Advertisement for orthosearch.org.uk
Results 1 - 15 of 15
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 249 - 253
1 Feb 2014
Euler SA Hengg C Kolp D Wambacher M Kralinger F

Antegrade nailing of proximal humeral fractures using a straight nail can damage the bony insertion of the supraspinatus tendon and may lead to varus failure of the construct. In order to establish the ideal anatomical landmarks for insertion of the nail and their clinical relevance we analysed CT scans of bilateral proximal humeri in 200 patients (mean age 45.1 years (. sd. 19.6; 18 to 97) without humeral fractures. The entry point of the nail was defined by the point of intersection of the anteroposterior and lateral vertical axes with the cortex of the humeral head. The critical point was defined as the intersection of the sagittal axis with the medial limit of the insertion of the supraspinatus tendon on the greater tuberosity. The region of interest, i.e. the biggest entry hole that would not encroach on the insertion of the supraspinatus tendon, was calculated setting a 3 mm minimal distance from the critical point. This identified that 38.5% of the humeral heads were categorised as ‘critical types’, due to morphology in which the predicted offset of the entry point would encroach on the insertion of the supraspinatus tendon that may damage the tendon and reduce the stability of fixation. We therefore emphasise the need for ‘fastidious’ pre-operative planning to minimise this risk. Cite this article: Bone Joint J 2014;96-B:249–53


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 1 | Pages 33 - 41
1 Jan 2000
Hernigou P Cohen D

The risk of articular penetration during tibial nailing is well known, but the incidence of unrecognised damage to joint cartilage has not been described. We have identified this complication in the treatment of tibial fractures, described the anatomical structures at risk and examined the most appropriate site of entry for tibial nailing in relation to the shape of the bone, the design of the nail and the surgical approach. We studied the relationship between the intra-articular structures of the knee and the entry point used for nailing in 54 tibiae from cadavers. The results showed that the safe zone in some bones is smaller than the size of standard reamers and the proximal part of some nails. The structures at risk are the anterior horns of the medial and lateral menisci, the anterior part of the medial and lateral plateaux and the ligamentum transversum. This was confirmed by observations made after nailing 12 pairs of cadaver knees. A retrospective radiological analysis of 30 patients who had undergone tibial nailing identified eight at risk according to the entry point and the size of the nail. Unrecognised articular penetration and damage during surgery were confirmed in four. Although intramedullary nailing has been shown to be a successful method for treating fractures of the tibia, one of the most common problems after bony union is pain in the knee. Unrecognised intra-articular injury of the knee may be one cause of this


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 2 | Pages 239 - 243
1 Mar 2004
Cutler L Molloy A Dhukuram V Bass A

Distal tibial physeal fractures are the second most common growth plate injury and the most common cause of growth arrest and deformity. This study assesses the accuracy of pre-operative planning for placement of the screws in these fractures using either standard radiographs or CT scans. We studied 62 consecutive physeal fractures over a period of four years. An outline of a single cut of the CT scan was used for each patient. An ideal position for the screw was determined as being perpendicular to and at the midpoint of the fracture. The difference in entry point and direction of the screw between the ideal and the observers’ assessments were compared using the paired Student’s t-test. There was a statistically significant improvement (p < 0.0001) in the accuracy of the point of insertion and the direction of the screw on the pre-operative plan when CT scans were used rather than plain radiographs. We would, therefore, recommend that CT scans are routinely used in the pre-operative assessment and treatment of distal tibial physeal fractures


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 72 - 81
1 Jan 2023
Stake IK Ræder BW Gregersen MG Molund M Wang J Madsen JE Husebye EE

Aims

The aim of this study was to compare the functional and radiological outcomes and the complication rate after nail and plate fixation of unstable fractures of the ankle in elderly patients.

Methods

In this multicentre study, 120 patients aged ≥ 60 years with an acute unstable AO/OTA type 44-B fracture of the ankle were randomized to fixation with either a nail or a plate and followed for 24 months after surgery. The primary outcome measure was the American Orthopaedic Foot and Ankle Society (AOFAS) Ankle-Hindfoot score. Secondary outcome measures were the Manchester-Oxford Foot Questionnaire, the Olerud and Molander Ankle score, the EuroQol five-dimension questionnaire, a visual analogue score for pain, complications, the quality of reduction of the fracture, nonunion, and the development of osteoarthritis.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 294 - 298
1 Feb 2021
Hadeed MM Prakash H Yarboro SR Weiss DB

Aims

The aim of this study was to determine the immediate post-fixation stability of a distal tibial fracture fixed with an intramedullary nail using a biomechanical model. This was used as a surrogate for immediate weight-bearing postoperatively. The goal was to help inform postoperative protocols.

Methods

A biomechanical model of distal metaphyseal tibial fractures was created using a fourth-generation composite bone model. Three fracture patterns were tested: spiral, oblique, and multifragmented. Each fracture extended to within 4 cm to 5 cm of the plafond. The models were nearly-anatomically reduced and stabilized with an intramedullary nail and three distal locking screws. Cyclic loading was performed to simulate normal gait. Loading was completed in compression at 3,000 N at 1 Hz for a total of 70,000 cycles. Displacement (shortening, coronal and sagittal angulation) was measured at regular intervals.


The Bone & Joint Journal
Vol. 99-B, Issue 9 | Pages 1232 - 1236
1 Sep 2017
Dahill M McArthur J Roberts GL Acharya MR Ward AJ Chesser TJS

Aims

The anterior pelvic internal fixator is increasingly used for the treatment of unstable, or displaced, injuries of the anterior pelvic ring. The evidence for its use, however, is limited. The aim of this paper is to describe the indications for its use, how it is applied and its complications.

Patients and Methods

We reviewed the case notes and radiographs of 50 patients treated with an anterior pelvic internal fixator between April 2010 and December 2015 at a major trauma centre in the United Kingdom. The median follow-up time was 38 months (interquartile range 24 to 51).


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1095 - 1101
1 Aug 2017
Fuglesang HFS Flugsrud GB Randsborg PH Oord P Benth JŠ Utvåg SE

Aims

This is a prospective randomised controlled trial comparing the functional outcomes of plate fixation and elastic stable intramedullary nailing (ESIN) of completely displaced mid-shaft fractures of the clavicle in the active adult population.

Patients and Methods

We prospectively recruited 123 patients and randomised them to either plate fixation or ESIN. Patients completed the Quick Disabilities of the Arm, Shoulder and Hand (DASH) score at one to six weeks post-operatively. They were followed up at six weeks, three and six months and one year with radiographs, and their clinical outcome was assessed using both the DASH and the Constant Score.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 817 - 822
1 Jun 2014
Al-Nammari SS Dawson-Bowling S Amin A Nielsen D

Conventional methods of treating ankle fractures in the elderly are associated with high rates of complication. We describe the results of treating these injuries in 48 frail elderly patients with a long calcaneotalotibial nail.

The mean age of the group was 82 years (61 to 96) and 41 (85%) were women. All were frail, with multiple medical comorbidities and their mean American Society of Anaesthesiologists score was 3 (3 to 4). None could walk independently before their operation. All the fractures were displaced and unstable; the majority (94%, 45 of 48) were low-energy injuries and 40% (19 of 48) were open.

The overall mortality at six months was 35%. Of the surviving patients, 90% returned to their pre-injury level of function. The mean pre- and post-operative Olerud and Molander questionnaire scores were 62 and 57 respectively. Complications included superficial infection (4%, two of 48); deep infection (2%, one of 48); a broken or loose distal locking screw (6%, three of 48); valgus malunion (4%, two of 48); and one below-knee amputation following an unsuccessful vascular operation. There were no cases of nonunion, nail breakage or peri-prosthetic fracture.

A calcaneotalotibial nail is an excellent device for treating an unstable fracture of the ankle in the frail elderly patient. It allows the patient to mobilise immediately and minimises the risk of bone or wound problems. A long nail which crosses the isthmus of the tibia avoids the risk of peri-prosthetic fracture associated with shorter devices.

Cite this article: Bone Joint J 2014; 96-B:817–22.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1495 - 1497
1 Nov 2007
Jones BG Mehin R Young D

Intramedullary tibial nailing was performed in ten paired cadavers and the insertion of a medial-to-lateral proximal oblique locking screw was simulated in each specimen. Anatomical dissection was undertaken to determine the relationship of the common peroneal nerve to the cross-screw.

The common peroneal nerve was contacted directly in four tibiae and the cross-screw was a mean of 2.6 mm (1.0 to 10.7) away from the nerve in the remaining 16. Iatrogenic injury to the common peroneal nerve by medial-to-lateral proximal oblique locking screws is therefore a significant risk during tibial nailing.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 66 - 71
1 Jan 2008
McQueen MM Gelbke MK Wakefield A Will EM Gaebler C

We randomly allocated 60 consecutive patients with fractures of the waist of the scaphoid to percutaneous fixation with a cannulated Acutrak screw or immobilisation in a cast. The range of movement, the grip and pinch strength, the modified Green/O’Brien functional score, return to work and sports, and radiological evidence of union were evaluated at each follow-up visit. Patients were followed sequentially for one year.

Those undergoing percutaneous screw fixation showed a quicker time to union (9.2 weeks vs 13.9 weeks, p < 0.001) than those treated with a cast. There was a trend towards a higher rate of nonunion in the non-operative group, although this was not statistically significant. Patients treated by operation had a more rapid return of function and to sport and full work compared with those managed conservatively. There was a very low complication rate.

We recommend that all active patients should be offered percutaneous stabilisation for fractures of the waist of the scaphoid.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 940 - 942
1 Jul 2007
Chaarani MW

A single antegrade Rush nail was used to treat 37 fractures of the distal humerus extending to the olecranon fossa and up to 6.5 cm proximal to it. In order to stabilise the fracture, the bent tip of the Rush nail was directed into the lateral epicondylar ridge. This gave immediate stability. All the fractures united without residual stiffness of the shoulder or elbow.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1073 - 1078
1 Aug 2008
Little NJ Verma V Fernando C Elliott DS Khaleel A

We compared the outcome of patients treated for an intertrochanteric fracture of the femoral neck with a locked, long intramedullary nail with those treated with a dynamic hip screw (DHS) in a prospective randomised study.

Each patient who presented with an extra-capsular hip fracture was randomised to operative stabilisation with either a long intramedullary Holland nail or a DHS. We treated 92 patients with a Holland nail and 98 with a DHS. Pre-operative variables included the Mini Mental test score, patient mobility, fracture pattern and American Society of Anesthesiologists grading. Peri-operative variables were anaesthetic time, operating time, radiation time and blood loss. Post-operative variables were time to mobilising with a frame, wound infection, time to discharge, time to fracture union, and mortality.

We found no significant difference in the pre-operative variables. The mean anaesthetic and operation times were shorter in the DHS group than in the Holland nail group (29.7 vs 40.4 minutes, p < 0.001; and 40.3 vs 54 minutes, p < 0.001, respectively). There was an increased mean blood loss within the DHS group versus the Holland nail group (160 ml vs 78 ml, respectively, p < 0.001). The mean time to mobilisation with a frame was shorter in the Holland nail group (DHS 4.3 days, Holland nail 3.6 days, p = 0.012). More patients needed a post-operative blood transfusion in the DHS group (23 vs seven, p = 0.003) and the mean radiation time was shorter in this group (DHS 0.9 minutes vs Holland nail 1.56 minutes, p < 0.001). The screw of the DHS cut out in two patients, one of whom underwent revision to a Holland nail. There were no revisions in the Holland nail group. All fractures in both groups were united when followed up after one year.

We conclude that the DHS can be implanted more quickly and with less exposure to radiation than the Holland nail. However, the resultant blood loss and need for transfusion is greater. The Holland nail allows patients to mobilise faster and to a greater extent. We have therefore adopted the Holland nail as our preferred method of treating intertrochanteric fractures of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 358 - 361
1 Mar 2006
Prathapkumar KR Garg NK Bruce CE

Several techniques are available for the treatment of displaced fractures of the neck of the radius in children. We report our experience in 14 children treated by indirect reduction and fixation using an elastic stable intramedullary nail. The patients were followed up for a mean of 28 months. One developed asymptomatic avascular necrosis. The rest had excellent results. One had a neuropraxia of the posterior interosseous nerve, which recovered within six weeks. We advocate elastic stable intramedullary nailing for the closed reduction and fixation of these fractures in children.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1634 - 1641
1 Dec 2006
Stulik J Stehlik J Rysavy M Wozniak A

We describe the results of 287 intra-articular fractures of the calcaneum in 247 patients treated by minimally-invasive reduction and K-wire fixation between 1994 and 2003. There were 210 men (85%) and 37 women (15%). The most common cause of injury was a fall from a height in 237 patients (96%). Fracture classification was based on the method described by Sanders and Essex-Lopresti. All patients were operated on within 21 days of injury and 89% (220) within 48 hours. The reduction was graded as nearly anatomical (less than 2 mm residual articular displacement and satisfactory overall alignment) in 212 (73.9%) fractures.

There were 20 cases (7%) of superficial pin-track infection and five (1.7%) of deep infection. All healed at a mean of 6 weeks (3 to 19). Loss of reduction was observed in 13 fractures (4.5%) and a musculocutaneous flap was needed in three (1%).

The results were evaluated in 176 patients (205 fractures) with a mean age of 44.3 years (13 to 67), available for follow-up at a mean of 43.4 months (25 to 87) using the Creighton-Nebraska Health Foundation Assessment score. The mean score was 83.9 points (63 to 100). There were 29 (16.5%) excellent, 98 (55.7%) good, 26 (14.8%) fair and 23 (13%) poor results. A total of 130 patients (73.9%) were able to return to their original occupation at a mean of 5.6 months (3.2 to 12.5) after the injury.

Semi-open reduction and percutaneous fixation is an effective treatment for displaced intra-articular fractures of the calcaneum.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 68 - 75
1 Jan 2005
Miedel R Ponzer S Törnkvist H Söderqvist A Tidermark J

We studied 217 patients with an unstable trochanteric or subtrochanteric fracture who had been randomly allocated to treatment by either internal fixation with a standard Gamma nail (SGN) or a Medoff sliding plate (MSP, biaxial dynamisation mode). Their mean age was 84 years (65 to 99) and they were reviewed at four and 12 months after surgery. Assessments of outcome included general complications, technical failures, revision surgery, activities of daily living (ADL), hip function (Charnley score) and the health-related quality of life (HRQOL, EQ-5D).

The rate of technical failure in patients with unstable trochanteric fractures was 6.5% (6/93) (including intra-operative femoral fractures) in the SGN group and 5.2% (5/96) in the MSP group. In patients with subtrochanteric fractures, there were no failures in the SGN group (n = 16) and two in the MSP group (n = 12). In the SGN group, there were intra-operative femoral fractures in 2.8% (3/109) and no post-operative fractures. There was a reduced need for revision surgery in the SGN group compared with the MSP group (8.3%; 9/108; p = 0.072). The SGN group also showed a lower incidence of severe general complications (p < 0.05) and a trend towards a lower incidence of wound infections (p = 0.05). There were no differences between the groups regarding the outcome of ADL, hip function or the HRQOL. The reduction in the HRQOL (EQ-5Dindexscore) was significant in both groups compared with that before the fracture (p < 0.005).

Our findings indicate that the SGN showed good results in both trochanteric and subtrochanteric fractures. The limited number of intra-operative femoral fractures did not influence the outcome or the need for revision surgery. Moreover, the SGN group had a reduced number of serious general complications and wound infections compared with the MSP group. The MSP in the biaxial dynamisation mode had a low rate of failure in trochanteric fractures but an unacceptably high rate when used in the biaxial dynamisation mode in subtrochanteric fractures.

The negative influence of an unstable trochanteric or subtrochanteric fracture on the quality of life was significant regardless of the surgical method.