41.8 years; 37 to
53) using 2D and 3D registration technique, and modelled the knees
with a computer aided design program. Each subject squatted, kneeled,
and sat cross-legged. We evaluated the femoral rotation and varus-valgus
angle relative to the tibia and anteroposterior translation of the
medial and lateral side, using the transepicodylar axis as our femoral
reference relative to the perpendicular projection on to the tibial
plateau. This method evaluates the femur medially from what has
elsewhere been described as the extension facet centre, and differs
from the method classically applied. . Results. During squatting and kneeling, the knees displayed femoral external
rotation. When sitting cross-legged, femurs displayed internal rotation
from 10° to 100°. From 100°, femoral external rotation was observed.
No significant difference in varus-valgus angle was seen between
squatting and kneeling, whereas a varus position was observed from
140° when sitting cross-legged. The measure kinematic pathway using
our methodology found during squatting a medial pivoting pattern
from 0° to 40° and bicondylar rollback from 40° to 150°. During
kneeling, a medial pivot pattern was evident. When sitting cross-legged,
a lateral pivot pattern was seen from 0° to 100°, and a medial pivot
pattern beyond 100°. Conclusion. The kinematics of normal knees during
The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients. A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with radiostereometry (RSA) at three and six months and one, two, and five years postoperatively. PROMs were collected preoperatively and at all periods of follow-up.Aims
Methods
The aims of this study were to investigate the ability to kneel after total knee arthroplasty (TKA) without patellar resurfacing, and its effect on patient-reported outcome measures (PROMs). Secondary aims included identifying which kneeling positions were most important to patients, and the influence of radiological parameters on the ability to kneel before and after TKA. This prospective longitudinal study involved 209 patients who underwent single radius cruciate-retaining TKA without patellar resurfacing. Preoperative EuroQol five-dimension questionnaire (EQ-5D), Oxford Knee Score (OKS), and the ability to achieve four kneeling positions were assessed including a single leg kneel, a double leg kneel, a high-flexion kneel, and a praying position. The severity of radiological osteoarthritis (OA) was graded and the pattern of OA was recorded intraoperatively. The flexion of the femoral component, posterior condylar offset, and anterior femoral offset were measured radiologically. At two to four years postoperatively, 151 patients with a mean age of 70.0 years (SD 9.44) were included. Their mean BMI was 30.4 kg/m2 (SD 5.36) and 60 were male (40%). They completed EQ-5D, OKS, and Kujala scores, assessments of the ability to kneel, and a visual analogue scale for anterior knee pain and satisfaction.Aims
Methods
The patient-acceptable symptom state (PASS) is a level of wellbeing, which is measured by the patient. The aim of this study was to determine if the proportion of patients who achieved an acceptable level of function (PASS) after medial unicompartmental knee arthroplasty (UKA) was different based on the status of the anterior cruciate ligament (ACL) at the time of surgery. A total of 114 patients who underwent UKA for isolated medial osteoarthritis (OA) of the knee were included in the study. Their mean age was 65 years (SD 10). No patient underwent a bilateral procedure. Those who had undergone ACL reconstruction during the previous five years were excluded. The Knee injury Osteoarthritis Outcome Score Activities of Daily Living (KOOS ADL) function score was used as the primary outcome measure with a PASS of 87.5, as described for total knee arthroplasty (TKA). Patients completed all other KOOS subscales, Lysholm score, the Western Ontario and McMaster Universities Osteoarthritis Index, and the Veterans Rand 12-item health survey score. Failure was defined as conversion to TKA.Aims
Methods
The optimal management of the tibial slope in
achieving a
Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.Aims
Methods
To report mid- to long-term results of Oxford mobile bearing domed lateral unicompartmental knee arthroplasty (UKA), and determine the effect of potential contraindications on outcome. A total of 325 consecutive domed lateral UKAs undertaken for the recommended indications were included, and their functional and survival outcomes were assessed. The effects of age, weight, activity, and the presence of full-thickness erosions of cartilage in the patellofemoral joint on outcome were evaluated.Aims
Methods
It remains controversial whether patellofemoral joint pathology is a contraindication to lateral unicompartmental knee arthroplasty (UKA). This study aimed to evaluate the effect of preoperative radiological degenerative changes and alignment on patient-reported outcome scores (PROMs) after lateral UKA. Secondarily, the influence of lateral UKA on the alignment of the patellofemoral joint was studied. A consecutive series of patients who underwent robotic arm-assisted fixed-bearing lateral UKA with at least two-year follow-up were retrospectively reviewed. Radiological evaluation was conducted to obtain a Kellgren Lawrence (KL) grade, an Altman score, and alignment measurements for each knee. Postoperative PROMs were assessed using the Kujala (Anterior Knee Pain Scale) score, Knee Injury and Osteoarthritis Outcome Score Joint Replacement (KOOS JR), and satisfaction levels.Aims
Methods
Bicruciate-stabilised total knee replacement (TKR) aims to restore normal kinematics by replicating the function of both cruciate ligaments. We performed a prospective, randomised controlled trial in which bicruciate- and posterior-stabilised TKRs were implanted in 13 and 15 osteo-arthritic knees, respectively. The mean age of the bicruciate-stabilised group was 63.9 years (. sd. 10.00) and that of the posterior-stabilised group 63.2 years (. sd. 6.7). A control group comprised 14 normal subjects with a mean age of 67.9 years (. sd. 7.9). The patellar tendon angle (PTA) was measured one week pre-operatively and at seven weeks post-operatively during knee extension, flexion and step-up exercises. At near full extension during step-up, the bicruciate-stabilised TKR produced a higher mean PTA than the posterior-stabilised TKR, indicating that the bicruciate design at least partially restored the kinematic role of the anterior cruciate ligament. The bicruciate-stabilised TKR largely restored the pre-operative kinematics, whereas the posterior-stabilised TKR resulted in a consistently lower PTA at all activities. The PTA in the pre-operative knees was higher than in the control group during the step-up and at near full knee extension. Overall, both groups generated a more normal PTA than that seen in previous studies in
Between 15% and 20% of patients remain dissatisfied following total knee arthroplasty (TKA). The SAIPH knee system (MatOrtho, Surrey, United Kingdom) is a medial ball and socket TKA that has been designed to replicate native knee kinematics in order to maximize the range of movement, stability, and function. This system is being progressively introduced in a stepwise fashion, with this study reporting the mid-term clinical and radiological outcomes. A retrospective review was undertaken of the first 100 consecutive patients with five-year follow-up following SAIPH TKA performed by the senior authors. The data that were collected included the demographics of the patients, clinical findings, the rate of intraoperative ligamentous release, patient-reported outcome measures (PROMS), radiological assessment, complications, and all-cause revision. Revision data were cross-checked with a national registry.Aims
Patients and Methods
Aims
Patients and Methods
The aim of this independent multicentre study was to assess the
mid-term results of mobile bearing unicondylar knee arthroplasty
(UKA) for isolated lateral osteoarthritis of the knee joint. We retrospectively evaluated 363 consecutive, lateral UKAs (346
patients) performed using the Oxford domed lateral prosthesis undertaken
in three high-volume knee arthroplasty centres between 2006 and
2014. Mean age of the patients at surgery was 65 years (36 to 88)
with a mean final follow-up of 37 months (12 to 93)Aims
Patients and Methods
We sought to establish whether an oxidised zirconium (OxZr) femoral
component causes less loss of polyethylene volume than a cobalt
alloy (CoCr) femoral component in total knee arthroplasty. A total of 20 retrieved tibial inserts that had articulated with
OxZr components were matched with 20 inserts from CoCr articulations
for patient age, body mass index, length of implantation, and revision
diagnosis. Changes in dimensions of the articular surfaces were compared
with those of pristine inserts using laser scanning. The differences
in volume between the retrieved and pristine surfaces of the two
groups were calculated and compared.Aims
Materials and Methods