Aseptic loosening of the femoral component is
an important indication for revision surgery in unicompartmental knee
replacement (UKR). A new design of femoral component with an additional
peg was introduced for the cemented Oxford UKR to increase its stability.
The purpose of this study was to compare the primary stability of
the two designs of component. Medial Oxford UKR was performed in 12 pairs of human cadaver
knees. In each pair, one knee received the single peg and one received
the twin peg design. Three dimensional micromotion and subsidence
of the component in relation to the bone was measured under cyclical
loading at flexion of 40° and 70° using an optical measuring system.
Wilcoxon matched pairs signed-rank test was performed to detect
differences between the two groups. There was no significant difference in the relative micromotion
(p = 0.791 and 0.380, respectively) and subsidence (p = 0.301 and
0.176, respectively) of the component between the two groups at
both angles of flexion. Both designs of component offered good strength
of fixation in this cadaver study. Cite this article:
Aims. Many surgeons choose to perform total knee arthroplasty (TKA) surgery with the aid of a tourniquet. A tourniquet is a device that fits around the leg and restricts blood flow to the limb. There is a need to understand whether tourniquets are safe, and if they benefit, or harm, patients. The aim of this study was to determine the benefits and harms of tourniquet use in TKA surgery. Methods. We searched MEDLINE, EMBASE, Cochrane Central Register of Controlled trials, and trial registries up to 26 March 2020. We included randomized controlled trials (RCTs), comparing TKA with a tourniquet versus without a tourniquet. Outcomes included: pain, function, serious adverse events (SAEs), blood loss,
The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components. Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).Aims
Methods
The Unified Classification System (UCS), or Vancouver system, is a validated and widely used classification system to guide the management of periprosthetic femoral fractures. It suggests that well-fixed stems (type B1) can be treated with fixation but that loose stems (types B2 and B3) should be revised. Determining whether a stem is loose can be difficult and some authors have questioned how to apply this classification system to polished taper slip stems which are, by definition, loose within their cement mantle. Recent evidence has challenged the common perception that revision surgery is preferable to fixation surgery for UCS-B periprosthetic fractures around cemented polished taper slip stems. Indications for fixation include an anatomically reducible fracture and cement mantle, a well-fixed femoral bone-cement interface, and a well-functioning acetabular component. However, not all type B fractures can or should be managed with fixation due to the risk of early failure. This annotation details specific fracture patterns that should not be managed with fixation alone. Cite this article:
To investigate the extent of bone development around the scaffold of custom triflange acetabular components (CTACs) over time. We performed a single-centre historical prospective cohort study, including all patients with revision THA using the aMace CTAC between January 2017 and March 2021. A total of 18 patients (18 CTACs) were included. Models of the hemipelvis and the scaffold component of the CTACs were created by segmentation of CT scans. The CT scans were performed immediately postoperatively and at least one year after surgery. The amount of bone in contact with the scaffold was analyzed at both times, and the difference was calculated.Aims
Methods
The first aim of this study was to evaluate whether preoperative renal function is associated with postoperative changes in whole blood levels of metal ions in patients who have undergone a Birmingham Hip Resurfacing (BHR) arthroplasty with a metal-on-metal bearing. The second aim was to evaluate whether exposure to increased cobalt (Co) and chromium (Cr) levels for ten years adversely affected renal function. As part of a multicentre, prospective post-approval study, whole blood samples were sent to a single specialized laboratory to determine Co and Cr levels, and the estimated glomerular filtration rate (eGFR). The study included patients with 117 unrevised unilateral BHRs. There were 36 females (31%). The mean age of the patients at the time of surgery of 51.3 years (SD 6.5), and they all had preoperative one-, four-, five-, and ten-year laboratory data. The mean follow-up was 10.1 years (SD 0.2).Aims
Methods
Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.Aims
Methods
The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling. Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.Aims
Methods
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation. A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.Aims
Methods
This systematic review aims to compare the precision of component positioning, patient-reported outcome measures (PROMs), complications, survivorship, cost-effectiveness, and learning curves of MAKO robotic arm-assisted unicompartmental knee arthroplasty (RAUKA) with manual medial unicompartmental knee arthroplasty (mUKA). Searches of PubMed, MEDLINE, and Google Scholar were performed in November 2021 according to the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “unicompartmental”, “knee”, and “arthroplasty”. Published clinical research articles reporting the learning curves and cost-effectiveness of MAKO RAUKA, and those comparing the component precision, functional outcomes, survivorship, or complications with mUKA, were included for analysis.Aims
Methods
Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA. A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures.Aims
Methods
Arthroplasty is being increasingly used for the management of distal humeral fractures (DHFs) in elderly patients. Arthroplasty options include total elbow arthroplasty (TEA) and hemiarthroplasty (HA); both have unique complications and there is not yet a consensus on which implant is superior. This systematic review asked: in patients aged over 65 years with unreconstructable DHFs, what differences are there in outcomes, as measured by patient-reported outcome measures (PROMs), range of motion (ROM), and complications, between distal humeral HA and TEA? A systematic review of the literature was performed via a search of MEDLINE and Embase. Two reviewers extracted data on PROMs, ROM, and complications. PROMs and ROM results were reported descriptively and a meta-analysis of complications was conducted. Quality of methodology was assessed using Wylde’s non-summative four-point system. The study was registered with PROSPERO (CRD42021228329).Aims
Methods
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase
The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up. A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.Aims
Methods
We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty. We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.Aims
Methods
This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems. Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.Aims
Methods
The aims of this study were to develop an in vivo model of periprosthetic joint infection (PJI) in cemented hip hemiarthroplasty, and to monitor infection and biofilm formation in real-time. Sprague-Dawley rats underwent cemented hip hemiarthroplasty via the posterior approach with pre- and postoperative gait assessments. Infection with Aims
Methods
The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported.Aims
Methods
We studied the bone mineral density (BMD) and
the bone mineral content (BMC) of the proximal tibia in patients with
a well-functioning uncemented Oxford medial compartment arthroplasty
using the Lunar iDXA bone densitometer. Our hypothesis was that
there would be decreased BMD and BMC adjacent to the tibial base
plate and increased BMD and BMC at the tip of the keel. There were 79 consecutive patients (33 men, 46 women) with a
mean age of 65 years (44 to 84) with a minimum two-year follow-up
(mean 2.6 years (2.0 to 5.0)) after unilateral arthroplasty, who
were scanned using a validated standard protocol where seven regions
of interest (ROI) were examined and compared with the contralateral
normal knee. All had well-functioning knees with a mean Oxford knee
score of 43 (14 to 48) and mean Knee Society function score of 90
(20 to 100), showing a correlation with the increasing scores and
higher BMC and BMD values in ROI 2 in the non-implanted knee relative
to the implanted knee (p = 0.013 and p = 0.015, respectively). The absolute and percentage changes in BMD and BMC were decreased
in all ROIs in the implanted knee compared with the non-implanted
knee, but this did not reach statistical significance. Bone loss
was markedly less than reported losses with total knee replacement. There was no significant association with side, although there
was a tendency for the BMC to decrease with age in men. The BMC
was less in the implanted side relative to the non-implanted side
in men compared with women in ROI 2 (p = 0.027), ROI 3 (p = 0.049)
and ROI 4 (p = 0.029). The uncemented Oxford medial compartment arthroplasty appears
to allow relative preservation of the BMC and BMD of the proximal
tibia, suggesting that the implant acts more physiologically than
total knee replacement. Peri-prosthetic bone loss is an important
factor in assessing long-term
It has been hypothesized that proximal radial neck resorption (PRNR) following press-fit radial head arthroplasty (RHA) is due to stress-shielding. We compared two different press-fit stems by means of radiographs to investigate whether the shape and size of the stems are correlated with the degree of PRNR. The radiographs of 52 RHAs were analyzed both at 14 days postoperatively and after two years. A cylindrical stem and a conical stem were implanted in 22 patients (group 1) and 30 patients (group 2), respectively. The PRNR was measured in the four quadrants of the radial neck and the degree of stem filling was calculated by analyzing the ratio between the prosthetic stem diameter (PSD) and the medullary canal diameter (MCD) at the proximal portion of the stem (level A), halfway along the stem length (level B), and distally at the stem tip (level C).Aims
Methods
Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising coating for metallic surfaces, and several types of silver-coated arthroplasty implants are in clinical use today. However, silver can also exert toxic effects on eukaryotic cells both in the immediate vicinity of the coated implants and systemically. In most clinically-used implants, silver coatings are applied on bulk components that are not in direct contact with bone, such as in partial or total long bone arthroplasties used in tumour or complex revision surgery. These implants differ considerably in the coating method, total silver content, and silver release rates. Safety issues, such as the occurrence of argyria, have been a cause for concern, and the efficacy of silver coatings in terms of preventing PJI is also controversial. The application of silver coatings is uncommon on parts of implants intended for cementless fixation in host bone, but this option might be highly desirable since the modification of implant surfaces in order to improve osteoconductivity can also increase bacterial adhesion. Therefore, an optimal silver content that inhibits bacterial colonization while maintaining osteoconductivity is crucial if silver were to be applied as a coating on parts intended for bone contact. This review summarizes the different methods used to apply silver coatings to arthroplasty components, with a focus on the amount and duration of silver release from the different coatings; the available experience with silver-coated implants that are in clinical use today; and future strategies to balance the effects of silver on bacteria and eukaryotic cells, and to develop silver-coated titanium components suitable for bone ingrowth. Cite this article:
Optimal exposure through the direct anterior approach (DAA) for total hip arthroplasty (THA) conducted on a regular operating theatre table is achieved with a standardized capsular releasing sequence in which the anterior capsule can be preserved or resected. We hypothesized that clinical outcomes and implant positioning would not be different in case a capsular sparing (CS) technique would be compared to capsular resection (CR). In this prospective trial, 219 hips in 190 patients were randomized to either the CS (n = 104) or CR (n = 115) cohort. In the CS cohort, a medial based anterior flap was created and sutured back in place at the end of the procedure. The anterior capsule was resected in the CR cohort. Primary outcome was defined as the difference in patient-reported outcome measures (PROMs) after one year. PROMs (Harris Hip Score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), and Short Form 36 Item Health Survey (SF-36)) were collected preoperatively and one year postoperatively. Radiological parameters were analyzed to assess implant positioning and implant ingrowth. Adverse events were monitored.Aims
Methods
Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.Aims
Methods
The osseo-integration of an uncemented acetabular component depends on its initial stability. This is usually provided by under-reaming of the acetabulum. We have assessed the fixation of 52 mm porous-coated hemispherical prostheses inserted into cadaveric acetabula under-reamed by 1, 2, 3 and 4 mm. We tested the torsional stability of fixation, after preloading with 686 N in compression, by measuring the torque required to produce 1 degree and 2 degrees of rotation. Under-reaming by 2 mm and 3 mm gave significantly better fixation than 1 mm (p less than 0.01, p less than 0.02). Insertion after under-reaming of 4 mm caused some fractures. To obtain maximum interference fit and optimal
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions. Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.Aims
Methods
This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an uncemented component in acetabular revisions performed in a standardized manner since 1993. Between 1993 and 2013, 370 patients with a median age of 72 years (interquartile range (IQR) 63 to 79 years) underwent acetabular revision surgery. Of these, 229 were more than ten years following surgery and 137 were more than 15 years. All revisions were performed with extensive use of morcellized allograft firmly impacted into the entire acetabular cavity, followed by insertion of an uncemented component with supplementary screw fixation. All types of reoperation were captured using review of radiographs and medical charts, combined with data from the local surgical register and the Swedish Hip Arthroplasty Register.Aims
Methods
We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression.Aims
Patients and Methods
Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell ( This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m2 and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (Aims
Patients and Methods
We analyzed the acetabular morphology of Crowe type IV hips using CT data to identify a landmark for the ideal placement of the centre of the acetabular component, as assessed by morphometric geometrical analysis, and its reliability. A total of 52 Crowe IV hips (42 patients; seven male, 35 female; mean age 68.5 years (32 to 82)) and 50 normal hips (50 patients; eight male, 42 female; mean age 60.7 years (34 to 86)) undergoing total hip arthroplasty were retrospectively identified. In this CT-based simulation study, the acetabular component was positioned at the true acetabulum with a radiological inclination of 40° and anteversion of 20°. Acetabular shape and the position of the centre of the acetabular component were analyzed by morphometric geometrical analysis using the generalized Procrustes analysis.Aims
Patients and Methods
Severe acetabular bone loss and pelvic discontinuity (PD) present particular challenges in revision total hip arthroplasty. To deal with such complex situations, cup-cage reconstruction has emerged as an option for treating this situation. We aimed to examine our success in using this technique for these anatomical problems. We undertook a retrospective, single-centre series of 35 hips in 34 patients (seven male, 27 female) treated with a cup-cage construct using a trabecular metal shell in conjunction with a titanium cage, for severe acetabular bone loss between 2011 and 2015. The mean age at the time of surgery was 70 years (42 to 85) and all patients had an acetabular defect graded as Paprosky Type 2C through to 3B, with 24 hips (69%) having PD. The mean follow-up was 47 months (25 to 84).Aims
Patients and Methods
This study aimed to evaluate implant survival of reverse hybrid
total hip arthroplasty (THA) at medium-term follow-up. A consecutive series of 1082 THAs in 982 patients with mean follow-up
of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21
to 94). Of these, 194 (17.9%) were in patients under 60 years, 663
(61.3%) in female patients and 348 (32.2%) performed by a trainee.
Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%).
Survival analysis was performed and subgroups compared using log
rank tests.Aims
Patients and Methods
Advocates of debridement, antibiotics and implant retention (DAIR)
in hip periprosthetic joint infection (PJI) argue that a procedure
not disturbing a sound prosthesis-bone interface is likely to lead
to better survival and functional outcome compared with revision.
This case-control study aims were to compare outcome of DAIRs for
infected primary total hip arthroplasty (THA) with outcomes following
primary THA and two-stage revision of infected primary THAs. We retrospectively reviewed all DAIRs, performed for confirmed
infected primary hip arthropasty (n = 82) at out institution, between
1997 and 2013. Data recorded included full patient information and
type of surgery. Outcome measures included complications, mortality,
implant survivorship and functional outcome. Outcome was compared with
two control groups matched for gender and age; a cohort of primary
THAs (n = 120) and a cohort of two-stage revisions for infection
(n = 66).Aims
Patients and Methods
The purpose of this study was to compare the clinical and radiographic
outcomes of total ankle arthroplasty (TAA) in patients with pre-operatively
moderate and severe arthritic varus ankles to those achieved for
patients with neutral ankles. A total of 105 patients (105 ankles), matched for age, gender,
body mass index, and follow-up duration, were divided into three
groups by pre-operative coronal plane tibiotalar angle; neutral
(<
5°), moderate (5° to 15°) and severe (>
15°) varus deformity.
American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot
score, a visual analogue scale (VAS), and Short Form (SF)-36 score
were used to compare the clinical outcomes after a mean follow-up period
of 51 months (24 to 147).Aims
Patients and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
The Unified Classification System (UCS) emphasises
the key principles in the assessment and management of peri-prosthetic
fractures complicating partial or total joint replacement. We tested the inter- and intra-observer agreement for the UCS
as applied to the pelvis and femur using 20 examples of peri-prosthetic
fracture in 17 patients. Each subtype of the UCS was represented
by at least one case. Specialist orthopaedic surgeons (experts)
and orthopaedic residents (pre-experts) assessed reliability on
two separate occasions. For the pelvis, the UCS showed inter-observer agreement of 0.837
(95% confidence intervals (CI) 0.798 to 0.876) for the experts and
0.728 (95% CI 0.689 to 0.767) for the pre-experts. The intra-observer
agreement for the experts was 0.861 (95% CI 0.760 to 0.963) and
0.803 (95% 0.688 to 0.918) for the pre-experts. For the femur, the
UCS showed an inter-observer kappa value of 0.805 (95% CI 0.765
to 0.845) for the experts and a value of 0.732 (95% CI 0.690 to 0.773)
for the pre-experts. The intra-observer agreement was 0.920 (95%
CI 0.867 to 0.973) for the experts, and 0.772 (95% CI 0.652 to 0.892)
for the pre-experts. This corresponds to a substantial and ‘almost
perfect’ inter- and intra-observer agreement for the UCS for peri-prosthetic
fractures of the pelvis and femur. We hope that unifying the terminology of these injuries will
assist in their assessment, treatment and outcome. Cite this article:
To assess the extent of osteointegration in two designs of shoulder
resurfacing implants. Bony integration to the Copeland cylindrical
central stem design and the Epoca RH conical-crown design were compared. Implants retrieved from six patients in each group were pair-matched.
Mean time to revision surgery of Copeland implants was 37 months
(standard deviation (Aims
Patients and Methods
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
We are currently facing an epidemic of periprosthetic
fractures around the hip. They may occur either during surgery or
post-operatively. Although the acetabulum may be involved, the femur
is most commonly affected. We are being presented with new, difficult
fracture patterns around cemented and cementless implants, and we
face the challenge of an elderly population who may have grossly
deficient bone and may struggle to rehabilitate after such injuries.
The correct surgical management of these fractures is challenging.
This article will review the current choices of implants and techniques
available to deal with periprosthetic fractures of the femur. Cite this article:
Pre-operative planning for total hip replacement
(THR) is challenging in hips with severe acetabular deformities, including
those with a hypoplastic acetabulum or severe defects and in the
presence of arthrodesis or ankylosis. We evaluated whether a Rapid
Prototype (RP) model, which is a life-sized reproduction based on
three-dimensional CT scans, can determine the feasibility of THR
and provide information about the size and position of the acetabular component
in severe acetabular deformities. THR was planned using an RP model
in 21 complex hips in five men (five hips) and 16 women (16 hips)
with a mean age of 47.7 years (24 to 70) at operation. An acetabular
component was implanted successfully and THR completed in all hips.
The acetabular component used was within 2 mm of the predicted size
in 17 hips (80.9%). All of the acetabular components and femoral
stems had radiological evidence of bone ingrowth and stability at
the final follow-up, without any detectable wear or peri-prosthetic
osteolysis. The RP model allowed a simulated procedure pre-operatively
and was helpful in determining the feasibility of THR pre-operatively,
and to decide on implant type, size and position in complex THRs. Cite this article:
Management of bisphosphonate-associated subtrochanteric
fractures remains opinion- or consensus-based. There are limited
data regarding the outcomes of this fracture. We retrospectively reviewed 33 consecutive female patients with
a mean age of 67.5 years (47 to 91) who were treated surgically
between May 2004 and October 2009. The mean follow-up was 21.7 months
(0 to 53). Medical records and radiographs were reviewed to determine
the post-operative ambulatory status, time to clinical and radiological
union and post-fixation complications such as implant failure and
need for second surgery. The predominant fixation method was with an extramedullary device
in 23 patients. 25 (75%) patients were placed on wheelchair mobilisation
or no weight-bearing initially. The mean time to full weight-bearing
was 7.1 months (2.2 to 29.7). The mean time for fracture site pain
to cease was 6.2 months (1.2 to 17.1). The mean time to radiological
union was 10.0 months (2.2 to 27.5). Implant failure was seen in
seven patients (23%, 95 confidence interval (CI) 11.8 to 40.9).
Revision surgery was required in ten patients (33%, 95 CI 19.2 to
51.2). A large proportion of the patients required revision surgery
and suffered implant failure. This fracture is associated with slow
healing and prolonged post-operative immobility. Cite this article:
Down’s syndrome is associated with a number of
musculoskeletal abnormalities, some of which predispose patients
to early symptomatic arthritis of the hip. The purpose of the present
study was to review the general and hip-specific factors potentially
compromising total hip replacement (THR) in patients with Down’s
syndrome, as well as to summarise both the surgical techniques that
may anticipate the potential adverse impact of these factors and
the clinical results reported to date. A search of the literature
was performed, and the findings further informed by the authors’
clinical experience, as well as that of the hip replacement in Down
Syndrome study group. The general factors identified include a high
incidence of ligamentous laxity, as well as associated muscle hypotonia
and gait abnormalities. Hip-specific factors include: a high incidence
of hip dysplasia, as well as a number of other acetabular, femoral
and combined femoroacetabular anatomical variations. Four studies
encompassing 42 hips, which reported the clinical outcomes of THR
in patients with Down’s syndrome, were identified. All patients
were successfully treated with standard acetabular and femoral components.
The use of supplementary acetabular screw fixation to enhance component
stability was frequently reported. The use of constrained liners
to treat intra-operative instability occurred in eight hips. Survival
rates of between 81% and 100% at a mean follow-up of 105 months
(6 to 292) are encouraging. Overall, while THR in patients with
Down’s syndrome does present some unique challenges, the overall
clinical results are good, providing these patients with reliable
pain relief and good function. Cite this article:
Non-modular tapered fluted, titanium stems are
available for use in femoral revision. The combination of taper
and flutes on the stem provides axial and rotational stability,
respectively. The material and surface properties of the stem promotes
bone on-growth. If the surgeon is confident and reasonably experienced
in the surgical use of this sort of design and the case is relatively
straightforward, a non-modular design is effective. It also potentially reduces
implant inventory, and circumvents the potential problems of taper
junction corrosion and fatigue fracture. There are reports of excellent
survival, good clinical and functional results and evidence of subsequent
increase in proximal bone stock. Cite this article: Bo
In developmental dysplasia of the hip (DDH),
a bone defect is often observed superior to the acetabulum after
the reconstruction at the level of the true acetabulum during total
hip replacement (THR). However, the essential amount of uncemented
acetabular component coverage required for a satisfactory outcome
remains controversial. The purpose of this study was to assess the
stability and function of acetabular components with a lack of coverage >
30% (31% to 50%). A total of 760 DDH patients underwent THR with
acetabular reconstruction at the level of the true floor. Lack of
coverage above the acetabular component of >
30% occurred in 56
patients. Intra-operatively, autogenous morcellised bone grafts
were used to fill the uncovered portion. Other than two screws inserted through
the acetabular shell, no additional structural supports were used
in these hips. In all, four patients were lost to follow-up. Therefore,
52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1
years (42 to 78) were available for this study at a mean of 4.8
years (3 to 7). There were no instances of prosthesis revision or
marked loosening during the follow-up. The Harris hip score improved
from a mean of 40.7 points ( Cite this article:
We undertook a randomised controlled trial to
compare bipolar hemiarthroplasty (HA) with a novel total hip replacement
(THR) comprising a polycarbonate–urethane (PCU) acetabular component
coupled with a large-diameter metal femoral head for the treatment
of displaced fractures of the femoral neck in elderly patients. Functional
outcome, assessed with the Harris hip score (HHS) at three months
and then annually after surgery, was the primary endpoint. Rates
of revision and complication were secondary endpoints. Based on a power analysis, 96 consecutive patients aged >
70
years were randomised to receive either HA (49) or a PCU-THR (47).
The mean follow-up was 30.1 months (23 to 50) and 28.6 months (22
to 52) for the HA and the PCU group, respectively. The HHS showed no statistically significant difference between
the groups at every follow-up. Higher pain was recorded in the PCU
group at one and two years’ follow-up
(p = 0.006 and p = 0.019, respectively). In the HA group no revision
was performed. In the PCU-THR group six patients underwent revision
and one patient is currently awaiting
re-operation. The three-year survival rate of the PCU-THR group
was 0.841 (95% confidence interval 0.680 to 0.926). Based on our findings we do not recommend the use of the PCU
acetabular component as part of the treatment of patients with fractures
of the femoral neck. Cite this article:
The use of short stems has become more popular
with the increased interest in less invasive approaches for total hip
arthroplasty. The curved broaches and stem can be inserted along
a curved track to avoid the abductor attachments. Short stems have
the potential of being more bone conserving by allowing for higher
neck retention, maintenance of the medial greater trochanter, and
preferential stress transfer to the proximal femur. An initial experience with 500 new short stems (Fitmore, Zimmer
Orthopedics) stems used for total hip arthroplasty is reported.
No stems have been revised for aseptic loosening.