The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 44 - 48
1 Jan 2006
Keene G Simpson D Kalairajah Y

Twenty patients underwent simultaneous bilateral medial unicompartmental knee arthroplasty. Pre-operative hip-knee-ankle alignment and valgus stress radiographs were used to plan the desired post-operative alignment of the limb in accordance with established principles for unicompartmental arthroplasty. In each patient the planned alignment was the same for both knees. Overall, the mean planned post-operative alignment was to 2.3° of varus (0° to 5°).

The side and starting order of surgery were randomised, using conventional instrumentation for one knee and computer-assisted surgery for the opposite side.

The mean variation between the pre-operative plan and the achieved correction in the navigated and the non-navigated limb was 0.9° (sd 1.1; 0° to 4°) and 2.8° (sd 1.4; 1° to 7°), respectively. Using the Wilcoxon signed rank test, we found the difference in variation statistically significant (p < 0.001).

Assessment of lower limb alignment in the non-navigated group revealed that 12 (60%) were within ± 2° of the pre-operative plan, compared to 17 (87%) of the navigated cases.

Computer-assisted surgery significantly improves the post-operative alignment of medial unicompartmental knee arthroplasty compared to conventional techniques in patients undergoing bilateral simultaneous arthroplasty. Improved alignment after arthroplasty is associated with better function and increased longevity.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 3 | Pages 515 - 516
1 May 1991
Saleh M Harriman P Edwards D


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 74 - 80
1 Jun 2021
Deckey DG Rosenow CS Verhey JT Brinkman JC Mayfield CK Clarke HD Bingham JS

Aims. Robotic-assisted total knee arthroplasty (RA-TKA) is theoretically more accurate for component positioning than TKA performed with mechanical instruments (M-TKA). Furthermore, the ability to incorporate soft-tissue laxity data into the plan prior to bone resection should reduce variability between the planned polyethylene thickness and the final implanted polyethylene. The purpose of this study was to compare accuracy to plan for component positioning and precision, as demonstrated by deviation from plan for polyethylene insert thickness in measured-resection RA-TKA versus M-TKA. Methods. A total of 220 consecutive primary TKAs between May 2016 and November 2018, performed by a single surgeon, were reviewed. Planned coronal plane component alignment and overall limb alignment were all 0° to the mechanical axis; tibial posterior slope was 2°; and polyethylene thickness was 9 mm. For RA-TKA, individual component position was adjusted to assist gap-balancing but planned coronal plane alignment for the femoral and tibial components and overall limb alignment remained 0 ± 3°; planned tibial posterior slope was 1.5°. Mean deviations from plan for each parameter were compared between groups for positioning and size and outliers were assessed. Results. In all, 103 M-TKAs and 96 RA-TKAs were included. In RA-TKA versus M-TKA, respectively: mean femoral positioning (0.9° (SD 1.2°) vs 1.7° (SD 1.1°)), mean tibial positioning (0.3° (SD 0.9°) vs 1.3° (SD 1.0°)), mean posterior tibial slope (-0.3° (SD 1.3°) vs 1.7° (SD 1.1°)), and mean mechanical axis limb alignment (1.0° (SD 1.7°) vs 2.7° (SD 1.9°)) all deviated significantly less from the plan (all p < 0.001); significantly fewer knees required a distal femoral recut (10 (10%) vs 22 (22%), p = 0.033); and deviation from planned polyethylene thickness was significantly less (1.4 mm (SD 1.6) vs 2.7 mm (SD 2.2), p < 0.001). Conclusion. RA-TKA is significantly more accurate and precise in planning both component positioning and final polyethylene insert thickness. Future studies should investigate whether this increased accuracy and precision has an impact on clinical outcomes. The greater accuracy and reproducibility of RA-TKA may be important as precise new goals for component positioning are developed and can be further individualized to the patient. Cite this article: Bone Joint J 2021;103-B(6 Supple A):74–80


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 442 - 448
1 Apr 2020
Kayani B Konan S Ahmed SS Chang JS Ayuob A Haddad FS

Aims. The objectives of this study were to assess the effect of anterior cruciate ligament (ACL) resection on flexion-extension gaps, mediolateral soft tissue laxity, maximum knee extension, and limb alignment during primary total knee arthroplasty (TKA). Methods. This prospective study included 140 patients with symptomatic knee osteoarthritis undergoing primary robotic-arm assisted TKA. All operative procedures were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess study outcomes pre- and post-ACL resection with knee extension and 90° knee flexion. This study included 76 males (54.3%) and 64 females (45.7%) with a mean age of 64.1 years (SD 6.8) at time of surgery. Mean preoperative hip-knee-ankle deformity was 6.1° varus (SD 4.6° varus). Results. ACL resection increased the mean extension gap significantly more than the flexion gap in the medial (mean 1.2 mm (SD 1.0) versus mean 0.2 mm (SD 0.7) respectively; p < 0.001) and lateral (mean 1.1 mm (SD 0.9) versus mean 0.2 mm (SD 0.6) respectively; p < 0.001) compartments. The mean gap differences following ACL resection did not create any significant mediolateral soft tissue laxity in extension (gap difference: mean 0.1 mm (SD 2.4); p = 0.89) or flexion (gap difference: mean 0.2 mm (SD 3.1); p = 0.40). ACL resection did not significantly affect maximum knee extension (change in maximum knee extension = mean 0.2° (SD 0.7°); p = 0.23) or fixed flexion deformity (mean 4.2° (SD 3.2°) pre-ACL release versus mean 3.9° (SD 3.7°) post-ACL release; p = 0.61). ACL resection did not significantly affect overall limb alignment (change in alignment = mean 0.2° valgus (SD 1.0° valgus; p = 0.11). Conclusion. ACL resection creates flexion-extension mismatch by increasing the extension gap more than the flexion gap. However, gap differences following ACL resection do not create any mediolateral soft tissue laxity in extension or flexion. ACL resection does not affect maximum knee extension or overall limb alignment. Cite this article: Bone Joint J 2020;102-B(4):442–448


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1230 - 1237
1 Oct 2019
Kayani B Konan S Horriat S Ibrahim MS Haddad FS

Aims. The aim of this study was to assess the effect of posterior cruciate ligament (PCL) resection on flexion-extension gaps, mediolateral soft-tissue laxity, fixed flexion deformity (FFD), and limb alignment during posterior-stabilized (PS) total knee arthroplasty (TKA). Patients and Methods. This prospective study included 110 patients with symptomatic osteoarthritis of the knee undergoing primary robot-assisted PS TKA. All operations were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess gaps before and after PCL resection in extension and 90° knee flexion. Measurements were made after excision of the anterior cruciate ligament and prior to bone resection. There were 54 men (49.1%) and 56 women (50.9%) with a mean age of 68 years (. sd. 6.2) at the time of surgery. The mean preoperative hip-knee-ankle deformity was 4.1° varus (. sd. 3.4). Results. PCL resection increased the mean flexion gap significantly more than the extension gap in the medial (2.4 mm (. sd. 1.5) vs 1.3 mm (. sd. 1.0); p < 0.001) and lateral (3.3 mm (. sd. 1.6) vs 1.2 mm (. sd. 0.9); p < 0.01) compartments. The mean gap differences after PCL resection created significant mediolateral laxity in flexion (gap difference: 1.1 mm (. sd. 2.5); p < 0.001) but not in extension (gap difference: 0.1 mm (. sd. 2.1); p = 0.51). PCL resection significantly improved the mean FFD (6.3° (. sd. 4.4) preoperatively vs 3.1° (. sd. 1.5) postoperatively; p < 0.001). There was a strong positive correlation between the preoperative FFD and change in FFD following PCL resection (Pearson’s correlation coefficient = 0.81; p < 0.001). PCL resection did not significantly affect limb alignment (mean change in alignment: 0.2° valgus (. sd. 1.2); p = 0.60). Conclusion. PCL resection creates flexion-extension mismatch by increasing the flexion gap more than the extension gap. The increase in the lateral flexion gap is greater than the increase in the medial flexion gap, which creates mediolateral laxity in flexion. Improvements in FFD following PCL resection are dependent on the degree of deformity before PCL resection. Cite this article: Bone Joint J 2019;101-B:1230–1237


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 507 - 514
1 Mar 2021
Chang JS Kayani B Wallace C Haddad FS

Aims. Total knee arthroplasty (TKA) using functional alignment aims to implant the components with minimal compromise of the soft-tissue envelope by restoring the plane and obliquity of the non-arthritic joint. The objective of this study was to determine the effect of TKA with functional alignment on mediolateral soft-tissue balance as assessed using intraoperative sensor-guided technology. Methods. This prospective study included 30 consecutive patients undergoing robotic-assisted TKA using the Stryker PS Triathlon implant with functional alignment. Intraoperative soft-tissue balance was assessed using sensor-guided technology after definitive component implantation; soft-tissue balance was defined as intercompartmental pressure difference (ICPD) of < 15 psi. Medial and lateral compartment pressures were recorded at 10°, 45°, and 90° of knee flexion. This study included 18 females (60%) and 12 males (40%) with a mean age of 65.2 years (SD 9.3). Mean preoperative hip-knee-ankle deformity was 6.3° varus (SD 2.7°). Results. TKA with functional alignment achieved balanced medial and lateral compartment pressures at 10° (25.0 psi (SD 6.1) vs 23.1 psi (SD 6.7), respectively; p = 0.140), 45° (21.4 psi (SD 5.9) vs 20.6 psi (SD 5.9), respectively; p = 0.510), and 90° (21.2 psi (SD 7.1) vs 21.6 psi (SD 9.0), respectively; p = 0.800) of knee flexion. Mean ICPD was 6.1 psi (SD 4.5; 0 to 14) at 10°, 5.4 psi (SD 3.9; 0 to 12) at 45°, and 4.9 psi (SD 4.45; 0 to 15) at 90° of knee flexion. Mean postoperative limb alignment was 2.2° varus (SD 1.0°). Conclusion. TKA using the functional alignment achieves balanced mediolateral soft-tissue tension through the arc of knee flexion as assessed using intraoperative pressure-sensor technology. Further clinical trials are required to determine if TKA with functional alignment translates to improvements in patient satisfaction and outcomes compared to conventional alignment techniques. Cite this article: Bone Joint J 2021;103-B(3):507–514


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 59 - 66
1 Jun 2021
Abhari S Hsing TM Malkani MM Smith AF Smith LS Mont MA Malkani AL

Aims. Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment. Methods. A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019. Results. Mean overall patient satisfaction for the entire cohort was 4.7 (SE 0.1) on a 5-point Likert scale, with 93% being either very satisfied or satisfied compared with a Likert of 4.3 and patient satisfaction of 81% in the mechanical alignment group (p < 0.001 and p < 0.006 respectively). At mean follow-up of 17 months (11 to 27), the mean overall Likert, Knee Injury and Osteoarthritis Outcome Score for Joint Replacement, Western Ontario and McMaster Universities Osteoarthritis Index, Forgotten Joint Score, and Knee Society Knee and Function Scores were significantly better in the kinematic group than in the neutral mechanical alignment group. The most common complication in both groups was contracture requiring manipulation under anaesthesia, involving seven knees (6.1%) in the kinematic group and nine knees (7.8%) in the mechanical alignment group. Conclusion. With the advent of advanced technology, and the ability to obtain accurate bone cuts, the target limb alignment, and soft-tissue balance within millimetres, using a restricted kinematic alignment concept demonstrated excellent patient satisfaction following primary TKA. Longer-term analysis is required as to the durability of this method. Cite this article: Bone Joint J 2021;103-B(6 Supple A):59–66


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 113 - 122
1 Jan 2021
Kayani B Tahmassebi J Ayuob A Konan S Oussedik S Haddad FS

Aims. The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups. Methods. This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups. Results. Patients undergoing conventional TKA and robotic TKA had comparable changes in the postoperative systemic inflammatory and localized thermal response at six hours, day 1, day 2, and day 28 after surgery. Robotic TKA had significantly reduced levels of interleukin-6 (p < 0.001), tumour necrosis factor-α (p = 0.021), ESR (p = 0.001), CRP (p = 0.004), lactate dehydrogenase (p = 0.007), and creatine kinase (p = 0.004) at day 7 after surgery compared with conventional TKA. Robotic TKA was associated with significantly improved preservation of the periarticular soft tissue envelope (p < 0.001), and reduced femoral (p = 0.012) and tibial (p = 0.023) bone trauma compared with conventional TKA. Robotic TKA significantly improved the accuracy of achieving the planned limb alignment (p < 0.001), femoral component positioning (p < 0.001), and tibial component positioning (p < 0.001) compared with conventional TKA. Conclusion. Robotic TKA was associated with a transient reduction in the early (day 7) postoperative inflammatory response but there was no difference in the immediate (< 48 hours) or late (day 28) postoperative systemic inflammatory response compared with conventional TKA. Robotic TKA was associated with decreased iatrogenic periarticular soft tissue injury, reduced femoral and tibial bone trauma, and improved accuracy of component positioning compared with conventional TKA. Cite this article: Bone Joint J 2021;103-B(1):113–122


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 329 - 337
1 Feb 2021
MacDessi SJ Griffiths-Jones W Harris IA Bellemans J Chen DB

Aims. A comprehensive classification for coronal lower limb alignment with predictive capabilities for knee balance would be beneficial in total knee arthroplasty (TKA). This paper describes the Coronal Plane Alignment of the Knee (CPAK) classification and examines its utility in preoperative soft tissue balance prediction, comparing kinematic alignment (KA) to mechanical alignment (MA). Methods. A radiological analysis of 500 healthy and 500 osteoarthritic (OA) knees was used to assess the applicability of the CPAK classification. CPAK comprises nine phenotypes based on the arithmetic HKA (aHKA) that estimates constitutional limb alignment and joint line obliquity (JLO). Intraoperative balance was compared within each phenotype in a cohort of 138 computer-assisted TKAs randomized to KA or MA. Primary outcomes included descriptive analyses of healthy and OA groups per CPAK type, and comparison of balance at 10° of flexion within each type. Secondary outcomes assessed balance at 45° and 90° and bone recuts required to achieve final knee balance within each CPAK type. Results. There was similar frequency distribution between healthy and arthritic groups across all CPAK types. The most common categories were Type II (39.2% healthy vs 32.2% OA), Type I (26.4% healthy vs 19.4% OA) and Type V (15.4% healthy vs 14.6% OA). CPAK Types VII, VIII, and IX were rare in both populations. Across all CPAK types, a greater proportion of KA TKAs achieved optimal balance compared to MA. This effect was largest, and statistically significant, in CPAK Types I (100% KA vs 15% MA; p < 0.001), Type II (78% KA vs 46% MA; p = 0.018). and Type IV (89% KA vs 0% MA; p < 0.001). Conclusion. CPAK is a pragmatic, comprehensive classification for coronal knee alignment, based on constitutional alignment and JLO, that can be used in healthy and arthritic knees. CPAK identifies which knee phenotypes may benefit most from KA when optimization of soft tissue balance is prioritized. Further, it will allow for consistency of reporting in future studies. Cite this article: Bone Joint J 2021;103-B(2):329–337


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 49 - 58
1 Jun 2020
Mullaji A

Aims. The aims of this study were to determine the effect of osteophyte excision on deformity correction and soft tissue gap balance in varus knees undergoing computer-assisted total knee arthroplasty (TKA). Methods. A total of 492 consecutive, cemented, cruciate-substituting TKAs performed for varus osteoarthritis were studied. After exposure and excision of both cruciates and menisci, it was noted from operative records the corrective interventions performed in each case. Knees in which no releases after the initial exposure, those which had only osteophyte excision, and those in which further interventions were performed were identified. From recorded navigation data, coronal and sagittal limb alignment, knee flexion range, and medial and lateral gap distances in maximum knee extension and 90° knee flexion with maximal varus and valgus stresses, were established, initially after exposure and excision of both cruciate ligaments, and then also at trialling. Knees were defined as ‘aligned’ if the hip-knee-ankle axis was between 177° and 180°, (0° to 3° varus) and ‘balanced’ if medial and lateral gaps in extension and at 90° flexion were within 2 mm of each other. Results. Of 50 knees (10%) with no soft tissue releases (other than cruciate ligaments), 90% were aligned, 81% were balanced, and 73% were aligned and balanced. In 288 knees (59%) only osteophyte excision was performed by subperiosteally releasing the deep medial collateral ligament. Of these, 98% were aligned, 80% were balanced, and 79% were aligned and balanced. In 154 knees (31%), additional procedures were performed (reduction osteotomy, posterior capsular release, and semimembranosus release). Of these, 89% were aligned, 68% were balanced, and 66% were aligned and balanced. The superficial medial collateral ligament was not released in any case. Conclusion. Two-thirds of all knees could be aligned and balanced with release of the cruciate ligaments alone and excision of osteophytes. Excision of osteophytes can be a useful step towards achieving deformity correction and gap balance without having to resort to soft tissue release in varus knees while maintaining classical coronal and sagittal alignment of components. Cite this article: Bone Joint J 2020;102-B(6 Supple A):49–58


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 331 - 339
1 Mar 2019
McEwen P Balendra G Doma K

Aims. The results of kinematic total knee arthroplasty (KTKA) have been reported in terms of limb and component alignment parameters but not in terms of gap laxities and differentials. In kinematic alignment (KA), balance should reflect the asymmetrical balance of the normal knee, not the classic rectangular flexion and extension gaps sought with gap-balanced mechanical axis total knee arthroplasty (MATKA). This paper aims to address the following questions: 1) what factors determine coronal joint congruence as measured on standing radiographs?; 2) is flexion gap asymmetry produced with KA?; 3) does lateral flexion gap laxity affect outcomes?; 4) is lateral flexion gap laxity associated with lateral extension gap laxity?; and 5) can consistent ligament balance be produced without releases?. Patients and Methods. A total of 192 KTKAs completed by a single surgeon using a computer-assisted technique were followed for a mean of 3.5 years (2 to 5). There were 116 male patients (60%) and 76 female patients (40%) with a mean age of 65 years (48 to 88). Outcome measures included intraoperative gap laxity measurements and component positions, as well as joint angles from postoperative three-foot standing radiographs. Patient-reported outcome measures (PROMs) were analyzed in terms of alignment and balance: EuroQol (EQ)-5D visual analogue scale (VAS), Knee Injury and Osteoarthritis Outcome Score (KOOS), KOOS Joint Replacement (JR), and Oxford Knee Score (OKS). Results. Postoperative limb alignment did not affect outcomes. The standing hip-knee-ankle (HKA) angle was the sole positive predictor of the joint line convergence angle (JLCA) (p < 0.001). Increasing lateral flexion gap laxity was consistently associated with better outcomes. Lateral flexion gap laxity did not correlate with HKA angle, the JLCA, or lateral extension gap laxity. Minor releases were required in one third of cases. Conclusion. The standing HKA angle is the primary determinant of the JLCA in KTKA. A rectangular flexion gap is produced in only 11% of cases. Lateral flexion gap laxity is consistently associated with better outcomes and does not affect balance in extension. Minor releases are sometimes required as well, particularly in limbs with larger preoperative deformities. Cite this article: Bone Joint J 2019;101-B:331–339


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 354 - 359
1 Mar 2013
Chareancholvanich K Narkbunnam R Pornrattanamaneewong C

Patient-specific cutting guides (PSCGs) are designed to improve the accuracy of alignment of total knee replacement (TKR). We compared the accuracy of limb alignment and component positioning after TKR performed using PSCGs or conventional instrumentation. A total of 80 patients were randomised to undergo TKR with either of the different forms of instrumentation, and radiological outcomes and peri-operative factors such as operating time were assessed. No significant difference was observed between the groups in terms of tibiofemoral angle or femoral component alignment. Although the tibial component in the PSCGs group was measurably closer to neutral alignment than in the conventional group, the size of the difference was very small (89.8° (. sd. 1.2) vs 90.5° (. sd. 1.6); p = 0.030). This new technology slightly shortened the bone-cutting time by a mean of 3.6 minutes (p < 0.001) and the operating time by a mean 5.1 minutes (p = 0.019), without tangible differences in post-operative blood loss (p = 0.528) or need for blood transfusion (p = 0.789). This study demonstrated that both PSCGs and conventional instrumentation restore limb alignment and place the components with the similar accuracy. The minimal advantages of PSCGs in terms of consistency of alignment or operative time are unlikely to be clinically relevant. Cite this article: Bone Joint J 2013;95-B:354–9


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1033 - 1042
1 Aug 2018
Kayani B Konan S Pietrzak JRT Huq SS Tahmassebi J Haddad FS

Aims. The primary aim of this study was to determine the surgical team’s learning curve for introducing robotic-arm assisted unicompartmental knee arthroplasty (UKA) into routine surgical practice. The secondary objective was to compare accuracy of implant positioning in conventional jig-based UKA versus robotic-arm assisted UKA. Patients and Methods. This prospective single-surgeon cohort study included 60 consecutive conventional jig-based UKAs compared with 60 consecutive robotic-arm assisted UKAs for medial compartment knee osteoarthritis. Patients undergoing conventional UKA and robotic-arm assisted UKA were well-matched for baseline characteristics including a mean age of 65.5 years (. sd. 6.8) vs 64.1 years (. sd. 8.7), (p = 0.31); a mean body mass index of 27.2 kg.m2 (. sd. 2.7) vs 28.1 kg.m2 (. sd. 4.5), (p = 0.25); and gender (27 males: 33 females vs 26 males: 34 females, p = 0.85). Surrogate measures of the learning curve were prospectively collected. These included operative times, the Spielberger State-Trait Anxiety Inventory (STAI) questionnaire to assess preoperative stress levels amongst the surgical team, accuracy of implant positioning, limb alignment, and postoperative complications. Results. Robotic-arm assisted UKA was associated with a learning curve of six cases for operating time (p < 0.001) and surgical team confidence levels (p < 0.001). Cumulative robotic experience did not affect accuracy of implant positioning (p = 0.52), posterior condylar offset ratio (p = 0.71), posterior tibial slope (p = 0.68), native joint line preservation (p = 0.55), and postoperative limb alignment (p = 0.65). Robotic-arm assisted UKA improved accuracy of femoral (p < 0.001) and tibial (p < 0.001) implant positioning with no additional risk of postoperative complications compared to conventional jig-based UKA. Conclusion. Robotic-arm assisted UKA was associated with a learning curve of six cases for operating time and surgical team confidence levels but no learning curve for accuracy of implant positioning. Cite this article: Bone Joint J 2018;100-B:1033–42


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 498 - 502
1 Apr 2015
Deep K Eachempati KK Apsingi S

The restoration of knee alignment is an important goal during total knee arthroplasty (TKA). In the past surgeons aimed to restore neutral limb alignment during surgery. However, previous studies have demonstrated alignment to be dynamic, varying depending on the position of the limb and the degree of weight-bearing, and between patients. We used a validated computer navigation system to measure the femorotibial mechanical angle (FTMA) in 264 knees in 77 male and 55 female healthy volunteers aged 18 to 35 years (mean 26.2). We found the mean supine alignment to be a varus angle of 1.2° (standard deviation (. sd. ) 4), with few patients having neutral alignment. FTMA differs significantly between males and females (with a mean varus of 1.7° (. sd. 4) and 0.4° (. sd. 3.9), respectively; p = 0.008). It changes significantly with posture, the knee hyperextending by a mean of 5.6°, and coronal plane alignment becoming more varus by 2.2° (. sd. 3.6) on standing compared with supine. Knee alignment is different in different individuals and is dynamic in nature, changing with different postures. This may have implications for the assessment of alignment in TKA, which is achieved in non-weight-bearing conditions and which may not represent the situation observed during weight-bearing. Cite this article: Bone Joint J 2015; 97-B:498–502


The Bone & Joint Journal
Vol. 106-B, Issue 7 | Pages 680 - 687
1 Jul 2024
Mancino F Fontalis A Grandhi TSP Magan A Plastow R Kayani B Haddad FS

Aims

Robotic arm-assisted surgery offers accurate and reproducible guidance in component positioning and assessment of soft-tissue tensioning during knee arthroplasty, but the feasibility and early outcomes when using this technology for revision surgery remain unknown. The objective of this study was to compare the outcomes of robotic arm-assisted revision of unicompartmental knee arthroplasty (UKA) to total knee arthroplasty (TKA) versus primary robotic arm-assisted TKA at short-term follow-up.

Methods

This prospective study included 16 patients undergoing robotic arm-assisted revision of UKA to TKA versus 35 matched patients receiving robotic arm-assisted primary TKA. In all study patients, the following data were recorded: operating time, polyethylene liner size, change in haemoglobin concentration (g/dl), length of inpatient stay, postoperative complications, and hip-knee-ankle (HKA) alignment. All procedures were performed using the principles of functional alignment. At most recent follow-up, range of motion (ROM), Forgotten Joint Score (FJS), and Oxford Knee Score (OKS) were collected. Mean follow-up time was 21 months (6 to 36).


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 525 - 531
1 Jun 2024
MacDessi SJ van de Graaf VA Wood JA Griffiths-Jones W Bellemans J Chen DB

The aim of mechanical alignment in total knee arthroplasty is to align all knees into a fixed neutral position, even though not all knees are the same. As a result, mechanical alignment often alters a patient’s constitutional alignment and joint line obliquity, resulting in soft-tissue imbalance. This annotation provides an overview of how the Coronal Plane Alignment of the Knee (CPAK) classification can be used to predict imbalance with mechanical alignment, and then offers practical guidance for bone balancing, minimizing the need for soft-tissue releases.

Cite this article: Bone Joint J 2024;106-B(6):525–531.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 914 - 922
1 Jul 2014
Lee SY Bae JH Kim JG Jang KM Shon WY Kim KW Lim HC

The aim of this study was to evaluate the risk factors for dislocation of the bearing after a mobile-bearing Oxford medial unicompartmental knee replacement (UKR) and to test the hypothesis that surgical factors, as measured from post-operative radiographs, are associated with its dislocation. From a total of 480 UKRs performed between 2001 and 2012, in 391 patients with a mean age of 66.5 years (45 to 82) (316 female, 75 male), we identified 17 UKRs where bearing dislocation occurred. The post-operative radiological measurements of the 17 UKRs and 51 matched controls were analysed using conditional logistic regression analysis. The post-operative radiological measurements included post-operative change in limb alignment, the position of the femoral and tibial components, the resection depth of the proximal tibia, and the femoral component-posterior condyle classification. We concluded that a post-operative decrease in the posterior tibial slope relative to the pre-operative value was the only significant determinant of dislocation of the bearing after medial Oxford UKR (odds ratio 1.881; 95% confidence interval 1.272 to 2.779). A post-operative posterior tibial slope < 8.45° and a difference between the pre-operative and post-operative posterior tibial slope of > 2.19° may increase the risk of dislocation. Cite this article: Bone Joint J 2014; 96-B:914–22


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1259 - 1264
1 Dec 2023
Hurley ET Hughes AJ Savage-Elliott I Dejour D Campbell KA Mulcahey MK Wittstein JR Jazrawi LM

Aims

The aim of this study was to establish consensus statements on the diagnosis, nonoperative management, and indications, if any, for medial patellofemoral complex (MPFC) repair in patients with patellar instability, using the modified Delphi approach.

Methods

A total of 60 surgeons from 11 countries were invited to develop consensus statements based on their expertise in this area. They were assigned to one of seven working groups defined by subtopics of interest within patellar instability. Consensus was defined as achieving between 80% and 89% agreement, strong consensus was defined as between 90% and 99% agreement, and 100% agreement was considered to be unanimous.


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 301 - 306
1 Mar 2023
Jennison T Ukoumunne O Lamb S Sharpe I Goldberg AJ

Aims

Despite the increasing numbers of ankle arthroplasties, there are limited studies on their survival and comparisons between different implants. The primary aim of this study was to determine the failure rates of primary ankle arthroplasties commonly used in the UK.

Methods

A data linkage study combined National Joint Registry (NJR) data and NHS Digital data. The primary outcome of failure was defined as the removal or exchange of any components of the implanted device. Life tables and Kaplan-Meier survival charts were used to illustrate survivorship. Cox proportional hazards regression models were fitted to compare failure rates between 1 April 2010 and 31 December 2018.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 786 - 792
1 Jun 2016
Schotanus MGM Sollie R van Haaren EH Hendrickx RPM Jansen EJP Kort NP

Aims. This prospective randomised controlled trial was designed to evaluate the outcome of both the MRI- and CT-based patient-specific matched guides (PSG) from the same manufacturer. Patients and Methods. A total of 137 knees in 137 patients (50 men, 87 women) were included, 67 in the MRI- and 70 in the CT-based PSG group. Their mean age was 68.4 years (47.0 to 88.9). Outcome was expressed as the biomechanical limb alignment (centre hip-knee-ankle: HKA-axis) achieved post-operatively, the position of the individual components within 3° of the pre-operatively planned alignment, correct planned implant size and operative data (e.g. operating time and blood loss). Results. The patient demographics (e.g. age, body mass index), correct planned implant size and operative data were not significantly different between the two groups. The proportion of outliers in the coronal and sagittal plane ranged from 0% to 21% in both groups. Only the number of outliers for the posterior slope of the tibial component showed a significant difference (p = 0.004) with more outliers in the CT group (n = 9, 13%) than in the MRI group (0%). . Conclusion. The post-operative HKA-axis was comparable in the MRI- and CT-based PSGs, but there were significantly more outliers for the posterior slope in the CT-based PSGs. Take home message: Alignment with MRI-based PSG is at least as good as, if not better, than that of the CT-based PSG, and is the preferred imaging modality when performing TKA with use of PSG. Cite this article: Bone Joint J 2016;98-B:786–92


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 541 - 548
1 May 2022
Zhang J Ng N Scott CEH Blyth MJG Haddad FS Macpherson GJ Patton JT Clement ND

Aims

This systematic review aims to compare the precision of component positioning, patient-reported outcome measures (PROMs), complications, survivorship, cost-effectiveness, and learning curves of MAKO robotic arm-assisted unicompartmental knee arthroplasty (RAUKA) with manual medial unicompartmental knee arthroplasty (mUKA).

Methods

Searches of PubMed, MEDLINE, and Google Scholar were performed in November 2021 according to the Preferred Reporting Items for Systematic Review and Meta-­Analysis statement. Search terms included “robotic”, “unicompartmental”, “knee”, and “arthroplasty”. Published clinical research articles reporting the learning curves and cost-effectiveness of MAKO RAUKA, and those comparing the component precision, functional outcomes, survivorship, or complications with mUKA, were included for analysis.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1686 - 1694
1 Nov 2021
Yang H Kwak W Kang SJ Song E Seon J

Aims

To determine the relationship between articular cartilage status and clinical outcomes after medial opening-wedge high tibial osteotomy (MOHTO) for medial compartmental knee osteoarthritis at intermediate follow-up.

Methods

We reviewed 155 patients (155 knees) who underwent MOHTO from January 2008 to December 2016 followed by second-look arthroscopy with a mean 5.3-year follow-up (2.0 to 11.7). Arthroscopic findings were assessed according to the International Cartilage Repair Society (ICRS) Cartilage Repair Assessment (CRA) grading system. Patients were divided into two groups based on the presence of normal or nearly normal quality cartilage in the medial femoral condyle: good (second-look arthroscopic) status (ICRS grade I or II; n = 70), and poor (second-look arthroscopic) status (ICRS grade III or IV; n = 85) groups at the time of second-look arthroscopy. Clinical outcomes were assessed using the International Knee Documentation Committee (IKDC) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 36-Item Short Form survey.


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 657 - 662
1 Jun 2022
Barlow T Coco V Shivji F Grassi A Asplin L Thompson P Metcalfe A Zaffagnini S Spalding T

Aims

Meniscal allograft transplantation (MAT) for patients with symptomatic meniscal loss has demonstrated good clinical results and survivorship. Factors that affect both functional outcome and survivorship have been reported in the literature. These are typically single-centre case series with relatively small numbers and conflicting results. Our aim was to describe an international, two-centre case series, and identify factors that affect both functional outcome and survival.

Methods

We report factors that affect outcome on 526 patients undergoing MAT across two sites (one in the UK and one in Italy). Outcomes of interest were the Knee injury and Osteoarthritis Outcome Score four (KOOS4) at two years and failure rates. We performed multiple regression analysis to examine for factors affecting KOOS, and Cox proportional hazards models for survivorship.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 34 - 44
1 Jan 2022
Beckers L Dandois F Ooms D Berger P Van Laere K Scheys L Vandenneucker H

Aims

Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs.

Methods

In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 5 | Pages 715 - 720
1 Sep 1991
Cooke T Scudamore R Bryant J Sorbie C Siu D Fisher B

A method is described which provides standardised reproducible radiographic images of the lower limb. Anteroposterior and lateral radiographs are digitised and processed by computer to provide graphic/numeric displays of angles and linear measurements, relating the centre points of the hip, knee, and ankle. Two cases illustrate how surgical planning is facilitated when standardised data are available. These data confirm the close relationship between postoperative limb alignment and positioning of prosthetic elements


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 137 - 144
1 Jun 2021
Lachiewicz PF Steele JR Wellman SS

Aims

To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period.

Methods

This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 338 - 346
1 Feb 2021
Khow YZ Liow MHL Lee M Chen JY Lo NN Yeo SJ

Aims

This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA.

Methods

Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 3 | Pages 358 - 364
1 May 1993
Eyres K Bell M Kanis J

We studied the quantity and rate of formation of new bone during lengthening of 17 limb segments in 10 patients using dual-energy X-ray absorptiometry (DEXA), ultrasonography and radiography. Whereas new bone was detected by both DEXA and ultrasonography within 1 to 2 weeks of distraction, it was not visible on the radiographs until 4 to 8 weeks. Limb alignment and gap measurement were accurately assessed by DEXA without the need for standard radiographs or scanograms. With ultrasound the distraction gap appeared as an echolucent window which narrowed progressively producing a hyper-reflecting line after which further consolidation could not be assessed. As measured by DEXA the density of the new bone at this stage was approximately 45% of control values and did not represent normal cortication. Whereas ultrasound could be used to identify defects in mineralisation and to determine when to dynamise the fixator system, DEXA could measure the quantity and rate of formation of bone throughout lengthening


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 58 - 63
1 Jan 2009
Hooper G Rothwell A Frampton C

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1367 - 1372
1 Aug 2021
Plancher KD Brite JE Briggs KK Petterson SC

Aims

The patient-acceptable symptom state (PASS) is a level of wellbeing, which is measured by the patient. The aim of this study was to determine if the proportion of patients who achieved an acceptable level of function (PASS) after medial unicompartmental knee arthroplasty (UKA) was different based on the status of the anterior cruciate ligament (ACL) at the time of surgery.

Methods

A total of 114 patients who underwent UKA for isolated medial osteoarthritis (OA) of the knee were included in the study. Their mean age was 65 years (SD 10). No patient underwent a bilateral procedure. Those who had undergone ACL reconstruction during the previous five years were excluded. The Knee injury Osteoarthritis Outcome Score Activities of Daily Living (KOOS ADL) function score was used as the primary outcome measure with a PASS of 87.5, as described for total knee arthroplasty (TKA). Patients completed all other KOOS subscales, Lysholm score, the Western Ontario and McMaster Universities Osteoarthritis Index, and the Veterans Rand 12-item health survey score. Failure was defined as conversion to TKA.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 610 - 618
1 Apr 2021
Batailler C Bordes M Lording T Nigues A Servien E Calliess T Lustig S

Aims

Ideal component sizing may be difficult to achieve in unicompartmental knee arthroplasty (UKA). Anatomical variants, incremental implant size, and a reduced surgical exposure may lead to over- or under-sizing of the components. The purpose of this study was to compare the accuracy of UKA sizing with robotic-assisted techniques versus a conventional surgical technique.

Methods

Three groups of 93 medial UKAs were assessed. The first group was performed by a conventional technique, the second group with an image-free robotic-assisted system (Image-Free group), and the last group with an image-based robotic arm-assisted system, using a preoperative CT scan (Image-Based group). There were no demographic differences between groups. We compared six parameters on postoperative radiographs to assess UKA sizing. Incorrect sizing was defined by an over- or under-sizing greater than 3 mm.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 7 | Pages 996 - 1000
1 Sep 2000
Weale AE Murray DW Baines J Newman JH

Failure of a unicompartmental knee replacement (UKR) may be caused by progressive osteoarthritis of the knee and/or failure of the prosthesis. Limb alignment can influence both of these factors. We have examined the fate of the other compartments and measured changes in leg alignment after UKR. A total of 50 UKRs was carried out on 45 carefully selected patients between 1989 and 1992. At operation, deliberate attempts were made to avoid overcorrection of the deformity. Four patients died, one patient was lost to follow-up and two knees were revised before review which was at a minimum of five years. Standard long-leg weight-bearing anteroposterior views of the knee and skyline views of the patellofemoral joint were taken before and at eight months and five years after operation. The radiographs of the remaining 43 knees were reviewed twice by blind and randomised assessment to measure the progression of osteoarthritis within the joints. Overcorrection of the deformity in the coronal plane was avoided in all but two knees. Only one showed evidence of progression of osteoarthritis within the patellofemoral joint, and this was only identified in one of the four assessments. Deterioration in the state of the opposite tibiofemoral compartment was not seen. Varus deformity tended to recur. Recurrent varus of 2° was observed between eight months and five years after operation. There was no correlation between the postoperative tibiofemoral angle and the extent of recurrent varus recorded at five years. Changes in alignment may be indicative of minor polyethylene wear or of subsidence of the tibial component. The incidence of progressive osteoarthritis within the knee was very low after UKR. Patients should be carefully selected and overcorrection of the deformity be avoided


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 67 - 73
1 Jun 2021
Lee G Wakelin E Randall A Plaskos C

Aims

Neither a surgeon’s intraoperative impression nor the parameters of computer navigation have been shown to be predictive of the outcomes following total knee arthroplasty (TKA). The aim of this study was to determine whether a surgeon, with robotic assistance, can predict the outcome as assessed using the Knee Injury and Osteoarthritis Outcome Score (KOOS) for pain (KPS), one year postoperatively, and establish what factors correlate with poor KOOS scores in a well-aligned and balanced TKA.

Methods

A total of 134 consecutive patients who underwent TKA using a dynamic ligament tensioning robotic system with a tibia first resection technique and a cruciate sacrificing ultracongruent TKA system were enrolled into a prospective study. Each TKA was graded based on the final mediolateral ligament balance at 10° and 90° of flexion: 1) < 1 mm difference in the thickness of the tibial insert and that which was planned (n = 75); 2) < 1 mm difference (n = 26); 3) between 1 mm to 2 mm difference (n = 26); and 4) > 2 mm difference (n = 7). The mean one-year KPS score for each grade of TKA was compared and the likelihood of achieving an KPS score of > 90 was calculated. Finally, the factors associated with lower KPS despite achieving a high-grade TKA (grade A and B) were analyzed.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 276 - 279
1 Mar 2020
Oussedik S Abdel MP Victor J Pagnano MW Haddad FS

Dissatisfaction following total knee arthroplasty is a well-documented phenomenon. Although many factors have been implicated, including modifiable and nonmodifiable patient factors, emphasis over the past decade has been on implant alignment and stability as both a cause of, and a solution to, this problem. Several alignment targets have evolved with a proliferation of techniques following the introduction of computer and robotic-assisted surgery. Mechanical alignment targets may achieve mechanically-sound alignment while ignoring the soft tissue envelope; kinematic alignment respects the soft tissue envelope while ignoring the mechanical environment. Functional alignment is proposed as a hybrid technique to allow mechanically-sound, soft tissue-friendly alignment targets to be identified and achieved.

Cite this article: Bone Joint J 2020;102-B(3):276–279.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 43 - 48
1 Jun 2020
D’Lima DP Huang P Suryanarayan P Rosen A D’Lima DD

Aims

The extensive variation in axial rotation of tibial components can lead to coronal plane malalignment. We analyzed the change in coronal alignment induced by tray malrotation.

Methods

We constructed a computer model of knee arthroplasty and used a virtual cutting guide to cut the tibia at 90° to the coronal plane. The virtual guide was rotated axially (15° medial to 15° lateral) and with posterior slopes (0° to 7°). To assess the effect of axial malrotation, we measured the coronal plane alignment of a tibial tray that was axially rotated (25° internal to 15° external), as viewed on a standard anteroposterior (AP) radiograph.


Aims

Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) versus components placed using computer-assisted surgery targeting neutral mechanical alignment (MA+CAS). Tibial component migration measured by radiostereometric analysis was the primary outcome measure (compared longitudinally between groups and to published acceptable thresholds). Secondary outcome measures were inducible displacement after one year and patient-reported outcome measures (PROMS) over two years. The secondary objective was to assess the relationship between alignment and both tibial component migration and inducible displacement.

Patients and Methods

A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (sd 8). In the MA+CAS group, there were 17 female and six male patients with a mean age of 63 years (sd 7). Surgery was performed using cemented, cruciate-retaining Triathlon total knees with patellar resurfacing, and patients were followed up for two years. The effect of alignment on tibial component migration and inducible displacement was analyzed irrespective of study group.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 129 - 137
1 Jun 2020
Knowlton CB Lundberg HJ Wimmer MA Jacobs JJ

Aims

A retrospective longitudinal study was conducted to compare directly volumetric wear of retrieved polyethylene inserts to predicted volumetric wear modelled from individual gait mechanics of total knee arthroplasty (TKA) patients.

Methods

In total, 11 retrieved polyethylene tibial inserts were matched with gait analysis testing performed on those patients. Volumetric wear on the articular surfaces was measured using a laser coordinate measure machine and autonomous reconstruction. Knee kinematics and kinetics from individual gait trials drove computational models to calculate medial and lateral tibiofemoral contact paths and forces. Sliding distance along the contact path, normal forces and implantation time were used as inputs to Archard’s equation of wear to predict volumetric wear from gait mechanics. Measured and modelled wear were compared for each component.


The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 24 - 33
1 Jan 2019
Kayani B Konan S Tahmassebi J Rowan FE Haddad FS

Aims

The objectives of this study were to compare postoperative pain, analgesia requirements, inpatient functional rehabilitation, time to hospital discharge, and complications in patients undergoing conventional jig-based unicompartmental knee arthroplasty (UKA) versus robotic-arm assisted UKA.

Patients and Methods

This prospective cohort study included 146 patients with symptomatic medial compartment knee osteoarthritis undergoing primary UKA performed by a single surgeon. This included 73 consecutive patients undergoing conventional jig-based mobile bearing UKA, followed by 73 consecutive patients receiving robotic-arm assisted fixed bearing UKA. All surgical procedures were performed using the standard medial parapatellar approach for UKA, and all patients underwent the same postoperative rehabilitation programme. Postoperative pain scores on the numerical rating scale and opiate analgesia consumption were recorded until discharge. Time to attainment of predefined functional rehabilitation outcomes, hospital discharge, and postoperative complications were recorded by independent observers.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 319 - 328
1 Mar 2020
St Mart J de Steiger RN Cuthbert A Donnelly W

Aim

There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several national joint registry studies documenting high UKA revision rates when compared to total knee arthroplasty (TKA). With the recent introduction of robotically assisted UKA procedures, it is hoped that outcomes improve. This study examines the cumulative revision rate of UKA procedures implanted with a newly introduced robotic system and compares the results to one of the best performing non-robotically assisted UKA prostheses, as well as all other non-robotically assisted UKA procedures.

Methods

Data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR) for all UKA procedures performed for osteoarthritis (OA) between 2015 and 2018 were analyzed. Procedures using the Restoris MCK UKA prosthesis implanted using the Mako Robotic-Arm Assisted System were compared to non-robotically assisted Zimmer Unicompartmental High Flex Knee System (ZUK) UKA, a commonly used UKA with previously reported good outcomes and to all other non-robotically assisted UKA procedures using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 565 - 572
1 May 2019
Teeter MG Marsh JD Howard JL Yuan X Vasarhelyi EM McCalden RW Naudie DDR

Aims

The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs.

Patients and Methods

The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (sd 8.4) in the PSI group and 69.4 years (sd 8.4) in the CSI group. All patients received the same TKA implant. Intraoperative surgical resources and any surgical waste generated were recorded. Patients underwent radiostereometric analysis (RSA) studies to measure femoral and tibial component migration over two years. Outcome measures were recorded pre- and postoperatively. Overall costs were calculated for each group.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1168 - 1176
1 Sep 2019
Calder PR McKay JE Timms AJ Roskrow T Fugazzotto S Edel P Goodier WD

Aims

The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients.

Patients and Methods

A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 915 - 921
1 Aug 2019
Beckers L Ooms D Berger P Van Laere K Scheys L Vandenneucker H

Aims

Altered alignment and biomechanics are thought to contribute to the progression of osteoarthritis (OA) in the native compartments after medial unicompartmental knee arthroplasty (UKA). The aim of this study was to evaluate the bone activity and remodelling in the lateral tibiofemoral and patellofemoral compartment after medial mobile-bearing UKA.

Patients and Methods

In total, 24 patients (nine female, 15 male) with 25 medial Oxford UKAs (13 left, 12 right) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively and at one and two years postoperatively, along with standard radiographs and clinical outcome scores. The mean patient age was 62 years (40 to 78) and the mean body mass index (BMI) was 29.7 kg/m2 (23.6 to 42.2). Mean osteoblastic activity was evaluated using a tracer localization scheme with volumes of interest (VOIs). Normalized mean tracer values were calculated as the ratio between the mean tracer activity in a VOI and background activity in the femoral diaphysis.


Aims

The objective of this study was to compare early postoperative functional outcomes and time to hospital discharge between conventional jig-based total knee arthroplasty (TKA) and robotic-arm assisted TKA.

Patients and Methods

This prospective cohort study included 40 consecutive patients undergoing conventional jig-based TKA followed by 40 consecutive patients receiving robotic-arm assisted TKA. All surgical procedures were performed by a single surgeon using the medial parapatellar approach with identical implant designs and standardized postoperative inpatient rehabilitation. Inpatient functional outcomes and time to hospital discharge were collected in all study patients.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 59 - 65
1 Jan 2019
Katchky AM Jones CW Walter WL Shimmin AJ

Aims

Between 15% and 20% of patients remain dissatisfied following total knee arthroplasty (TKA). The SAIPH knee system (MatOrtho, Surrey, United Kingdom) is a medial ball and socket TKA that has been designed to replicate native knee kinematics in order to maximize the range of movement, stability, and function. This system is being progressively introduced in a stepwise fashion, with this study reporting the mid-term clinical and radiological outcomes.

Patients and Methods

A retrospective review was undertaken of the first 100 consecutive patients with five-year follow-up following SAIPH TKA performed by the senior authors. The data that were collected included the demographics of the patients, clinical findings, the rate of intraoperative ligamentous release, patient-reported outcome measures (PROMS), radiological assessment, complications, and all-cause revision. Revision data were cross-checked with a national registry.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 435 - 442
1 Apr 2019
Zambianchi F Franceschi G Rivi E Banchelli F Marcovigi A Nardacchione R Ensini A Catani F

Aims

The purpose of this multicentre observational study was to investigate the association between intraoperative component positioning and soft-tissue balancing on short-term clinical outcomes in patients undergoing robotic-arm assisted unicompartmental knee arthroplasty (UKA).

Patients and Methods

Between 2013 and 2016, 363 patients (395 knees) underwent robotic-arm assisted UKAs at two centres. Pre- and postoperatively, patients were administered Knee Injury and Osteoarthritis Score (KOOS) and Forgotten Joint Score-12 (FJS-12). Results were stratified as “good” and “bad” if KOOS/FJS-12 were more than or equal to 80. Intraoperative, post-implantation robotic data relative to CT-based components placement were collected and classified. Postoperative complications were recorded.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 450 - 454
1 Apr 2018
Chalmers BP Mehrotra KG Sierra RJ Pagnano MW Taunton MJ Abdel MP

Aims

Primary (or spontaneous) and secondary osteonecrosis of the knee can lead to severe joint degeneration, for which either total or unicompartmental arthroplasty may be considered. However, there are limited studies analyzing outcomes of unicompartmental knee arthroplasties (UKAs) for osteonecrosis involving an isolated compartment of the knee. The aims of this study were to analyze outcomes of UKAs for osteonecrosis with specific focus on 1) survivorship free of any revision or reoperation, 2) risk factors for failure, 3) clinical outcomes, and 4) complications.

Patients and Methods

A total of 45 patients underwent 46 UKAs for knee osteonecrosis between 2002 and 2014 at our institution (The Mayo Clinic, Rochester, Minnesota). Twenty patients (44%) were female; the mean age of the patients was 66 years, and mean body mass index (BMI) was 31 kg/m2. Of the 46 UKAs, 44 (96%) were medial UKAs, and 35 (76%) were fixed-bearing design. Mean mechanical axis postoperatively was 1.5° varus (0° to 5° varus); 41 UKAs (89%) were performed for primary osteonecrosis. Mean follow-up was five years (2 to 12)


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1276 - 1282
1 Sep 2016
Donnan LT Gomes B Donnan A Harris C Torode† I Heidt C

Aims

We wished to examine the effectiveness of tibial lengthening using a two ring Ilizarov frame in skeletally immature patients. This is a potentially biomechanically unstable construct which risks the loss of axial control.

Patients and Methods

We retrospectively reviewed a consecutive series of 24 boys and 26 girls, with a mean age of 8.6 years (4 to 14), who underwent 52 tibial lengthening procedures with a mean follow-up of 4.3 years (4.0 to 16.9). Tibial alignment was measured before and after treatment using joint orientation lines from the knee and a calculation of the oblique plane axis.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 640 - 646
1 May 2017
Matsumoto T Takayama K Ishida K Hayashi S Hashimoto S Kuroda R

Aims

The aim of this study was to compare the post-operative radiographic and clinical outcomes between kinematically and mechanically aligned total knee arthroplasties (TKAs).

Patients and Methods

A total of 60 TKAs (30 kinematically and 30 mechanically aligned) were performed in 60 patients with varus osteoarthritis of the knee using a navigation system. The angles of orientation of the joint line in relation to the floor, the conventional and true mechanical axis (tMA) (the line from the centre of the hip to the lowest point of the calcaneus) were compared, one year post-operatively, on single-leg and double-leg standing long leg radiographs between the groups. The range of movement and 2011 Knee Society Scores were also compared between the groups at that time.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 65 - 69
1 Jan 2017
Thienpont E

Objectives

Unicompartmental knee arthroplasty (UKA) is a potential treatment for isolated bone on bone osteoarthritis when limited to a single compartment. The risk for revision of UKA is three times higher than for total knee arthroplasty (TKA). The aim of this review was to discuss the different revision options after UKA failure.

Materials and Methods

A search was performed for English language articles published between 2006 and 2016. After reviewing titles and abstracts, 105 papers were selected for further analysis. Of these, 39 papers were deemed to contain clinically relevant data to be included in this review.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 159 - 170
1 Feb 2017
Clark D Metcalfe A Wogan C Mandalia V Eldridge J

Patellar instability most frequently presents during adolescence. Congenital and infantile dislocation of the patella is a distinct entity from adolescent instability and measurable abnormalities may be present at birth. In the normal patellofemoral joint an increase in quadriceps angle and patellar height are matched by an increase in trochlear depth as the joint matures. Adolescent instability may herald a lifelong condition leading to chronic disability and arthritis.

Restoring normal anatomy by trochleoplasty, tibial tubercle transfer or medial patellofemoral ligament (MPFL) reconstruction in the young adult prevents further instability. Although these techniques are proven in the young adult, they may cause growth arrest and deformity where the physis is open. A vigorous non-operative strategy may permit delay of surgery until growth is complete. Where non-operative treatment has failed a modified MPFL reconstruction may be performed to maintain stability until physeal closure permits anatomical reconstruction. If significant growth remains an extraosseous reconstruction of the MPFL may impart the lowest risk to the physis. If minor growth remains image intensifier guided placement of femoral intraosseous fixation may impart a small, but acceptable, risk to the physis.

This paper presents and discusses the literature relating to adolescent instability and provides a framework for management of these patients.

Cite this article: Bone Joint J 2017;99-B:159–70.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 723 - 729
1 Jun 2016
Jones MH Williams AM

Osteochondritis Dissecans (OCD) is a condition for which the aetiology remains unknown. It affects subchondral bone and secondarily its overlying cartilage and is mostly found in the knee. It can occur in adults, but is generally identified when growth remains, when it is referred to as juvenile OCD. As the condition progresses, the affected subchondral bone separates from adjacent healthy bone, and can lead to demarcation and separation of its associated articular cartilage. Any symptoms which arise relate to the stage of the disease. Early disease without separation of the lesion results in pain. Separation of the lesion leads to mechanical symptoms and swelling and, in advanced cases, the formation of loose bodies.

Early identification of OCD is essential as untreated OCD can lead to the premature degeneration of the joint, whereas appropriate treatment can halt the disease process and lead to healing. Establishing the stability of the lesion is a key part of providing the correct treatment. Stable lesions, particularly in juvenile patients, have greater propensity to heal with non-surgical treatment, whereas unstable or displaced lesions usually require surgical management.

This article discusses the aetiology, clinical presentation and prognosis of OCD in the knee. It presents an algorithm for treatment, which aims to promote healing of native hyaline cartilage and to ensure joint congruity.

Take home message: Although there is no clear consensus as to the best treatment of OCD, every attempt should be made to retain the osteochondral fragment when possible as, with a careful surgical technique, there is potential for healing even in chronic lesions

Cite this article: Bone Joint J 2016;98-B:723–9.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 101 - 104
1 Jan 2016
Mullaji AB Shetty GM

Collateral ligament release is advocated in total knee arthroplasty (TKA) to deal with significant coronal plane deformities, but is also associated with significant disadvantages.

We describe steps to avoid release of the collateral (superficial medial and lateral collateral) ligaments during TKA in severely deformed knees, while correcting deformity and balancing the knee.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):101–4.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 81 - 83
1 Jan 2016
Allen MM Pagnano MW

The cause of dissatisfaction following total knee arthroplasty (TKA) remains elusive. Much attention has been focused on static mechanical alignment as a basis for surgical success and optimising outcomes. More recently, research on both normal and osteoarthritic knees, as well as kinematically aligned TKAs, has suggested that other specific and dynamic factors may be more important than a generic target of 0 ± 3º of a neutral axis. Consideration of these other variables is necessary to understand ideal targets and move beyond generic results.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):81–3.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1491 - 1497
1 Nov 2014
Howells NR Salmon L Waller A Scanelli J Pinczewski LA

The aim of this study was to examine the functional outcome at ten years following lateral closing wedge high tibial osteotomy for medial compartment osteoarthritis of the knee and to define pre-operative predictors of survival and determinants of functional outcome.

164 consecutive patients underwent high tibial osteotomy between 2000 and 2002. A total of 100 patients (100 knees) met the inclusion criteria and 95 were available for review at ten years. Data were collected prospectively and included patient demographics, surgical details, long leg alignment radiographs, Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and Knee Society scores (KSS) pre-operatively and at five and ten years follow-up.

At ten years, 21 patients had been revised at a mean of five years. Overall Kaplan–Meier survival was 87% (95% confidence interval (CI) 81 to 94) and 79% (95% CI 71 to 87) at five and ten years, respectively. When compared with unrevised patients, those who had been revised had significantly lower mean pre-operative WOMAC Scores (47 (21 to 85) vs 65 (32 to 99), p < 0.001), higher mean age (54 yrs (42 to 61) vs 49 yrs (26 to 66), p = 0.006) and a higher mean BMI (30.2; 25 to 39 vs 27.9; 21 to 36, p = 0.005). Each were found to be risk factors for revision, with hazard ratios of 10.7 (95% CI 4 to 28.6; pre-operative WOMAC < 45), 6.5 (95% CI 2.4 to 17.7; age > 55) and 3.0 (95%CI 1.2 to 7.6; BMI > 30). Survival of patients with pre-operative WOMAC > 45, age < 55 and BMI < 30 was 97% at five and ten years. WOMAC and KSS in surviving patients improved significantly between pre-operative (mean 61; 32 to 99) and five (mean 88; 35 to 100, p = 0.001) and ten years (mean 84; 38 to 100, p = 0.001). Older patients had better functional outcomes overall, despite their higher revision rate.

This study has shown that improved survival is associated with age < 55 years, pre-operative WOMAC scores > 45 and, a BMI < 30. In patients over 55 years of age with adequate pre-operative functional scores, survival can be good and functional outcomes can be significantly better than their younger counterparts. We recommend the routine use of pre-operative functional outcome scores to guide decision-making when considering suitability for high tibial osteotomy.

Cite this article: Bone Joint J 2014;96-B:1491–7.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 84 - 88
1 Jan 2016
Vince K

The term mid-flexion instability has entered the orthopaedic literature as a concept, but has not been confirmed as a distinct clinical entity. The term is used freely, sometimes as a synonym for flexion instability. However, the terms need to be clearly separated. A cadaver study published in 1990 associated joint line elevation with decreased stability at many angles of flexion, but that model was not typical of clinical scenarios. The literature is considered and it is proposed that the more common entity of an uncorrected flexion contracture after a measured resection arthroplasty technique is more likely to produce clinical findings that suggest instability mid-flexion.

It is proposed that the clinical scenario encountered is generalised instability, with the appearance of stability in full extension from tight posterior structures.

This paper seeks to clarify whether mid-flexion instability exists as an entity distinct from other commonly recognised forms of instability.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):84–8.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1348 - 1353
1 Oct 2013
Valenzuela GA Jacobson NA Buzas D Korecki TD Valenzuela RG Teitge RA

The outcome of high tibial osteotomy (HTO) deteriorates with time, and additional procedures may be required. The aim of this study was to compare the clinical and radiological outcomes between unicompartmental knee replacement (UKR) and total knee replacement (TKR) after HTO as well as after primary UKR. A total of 63 patients (63 knees) were studied retrospectively and divided into three groups: UKR after HTO (group A; n = 22), TKR after HTO (group B; n = 18) and primary UKR (group C; n = 22). The Oxford knee score (OKS), Knee Society score (KSS), hip–knee–ankle angles, mechanical axis and patellar height were evaluated pre- and post-operatively. At a mean of 64 months (19 to 180) post-operatively the mean OKS was 43.8 (33 to 49), 43.3 (30 to 48) and 42.5 (29 to 48) for groups A, B and C, respectively (p = 0.73). The mean KSS knee score was 88.8 (54 to 100), 88.11 (51 to 100) and 85.3 (45 to 100) for groups A, B and C, respectively (p = 0.65), and the mean KSS function score was 85.0 (50 to 100) in group A, 85.8 (20 to 100) in group B and 79.3 (50 to 100) in group C (p = 0.48). Radiologically the results were comparable for all groups except for patellar height, with a higher incidence of patella infra following a previous HTO (p = 0.02).

Cite this article: Bone Joint J 2013;95-B:1348–53.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 151 - 152
1 Nov 2012
Lackey WG Berend ME


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 112 - 114
1 Nov 2014
Abdel MP Haas SB

Instability after total knee replacement (TKR) accounts for 10% to 22% of revision procedures. All patients who present for evaluation of instability require a thorough history to be taken and physical examination, as well as appropriate imaging. Deep periprosthetic infection must be ruled out by laboratory testing and an aspiration of the knee must be carried out. The three main categories of instability include flexion instability, extension instability (symmetric and asymmetric), and genu recurvatum. Most recently, the aetiologies contributing to, and surgical manoeuvres required to correct, flexion instability have been elucidated. While implant design and patient-related factors may certainly contribute to the aetiology, surgical technique is also a significant factor in all forms of post-operative instability.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):112–4.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 50 - 55
1 Jan 2015
Zuiderbaan HA Khamaisy S Thein R Nawabi DH Pearle AD

Progressive degenerative changes in the medial compartment of the knee following lateral unicompartmental arthroplasty (UKA) remains a leading indication for revision surgery. The purpose of this study is to evaluate changes in the congruence and joint space width (JSW) of the medial compartment following lateral UKA. The congruence of the medial compartment of 53 knees (24 men, 23 women, mean age 13.1 years; sd 62.1) following lateral UKA was evaluated pre-operatively and six weeks post-operatively, and compared with 41 normal knees (26 men, 15 women, mean age 33.7 years; sd 6.4), using an Interactive closest point algorithm which calculated the congruence index (CI) by performing a rigid transformation that best aligns the digitised tibial and femoral surfaces. Inner, middle and outer JSWs were measured by sub-dividing the medial compartment into four quarters on pre- and post-operative, weight bearing tunnel view radiographs. The mean CI of knees following lateral UKA significantly improved from 0.92 (sd 0.06) pre-operatively to 0.96 (sd 0.02) (p < 0.001) six weeks post-operatively. The mean CI of the healthy control group was 0.99 sd 0.01. Post-operatively, the mean inner JSW increased (p = 0.006) and the outer decreased (p = 0.002). The JSW was restored post-operatively as no significant differences were noted in all three locations compared with the control group (inner JSW p = 0.43; middle JSW p = 0.019, outer JSW p = 0.51).

Our data suggest that a well conducted lateral UKA may improve the congruence and normalise the JSW of the medial compartment, potentially preventing progression of degenerative change.

Cite this article: Bone Joint J 2015;97-B:50–5.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 101 - 104
1 Nov 2014
Lombardi Jr AV Berend KR Adams JB

Previous studies of failure mechanisms leading to revision total knee replacement (TKR) performed between 1986 and 2000 determined that many failed early, with a disproportionate amount accounted for by infection and implant-associated factors including wear, loosening and instability. Since then, efforts have been made to improve implant performance and instruct surgeons in best practice. Recently our centre participated in a multi-centre evaluation of 844 revision TKRs from 2010 to 2011. The purpose was to report a detailed analysis of failure mechanisms over time and to see if failure modes have changed over the past 10 to 15 years. Aseptic loosening was the predominant mechanism of failure (31.2%), followed by instability (18.7%), infection (16.2%), polyethylene wear (10.0%), arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to failure was 5.9 years (ten days to 31 years), 35.3% of all revisions occurred at less than two years, and 60.2% in the first five years. With improvements in implant and polyethylene manufacture, polyethylene wear is no longer a leading cause of failure. Early mechanisms of failure are primarily technical errors. In addition to improving implant longevity, industry and surgeons must work together to decrease these technical errors. All reports on failure of TKR contain patients with unexplained pain who not infrequently have unmet expectations. Surgeons must work to achieve realistic patient expectations pre-operatively, and therefore, improve patient satisfaction post-operatively.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):101–4.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 337 - 345
1 Mar 2015
Jaroma AVJ Soininvaara TA Kröger H

Total knee arthroplasty (TKA) is known to lead to a reduction in periprosthetic bone mineral density (BMD). In theory, this may lead to migration, instability and aseptic loosening of the prosthetic components. Bisphosphonates inhibit bone resorption and may reduce this loss in BMD. We hypothesised that treatment with bisphosphonates and calcium would lead to improved BMD and clinical outcomes compared with treatment with calcium supplementation alone following TKA. A total of 26 patients, (nine male and 17 female, mean age 67 years) were prospectively randomised into two study groups: alendronate and calcium (bisphosphonate group, n = 14) or calcium only (control group, n = 12). Dual energy X-ray absorptiometry (DEXA) measurements were performed post-operatively, and at three months, six months, one, two, four, and seven years post-operatively.

Mean femoral metaphyseal BMD was significantly higher in the bisphosphonate group compared with controls, up to four years following surgery in some areas of the femur (p = 0.045). BMD was observed to increase in the lateral tibial metaphysis in the bisphosphonate group until seven years (p = 0.002), and was significantly higher than that observed in the control group throughout (p = 0.024). There were no significant differences between the groups in the central femoral metaphyseal, tibial medial metaphyseal or diaphyseal regions of interest (ROI) of either the femur or tibia.

Bisphosphonate treatment after TKA may be of benefit for patients with poor bone quality. However, further studies with a larger number of patients are necessary to assess whether this is clinically beneficial.

Cite this article: Bone Joint J 2015;97-B:337–45.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1499 - 1505
1 Nov 2012
Kim J Kasat NS Bae J Kim S Oh S Lim H

The purpose of this study was to measure the radiological parameters of femoral component alignment of the Oxford Phase 3 unicompartmental knee replacement (UKR), and evaluate their effect on clinical outcome. Multiple regression analysis was used to examine the relative contributions of the radiological assessment of femoral component alignment in 189 consecutive UKRs performed by a single surgeon. The American Knee Society scores were compared between groups, defined as being within or outside recommended tolerances of the position of the femoral component. For the flexion/extension position 21 UKRs (11.1%) lay outside the recommended limits, and for posterior overhang of the femoral component nine (4.8%) lay outside the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest canal distance from the distal femoral entry point of the alignment jig and coronal entry-point position had significant effects on the flexion/extension position. Pre-operative HKA angle had a significant influence on posterior overhang of the femoral component. However, there was no significant difference in American Knee Society scores relative to the position of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1160 - 1169
1 Sep 2012
Bohm ER Tufescu TV Marsh JP

This review considers the surgical treatment of displaced fractures involving the knee in elderly, osteoporotic patients. The goals of treatment include pain control, early mobilisation, avoidance of complications and minimising the need for further surgery. Open reduction and internal fixation (ORIF) frequently results in loss of reduction, which can result in post-traumatic arthritis and the occasional conversion to total knee replacement (TKR). TKR after failed internal fixation is challenging, with modest functional outcomes and high complication rates. TKR undertaken as treatment of the initial fracture has better results to late TKR, but does not match the outcome of primary TKR without complications. Given the relatively infrequent need for late TKR following failed fixation, ORIF is the preferred management for most cases. Early TKR can be considered for those patients with pre-existing arthritis, bicondylar femoral fractures, those who would be unable to comply with weight-bearing restrictions, or where a single definitive procedure is required.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 642 - 647
1 May 2012
Mullaji A Lingaraju AP Shetty GM

We retrospectively reviewed the records of 1150 computer-assisted total knee replacements and analysed the clinical and radiological outcomes of 45 knees that had arthritis with a pre-operative recurvatum deformity. The mean pre-operative hyperextension deformity of 11° (6° to 15°), as measured by navigation at the start of the operation, improved to a mean flexion deformity of 3.1° (0° to 7°) post-operatively. A total of 41 knees (91%) were managed using inserts ≤ 12.5 mm thick, and none had mediolateral laxity > 2 mm from a mechanical axis of 0° at the end of the surgery. At a mean follow-up of 26.4 months (13 to 48) there was significant improvement in the mean Knee Society, Oxford knee and Western Ontario and McMaster Universities Osteoarthritis Index scores compared with the pre-operative values. The mean knee flexion improved from 105° (80° to 125°) pre-operatively to 131° (120° to 145°), and none of the limbs had recurrent recurvatum.

These early results show that total knee replacement using computer navigation and an algorithmic approach for arthritic knees with a recurvatum deformity can give excellent radiological and functional outcomes without recurrent deformity.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 857 - 862
1 Jul 2014
Abdel MP Oussedik S Parratte S Lustig S Haddad FS

Substantial healthcare resources have been devoted to computer navigation and patient-specific instrumentation systems that improve the reproducibility with which neutral mechanical alignment can be achieved following total knee replacement (TKR). This choice of alignment is based on the long-held tenet that the alignment of the limb post-operatively should be within 3° of a neutral mechanical axis. Several recent studies have demonstrated no significant difference in survivorship when comparing well aligned versus malaligned TKRs. Our aim was to review the anatomical alignment of the knee, the historical and contemporary data on a neutral mechanical axis in TKR, and the feasibility of kinematically-aligned TKRs. . Review of the literature suggests that a neutral mechanical axis remains the optimal guide to alignment. Cite this article: Bone Joint J 2014;96-B:857–62


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 471 - 476
1 Apr 2007
Kim Y Kim J Yoon S

Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiological and CT imaging was carried out to determine the alignment of the components. The mean follow-up was 2.3 years (2 to 3).

The operating and tourniquet times were significantly longer in the navigation group (p < 0.001). There were no significant pre- or post-operative differences between the knee scores of the two groups (p = 0.288 and p = 0.429, respectively). The results of imaging and the number of outliers for all radiological parameters were not statistically different (p = 0.109 to p = 0.920).

In this series computer-assisted navigated total knee replacement did not result in more accurate orientation and alignment of the components than that achieved by conventional total knee replacement.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 907 - 913
1 Jul 2014
Dossett HG Estrada NA Swartz GJ LeFevre GW Kwasman BG

We have previously reported the short-term radiological results of a randomised controlled trial comparing kinematically aligned total knee replacement (TKR) and mechanically aligned TKR, along with early pain and function scores. In this study we report the two-year clinical results from this trial. A total of 88 patients (88 knees) were randomly allocated to undergo either kinematically aligned TKR using patient-specific guides, or mechanically aligned TKR using conventional instruments. They were analysed on an intention-to-treat basis. The patients and the clinical evaluator were blinded to the method of alignment.

At a minimum of two years, all outcomes were better for the kinematically aligned group, as determined by the mean Oxford knee score (40 (15 to 48) versus 33 (13 to 48); p = 0.005), the mean Western Ontario McMaster Universities Arthritis index (WOMAC) (15 (0 to 63) versus 26 (0 to 73); p = 0.005), mean combined Knee Society score (160 (93 to 200) versus 137 (64 to 200); p= 0.005) and mean flexion of 121° (100 to 150) versus 113° (80 to 130) (p = 0.002). The odds ratio of having a pain-free knee at two years with the kinematically aligned technique (Oxford and WOMAC pain scores) was 3.2 (p = 0.020) and 4.9 (p = 0.001), respectively, compared with the mechanically aligned technique. Patients in the kinematically aligned group walked a mean of 50 feet further in hospital prior to discharge compared with the mechanically aligned group (p = 0.044).

In this study, the use of a kinematic alignment technique performed with patient-specific guides provided better pain relief and restored better function and range of movement than the mechanical alignment technique performed with conventional instruments.

Cite this article: Bone Joint J 2014;96-B:907–13.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 609 - 618
1 May 2014
Gøthesen Ø Espehaug B Havelin LI Petursson G Hallan G Strøm E Dyrhovden G Furnes O

We performed a randomised controlled trial comparing computer-assisted surgery (CAS) with conventional surgery (CONV) in total knee replacement (TKR). Between 2009 and 2011 a total of 192 patients with a mean age of 68 years (55 to 85) with osteoarthritis or arthritic disease of the knee were recruited from four Norwegian hospitals. At three months follow-up, functional results were marginally better for the CAS group. Mean differences (MD) in favour of CAS were found for the Knee Society function score (MD: 5.9, 95% confidence interval (CI) 0.3 to 11.4, p = 0.039), the Knee Injury and Osteoarthritis Outcome Score (KOOS) subscales for ‘pain’ (MD: 7.7, 95% CI 1.7 to 13.6, p = 0.012), ‘sports’ (MD: 13.5, 95% CI 5.6 to 21.4, p = 0.001) and ‘quality of life’ (MD: 7.2, 95% CI 0.1 to 14.3, p = 0.046). At one-year follow-up, differences favouring CAS were found for KOOS ‘sports’ (MD: 11.0, 95% CI 3.0 to 19.0, p = 0.007) and KOOS ‘symptoms’ (MD: 6.7, 95% CI 0.5 to 13.0, p = 0.035). The use of CAS resulted in fewer outliers in frontal alignment (> 3° malalignment), both for the entire TKR (37.9% vs 17.9%, p = 0.042) and for the tibial component separately (28.4% vs 6.3%, p = 0.002). Tibial slope was better achieved with CAS (58.9% vs 26.3%, p < 0.001). Operation time was 20 minutes longer with CAS. In conclusion, functional results were, statistically, marginally in favour of CAS. Also, CAS was more predictable than CONV for mechanical alignment and positioning of the prosthesis. However, the long-term outcomes must be further investigated.

Cite this article: Bone Joint J 2014; 96-B:609–18.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 57 - 61
1 Jan 2007
Lee ST Song HR Mahajan R Makwana V Suh SW Lee SH

Genu varum in the achondroplastic patient has a complex and multifactorial aetiology. There is little mention in the literature of the role of fibular overgrowth. Using the ratio of fibular to tibial length as a measurement of possible fibular overgrowth, we have related it to the development of genu varum. Full-length standing anteroposterior radiographs of 53 patients with achondroplasia were analysed. There were 30 skeletally-immature and 23 skeletally-mature patients. Regression analysis was performed in order to determine if there was a causal relationship between fibular overgrowth and the various indices of alignment of the lower limb.

Analysis showed that the fibular to tibial length ratio had a significant correlation with the medial proximal tibial angle and the mechanical axial deviation in the skeletally-immature group. We conclude that there is a significant relationship between fibular overgrowth and the development of genu varum in the skeletally-immature achondroplastic patient.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 906 - 910
1 Jul 2013
Lin S Chen C Fu Y Huang P Lu C Su J Chang J Huang H

Minimally invasive total knee replacement (MIS-TKR) has been reported to have better early recovery than conventional TKR. Quadriceps-sparing (QS) TKR is the least invasive MIS procedure, but it is technically demanding with higher reported rates of complications and outliers. This study was designed to compare the early clinical and radiological outcomes of TKR performed by an experienced surgeon using the QS approach with or without navigational assistance (NA), or using a mini-medial parapatellar (MP) approach. In all, 100 patients completed a minimum two-year follow-up: 30 in the NA-QS group, 35 in the QS group, and 35 in the MP group. There were no significant differences in clinical outcome in terms of ability to perform a straight-leg raise at 24 hours (p = 0.700), knee score (p = 0.952), functional score (p = 0.229) and range of movement (p = 0.732) among the groups. The number of outliers for all three radiological parameters of mechanical axis, frontal femoral component alignment and frontal tibial component alignment was significantly lower in the NA-QS group than in the QS group (p = 0.008), but no outlier was found in the MP group.

In conclusion, even after the surgeon completed a substantial number of cases before the commencement of this study, the supplementary intra-operative use of computer-assisted navigation with QS-TKR still gave inferior radiological results and longer operating time, with a similar outcome at two years when compared with a MP approach.

Cite this article: Bone Joint J 2013;95-B:906–10.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 601 - 605
1 May 2006
Pitto RP Graydon AJ Bradley L Malak SF Walker CG Anderson IA

The object of this study was to develop a method to assess the accuracy of an image-free total knee replacement navigation system in legs with normal or abnormal mechanical axes. A phantom leg was constructed with simulated hip and knee joints and provided a means to locate the centre of the ankle joint. Additional joints located at the midshaft of the tibia and femur allowed deformation in the flexion/extension, varus/valgus and rotational planes. Using a digital caliper unit to measure the coordinates precisely, a software program was developed to convert these local coordinates into a determination of actual leg alignment. At specific points in the procedure, information was compared between the digital caliper measurements and the image-free navigation system. Repeated serial measurements were undertaken. In the setting of normal alignment the mean error of the system was within 0.5°. In the setting of abnormal plane alignment in both the femur and the tibia, the error was within 1°. This is the first study designed to assess the accuracy of a clinically-validated navigation system. It demonstrates in vitro accuracy of the image-free navigation system in both normal and abnormal leg alignment settings.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 1 - 6
1 Jan 2010
Clifton R Ng CY Nutton RW

We have reviewed the literature to establish the role of lateral retinacular release in the management of disorders of the extensor apparatus of the knee. The scientific evidence for intervention is explored and reports on outcome are discussed.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1394 - 1399
1 Oct 2009
Oh C Song H Kim J Choi J Min W Park B

Ten patients, who were unsuitable for limb lengthening over an intramedullary nail, underwent lengthening with a submuscular locking plate. Their mean age at operation was 18.5 years (11 to 40). After fixing a locking plate submuscularly on the proximal segment, an external fixator was applied to lengthen the bone after corticotomy. Lengthening was at 1 mm/day and on reaching the target length, three or four screws were placed in the plate in the distal segment and the external fixator was removed. All patients achieved the pre-operative target length at a mean of 4.0 cm (3.2 to 5.5). The mean duration of external fixation was 61.6 days (45 to 113) and the mean external fixation index was 15.1 days/cm (13.2 to 20.5), which was less than one-third of the mean healing index (48 days/cm (41.3 to 55). There were only minor complications.

Lengthening with a submuscular locking plate can successfully permit early removal of the fixator with fewer complications and is a useful alternative in children or when nailing is difficult.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1164 - 1171
1 Sep 2009
Bae DK Song SJ Yoon KH

We assessed the reliability, accuracy and variability of closed-wedge high tibial osteotomy (HTO) using computer-assisted surgery compared to the conventional technique. A total of 50 closed-wedge HTO procedures were performed using the navigation system, and compared with 50 HTOs that had been performed with the conventional technique. In the navigation group, the mean mechanical axis prior to osteotomy was varus 8.2°, and the mean mechanical axis following fixation was valgus 3.6°. On the radiographs the mean pre-operative mechanical axis was varus 7.3°, and the mean post-operative mechanical axis was valgus 2.1°. There was a positive correlation between the measured data taken under navigation and by radiographs (r > 0.3, p < 0.05). The mean correction angle was significantly more accurate in the navigation group (p < 0.002). The variability of the correction was significantly lower in the navigation group (2.3° vs 3.7°, p = 0,012). We conclude that navigation provides reliable real-time intra-operative information, may increase accuracy, and improves the precision of a closed-wedge HTO.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1234 - 1240
1 Sep 2012
Willcox NMJ Clarke JV Smith BRK Deakin AH Deep K

We compared lower limb coronal alignment measurements obtained pre- and post-operatively with long-leg radiographs and computer navigation in patients undergoing primary total knee replacement (TKR). A series of 185 patients had their pre- and post-implant radiological and computer-navigation system measurements of coronal alignment compared using the Bland-Altman method. The study included 81 men and 104 women with a mean age of 68.5 years (32 to 87) and a mean body mass index of 31.7 kg/m2 (19 to 49). Pre-implant Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability coefficient of 9.0°. The Bland–Altman plot showed a tendency for the radiological measurement to indicate a higher level of pre-operative deformity than the corresponding navigation measurement. Post-implant limits of agreement were -5.0° to 5.4° with a repeatability coefficient of 5.2°. The tendency for valgus knees to have greater deformity on the radiograph was still seen, but was weaker for varus knees.

The alignment seen or measured intra-operatively during TKR is not necessarily the same as the deformity seen on a standing long-leg radiograph either pre- or post-operatively. Further investigation into the effect of weight-bearing and surgical exposure of the joint on the mechanical femorotibial angle is required to enable the most appropriate intra-operative alignment to be selected.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (sd 0.76) in the MB group and 0.66 mm (sd 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores.

For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (sd 0.014). This was significantly smaller than the wear rate of 0.11 mm/year (sd 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 265 - 271
1 Mar 2008
Mandalia V Eyres K Schranz P Toms AD

Evaluation of patients with painful total knee replacement requires a thorough clinical examination and relevant investigations in order to reach a diagnosis. Awareness of the common and uncommon problems leading to painful total knee replacement is useful in the diagnostic approach. This review article aims to act as a guide to the evaluation of patients with painful total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 908 - 913
1 Jul 2012
Seo JG Moon YW Park SH Lee JH Kang HM Kim SM

Peri-prosthetic patellar fracture following resurfacing as part of total knee replacement (TKR) is an infrequent yet challenging complication. This case-control study was performed to identify clinical, radiological and surgical factors that increase the risk of developing a spontaneous patellar fracture after TKR. Patellar fractures were identified in 74 patients (88 knees) from a series of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding those with a previous history of extensor mechanism realignment or a clear traumatic event, a metal-backed patella, any uncemented component or subsequent infection, the remaining 64 fractures were compared with a matched group of TKRs with an excellent outcome defined by the Knee Society score. The mean age of patients with a fracture was 70 years (51 to 81) at the time of TKR. Patellar fractures were detected at a mean of 13.4 months (2 to 84) after surgery. The incidence of patellar fracture was found to be strongly associated with the number of previous knee operations, greater pre-operative mechanical malalignment, smaller post-operative patellar tendon length, thinner post-resection patellar thickness, and a lower post-operative Insall-Salvati ratio.

An understanding of the risk factors associated with spontaneous patellar fracture following TKR provides a valuable insight into prevention of this challenging complication.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1296 - 1299
1 Oct 2011
Lang JE Mannava S Floyd AJ Goddard MS Smith BP Mofidi A M. Seyler T Jinnah RH

Robots have been used in surgery since the late 1980s. Orthopaedic surgery began to incorporate robotic technology in 1992, with the introduction of ROBODOC, for the planning and performance of total hip replacement. The use of robotic systems has subsequently increased, with promising short-term radiological outcomes when compared with traditional orthopaedic procedures. Robotic systems can be classified into two categories: autonomous and haptic (or surgeon-guided). Passive surgery systems, which represent a third type of technology, have also been adopted recently by orthopaedic surgeons.

While autonomous systems have fallen out of favour, tactile systems with technological improvements have become widely used. Specifically, the use of tactile and passive robotic systems in unicompartmental knee replacement (UKR) has addressed some of the historical mechanisms of failure of non-robotic UKR. These systems assist with increasing the accuracy of the alignment of the components and produce more consistent ligament balance. Short-term improvements in clinical and radiological outcomes have increased the popularity of robot-assisted UKR.

Robot-assisted orthopaedic surgery has the potential for improving surgical outcomes. We discuss the different types of robotic systems available for use in orthopaedics and consider the indication, contraindications and limitations of these technologies.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 648 - 655
1 May 2012
Jolles BM Grzesiak A Eudier A Dejnabadi H Voracek C Pichonnaz C Aminian K Martin E

This study compared the outcome of total knee replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses during the first post-operative year and at five years’ follow-up, using gait parameters as a new objective measure. This double-blind randomised controlled clinical trial included 55 patients with mobile-bearing (n = 26) and fixed-bearing (n = 29) prostheses of the same design, evaluated pre-operatively and post-operatively at six weeks, three months, six months, one year and five years. Each participant undertook two walking trials of 30 m and completed the EuroQol questionnaire, Western Ontario and McMaster Universities osteoarthritis index, Knee Society score, and visual analogue scales for pain and stiffness. Gait analysis was performed using five miniature angular rate sensors mounted on the trunk (sacrum), each thigh and calf. The study population was divided into two groups according to age (≤ 70 years versus > 70 years).

Improvements in most gait parameters at five years’ follow-up were greater for fixed-bearing TKRs in older patients (> 70 years), and greater for mobile-bearing TKRs in younger patients (≤ 70 years). These findings should be confirmed by an extended age controlled study, as the ideal choice of prosthesis might depend on the age of the patient at the time of surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design.

At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%).

In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 157 - 166
1 Nov 2012
Rosenberg AG Berend ME Berry DJ Della Valle CJ MacDonald SJ Minas T

This conversation represents an attempt by several arthroplasty surgeons to critique several abstracts presented over the last year as well as to use them as a jumping off point for trying to figure out where they fit in into our current understanding of multiple issues in modern hip and knee arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 120 - 125
1 Jan 2011
Lim H Bae J Song H Teoh SH Kim H Kum D

Medial open-wedge high tibial osteotomy has been gaining popularity in recent years, but adequate supporting material is required in the osteotomy gap for early weight-bearing and rapid union. The purpose of this study was to investigate whether the implantation of a polycaprolactone-tricalcium phosphate composite scaffold wedge would enhance healing of the osteotomy in a micro pig model. We carried out open-wedge high tibial osteotomies in 12 micro pigs aged from 12 to 16 months. A scaffold wedge was inserted into six of the osteotomies while the other six were left open. Bone healing was evaluated after three and six months using plain radiographs, CT scans, measurement of the bone mineral density and histological examination.

Complete bone union was obtained at six months in both groups. There was no collapse at the osteotomy site, loss of correction or failure of fixation in either group. Staining with haematoxylin and eosin demonstrated that there was infiltration of new bone tissue into the macropores and along the periphery of the implanted scaffold in the scaffold group. The CT scans and measurement of the bone mineral density showed that at six months specimens in the scaffold group had a higher bone mineral density than in the control group, although the implantation of the polycaprolactone-tricalcium phosphate composite scaffold wedge did not enhance healing of the osteotomy.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 345 - 350
1 Mar 2011
Huang T Hsu W Peng K Hsu RW

We conducted a retrospective study to investigate the effect of femoral bowing on the placement of components in total knee replacement (TKR), with regard to its effect on reestablishing the correct mechanical axis, as we hypothesised that computer-assisted total knee replacement (CAS-TKR) would produce more accurate alignment than conventional TKR. Between January 2006 and December 2009, 212 patients (306 knees) underwent TKR. The conventional TKR was compared with CAS-TKR for accuracy of placement of the components and post-operative alignment, as determined by five radiological measurements. There were significant differences in the reconstructed mechanical axes between the bowed and the non-bowed group after conventional TKR (176.2° (sd 3.4) vs 179.3° (sd 2.1), p < 0.001).

For patients with significant femoral bowing, the reconstructed mechanical axes were significantly closer to normal in the CAS group than in the conventional group (179.2° (sd 1.9) vs 176.2° (sd 3.4), p < 0.001). Femoral bowing resulted in inaccuracy when a conventional technique was used. CAS-TKR provides an effective method of restoring the mechanical axis in the presence of significant femoral bowing.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1631 - 1633
1 Dec 2008
Atinga M Hamer AJ

The Morquio syndrome is a rare disorder which presents with a number of musculoskeletal problems. The literature describing total knee replacement in these patients is sparse. We describe the management of a patient with bilateral instability and pain in the knees using bilateral constrained knee replacements, and followed up for five years with pre- and postoperative knee scores. We highlight the difficulties encountered and discuss the end results.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1180 - 1185
1 Sep 2008
Hamai S Miura H Higaki H Shimoto T Matsuda S Iwamoto Y

Mechanical failure because of wear or fracture of the polyethylene tibial post in posteriorly-stabilised total knee replacements has been extensively described. In this study of 12 patients with a clinically and radiologically successful NexGen LPS posteriorly-stabilised prosthesis impingement of the anterior tibial post was evaluated in vivo in three dimensions during gait using radiologically-based image-matching techniques.

Impingement was observed in all images of the patients during the stance phase, although the NexGen LPS was designed to accommodate 14° of hyperextension of the component before impingement occurred. Impingement arises as a result of posterior translation of the femur during the stance phase. Further attention must therefore be given to the configuration of the anterior portion of the femoral component and the polyethylene post when designing posteriorly-stabilised total knee replacements.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1446 - 1450
1 Nov 2008
Dahm DL Wulf CA Dajani KA Dobbs RE Levy BA Stuart MA

The records of patients aged 50 years or over who underwent primary reconstruction of the anterior cruciate ligament between 1990 and 2002 were reviewed. There were 35 knees in 34 patients that met the inclusion criteria. The mean age of the patients was 57 years (50 to 66) and the mean clinical follow-up was for 72 months (25 to 173). A total of 23 knees were reconstructed with patellar tendon allograft, and 12 with patellar tendon autograft. The mean pre-operative knee extension was 1° (−5° to 10°) and flexion was 129° (125° to 150°) and at follow-up these values were 0° (−5° to 5°) and 135° (120° to 150°), respectively. Pre-operatively there were 31 knees (89%) with a Lachman grade 2+ or 3+. Post-operatively, 33 knees (94%) were Lachman grade 0 or 1+. The mean pre- and post-operative International Knee Documentation Committee scores were 39 (23 to 72) and 90 (33 to 100) respectively. The mean pre- and post-operative Lysholm scores were 50 (18 to 68) and 92 (28 to 100) respectively and the mean University of California Los Angeles activity scores were 8.5 before injury (4 to 10), 4.3 (3 to 6) after injury and 8.3 (4 to 10) post-operatively. There were three graft failures (8.6%) requiring revision.

We conclude that reconstruction of the anterior cruciate ligament in carefully-selected patients aged 50 years or over can achieve similar results to those in younger patients, with no increased risk of complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 64 - 68
1 Jan 2009
Kim Y Choi Y Kim J

We wished to determine whether simultaneous bilateral sequential total knee replacement (TKR) carried increased rates of mortality and complications compared with unilateral TKR in low- and high-risk patients.

Our study included 2385 patients who had undergone bilateral sequential TKR under one anaesthetic and 719 who had unilateral TKR. There were no significant pre-operative differences between the groups in terms of age, gender, height, weight, body mass index, diagnosis, comorbidity and duration of follow-up, which was a mean of 10.2 years (5 to 14) in the bilateral and 10.4 years (5 to 14) in the unilateral group.

The peri-operative mortality rate (eight patients, 0.3%) of patients who had bilateral sequential TKR was similar to that (five patients, 0.7%) of those undergoing unilateral TKR. In bilateral cases the peri-operative mortality rate (three patients, 0.4%) of patients at high risk was similar to that (five patients, 0.3%) of patients at low risk as it was also in unilateral cases (two patients, 1.0% vs three patients, 0.6%). There was no significant difference (p = 0.735) in either the overall number of major complications between bilateral and unilateral cases or between low- (p = 0.57) and high-risk (p = 0.61) patients. Also, the overall number of minor complications was not significantly different between the bilateral and unilateral group (p = 0.143).

Simultaneous bilateral sequential TKR can be offered to patients at low and high risk and has an expected rate of complications similar to that of unilateral TKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1304 - 1310
1 Oct 2008
Kim T Lee D Bin S

We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired t-test, p < 0.05). At the last follow-up, the mean maximum flexion was 135° (110° to 150°). Two knees showed radiolucent lines, but revision was not required because the patients were asymptomatic. Revision was required in one case because of infection, but there were no prosthesis-related revisions. There were no other complications. The estimated survival rate at four years with revision for any reason and prosthesis-related problems was 99.6% and 100%, respectively.

This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1513 - 1518
1 Nov 2006
Henckel J Richards R Lozhkin K Harris S Baena FMRY Barrett ARW Cobb JP

Surgeons need to be able to measure angles and distances in three dimensions in the planning and assessment of knee replacement. Computed tomography (CT) offers the accuracy needed but involves greater radiation exposure to patients than traditional long-leg standing radiographs, which give very little information outside the plane of the image.

There is considerable variation in CT radiation doses between research centres, scanning protocols and individual scanners, and ethics committees are rightly demanding more consistency in this area.

By refining the CT scanning protocol we have reduced the effective radiation dose received by the patient down to the equivalent of one long-leg standing radiograph. Because of this, it will be more acceptable to obtain the three-dimensional data set produced by CT scanning. Surgeons will be able to document the impact of implant position on outcome with greater precision.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1077 - 1083
1 Aug 2007
Tsuchiya H Morsy AF Matsubara H Watanabe K Abdel-Wanis ME Tomita K

We present a retrospective study of patients suffering from a variety of benign tumours in whom external fixators were used to treat deformity and limb-length discrepancy, and for the reconstruction of bone defects. A total of 43 limbs in 31 patients (12 male and 19 female) with a mean age of 14 years (2 to 54) were treated.

The diagnosis was Ollier’s disease in 12 limbs, fibrous dysplasia in 11, osteochondroma in eight, giant cell tumour in five, osteofibrous dysplasia in five and non-ossifying fibroma in two. The lesions were treated in the tibia in 19 limbs, in the femur in 16, and in the forearm in eight. The Ilizarov frame was used in 25 limbs, the Taylor Spatial Frame in seven, the Orthofix fixator in six, the Monotube in four and the Heidelberg fixator in one. The mean follow-up was 72 months (22 to 221).

The mean external fixation period was 168 days (71 to 352). The mean external fixation index was 42 days/cm (22.2 to 102.0) in the 22 patients who required limb lengthening. The mean correction angle for those with angular deformity was 23° (7° to 45°).

At final follow-up all patients had returned to normal activities. Four patients required a second operation for recurrent deformity of further limb lengthening. Local recurrence occurred in one patient, requiring further surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 467 - 470
1 Apr 2007
Kim Y Kim J Kim D

We performed a prospective, randomised study to compare the results and rates of complications of primary total knee replacement performed using a quadriceps-sparing technique or a standard arthrotomy in 120 patients who had bilateral total knee replacements carried out under the same anaesthetic. The clinical results, pain scales, surgical and hospital data, post-operative complications and radiological results were compared.

No significant differences were found between the two groups with respect to the blood loss, knee score, function score, pain scale, range of movement or radiological findings. In contrast, the operating time (p = 0.0001) and the tourniquet time (p < 0.0001) were significantly longer in the quadriceps-sparing group, as was the rate of complications (p = 0.0468).

We therefore recommend the use of a standard arthrotomy with the shortest possible skin incision for total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 800 - 803
1 Jun 2005
Ramesh R Von Arx O Azzopardi T Schranz PJ

We assessed hyperextension of the knee and joint laxity in 169 consecutive patients who underwent an anterior cruciate ligament reconstruction between 2000 and 2002 and correlated this with a selected number of age- and gender-matched controls. In addition, the mechanism of injury in the majority of patients was documented. Joint laxity was present in 42.6% (72 of 169) of the patients and hyperextension of the knee in 78.7% (133 of 169). All patients with joint laxity had hyperextension of their knee. In the control group only 21.5% (14 of 65) had joint laxity and 37% (24 of 65) had hyperextension of the knee. Statistical analysis showed a significant correlation for these associations. We conclude that anterior cruciate ligament injury is more common in those with joint laxity and particularly so for those with hyperextension of the knee.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 904 - 910
1 Jul 2007
Kim Y Kim D Kim J

We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1).

The total knee score, pain score, functional score and range of movement were not statistically different (p > 0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 577 - 582
1 Apr 2005
Senavongse W Amis AA

Normal function of the patellofemoral joint is maintained by a complex interaction between soft tissues and articular surfaces. No quantitative data have been found on the relative contributions of these structures to patellar stability. Eight knees were studied using a materials testing machine to displace the patella 10 mm laterally and medially and measure the force required. Patellar stability was tested from 0° to 90° knee flexion with the quadriceps tensed to 175 N. Four conditions were examined: intact, vastus medialis obliquus relaxed, flat lateral condyle, and ruptured medial retinaculae. Abnormal trochlear geometry reduced the lateral stability by 70% at 30° flexion, while relaxation of vastus medialis obliquus caused a 30% reduction. Ruptured medial retinaculae had the largest effect at 0° flexion with 49% reduction. There was no effect on medial stability. There is a complex interaction between these structures, with their contributions to loss of lateral patellar stability varying with knee flexion.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 710 - 715
1 May 2005
van Huyssteen AL Hastings CJ Olesak M Hoffman EB

We reviewed 34 knees in 24 children after a double-elevating osteotomy for late-presenting infantile Blount’s disease. The mean age of patients was 9.1 years (7 to 13.5).

All knees were in Langenskiöld stages IV to VI. The operative technique corrected the depression of the medial joint line by an elevating osteotomy, and the remaining tibial varus and internal torsion by an osteotomy just below the apophysis. In the more recent patients (19 knees), a proximal lateral tibial epiphysiodesis was performed at the same time.

The mean pre-operative angle of depression of the medial tibial plateau of 49° (40° to 60°) was corrected to a mean of 26° (20° to 30°), which was maintained at follow-up. The femoral deformity was too small to warrant femoral osteotomy in any of our patients. The mean pre-operative mechanical varus of 30.6° (14° to 66°) was corrected to 0° to 5° of mechanical valgus in 29 knees. In five knees, there was an undercorrection of 2° to 5° of mechanical varus. At follow-up a further eight knees, in which lateral epiphysiodesis was delayed beyond five months, developed recurrent tibial varus associated with fusion of the medial proximal tibial physis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 646 - 655
1 May 2005
Victor J Banks S Bellemans J

We performed a prospective, randomised trial of 44 patients to compare the functional outcomes of a posterior-cruciate-ligament-retaining and posterior-cruciate-ligament-substituting total knee arthroplasty, and to gain a better understanding of the in vivo kinematic behaviour of both devices.

At follow-up at five years, no statistically significant differences were found in the clinical outcome measurements for either design. The prevalence of radiolucent lines and the survivorship were the same. In a subgroup of 15 knees, additional image-intensifier analysis in the horizontal and sagittal planes was performed during step-up and lunge activity. Our analysis revealed striking differences. Lunge activity showed a mean posterior displacement of both medial and lateral tibiofemoral contact areas (roll-back) which was greater and more consistent in the cruciate-substituting than in the cruciate-retaining group (medial p < 0.0001, lateral p = 0.011). The amount of posterior displacement could predict the maximum flexion which could be achieved (p = 0.018). Forward displacement of the tibiofemoral contact area in flexion during stair activity was seen more in the cruciate-retaining than in the cruciate-substituting group. This was attributed mainly to insufficiency of the posterior cruciate ligament and partially to that of the anterior cruciate ligament. We concluded that, despite similar clinical outcomes, there are significant kinematic differences between cruciate-retaining and cruciate-substituting arthroplasties.