Advertisement for orthosearch.org.uk
Results 1 - 50 of 75
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 76 - 81
1 Jan 2005
Pajarinen J Lindahl J Michelsson O Savolainen V Hirvensalo E

We treated 108 patients with a pertrochanteric femoral fracture using either the dynamic hip screw or the proximal femoral nail in this prospective, randomised series. We compared walking ability before fracture, intra-operative variables and return to their residence. Patients treated with the proximal femoral nail (n = 42) had regained their pre-operative walking ability significantly (p = 0.04) more often by the four-month review than those treated with the dynamic hip screw (n = 41). Peri-operative or immediate post-operative measures of outcome did not differ between the groups, with the exception of operation time. The dynamic hip screw allowed a significantly greater compression of the fracture during the four-month follow-up, but consolidation of the fracture was comparable between the two groups. Two major losses of reduction were observed in each group, resulting in a total of four revision operations. Our results suggest that the use of the proximal femoral nail may allow a faster postoperative restoration of walking ability, when compared with the dynamic hip screw


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 5 | Pages 652 - 654
1 Jul 2000
Tornetta P Tiburzi D

Retrograde femoral nailing is gaining in popularity. We report a prospective, randomised comparison of antegrade and retrograde procedures in 68 patients with 69 fractures of the femoral shaft. All nails were inserted after appropriate reaming. There was no difference in operating time, blood loss, technical complications, size of nail or reamer, or transfusion requirements. There were more problems of length and rotation using a retrograde technique on a radiolucent table than with an antegrade approach on a fracture table. All fractures in both groups healed and there was no difference in the time taken to achieve union. Although retrograde nailing is a promising technique the skills required need practice. A longer period of follow-up is necessary to determine whether there are long-term problems in the knee after such surgery


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 4 | Pages 660 - 661
1 Jul 1991
Knudsen C Grobler G Close R

We describe a simple method of inserting the distal screws in a locked femoral nail. The method requires no aiming device and no assistant. The only equipment needed is a 3 mm Kirschner wire and an air drill


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 4 | Pages 666 - 667
1 Jul 1995
Bowyer G Clarke N


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 4 | Pages 677 - 678
1 Jul 1994
Dewey P


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 485 - 489
1 May 1998
Clatworthy MG Clark DI Gray DH Hardy AE

We performed a randomised, prospective trial to evaluate the use of unreamed titanium nails for femoral fractures. Of 48 patients with 50 femoral fractures 45 were followed to union; 23 with an unreamed and 22 with a reamed nail. The study was stopped early because of a high rate of implant failure.

The fractures in the unreamed group were slower to unite (39.4 weeks) than those in the reamed group (28.5 weeks; p = 0.007). The time to union was over nine months in 57% of the unreamed group and in 18% of the reamed group.

In the unreamed group 14 secondary procedures were required in ten patients to enhance healing compared with three in three patients in the reamed group. Six implants (13%) failed, three in each group. Four of these six fractures showed evidence of delayed union.

To achieve quicker union and fewer implant failures we recommend the use of reamed nails of at least 12 mm in diameter for female patients and 13 mm in males.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 4 | Pages 689 - 690
1 Jul 1991
Tait G Danton M


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 5 | Pages 922 - 922
1 Sep 1990
Mahaisavariya B Suibnugarn C Laupattarakasem W Kowsuwon W


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 1 | Pages 66 - 68
1 Jan 1993
Mahaisavariya B Laupattarakasem W

We describe a method of closed, unlocked nailing for femoral fractures using ultrasound instead of an image intensifier. Radiography was used only to confirm that the guide wire had been passed into the intramedullary canal of both fragments. The method succeeded in 26 of 30 cases. The failures all occurred in fractures which could not be reduced within 20 minutes. The operating time in those nailed successfully with ultrasound control was not different from the time for 30 control cases using conventional methods with an image intensifier.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 114 - 121
1 Jan 2014
Pekmezci M McDonald E Buckley J Kandemir U

We investigated a new intramedullary locking nail that allows the distal interlocking screws to be locked to the nail. We compared fixation using this new implant with fixation using either a conventional nail or a locking plate in a laboratory simulation of an osteoporotic fracture of the distal femur. A total of 15 human cadaver femora were used to simulate an AO 33-A3 fracture pattern. Paired specimens compared fixation using either a locking or non-locking retrograde nail, and using either a locking retrograde nail or a locking plate. The constructs underwent cyclical loading to simulate single-leg stance up to 125 000 cycles. Axial and torsional stiffness and displacement, cycles to failure and modes of failure were recorded for each specimen. When compared with locking plate constructs, locking nail constructs had significantly longer mean fatigue life (75 800 cycles (. sd. 33 900) vs 12 800 cycles (. sd. 6100); p = 0.007) and mean axial stiffness (220 N/mm (. sd. 80) vs 70 N/mm (. sd. 18); p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (. sd. 0.9) vs 5.1 Nm/° (. sd. 1.5); p = 0.008). In addition, in the nail group the mode of failure was either cut-out of the distal screws or breakage of nails, and in the locking plate group breakage of the plate was always the mode of failure. Locking nail constructs had significantly longer mean fatigue life than non-locking nail constructs (78 900 cycles (. sd. 25 600) vs 52 400 cycles (. sd. 22 500); p = 0.04). The new locking retrograde femoral nail showed better stiffness and fatigue life than locking plates, and superior fatigue life to non-locking nails, which may be advantageous in elderly patients. Cite this article: Bone Joint J 2014;96-B:114–21


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 158 - 165
1 Feb 2024
Nasser AAHH Sidhu M Prakash R Mahmood A

Aims

Periprosthetic fractures (PPFs) around the knee are challenging injuries. This study aims to describe the characteristics of knee PPFs and the impact of patient demographics, fracture types, and management modalities on in-hospital mortality.

Methods

Using a multicentre study design, independent of registry data, we included adult patients sustaining a PPF around a knee arthroplasty between 1 January 2010 and 31 December 2019. Univariate, then multivariable, logistic regression analyses were performed to study the impact of patient, fracture, and treatment on mortality.


The Bone & Joint Journal
Vol. 105-B, Issue 4 | Pages 449 - 454
15 Mar 2023
Zhang C Wang C Duan N Zhou D Ma T

Aims

The aim of this study was to assess the safety and clinical outcome of patients with a femoral shaft fracture and a previous complex post-traumatic femoral malunion who were treated with a clamshell osteotomy and fixation with an intramedullary nail (IMN).

Methods

The study involved a retrospective analysis of 23 patients. All had a previous, operatively managed, femoral shaft fracture with malunion due to hardware failure. They were treated with a clamshell osteotomy between May 2015 and March 2020. The mean age was 42.6 years (26 to 62) and 15 (65.2%) were male. The mean follow-up was 2.3 years (1 to 5). Details from their medical records were analyzed. Clinical outcomes were assessed using the quality of correction of the deformity, functional recovery, the healing time of the fracture, and complications.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1382 - 1388
1 Oct 2016
Laubscher M Mitchell C Timms A Goodier D Calder P

Aims. Patients undergoing femoral lengthening by external fixation tolerate treatment less well when compared to tibial lengthening. Lengthening of the femur with an intramedullary device may have advantages. Patients and Methods. We reviewed all cases of simple femoral lengthening performed at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities, congenital limb deficiencies and lengthening with an unstable hip were excluded, leaving 33 cases (in 22 patients; 11 patients had bilateral procedures) for review. Healing index, implant tolerance and complications were compared. Results. In 20 cases (15 patients) the Precice lengthening nail was used and in 13 cases (seven patients) the LRS external fixator system. The desired length was achieved in all cases in the Precice group and in 12 of 13 cases in the LRS group. The mean healing index was 31.3 days/cm in the Precice and 47.1 days/cm in the LRS group (p < 0.001). This was associated with an earlier ability to bear full weight without aids in the Precice group. There were more complications with LRS lengthening, including pin site infections and regenerate deformity. Implant tolerance and the patients’ perception of the cosmetic result were better with the Precice treatment. Conclusion. Femoral lengthening with the Precice femoral nail achieved excellent functional results with fewer complications and greater patient satisfaction when compared with the LRS system in our patients. Cite this article: Bone Joint J 2016;98-B:1382–8


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 5 | Pages 788 - 790
1 Sep 1995
Mahaisavariya B Songcharoen P Chotigavanich C

We have compared the sonographic findings of six femoral fractures with soft-tissue interposition which required open reduction with those of a control group of 40 other femoral fractures. Ultrasound assessment before operation showed that the fractured end of the proximal fragment had penetrated the quadriceps muscle anteriorly while the distal fragment lay beneath it. Transverse scans showed less soft-tissue thickness over the end of the proximal fragment in the problem cases. Radiographic image intensification did not provide any additional information. Ultrasound is of value in demonstrating soft-tissue interposition at the fracture site before femoral nailing


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 1000 - 1004
1 Jul 2005
Baumgart R Bürklein D Hinterwimmer S Thaller P Mutschler W

Ollier’s disease is characterised by severe deformity of the extremities and retarded growth because of multiple enchondromas. For correction of deformity, the Ilizarov method has been used although it has many complications. A 17-year-old boy with Ollier’s disease had a limb-length discrepancy of 17.4 cm, with a valgus deformity of the right knee and recurvatum of the femur of 23°. He had undergone three unsuccessful attempts to correct the deformities by using external fixators. We used a fully implantable, motorised, lengthening and correction nail (Fitbone) to achieve full correction of all the deformities without complications. We decided to carry out the procedure in three stages. First, we lengthened the femur by 3.6 cm and the tibia by 4 cm. We then exchanged the femoral nail for a longer implant and achieved a further 6 cm of length. This reduced the shortening to 3.8 cm. When the boy has finished secondary school we will adjust the remaining discrepancy


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 1 | Pages 4 - 8
1 Jan 1997
Krettek C Miclau T Blauth M Lindsey RW Donow C Tscherne H

Rotational deformity following intramedullary nailing may cause symptoms and require surgical correction by osteotomy. Reamed, locked intramedullary nailing may be performed, but concern about cortical blood supply and potential pulmonary dysfunction from reaming have led many surgeons to limit this and use smaller diameter nails. Slotted nails are commonly used but are less stiff in torsion than the newer unslotted nails, particularly at the lower diameters. We report two cased of recurrent femoral rotational deformity after using statically interlocked slotted intramedullary nails to correct existing femoral rotational deformities. These patients show that small diameter statically interlocked femoral nails with diminished bone-nail contact must be stiff enough in rotation to avoid potential recurrence


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 86 - 94
1 Jan 2004
Schipper IB Steyerberg EW Castelein RM van der Heijden FHWM den Hoed PT Kerver AJH van Vugt AB

The proximal femoral nail (PFN) is a recently introduced intramedullary system, designed to improve treatment of unstable trochanteric fractures of the hip. In a multicentre prospective clinical study, the intra-operative use, complications and outcome of treatment using the PFN (n = 211) were compared with those using the gamma nail (GN) (n = 213). The intra-operative blood loss was lower with the PFN (220 ml v 287 ml, p = 0.001). Post-operatively, more lateral protrusion of the hip screws of the PFN (7.6%) was documented, compared with the gamma nail (1.6%, p = 0.02). Most local complications were related to suboptimal reduction of the fracture and/or positioning of the implant. Functional outcome and consolidation were equal for both implants. Generally, the results of treatment of unstable trochanteric fractures were comparable for the PFN and GN. The pitfalls and complications were similar, and mainly surgeon- or fracture-related, rather than implant-related


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 2 | Pages 207 - 209
1 Mar 1993
Vangsness C DeCampos J Merritt P Wiss D

We studied 47 patients with closed, displaced, diaphyseal fractures of the femur caused by blunt trauma, to determine the incidence of associated knee injuries, particularly of the meniscus. After femoral nailing, all patients had an examination under anaesthesia and an arthroscopy. There were 12 medial meniscal injuries (5 tears) and 13 injuries of the lateral meniscus (8 tears). Ten of the 13 tears were in the posterior third of the meniscus, and two patients had tears of both menisci. Synovitis was common at the meniscal attachments. Complex and radial tears were more common than peripheral or bucket-handle tears. Examination under anaesthesia revealed ligamentous laxity in 23 patients (49%), but meniscal injuries had a similar incidence in knees with and without ligament injury. Femoral shaft fractures are often associated with injuries to the ipsilateral knee, and a high index of suspicion is necessary to identify these lesions


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 356 - 361
1 Mar 1999
Giannoudis PV Smith RM Bellamy MC Morrison JF Dickson RA Guillou PJ

It has been suggested that reamed intramedullary nailing of the femur should be avoided in some patients with multiple injuries. We have studied prospectively the effect of femoral reaming on the inflammatory process as implicated in the pathogenesis of acute respiratory distress syndrome (ARDS) and multiple-organ failure (MOF). We studied changes in the levels of serum interleukin-6 (IL-6) (proinflammatory cytokine), neutrophil CD11b (C3) receptor expression (activated neutrophil adhesion molecule), serum soluble intracellular adhesion molecule (s-ICAM-1), serum soluble E-selectin (the soluble products of endothelial adhesion molecules) and plasma elastase (neutrophil protease) in a series of patients with femoral fractures treated by nailing. We have also compared reamed nailing with unreamed nailing. We found that the levels of serum IL-6 and elastase rose significantly during the nailing procedure indicating a measurable ‘second hit’. There was no clear response in leukocyte activation and no difference in the release of endothelial adhesion molecule markers. There was no significant difference between groups treated by reamed and unreamed nailing. Although clinically unremarkable, the one patient who died from ARDS was shown to be hyperstimulated after injury and again after nailing, suggesting the importance of an excessive inflammatory reaction in the pathogenesis of these serious problems. Our findings have shown that there is a second hit associated with femoral nailing and suggest that the degree of the inflammatory reaction may be important in the pathogenesis of ARDS and MOF


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 2 | Pages 293 - 296
1 Mar 1994
Simonian P Chapman Selznick H Benirschke S Claudi B Swiontkowski M

We have reviewed our experience of four iatrogenic femoral neck fractures in 315 consecutive closed intramedullary nailings with the AO/ASIF universal femoral nail. The average neck-shaft angle in the bones that fractured was 139.3 degrees +/- 1.2 degrees SD (136 to 141); in the whole series the average neck-shaft angle was 125.3 degrees +/- 8.6 degrees and only 11 had angles of more than 135 degrees. The upper ends of the nails in the four which fractured were 17 mm, 5 mm, 3 mm, and 1 mm below the tip of the greater trochanter. In the seven patients with neck-shaft angles greater than 135 degrees but with no fracture, none of the nails had been inserted beyond the tip of the greater trochanter. We nailed six cadaver femora to simulate the forces produced by intramedullary nailing. Despite deep insertion, only one of the six sustained a neck fracture. This specimen had a radiographic neck-shaft angle of 140 degrees against an average of 127.3 degrees +/- 4.0 degrees for the other five. We believe that the medial prong of the AO insertion jig, with its medial overhang of 2 to 3 mm, may impinge on the superior aspect of a valgus femoral neck during final impaction, causing a neck fracture. This may be avoided by leaving the end of the nail above the tip of the trochanter in such cases


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 775 - 781
1 Apr 2021
Mellema JJ Janssen S Schouten T Haverkamp D van den Bekerom MPJ Ring D Doornberg JN

Aims

This study evaluated variation in the surgical treatment of stable (A1) and unstable (A2) trochanteric hip fractures among an international group of orthopaedic surgeons, and determined the influence of patient, fracture, and surgeon characteristics on choice of implant (intramedullary nailing (IMN) versus sliding hip screw (SHS)).

Methods

A total of 128 orthopaedic surgeons in the Science of Variation Group evaluated radiographs of 30 patients with Type A1 and A2 trochanteric hip fractures and indicated their preferred treatment: IMN or SHS. The management of Type A3 (reverse obliquity) trochanteric fractures was not evaluated. Agreement between surgeons was calculated using multirater kappa. Multivariate logistic regression models were used to assess whether patient, fracture, and surgeon characteristics were independently associated with choice of implant.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 442 - 448
1 Mar 2021
Nikolaou VS Masouros P Floros T Chronopoulos E Skertsou M Babis GC

Aims

The aim of this study was to investigate the hypothesis that a single dose of tranexamic acid (TXA) would reduce blood loss and transfusion rates in elderly patients undergoing surgery for a subcapital or intertrochanteric (IT) fracture of the hip.

Methods

In this single-centre, randomized controlled trial, elderly patients undergoing surgery for a hip fracture, either hemiarthroplasty for a subcapital fracture or intramedullary nailing for an IT fracture, were screened for inclusion. Patients were randomly allocated to a study group using a sealed envelope. The TXA group consisted of 77 patients, (35 with a subcapital fracture and 42 with an IT fracture), and the control group consisted of 88 patients (29 with a subcapital fracture and 59 with an IT fracture). One dose of 15 mg/kg of intravenous (IV) TXA diluted in 100 ml normal saline (NS,) or one dose of IV placebo 100 ml NS were administered before the incision was made. The haemoglobin (Hb) concentration was measured before surgery and daily until the fourth postoperative day. The primary outcomes were the total blood loss and the rate of transfusion from the time of surgery to the fourth postoperative day.


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1168 - 1172
1 Jun 2021
Iliadis AD Wright J Stoddart MT Goodier WD Calder P

Aims

The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device.

Methods

This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1082 - 1087
1 Aug 2020
Yiğit Ş Arslan H Akar MS Şahin MA

Aims

Osteopetrosis (OP) is a rare hereditary disease that causes reduced bone resorption and increased bone density as a result of osteoclastic function defect. Our aim is to review the difficulties, mid-term follow-up results, and literature encountered during the treatment of OP.

Methods

This is a retrospective and observational study containing data from nine patients with a mean age of 14.1 years (9 to 25; three female, six male) with OP who were treated in our hospital between April 2008 and October 2018 with 20 surgical procedures due to 17 different fractures. Patient data included age, sex, operating time, length of stay, genetic type of the disease, previous surgery, fractures, complications, and comorbidity.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1048 - 1055
1 Aug 2020
Cox I Al Mouazzen L Bleibleh S Moldovan R Bintcliffe F Bache CE Thomas S

Aims

The Fassier Duval (FD) rod is a third-generation telescopic implant for children with osteogenesis imperfecta (OI). Threaded fixation enables proximal insertion without opening the knee or ankle joint. We have reviewed our combined two-centre experience with this implant.

Methods

In total, 34 children with a mean age of five years (1 to 14) with severe OI have undergone rodding of 72 lower limb long bones (27 tibial, 45 femoral) for recurrent fractures with progressive deformity despite optimized bone health and bisphosphonate therapy. Data were collected prospectively, with 1.5 to 11 years follow-up.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 530 - 538
1 Apr 2020
Rollick NC Gadinsky NE Klinger CE Kubik JF Dyke JP Helfet DL Wellman DS

Aims

Dual plating of distal femoral fractures with medial and lateral implants has been performed to improve construct mechanics and alignment, in cases where isolated lateral plating would be insufficient. This may potentially compromise vascularity, paradoxically impairing healing. This study investigates effects of single versus dual plating on distal femoral vascularity.

Methods

A total of eight cadaveric lower limb pairs were arbitrarily assigned to either 1) isolated lateral plating, or 2) lateral and medial plating of the distal femur, with four specimens per group. Contralateral limbs served as matched controls. Pre- and post-contrast MRI was performed to quantify signal intensity enhancement in the distal femur. Further evaluation of intraosseous vascularity was done with barium sulphate infusion with CT scan imaging. Specimens were then injected with latex medium and dissection was completed to assess extraosseous vasculature.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 394 - 399
1 Mar 2020
Parker MJ Cawley S

Aims

A lack of supporting clinical studies have been published to determine the ideal length of intramedullary nail in fixation of trochanteric fractures of the hip. Nevertheless, there has been a trend to use shorter intramedullary nails for the internal fixation of trochanteric hip fractures. Our aim was to determine if the length of nail affected the outcome.

Methods

We randomized 229 patients with a trochanteric hip fracture between two implants: a ‘standard’ nail of 220 mm and a shorter nail of 175 mm, which had decreased proximal angulation (4° vs 7°) and a reduced diameter at the level of the lesser trochanter. Patients were followed up for one year by a nurse blinded to the type of implant used to determine if there were differences in mobility and pain with two nail designs. Pain was assessed on a scale of 1 (none) to 8 (severe and constant) and mobility on a scale of 1 (full mobility) to 9 (immobile).


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1408 - 1415
1 Nov 2019
Hull PD Chou DTS Lewis S Carrothers AD Queally JM Allison A Barton G Costa ML

Aims

The aim of this study was to assess the feasibility of conducting a full-scale, appropriately powered, randomized controlled trial (RCT) comparing internal fracture fixation and distal femoral replacement (DFR) for distal femoral fractures in older patients.

Patients and Methods

Seven centres recruited patients into the study. Patients were eligible if they were greater than 65 years of age with a distal femoral fracture, and if the surgeon felt that they were suitable for either form of treatment. Outcome measures included the patients’ willingness to participate, clinicians’ willingness to recruit, rates of loss to follow-up, the ability to capture data, estimates of standard deviation to inform the sample size calculation, and the main determinants of cost. The primary clinical outcome measure was the EuroQol five-dimensional index (EQ-5D) at six months following injury.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1087 - 1092
1 Sep 2019
Garceau S Warschawski Y Dahduli O Alshaygy I Wolfstadt J Backstein D

Aims

The aim of this study was to assess the effects of transferring patients to a specialized arthroplasty centre between the first and second stages (interstage) of prosthetic joint infection (PJI) of the knee.

Patients and Methods

A search of our institutional database was performed to identify patients having undergone two-stage revision total knee arthroplasty (TKA) for PJI. Two cohorts were created: continuous care (CC) and transferred care (TC). Baseline characteristics and outcomes were collected and compared between cohorts.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1122 - 1128
1 Sep 2019
Yombi JC Putineanu DC Cornu O Lavand’homme P Cornette P Castanares-Zapatero D

Aims

Low haemoglobin (Hb) at admission has been identified as a risk factor for mortality for elderly patients with hip fractures in some studies. However, this remains controversial. This study aims to analyze the association between Hb level at admission and mortality in elderly patients with hip fracture undergoing surgery.

Patients and Methods

All consecutive patients (prospective database) admitted with hip fracture operated in a tertiary hospital between 2012 and 2016 were analyzed. We collected patient characteristics, time to surgery, duration and type of surgery, comorbidities, Hb at admission, nadir of Hb after surgery, the use and amount of red blood cells (RBCs) transfusion products, postoperative complications, and death. The main outcome measures were mortality at 30 days, 90 days, 180 days, and one year after surgery.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1168 - 1176
1 Sep 2019
Calder PR McKay JE Timms AJ Roskrow T Fugazzotto S Edel P Goodier WD

Aims

The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients.

Patients and Methods

A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 4 | Pages 673 - 678
1 Jul 1998
Yang KH Han DY Park HW Kang HJ Park JH

We report retrospective and prospective studies to identify the causes of fracture of the femoral neck associated with femoral shaft nailing on the same side. Of a total of 14 neck fractures in a series of 152 shaft nailings, eight were not visible on the initial pelvic radiographs. We used CT scans before and after operation, and fluoroscopy during the procedure in our prospective series, and reviewed abdominal CT scans retrospectively with the window set to bone level. Six of the eight undisplaced fractures were shown to have been present before operation, but two were iatrogenic. We recommend the preoperative use of CT scans of the femoral neck in high-risk patients such as those with associated fractures of the acetabulum, the distal femur or the patella. Early diagnosis will allow better general management and early fixation of the neck fracture


The Bone & Joint Journal
Vol. 100-B, Issue 5 | Pages 634 - 639
1 May 2018
Davda K Heidari N Calder P Goodier D

Aims

The management of a significant bony defect following excision of a diaphyseal atrophic femoral nonunion remains a challenge. We present the outcomes using a combined technique of acute femoral shortening, stabilized with a long retrograde intramedullary nail, accompanied by bifocal osteotomy compression and distraction osteogenesis with a temporary monolateral fixator.

Patients and Methods

Eight men and two women underwent the ‘rail and nail’ technique between 2008 and 2016. Proximal locking of the nail and removal of the external fixator was undertaken once the length of the femur had been restored and prior to full consolidation of the regenerate.


The Bone & Joint Journal
Vol. 100-B, Issue 5 | Pages 559 - 565
1 May 2018
Bartlett JD Lawrence JE Stewart ME Nakano N Khanduja V

Aims

The aim of this study was to assess the current evidence relating to the benefits of virtual reality (VR) simulation in orthopaedic surgical training, and to identify areas of future research.

Materials and Methods

A literature search using the MEDLINE, Embase, and Google Scholar databases was performed. The results’ titles, abstracts, and references were examined for relevance.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims

Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator.

Patients and Methods

A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 283 - 288
1 Feb 2017
Hughes A Heidari N Mitchell S Livingstone J Jackson M Atkins R Monsell F

Aims

Computer hexapod assisted orthopaedic surgery (CHAOS), is a method to achieve the intra-operative correction of long bone deformities using a hexapod external fixator before definitive internal fixation with minimally invasive stabilisation techniques.

The aims of this study were to determine the reliability of this method in a consecutive case series of patients undergoing femoral deformity correction, with a minimum six-month follow-up, to assess the complications and to define the ideal group of patients for whom this treatment is appropriate.

Patients and Methods

The medical records and radiographs of all patients who underwent CHAOS for femoral deformity at our institution between 2005 and 2011 were retrospectively reviewed. Records were available for all 55 consecutive procedures undertaken in 49 patients with a mean age of 35.6 years (10.9 to 75.3) at the time of surgery.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 817 - 822
1 Jun 2014
Al-Nammari SS Dawson-Bowling S Amin A Nielsen D

Conventional methods of treating ankle fractures in the elderly are associated with high rates of complication. We describe the results of treating these injuries in 48 frail elderly patients with a long calcaneotalotibial nail.

The mean age of the group was 82 years (61 to 96) and 41 (85%) were women. All were frail, with multiple medical comorbidities and their mean American Society of Anaesthesiologists score was 3 (3 to 4). None could walk independently before their operation. All the fractures were displaced and unstable; the majority (94%, 45 of 48) were low-energy injuries and 40% (19 of 48) were open.

The overall mortality at six months was 35%. Of the surviving patients, 90% returned to their pre-injury level of function. The mean pre- and post-operative Olerud and Molander questionnaire scores were 62 and 57 respectively. Complications included superficial infection (4%, two of 48); deep infection (2%, one of 48); a broken or loose distal locking screw (6%, three of 48); valgus malunion (4%, two of 48); and one below-knee amputation following an unsuccessful vascular operation. There were no cases of nonunion, nail breakage or peri-prosthetic fracture.

A calcaneotalotibial nail is an excellent device for treating an unstable fracture of the ankle in the frail elderly patient. It allows the patient to mobilise immediately and minimises the risk of bone or wound problems. A long nail which crosses the isthmus of the tibia avoids the risk of peri-prosthetic fracture associated with shorter devices.

Cite this article: Bone Joint J 2014; 96-B:817–22.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 814 - 817
1 Jun 2015
Bose D Kugan R Stubbs D McNally M

Infected nonunion of a long bone continues to present difficulties in management. In addition to treating the infection, it is necessary to establish bony stability, encourage fracture union and reconstruct the soft-tissue envelope.

We present a series of 67 infected nonunions of a long bone in 66 patients treated in a multidisciplinary unit. The operative treatment of patients suitable for limb salvage was performed as a single procedure. Antibiotic regimes were determined by the results of microbiological culture.

At a mean follow-up of 52 months (22 to 97), 59 patients (88%) had an infection-free united fracture in a functioning limb. Seven others required amputation (three as primary treatment, three after late failure of limb salvage and one for recalcitrant pain after union).

The initial operation achieved union in 54 (84%) of the salvaged limbs at a mean of nine months (three to 26), with recurrence of infection in 9%. Further surgery in those limbs that remained ununited increased the union rate to 62 (97%) of the 64 limbs treated by limb salvage at final follow-up. The use of internal fixation was associated with a higher risk of recurrent infection than external fixation.

Cite this article: Bone Joint J 2015; 97-B:814–17.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1073 - 1078
1 Aug 2008
Little NJ Verma V Fernando C Elliott DS Khaleel A

We compared the outcome of patients treated for an intertrochanteric fracture of the femoral neck with a locked, long intramedullary nail with those treated with a dynamic hip screw (DHS) in a prospective randomised study.

Each patient who presented with an extra-capsular hip fracture was randomised to operative stabilisation with either a long intramedullary Holland nail or a DHS. We treated 92 patients with a Holland nail and 98 with a DHS. Pre-operative variables included the Mini Mental test score, patient mobility, fracture pattern and American Society of Anesthesiologists grading. Peri-operative variables were anaesthetic time, operating time, radiation time and blood loss. Post-operative variables were time to mobilising with a frame, wound infection, time to discharge, time to fracture union, and mortality.

We found no significant difference in the pre-operative variables. The mean anaesthetic and operation times were shorter in the DHS group than in the Holland nail group (29.7 vs 40.4 minutes, p < 0.001; and 40.3 vs 54 minutes, p < 0.001, respectively). There was an increased mean blood loss within the DHS group versus the Holland nail group (160 ml vs 78 ml, respectively, p < 0.001). The mean time to mobilisation with a frame was shorter in the Holland nail group (DHS 4.3 days, Holland nail 3.6 days, p = 0.012). More patients needed a post-operative blood transfusion in the DHS group (23 vs seven, p = 0.003) and the mean radiation time was shorter in this group (DHS 0.9 minutes vs Holland nail 1.56 minutes, p < 0.001). The screw of the DHS cut out in two patients, one of whom underwent revision to a Holland nail. There were no revisions in the Holland nail group. All fractures in both groups were united when followed up after one year.

We conclude that the DHS can be implanted more quickly and with less exposure to radiation than the Holland nail. However, the resultant blood loss and need for transfusion is greater. The Holland nail allows patients to mobilise faster and to a greater extent. We have therefore adopted the Holland nail as our preferred method of treating intertrochanteric fractures of the hip.


The Bone & Joint Journal
Vol. 98-B, Issue 2 | Pages 266 - 270
1 Feb 2016
Stevenson JD McNair M Cribb GL Cool WP

Aims

Surgical intervention in patients with bone metastases from breast cancer is dependent on the estimated survival of the patient. The purpose of this paper was to identify factors that would predict survival so that specific decisions could be made in terms of surgical (or non-surgical) management.

Methods

The records of 113 consecutive patients (112 women) with metastatic breast cancer were analysed for clinical, radiological, serological and surgical outcomes. Their median age was 61 years (interquartile range 29 to 90) and the median duration of follow-up was 1.6 years (standard deviation (sd) 1.9, 95% confidence interval (CI) 0 to 5.9). The cumulative one- and five-year rates of survival were 68% and 16% (95% Cl 60 to 77 and 95% CI 10 to 26, respectively).


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 393 - 399
1 Mar 2008
Morley JR Smith RM Pape HC MacDonald DA Trejdosiewitz LK Giannoudis PV

We have undertaken a prospective study in patients with a fracture of the femoral shaft requiring intramedullary nailing to test the hypothesis that the femoral canal could be a potential source of the second hit phenomenon. We determined the local femoral intramedullary and peripheral release of interleukin-6 (IL-6) after fracture and subsequent intramedullary reaming.

In all patients, the fracture caused a significant increase in the local femoral concentrations of IL-6 compared to a femoral control group. The concentration of IL-6 in the local femoral environment was significantly higher than in the patients own matched blood samples from their peripheral circulation. The magnitude of the local femoral release of IL-6 after femoral fracture was independent of the injury severity score and whether the fracture was closed or open.

In patients who underwent intramedullary reaming of the femoral canal a further significant local release of IL-6 was demonstrated, providing evidence that intramedullary reaming can cause a significant local inflammatory reaction.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 658 - 664
1 May 2014
Teo BJX Koh JSB Goh SK Png MA Chua DTC Howe TS

Management of bisphosphonate-associated subtrochanteric fractures remains opinion- or consensus-based. There are limited data regarding the outcomes of this fracture.

We retrospectively reviewed 33 consecutive female patients with a mean age of 67.5 years (47 to 91) who were treated surgically between May 2004 and October 2009. The mean follow-up was 21.7 months (0 to 53). Medical records and radiographs were reviewed to determine the post-operative ambulatory status, time to clinical and radiological union and post-fixation complications such as implant failure and need for second surgery.

The predominant fixation method was with an extramedullary device in 23 patients. 25 (75%) patients were placed on wheelchair mobilisation or no weight-bearing initially. The mean time to full weight-bearing was 7.1 months (2.2 to 29.7). The mean time for fracture site pain to cease was 6.2 months (1.2 to 17.1). The mean time to radiological union was 10.0 months (2.2 to 27.5). Implant failure was seen in seven patients (23%, 95 confidence interval (CI) 11.8 to 40.9). Revision surgery was required in ten patients (33%, 95 CI 19.2 to 51.2).

A large proportion of the patients required revision surgery and suffered implant failure. This fracture is associated with slow healing and prolonged post-operative immobility.

Cite this article: Bone Joint J 2014;96-B:658–64.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1406 - 1409
1 Oct 2013
Wähnert D Lange JH Schulze M Gehweiler D Kösters C Raschke MJ

The augmentation of fixation with bone cement is increasingly being used in the treatment of severe osteoporotic fractures. We investigated the influence of bone quality on the mechanics of augmentation of plate fixation in a distal femoral fracture model (AO 33 A3 type). Eight osteoporotic and eight non-osteoporotic femoral models were randomly assigned to either an augmented or a non-augmented group. Fixation was performed using a locking compression plate. In the augmented group additionally 1 ml of bone cement was injected into the screw hole before insertion of the screw. Biomechanical testing was performed in axial sinusoidal loading. Augmentation significantly reduced the cut-out distance in the osteoporotic models by about 67% (non-augmented mean 0.30 mm (sd 0.08) vs augmented 0.13 mm (sd 0.06); p = 0.017). There was no statistical reduction in this distance following augmentation in the non-osteoporotic models (non-augmented mean 0.15 mm (sd 0.02) vs augmented 0.15 mm (sd 0.07); p = 0.915). In the osteoporotic models, augmentation significantly increased stability (p = 0.017).

Cite this article: Bone Joint J 2013;95-B:1406–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 576 - 580
1 May 2006
Katsoulis E Court-Brown C Giannoudis PV


The Bone & Joint Journal
Vol. 95-B, Issue 6 | Pages 764 - 769
1 Jun 2013
Roche JJW Jones CDS Khan RJK Yates PJ

The piriformis muscle is an important landmark in the surgical anatomy of the hip, particularly the posterior approach for total hip replacement (THR). Standard orthopaedic teaching dictates that the tendon must be cut in to allow adequate access to the superior part of the acetabulum and the femoral medullary canal. However, in our experience a routine THR can be performed through a posterior approach without sacrificing this tendon.

We dissected the proximal femora of 15 cadavers in order to clarify the morphological anatomy of the piriformis tendon. We confirmed that the tendon attaches on the crest of the greater trochanter, in a position superior to the trochanteric fossa, away from the entry point for broaching the intramedullary canal during THR. The tendon attachment site encompassed the summit and medial aspect of the greater trochanter as well as a variable attachment to the fibrous capsule of the hip joint. In addition we dissected seven cadavers resecting all posterior attachments except the piriformis muscle and tendon in order to study their relations to the hip joint, as the joint was flexed. At flexion of 90° the piriformis muscle lay directly posterior to the hip joint.

The piriform fossa is a term used by orthopaedic surgeons to refer the trochanteric fossa and normally has no relation to the attachment site of the piriformis tendon. In hip flexion the piriformis lies directly behind the hip joint and might reasonably be considered to contribute to the stability of the joint.

We conclude that the anatomy of the piriformis muscle is often inaccurately described in the current surgical literature and terms are used and interchanged inappropriately.

Cite this article: Bone Joint J 2013;95-B:764–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 639 - 643
1 May 2011
Burghardt RD Herzenberg JE Specht SC Paley D

Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device.

The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1198 - 1201
1 Sep 2014
Magu NK Singla R Rohilla R Gogna P Mukhopadhyay R Singh A

We report the outcome of 39 patients who underwent a modified Pauwels’ intertrochanteric osteotomy for nonunion of a femoral neck fracture following failed osteosynthesis. There were 31 men and eight women with a mean age of 47.2 years (34 to 59). By Pauwels’ classification, there were 11 Type II fractures and 28 Type III fractures. The mean follow-up was 7.9 years (2 to 19). In the 11 patients whose initial treatment had been osteotomy, union was achieved in nine (81.8%). In 28 patients whose initial treatment had been with a lag screw or a dynamic hip screw, union was achieved in 27 (96.4%). Limb lengths were equalised in 14 of 16 patients (87.5%) with pre-operative shortening. The mean neck-shaft angle improved significantly from 100.5° (80° to 120°) to 131.6° (120° to 155°) (p = 0.004). The mean modified Harris hip score was 85.6 points (70 to 97) and the mean modified Merle d’Aubigné score was 14.3 (11 to 18). Good to excellent functional outcomes were achieved in 32 patients (88.8%). A modified Pauwels’ intertrochanteric osteotomy is a reliable method of treating ununited fractures of the femoral neck following failed osteosynthesis: coxa vara and shortening can also simultaneously be addressed.

Cite this article: Bone Joint J 2014;96-B:1198–1201.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 23 - 27
1 Jan 2012
Uzoigwe CE Middleton RG

Radiological imaging is necessary in a wide variety of trauma and elective orthopaedic operations. The evolving orthopaedic workforce includes an increasing number of pregnant workers. Current legislation in the United Kingdom, Europe and United States allows them to choose their degree of participation, if any, with fluoroscopic procedures. For those who wish to engage in radiation-prone procedures, specific regulations apply to limit the radiation dose to the pregnant worker and unborn child.

This paper considers those aspects of radiation protection, the potential effects of exposure to radiation in pregnancy and the dose of radiation from common orthopaedic procedures, which are important for safe clinical practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 722 - 725
1 Jun 2011
Buckley R Leighton R Trask K

The Canadian Orthopaedic Trauma Society was started in an endeavour to answer the difficult problem of obtaining enough patients to perform top-quality research into fractures. By maintaining a high standard, including randomised study design, inclusivity, open discussion among surgeons and excellent long-term follow-up, this group has become a leader in the orthopaedic research community. This annotation describes the short history, important components and spirit necessary to build a research community or team which will function well despite the difficult research environment facing individual surgeons.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1144 - 1148
1 Aug 2013
Sternheim A Saidi K Lochab J O’Donnell PW Eward WC Griffin A Wunder JS Ferguson P

We investigated the clinical outcome of internal fixation for pathological fracture of the femur after primary excision of a soft-tissue sarcoma that had been treated with adjuvant radiotherapy.

A review of our database identified 22 radiation-induced fractures of the femur in 22 patients (seven men, 15 women). We noted the mechanism of injury, fracture pattern and any complications after internal fixation, including nonunion, hardware failure, secondary fracture or deep infection.

The mean age of the patients at primary excision of the tumour was 58.3 years (39 to 86). The mean time from primary excision to fracture was 73.2 months (2 to 195). The mean follow-up after fracture fixation was 65.9 months (12 to 205). Complications occurred in 19 patients (86%). Nonunion developed in 18 patients (82%), of whom 11 had a radiological nonunion at 12 months, five a nonunion and hardware failure and two an infected nonunion. One patient developed a second radiation-associated fracture of the femur after internal fixation and union of the initial fracture. A total of 13 patients (59%) underwent 24 revision operations.

Internal fixation of a pathological fracture of the femur after radiotherapy for a soft-tissue sarcoma has an extremely high rate of complication and requires specialist attention.

Cite this article: Bone Joint J 2013;95-B:1144–8.