Advertisement for orthosearch.org.uk
Results 1 - 50 of 110
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 66-B, Issue 2 | Pages 172 - 174
1 Mar 1984
Leyshon R Matthews J

Nine cases of acetabular erosion following insertion of the Monk "hard top" hip prosthesis for subcapital fractures are presented. All of these patients had significant symptoms and represented 5% of the total group. In our experience acetabular erosion has proved to be a significant complication of the Monk "hard top" prosthesis


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 5 | Pages 725 - 728
1 Sep 1992
Wilson L Nolan J Heywood-Waddington M

We report the clinical and operative details of seven cases of fracture of the femoral stem of the Ring TiMESH cementless hip prosthesis (two were cemented and five uncemented). Six fractures occurred in the proximal one-third of the stem and one at mid-stem. The failures are attributed mainly to two defects in stem design, the narrowness of the anteroposterior dimensions and the depth of the recess for the titanium mesh pads. Great difficulty was experienced in removing the osseo-integrated distal fragments of the broken stems


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1427 - 1430
1 Oct 2011
Lindgren JU Brismar BH Wikstrom AC

A 70-year-old man with an uncemented metal-on-polyethylene total hip prosthesis underwent revision arthroplasty 33 months later because of pain, swelling and recurrent dislocation. There appeared to be corrosion and metal release from the prosthetic head, resulting in pseudotumour formation and severe local soft-tissue destruction. The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 310 - 314
1 Mar 2006
Sexton SA Stossel CA Haddad FS

The Kent hip is a distally-locked femoral stem which was developed to address severe proximal bone loss, severe bony deformity and peri-prosthetic fracture.

We reviewed the results of 145 consecutive Kent hips implanted into 141 patients between 1987 and 2000. The indications for implantation were aseptic loosening (75 hips), septic loosening (two), peri-prosthetic and prosthetic fracture (37), severe bony deformity (24), and fracture through a proximal femoral metastasis (seven).

The median time to full weight-bearing after surgery was two days and the mean length of follow-up was 5.1 years (2 to 15). Further revisions were required for 13 femoral stems. With removal of the stem for any reason as an end-point, the cumulative survival at five, ten and 15 years was 93%, 89% and 77%, respectively. In patients aged ≥ 70 years, the cumulative survival at 15 years was 92%, compared with 68% in those aged < 70 years. Because of these findings, we recommend the use of interlocking stems in patients aged ≥ 70 years, particularly in those with a peri-prosthetic fracture, for whom alternative methods are limited. Outcome scores and survival data, compared with other systems, indicate that the Kent hip should be used with caution in younger patients.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1035 - 1041
1 Sep 2019
Markel DC Bou-Akl T Rossi MD Pizzimenti N Wu B Ren W

Aims

The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components.

Patients and Methods

A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 152 - 153
1 Jan 1996
Muñoz Vives JM Garin DM Montllor RG


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 85 - 91
1 Jan 1996
Önsten I Carlsson ÅS Sanzén L Besjakov J

A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems.

Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear.

We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 159 - 160
1 Jan 1992
Lacroix H du Bois N van Biezen F


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 83 - 87
1 Jan 1992
Kim Y Kim V

We reviewed 82 primary arthroplasties (in 71 patients) in which cementless porous-coated hip prostheses were used. The mean age of the patients at operation was 52 years (24 to 86); they were followed up for an average of 62.1 months (60 to 66). The diagnosis was avascular necrosis of the femoral head in 35%, fracture of the femoral neck in 24%, primary osteoarthritis in 16% and miscellaneous in 25%. The average preoperative Harris hip score was 56.7 points and the average postoperative score was 83.3 points. Eight hips (10%) had component loosening; four had been revised and four were awaiting revision. In 27 hips (33%) there was a radiolucent line wider than 2 mm in zones 1 and 7. In 55 hips (67%) there was calcar resorption of more than 10 mm. Twenty patients (28%) complained of thigh pain although they had no radiographic evidence of loosening of a component. Factors that may have contributed to the poor clinical and radiographic results were: 1) inadequate surface area for bone ingrowth, particularly on the lateral aspect of the upper part of the prosthesis, 2) poor initial fit of the stem in the metaphysis, which resulted in cantilever motion of the proximal part of the stem about the well-fixed distal stem, and 3) the collar of the prosthesis, which prevented it from subsiding to a naturally stable position and caused damage to the calcar.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 623 - 627
1 May 2005
Kim Y

Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20).

The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 282 - 283
1 Mar 1998
Verburg AD

In a frail elderly patient a pathological fracture of the femur at the lower end of a loose femoral stem was treated by the retrograde insertion of an intramedullary nail over the tip of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 6 | Pages 822 - 824
1 Nov 1992
Godsiff S Emery R Heywood-Waddington M Thomas T

A prospective, randomised and independently assessed trial of the Ring UPM total hip replacement showed that the quality of the early result was better if the femoral prosthesis was cemented than if it was not. More patients with cemented prostheses were painfree at four months (58% cemented:42% uncemented) and at one year (63% cemented:50% uncemented), but at two years pain relief was equal in both groups. At two years significantly more patients with cemented prostheses could walk without support (96% cemented:62% uncemented, p = 0.01 to 0.05). There is a need for more similar trials to compare the results of contemporary designs of cemented and uncemented total hip prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 11 - 13
1 Nov 2012
Cuckler JM

Hip implant retrieval analysis is the most important source of insight into the performance of new materials and designs of hip arthroplasties. Even the most rigorous in vitro testing will not accurately simulate the behavior of implant materials and new designs of prosthetic arthroplasties. Retrieval analysis has revealed such factors as the effects of gamma-in-air sterilisation of polyethylene, fatigue failure mechanisms of polymethylmethacrylate bone cement, fretting corrosion of Morse taper junctions, third body wear effects of both hard-on-hard and hard-on-soft bearing couples, and the effects of impingement of components on the full spectrum of bearing surfaces, none of which was predicted by pre-implantation in vitro testing of these materials and combinations. The temporal sequence of the retrieval process is approximately six years from first implantation through retrieval analysis, laboratory investigation, and publication of results, and thus, in addition to rigorous clinical evaluation, represents the true development and insight cycle for new designs and materials.


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 369 - 373
1 May 1995
Savilahti S Myllyneva I Lindholm T Pajamaki K Nevalainen J Laippala P

We report the clinical outcome and survival of 100 total hip arthroplasties with Link RS cementless hip prostheses after a mean of 63 months (51 to 93). The average preoperative Merle d'Aubigne hip score was 9.4 points and the average score at the latest follow-up was 15.3 points. Thirteen prostheses have been revised and seven await revision. Survivorship analysis based on intention to revise showed only 68% survival at seven years (95% CI 53% to 81%). The survivals of femoral and acetabular components verified to be loose at surgery were 78% and 89% respectively at seven years. A number of other patients have pain and it seems likely that there will be substantial reduction in these survival figures in the future.


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 4 | Pages 671 - 672
1 Jul 1994
Hettfleisch J Wissenbach R


The Journal of Bone & Joint Surgery British Volume
Vol. 66-B, Issue 5 | Pages 635 - 638
1 Nov 1984
Sharp D Porter K Duke R

Seven hips with protrusio acetabuli which showed complete or partial resolution of the protrusion after replacement with Ring's prosthesis are reported. It is suggested that this prosthesis offers a simple and effective method of treating painful protrusion into the pelvis; it allows healing of the medial wall of the acetabulum while avoiding many of the hazards of other methods of treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 61-B, Issue 4 | Pages 422 - 423
1 Nov 1979
Hunter G

Sixty-five total hip arthroplastics were reinserted after sepsis around the hip, positive cultures being obtained from fifty-six. Although 65 per cent of patients still have their implant in position, only sixteen of sixty-five (25 per cent) show an excellent or good result on a Harris rating. Twenty-three of sixty-five (35 per cent) subsequently required an excision arthroplasty. The indications and contraindications for this procedure are given.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 479 - 485
1 Apr 2014
Pedersen AB Mehnert F Sorensen HT Emmeluth C Overgaard S Johnsen SP

We examined the risk of thrombotic and major bleeding events in patients undergoing total hip and knee replacement (THR and TKR) treated with thromboprophylaxis, using nationwide population-based databases. We identified 83 756 primary procedures performed between 1997 and 2011. The outcomes were symptomatic venous thromboembolism (VTE), myocardial infarction (MI), stroke, death and major bleeding requiring hospitalisation within 90 days of surgery.

A total of 1114 (1.3%) and 483 (0.6%) patients experienced VTE and bleeding, respectively. The annual risk of VTE varied between 0.9% and 1.6%, and of bleeding between 0.4% and 0.8%. The risk of VTE and bleeding was unchanged over a 15-year period. A total of 0.7% of patients died within 90 days, with a decrease from 1% in 1997 to 0.6% in 2011 (p < 0.001). A high level of comorbidity and general anaesthesia were strong risk factors for both VTE and bleeding, with no difference between THR and TKR patients. The risk of both MI and stroke was 0.5%, which remained unchanged during the study period.

In this cohort study of patients undergoing THR and TKR patients in routine clinical practice, approximately 3% experienced VTE, MI, stroke or bleeding. These risks did not decline during the 15-year study period, but the risk of dying fell substantially.

Cite this article: Bone Joint J 2014;96-B:479–85.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 2 | Pages 305 - 305
1 Mar 2000
OVERGAARD S SØBALLE K


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 4 | Pages 582 - 589
1 Jul 1999
Røkkum M Brandt M Bye K Hetland KR Waage S Reigstad A

We have followed up for a period of seven to nine years 100 consecutive arthroplasties of the hip in which an entirely HA-coated implant had been used.

The clinical results were excellent and bony incorporation was extensive in all components. No stem became loose or subsided but five cups were revised because of loosening after 3.8 to 5.5 years, having functioned painlessly and shown radiological ingrowth.

Revision procedures because of excessive polyethylene wear have been performed on 18 hips and are planned for six more. Two eroded metal backings with worn-through polyethylene were exchanged; six hips showed metallosis without polyethylene wear-through. There were two cases of granulomatous cysts in the groin and 66 hips had osteolysis located periarticularly, in the greater trochanter or in the acetabulum.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 6 | Pages 884 - 891
1 Nov 1996
Malchau H Herberts P Wang YX Kärrholm J Romanus B

We enrolled 98 patients (107 hips) with a mean age of 47 years (SD 8.6) into a prospective study of the Madreporic Lord THR; 34 hips had primary and 73 secondary osteoarthritis.

After ten years, the survival rate using revision as the endpoint for failure was 70% (±9) for the cup and 98% (±0.3) for the stem. The combined clinical and radiological survival rates were 46% (±11) and 81% (±10), respectively. Osteoporosis due to stress-shielding was observed in the proximal femur. Hips with radiologically dense bone postoperatively showed the most pronounced bone loss.

We recommend continued radiological follow-up of patients with this type of implant to allow revision to be performed before there is severe bony destruction of the pelvis.


The Journal of Bone & Joint Surgery British Volume
Vol. 59-B, Issue 3 | Pages 298 - 301
1 Aug 1977
Clegg J

Twenty-nine patients with thirty pseudarthroses after removal of infected total hip prostheses have been reviewed one to six years later. Six hips still had a discharging sinus. Complete removal of all cement was found to be essential for healing, and various points in the technique of its clearance have been made. In this respect radio-opaque cement was a great advantage, and lateral guttering of the femur was most effective. Compared with the situation before replacement considerable relief of pain was obtained in most patients but there was much less improvement in function.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 781 - 785
1 Jun 2005
Temmerman OPP Raijmakers PGHM Berkhof J Hoekstra OS Teule GJJ Heyligers IC

In this meta-analysis we included 32 English-language articles published between January 1975 and June 2004 on the diagnostic performance of plain radiography, subtraction arthrography, nuclear arthrography and bone scintigraphy in detecting aseptic loosening of the femoral component, using criteria based on the Cochrane systematic review of screening and diagnostic tests.

The mean sensitivity and specificity were, respectively, 82% (95% confidence interval (CI) 76 to 87) and 81% (95% CI 73 to 87) for plain radiography and 85% (95% CI 75 to 91) and 83% (95% CI 75 to 89) for nuclear arthrography. Pooled sensitivity and specificity were, respectively, 86% (95% CI 74 to 93) and 85% (95% CI 77 to 91) for subtraction arthrography and 85% (95% CI 79 to 89) and 72% (95% CI 64 to 79) for bone scintigraphy. Although the diagnostic performance of the imaging techniques was not significantly different, plain radiography and bone scintigraphy are preferred for the assessment of a femoral component because of their efficacy and lower risk of patient morbidity.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 4 | Pages 598 - 603
1 May 2001
Nevelos JE Prudhommeaux F Hamadouche M Doyle C Ingham E Meunier A Nevelos AB Sedel L Fisher J

We compared and quantified the modes of failure and patterns of wear of 11 Mittelmeier and 11 Ceraver-Ostal retrieved alumina-alumina hip prostheses with reference to the corresponding clinical and radiological histories.

Macroscopic wear was assessed using a three-dimensional co-ordinate measuring machine. Talysurf contacting profilometry was used to measure surface roughness on a microscopic scale and SEM to determine mechanisms of wear at the submicron level.

The components were classified into one of three categories of wear: low (no visible/measurable wear), stripe (elliptical wear stripe on the heads and larger worn areas on the cups) and severe (macroscopic wear, large volumes of material lost). Overall, the volumetric wear of the alumina-alumina prostheses was substantially less than the widely used metal and ceramic-on-polyethylene combinations. By identifying and eliminating the factors which accelerate wear, it is expected that the lifetime of these devices can be further increased.


The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used.

The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm3 (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r² = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient’s mass, age and body mass index) did not correlate with the amount of material loss observed (p >  0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis.

Cite this article: Bone Joint J 2015;97-B:1350–7.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1003 - 1009
1 Aug 2020
Mononen H Sund R Halme J Kröger H Sirola J

Aims. There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision. Methods. Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis. Results. Prior lumbar fusion surgery was associated with increased risk of prosthetic dislocation (hazard ratio (HR) = 2.393, p < 0.001) and revision (HR = 1.528, p < 0.001). Head components larger than 28 mm were associated with lower dislocation rates compared to the 28 mm head (32 mm: HR = 0.712, p < 0.001; 36 mm: HR = 0.700, p < 0.001; 38 mm: HR = 0.808, p < 0.140; and 40 mm: HR = 0.421, p < 0.001). Heads of 38 mm (HR = 1.288, p < 0.001) and 40 mm (HR = 1.367, p < 0.001) had increased risk of revision compared to the 28 mm head. Conclusion. Lumbar fusion surgery was associated with higher rate of hip prosthesis dislocation and higher risk of revision surgery. Femoral head component of 32 mm (or larger) associates with lower risk of dislocation in patients with previous lumbar fusion. Cite this article: Bone Joint J 2020;102-B(8):1003–1009


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 1020 - 1029
1 Sep 2023
Trouwborst NM ten Duis K Banierink H Doornberg JN van Helden SH Hermans E van Lieshout EMM Nijveldt R Tromp T Stirler VMA Verhofstad MHJ de Vries JPPM Wijffels MME Reininga IHF IJpma FFA

Aims

The aim of this study was to investigate the association between fracture displacement and survivorship of the native hip joint without conversion to a total hip arthroplasty (THA), and to determine predictors for conversion to THA in patients treated nonoperatively for acetabular fractures.

Methods

A multicentre cross-sectional study was performed in 170 patients who were treated nonoperatively for an acetabular fracture in three level 1 trauma centres. Using the post-injury diagnostic CT scan, the maximum gap and step-off values in the weightbearing dome were digitally measured by two trauma surgeons. Native hip survival was reported using Kaplan-Meier curves. Predictors for conversion to THA were determined using Cox regression analysis.


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1180 - 1188
1 Oct 2022
Qu H Mou H Wang K Tao H Huang X Yan X Lin N Ye Z

Aims

Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation.

Methods

A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 775 - 782
1 Jul 2023
Koper MC Spek RWA Reijman M van Es EM Baart SJ Verhaar JAN Bos PK

Aims

The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate.

Methods

A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.


The Bone & Joint Journal
Vol. 104-B, Issue 12 | Pages 1292 - 1303
1 Dec 2022
Polisetty TS Jain S Pang M Karnuta JM Vigdorchik JM Nawabi DH Wyles CC Ramkumar PN

Literature surrounding artificial intelligence (AI)-related applications for hip and knee arthroplasty has proliferated. However, meaningful advances that fundamentally transform the practice and delivery of joint arthroplasty are yet to be realized, despite the broad range of applications as we continue to search for meaningful and appropriate use of AI. AI literature in hip and knee arthroplasty between 2018 and 2021 regarding image-based analyses, value-based care, remote patient monitoring, and augmented reality was reviewed. Concerns surrounding meaningful use and appropriate methodological approaches of AI in joint arthroplasty research are summarized. Of the 233 AI-related orthopaedics articles published, 178 (76%) constituted original research, while the rest consisted of editorials or reviews. A total of 52% of original AI-related research concerns hip and knee arthroplasty (n = 92), and a narrative review is described. Three studies were externally validated. Pitfalls surrounding present-day research include conflating vernacular (“AI/machine learning”), repackaging limited registry data, prematurely releasing internally validated prediction models, appraising model architecture instead of inputted data, withholding code, and evaluating studies using antiquated regression-based guidelines. While AI has been applied to a variety of hip and knee arthroplasty applications with limited clinical impact, the future remains promising if the question is meaningful, the methodology is rigorous and transparent, the data are rich, and the model is externally validated. Simple checkpoints for meaningful AI adoption include ensuring applications focus on: administrative support over clinical evaluation and management; necessity of the advanced model; and the novelty of the question being answered.

Cite this article: Bone Joint J 2022;104-B(12):1292–1303.


The Bone & Joint Journal
Vol. 104-B, Issue 12 | Pages 1369 - 1378
1 Dec 2022
van Rijckevorsel VAJIM de Jong L Verhofstad MHJ Roukema GR

Aims

Factors associated with high mortality rates in geriatric hip fracture patients are frequently unmodifiable. Time to surgery, however, might be a modifiable factor of interest to optimize clinical outcomes after hip fracture surgery. This study aims to determine the influence of postponement of surgery due to non-medical reasons on clinical outcomes in acute hip fracture surgery.

Methods

This observational cohort study enrolled consecutively admitted patients with a proximal femoral fracture, for which surgery was performed between 1 January 2018 and 11 January 2021 in two level II trauma teaching hospitals. Patients with medical indications to postpone surgery were excluded. A total of 1,803 patients were included, of whom 1,428 had surgery < 24 hours and 375 had surgery ≥ 24 hours after admission.


The Bone & Joint Journal
Vol. 104-B, Issue 8 | Pages 997 - 1008
1 Aug 2022

Aims

The aim of this study was to describe the management and associated outcomes of patients sustaining a femoral hip periprosthetic fracture (PPF) in the UK population.

Methods

This was a multicentre retrospective cohort study including adult patients who presented to 27 NHS hospitals with 539 new PPFs between 1 January 2018 and 31 December 2018. Data collected included: management strategy (operative and nonoperative), length of stay, discharge destination, and details of post-treatment outcomes (reoperation, readmission, and 30-day and 12-month mortality). Descriptive analysis by fracture type was performed, and predictors of PPF management and outcomes were assessed using mixed-effects logistic regression.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 9 - 16
1 Jul 2021
Hadden WJ Ibrahim M Taha M Ure K Liu Y Paish ADM Holdsworth DW Abdelbary H

Aims

The aims of this study were to develop an in vivo model of periprosthetic joint infection (PJI) in cemented hip hemiarthroplasty, and to monitor infection and biofilm formation in real-time.

Methods

Sprague-Dawley rats underwent cemented hip hemiarthroplasty via the posterior approach with pre- and postoperative gait assessments. Infection with Staphylococcus aureus Xen36 was monitored with in vivo photoluminescent imaging in real-time. Pre- and postoperative gait analyses were performed and compared. Postmortem micro (m) CT was used to assess implant integration; field emission scanning electron microscopy (FE-SEM) was used to assess biofilm formation on prosthetic surfaces.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 820 - 821
1 Jun 2009
Dabasia H Kokkinakis M El-Guindi M

A 66-year-old man presented with an infected resurfacing hip replacement in the immediate post-operative period following an uneventful transurethral resection of the prostate. Prophylactic antibiotics had been administered on induction of anaesthesia. The hip prosthesis had been inserted seven years previously and had been hitherto asymptomatic. The hip was washed out and microbiological examination identified Enterococcus faecalis as the infecting micro-organism. Despite current recommendations, clinicians undertaking invasive procedures which can lead to bacteraemia in patients with prosthetic joint replacements should be aware of the risk of haematogenous seeding in such prostheses, which although low, has potentially disastrous consequences


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 234 - 244
1 Feb 2021
Gibb BP Hadjiargyrou M

Antibiotic resistance represents a threat to human health. It has been suggested that by 2050, antibiotic-resistant infections could cause ten million deaths each year. In orthopaedics, many patients undergoing surgery suffer from complications resulting from implant-associated infection. In these circumstances secondary surgery is usually required and chronic and/or relapsing disease may ensue. The development of effective treatments for antibiotic-resistant infections is needed. Recent evidence shows that bacteriophage (phages; viruses that infect bacteria) therapy may represent a viable and successful solution. In this review, a brief description of bone and joint infection and the nature of bacteriophages is presented, as well as a summary of our current knowledge on the use of bacteriophages in the treatment of bacterial infections. We present contemporary published in vitro and in vivo data as well as data from clinical trials, as they relate to bone and joint infections. We discuss the potential use of bacteriophage therapy in orthopaedic infections. This area of research is beginning to reveal successful results, but mostly in nonorthopaedic fields. We believe that bacteriophage therapy has potential therapeutic value for implant-associated infections in orthopaedics.

Cite this article: Bone Joint J 2021;103-B(2):234–244.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 644 - 649
1 Apr 2021
Alsousou J Oragu E Martin A Strickland L Newman S Kendrick B Taylor A Glyn-Jones S

Aims

The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA).

Methods

A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 874 - 876
1 Jul 2013
Kallala R Anderson P Morris S Haddad FS

In a time of limited resources, the debate continues over which types of hip prosthesis are clinically superior and more cost-effective. Orthopaedic surgeons increasingly need robust economic evidence to understand the full value of the operation, and to aid decision making on the ‘package’ of procedures that are available and to justify their practice beyond traditional clinical preference. . In this paper we explore the current economic debate about the merits of cemented and cementless total hip replacement, an issue that continues to divide the orthopaedic community. Cite this article: Bone Joint J 2013;95-B:874–6


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 270 - 273
1 Feb 2014
Ogura K Miyamoto S Sakuraba M Chuman H Fujiwara T Kawai A

Resection of malignant bony tumours of the pelvis creates large bone and soft-tissue defects, and is frequently associated with complications such as wound dehiscence and deep infection. We present the results of six patients in whom a rectus abdominis myocutaneous (RAM) flap was used following resection of a malignant tumour of the pelvis. Bony reconstruction was performed using a constrained hip tumour prosthesis in three patients, vascularised fibular graft in two and frozen autograft in one. At a mean follow-up of 63 months (16 to 115), no patients had a problem with the wound. Immediate reconstruction using a RAM flap may be used after resection of a malignant tumour of the pelvis to provide an adequate volume of tissue to eliminate the dead space, cover the exposed bone or implants with well-vascularised soft tissue and to reduce the risk of complications. Cite this article: Bone Joint J 2014;96-B:270–3


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 949 - 950
1 Jul 2006
Lloyd JM Calder JDF

The Thompson hemiarthroplasty is a popular hip prosthesis. We present two case reports highlighting a significant alteration in the design of the implant which compromised the success of the operations. In recent years the manufacturing process of this prosthesis has changed, with a resultant increase in the volume of the stem of 10 ml. It is essential that manufacturers inform orthopaedic surgeons of any alteration in the design of the implant and supply compatible instrumentation to minimise surgical errors. Surgeons must remain vigilant when checking the compatibility of the trial and definitive prostheses


The Journal of Bone & Joint Surgery British Volume
Vol. 67-B, Issue 3 | Pages 413 - 415
1 May 1985
Nielsen H Veth R Oldhoff J Koops H Scales J

We report a patient with a peri-acetabular chondrosarcoma which was treated by resection of part of the pelvis and the femoral head. This was followed by replacement with a custom-made titanium alloy pelvic portion and a conventional total hip prosthesis. The operation was carried out in two stages, and yielded a good functional and cosmetic result


The Journal of Bone & Joint Surgery British Volume
Vol. 66-B, Issue 3 | Pages 337 - 339
1 May 1984
Pazzaglia U Byers P

This paper reports a singular case of pathological fracture through a large osteolytic lesion of the mid-femur consequent on reaction to a hip prosthesis. Not only metal particles but also polyethylene fibres were found at this site, some 15 cm from the prosthesis. The case sheds an interesting light on the reaction to prostheses


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 709 - 715
1 Jun 2020
Abdelsamie KR Elhawary I Ali H Ali M EL-Shafie M Dominic Meek RM

Aims

Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs.

Methods

A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 5 | Pages 802 - 805
1 Sep 1995
Otfinowski J Pawelec A

We performed thermal analysis of polyethylene samples obtained from 73 new cups of the Weller hip prosthesis. There were marked individual differences in the degree of their crystallinity, ranging from 37.8% to 67.2% with a mean of 53%. Analysis of polyethylene from removed cups also showed differences in crystallinity in individual cups, with the mean value being higher in the removed implants than in the new ones. This difference in crystallinity between the groups was statistically significant. Marked individual differences in the degree of cup wear may result from these changes


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 1 | Pages 58 - 62
1 Jan 1989
Langlais F Vielpeau C

Four en-bloc resections for malignant tumours of the hip, the peri-acetabular region and the iliac wing were reconstructed using an irradiated hemipelvic allograft together with a total hip prosthesis. Technical aspects include the use of an anterior Enneking approach which excises the previous biopsy site, division and re-attachment of the iliac crest and fixation of the prosthesis using a modified acetabular cup and three polypropylene artificial ligaments to increase the stability of the joint. Weight-bearing was allowed at three months. The oncological and clinical results were satisfactory after a mean follow-up of 19 months, with walking distances of 500 metres to two kilometres without pain


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 34 - 38
1 Jan 2004
Duffy P Sher JL Partington PF

The ABG I cementless hip prosthesis has demonstrated unacceptably high rates of wear and osteolysis in our patients. We performed a retrospective study of 97 hips implanted between 1992 and 1998. Radiographic analysis revealed high rates of wear of the polyethylene liner with marked periacetabular osteolysis. Clinical examination indicated that many of these patients were initially asymptomatic. Wear-related problems have required ten hips to be revised and a furher 13 are awaiting revision. This gives a failure rate of 24% at a mean follow-up of 69 months. Contributing factors are likely to include poor wear characteristics of the polyethylene liners which were gamma irradiated in air, and increased wear debris caused by a poor fit of the polyethylene liner within the shell. We believe that all ABG I implants should be immediately reviewed and remain under careful, long-term follow-up


The Journal of Bone & Joint Surgery British Volume
Vol. 59-B, Issue 3 | Pages 293 - 297
1 Aug 1977
Hunter G Dandy D

Deep infection, the most serious local complication of total hip replacement, prompted a study of the records of 135 patients (137 hips) thus afflicted in a nationwide survey of Canada. Particular attention has been paid to the natural history of the infection, and the problems of diagnosis are described. Twenty-one patients died after the insertion, or removal, of the prosthesis, and of the survivors of the original 135 patients only eighteen have been able to retain the prosthesis without further problems with the wound. The remaining patients had the prosthesis removed, and most dry wounds. Certain suggestions are made on management. The advice that a second total hip prosthesis should be inserted after a deep infection of the first implant is not supported


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 3 | Pages 468 - 474
1 May 1993
Schuller H Dalstra M Huiskes R Marti R

In acetabular dysplasia, fixation of the acetabular component of a cemented total hip prosthesis may be insecure and superolateral bone grafts are often used to augment the acetabular roof. We used finite element analysis to study the mechanical importance of the lateral acetabular roof and found that the lateral acetabular rim plays an important role in the load transfer of the pelvic bone. When the superlateral rim was lacking, the load shifted to the posterosuperior rim and to the area of pubic support, and the stresses in all materials, especially in the cement and in the trabecular bone, increased greatly. At the cement-bone interface the tilting component of the shear stress increased threefold. In a model in which the dysplastic acetabulum was augmented by a rigidly fixed, load-transmitting bone graft, the stresses were considerably diminished


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1447 - 1458
1 Nov 2019
Chatziagorou G Lindahl H Kärrholm J

Aims

We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis.

Patients and Methods

The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 8 | Pages 1202 - 1206
1 Nov 2003
Fiorito S Magrini L Goalard C

We investigated the circulating levels of the main cytokines involved in bone resorption (IL-1β, IL-6, TNF-α), prostaglandins (PGE. 2. ) and metalloproteases (MMP-1), as possible early markers of osteolysis, in the serum of eight patients with periprosthetic osteolysis and ten patients without osteolysis. All had received a cementless hip prosthesis (ABG-1). We also assessed the serum levels of IL-11 and TGF-β anti-inflammatory cytokines exerting protective effect on bone resorption. The mean serum levels of IL-1β, IL-6, TNF-α, TGF-β, MMP-1, and PGE. 2. in patients with periprosthetic osteolysis did not differ significantly from those of patients without osteolysis or from those of normal controls. IL-11 serum levels were not detectable at all in any of the patients, while they were detected within normal reference values in the control subjects (significant inverse correlation). We believe that circulating cytokines cannot be regarded as markers of osteolysis, a condition characterised by a local inflammation without systemic signs of inflammation. On the contrary, the undetectable levels of IL-11 in implanted patients could provide evidence for a lack of balance between pro- and anti-inflammatory cytokines in these patients


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 329 - 335
1 Mar 2020
Fink B Schuster P Braun R Tagtalianidou E Schlumberger M

Aims

Biopsy of the periprosthetic tissue is an important diagnostic tool for prosthetic joint infection (PJI) as it enables the detection of the responsible microorganism with its sensitivity to antibiotics. We aimed to investigate how often the bacteria identified in the tissue analysis differed between samples obtained from preoperative biopsy and intraoperative revision surgery in cases of late PJI; and whether there was a therapeutic consequence.

Methods

A total of 508 patients who required revision surgery of total hip arthroplasty (THA) (n = 231) or total knee arthroplasty (TKA) (n = 277) because of component loosening underwent biopsy before revision surgery. The tissue samples collected at biopsy and during revision surgery were analyzed according to the criteria of the Musculoskeletal Infection Society (MSIS).