Advertisement for orthosearch.org.uk
Results 1 - 20 of 221
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 5 | Pages 573 - 579
1 May 2020
Krueger DR Guenther K Deml MC Perka C

Aims. We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed. . Methods. All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation. Results. There were 113 recorded cases with fracture at the modular junction, resulting in a calculated fracture rate of 0.30% (113/37,600). The fracture rate of the implant without signs of improper use was 0.11% (41/37,600). In 79% (89/113) of cases with a failed implant, either a lateralized (high offset) neck segment, an extralong head, or the combination of both were used. Logistic regression analysis revealed male sex, high body mass index (BMI), straight component design, and small neck segments were significant risk factors for failure. Investigation of the implants (76/113) showed at least one sign of improper use in 72 cases. Conclusion. Implant failure at the modular junction is associated with patient- and implant-specific risk factors as well as technical errors during implantation. Whenever possible, the use of short and lateralized neck segments should be avoided with this revision system. Implantation instructions and contraindications need to be adhered to and respected. Cite this article: Bone Joint J 2020;102-B(5):573–579


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 1 | Pages 26 - 31
1 Jan 1990
Davis T Sher J Horsman A Simpson M Porter B Checketts R

In a prospective study we assessed the causes of mechanical failure in a series of 230 intertrochanteric femoral fractures which had been internally fixed with either a sliding hip screw or a Kuntscher Y-nail. The overall rate of mechanical failure was 16.5%; cutting-out of the implant from the femoral head was the cause in three-quarters of the instances. Implants placed posteriorly in the femoral head cut out more often (27%) than those placed centrally (7%). The cut-out rate was also determined by the quality of the fracture reduction, but age, walking ability and bone density (assessed by the Singh grade and metacarpal indices) had no significant influence. We conclude that these fractures should be reduced as accurately as possible and it is imperative that the implant is placed centrally within the femoral head


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 639 - 643
1 May 2011
Burghardt RD Herzenberg JE Specht SC Paley D

Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 801 - 810
1 Jul 2022
Krull P Steinbrück A Grimberg AW Melsheimer O Morlock M Perka C

Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model. Results. Median follow-up was 868 days (interquartile range 418 to 1,364). The offset liner design reduced the risk of revision (hazard ratio (HR) 0.68 (95% confidence interval (CI) 0.50 to 0.92)), while the angulated/offset liner increased the risk of revision for mechanical failure (HR 1.81 (95% CI 1.38 to 2.36)). The cumulative incidence of revision was lowest for the offset liner at one and seven years (1.0% (95% CI 0.7 to 1.3) and 1.8% (95% CI 1.0 to 3.0)). No difference was found between standard, lipped, and angulated liner designs. Higher age at index primary THA and an Elixhauser Comorbidity Index greater than 0 increased the revision risk in the first year after surgery. Implantation of a higher proportion of a single design of liner in a hospital reduced revision risk slightly but significantly (p = 0.001). Conclusion. The use of standard acetabular component liners remains a good choice in primary uncemented THA, as most modified liner designs were not associated with a reduced risk of revision for mechanical failure. Offset liner designs were found to be beneficial and angulated/offset liner designs were associated with higher risks of revision. Cite this article: Bone Joint J 2022;104-B(7):801–810


The Journal of Bone & Joint Surgery British Volume
Vol. 65-B, Issue 3 | Pages 274 - 275
1 May 1983
Zimmerman K Klasen H

The case reports of four patients with breakage of the intramedullary nail of the femur are presented. In all four patients the nail broke after consolidation of the fracture of the femoral shaft. All the nails broke at the same place: the junction between the round cross-section and the cloverleaf cross-section.


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 277 - 283
1 Mar 2023
Gausden EB Puri S Chiu Y Figgie MP Sculco TP Westrich G Sculco PK Chalmers BP

Aims. The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. Methods. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time. Results. The estimated survivorship free of all-cause revision was 98% (95% confidence interval (CI) 97% to 98%), 95% (95% CI 95% to 96%), and 86% (95% CI 83% to 88%) at two, five, and ten years, respectively. In 209/369 revisions there was a consistent constellation of findings with varying severity that included polyethylene wear and associated synovitis, osteolysis, and component loosening. This failure mode, which we refer to as aseptic mechanical failure, was the most common revision indication. The mean time from primary TKA to revision for aseptic mechanical failure was five years (5 months to 11 years). Conclusion. In this series of nearly 8,000 primary TKAs performed with a specific implant, we identified a lower-than-expected mid-term survivorship and a high number of revisions with a unique presentation. This study, along with the recent recall of the implant, confirms the need for frequent monitoring of patients with Optetrak TKAs given the incidence of polyethylene failure, osteolysis, and component loosening. Cite this article: Bone Joint J 2023;105-B(3):277–283


The Bone & Joint Journal
Vol. 106-B, Issue 10 | Pages 1067 - 1073
1 Oct 2024
Lodge CJ Adlan A Nandra RS Kaur J Jeys L Stevenson JD

Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected knee prosthesis between September 2007 and January 2020 at a single institution. Radiographs and clinical records were used to assess and classify the incidence and causes of static spacer failure. Of the 94 cases, there were 19 primary total knee arthroplasties (TKAs), ten revision TKAs (varus-valgus constraint), 20 hinged TKAs, one arthrodesis (nail), one failed spacer (performed elsewhere), 21 distal femoral endoprosthetic arthroplasties, and 22 proximal tibial arthroplasties. Results. A total of 35/94 patients (37.2%) had spacer-related complications, of which 26/35 complications (74.3%) were because of mechanical failure of the spacer construct, while 9/35 (25.7%) were due to recurrence of infection. Risk factors for internal failure were a construct where the total intramedullary spacer length was less than twice the length of the central osseous defect (p = 0.009), where proximal or distal intraosseous spacer contact was < 10%, and after tibial tubercle osteotomy (p = 0.005). The incidence of spacer complications significantly increased the time to second stage: mean 157 days (42 to 458) in those without complications versus 227 days (11 to 528) with complications (p = 0.014). Conclusion. The failure rate of static antibiotic-loaded cement spacers is much higher than anticipated. Complications of the spacer significantly increased the time to second-stage revision. The risk of mechanical failure is significantly increased if the spacer is less than double the size of the segmental defect, or if inadequate reinforcement is inserted into the residual bone. These findings serve as a guide for surgeons to avoid mechanical complications with static spacers. Cite this article: Bone Joint J 2024;106-B(10):1067–1073


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 170 - 176
1 Feb 2020
Bernthal NM Burke ZDC Hegde V Upfill-Brown A Chen CJ Hwang R Eckardt JJ

Aims. We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. Methods. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up. Results. A total of 278 patients were identified with 289 all-polyethylene tibial components. Mechanical survival was 98.4%, 91.1%, and 85.2% at five, ten and 15 years, respectively. A total of 15 mechanical failures were identified at the final follow-up. Of the 13 all-polyethylene tibial components used for revision of a previous tibial component, five (38.5%) failed mechanically. Younger patients (< 18 years vs > 18 years; p = 0.005) and those used as revision components (p < 0.001) had significantly increased rates of failure. Multivariate logistic regression modelling showed revision status to be a positive risk factor for failure (odds ratio (OR) 19.498, 95% confidence interval (CI) 4.598 to 82.676) and increasing age was a negative risk factor for failure (OR 0.927, 95% CI 0.872 to 0.987). Age-stratified risk analysis showed that age > 24 years was no longer a statistically significant risk factor for failure. The final mean MSTS score for all patients was 89% (8.5% to 100.0%). Conclusion. The long-term mechanical survivorship of all-polyethylene tibial components when used for tumour endoprostheses was excellent. Tumour surgeons should consider using these components for their durability and the secondary benefits of reduced cost and ease of removal and revision. However, caution should be taken when using all-polyethylene tibial components in the revision setting as a significantly higher rate of mechanical failure was seen in this group of patients. Cite this article: Bone Joint J. 2020;102-B(2):170–176


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 724 - 731
1 Jun 2019
Bernthal NM Upfill-Brown A Burke ZDC Ishmael CR Hsiue P Hori K Hornicek F Eckardt JJ

Aims. Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. Patients and Methods. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1). Results. The median follow-up was 132 months (interquartile range (IQR) 44 to 189). In all, 20 stems required revision: eight for infection, five for structural failure, five for aseptic loosening, and two for tumour progression. Mechanical survivorship at five, ten, and 15 years was 84%, 78%, and 78%, respectively. Mechanical failure rate varied by location, with no mechanical failures of proximal femoral constructs and distal femoral survivorship of 82%, 77%, and 77% at five, ten, and 15 years. The survivorship of primary constructs at five years was 74%, with no failure after 40 months, while the survivorship for revision constructs was 89%, 80%, and 80% at five, ten, and 15 years. . Conclusion. The rate of mechanical survivorship in our series is similar to those reported for other methods of reconstruction for short diaphyseal segments, such as compressive osseointegration. The mechanical failure rate differed by location, while there was no substantial difference in long-term survival between primary and revision reconstructions. Overall, custom cross-pin fixation is a viable option for endoprosthetic reconstruction of short metaphyseal segments with an acceptable rate of mechanical failure. Cite this article: Bone Joint J 2019;101-B:724–731


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 635 - 643
1 Apr 2021
Ross LA Keenan OJF Magill M Brennan CM Clement ND Moran M Patton JT Scott CEH

Aims. Debate continues regarding the optimum management of periprosthetic distal femoral fractures (PDFFs). This study aims to determine which operative treatment is associated with the lowest perioperative morbidity and mortality when treating low (Su type II and III) PDFFs comparing lateral locking plate fixation (LLP-ORIF) or distal femoral arthroplasty (DFA). Methods. This was a retrospective cohort study of 60 consecutive unilateral (PDFFs) of Su types II (40/60) and III (20/60) in patients aged ≥ 60 years: 33 underwent LLP-ORIF (mean age 81.3 years (SD 10.5), BMI 26.7 (SD 5.5); 29/33 female); and 27 underwent DFA (mean age 78.8 years (SD 8.3); BMI 26.7 (SD 6.6); 19/27 female). The primary outcome measure was reoperation. Secondary outcomes included perioperative complications, calculated blood loss, transfusion requirements, functional mobility status, length of acute hospital stay, discharge destination and mortality. Kaplan-Meier survival analysis was performed. Cox multivariate regression analysis was performed to identify risk factors for reoperation after LLP-ORIF. Results. Follow-up was at mean 3.8 years (1.0 to 10.4). One-year mortality was 13% (8/60). Reoperation was more common following LLP-ORIF: 7/33 versus 0/27 (p = 0.008). Five-year survival for reoperation was significantly better following DFA; 100% compared to 70.8% (95% confidence interval (CI) 51.8% to 89.8%, p = 0.006). There was no difference for the endpoint mechanical failure (including radiological loosening); ORIF 74.5% (56.3 to 92.7), and DFA 78.2% (52.3 to 100, p = 0.182). Reoperation following LLP-ORIF was independently associated with medial comminution; hazard ratio (HR) 10.7 (1.45 to 79.5, p = 0.020). Anatomical reduction was protective against reoperation; HR 0.11 (0.013 to 0.96, p = 0.046). When inadequately fixed fractures were excluded, there was no difference in five-year survival for either reoperation (p = 0.156) or mechanical failure (p = 0.453). Conclusion. Absolute reoperation rates are higher following LLP fixation of low PDFFs compared to DFA. Where LLP-ORIF was well performed with augmentation of medial comminution, there was no difference in survival compared to DFA. Though necessary in very low fractures, DFA should be used with caution in patients with greater life expectancies due to the risk of longer term aseptic loosening. Cite this article: Bone Joint J 2021;103-B(4):635–643


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1247 - 1253
1 Jul 2021
Slullitel PA Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims. There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA. Methods. We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality. Results. The cumulative incidence of septic failure was 8% (95% confidence interval (CI) 3.5 to 15) at one year, 13.8% (95% CI 7.6 to 22) at two years, and 19.7% (95% CI 12 to 28.6) at five and ten years of follow-up. A femoral bone defect worse than Paprosky IIIA (hazard ratio (HR) 13.58 (95% CI 4.86 to 37.93); p < 0.001) and obesity (BMI ≥ 30 kg/m. 2. ; HR 3.88 (95% CI 1.49 to 10.09); p = 0.005) were significantly associated with septic failure. Instability and periprosthetic fracture were the most common reasons for mechanical failure (5.7% and 4.5%, respectively). The cumulative incidence of aseptic failure was 2% (95% CI 0.4 to 7) at two years, 9% (95% CI 4 to 17) at five years, and 12% (95% CI 5 to 22) at ten years. A previous revision to treat PJI was significantly associated with non-septic failure (HR 9.93 (95% CI 1.77 to 55.46); p = 0.009). At the five-year timepoint, 93% of the patients were alive (95% CI 84% to 96%), which fell to 86% (95% CI 75% to 92%) at ten-year follow-up. Conclusion. Massive femoral bone loss was associated with greater chances of developing a further septic failure. All septic failures occurred within the first five years following the one-stage exchange. Surgeons should be aware of instability and periprosthetic fracture being potential causes of further aseptic revision surgery. Cite this article: Bone Joint J 2021;103-B(7):1247–1253


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 78 - 83
1 Jul 2021
Roedel GG Kildow BJ Sveom DS Garvin KL

Aims. Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant. Methods. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with cementless fixation with a median head size of 28 mm. HXLPE was the acetabular bearing for all hips. Radiographs were evaluated for polyethylene wear, radiolucent lines, and osteolysis. Results. Clinical outcomes showed significant improvement of mean Harris Hip Scores from 52.8 (SD 13.5) preoperatively to 94.8 (SD 7.6) postoperatively. One hip was revised for recurrent instability, and there were no infections. No hips were revised for mechanical loosening or osteolysis. Mean polyethylene linear wear was 0.04 mm/year and volumetric wear was 6.22 mm. 3. /year, with no significant differences between head size or material. Osteolysis was not present in any of the hips. Conclusion. The use of HXLPE in THA for patients aged 50 years and younger has performed exceptionally well without evidence of significant wear causing mechanical loosening or necessitating revision. The radiolucent lines of the acetabular component must be followed to determine the prognostic significance. This investigation represents the longest clinical follow-up of a large, consecutive cohort of patients aged 50 years or younger with THA using HXLPE. This long-term analysis found negligible polyethylene wear, no incidence of aseptic loosening, and excellent clinical outcomes at and beyond 15 years of follow-up. Cite this article: Bone Joint J 2021;103-B(7 Supple B):78–83


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1333 - 1338
2 Aug 2021
Kankanalu P Borton ZM Morgan ML Cresswell T Espag MP Tambe AA Clark DI

Aims. Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. Methods. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up. Results. The implant was revised in ten shoulders (5.2%) with a median time to revision of 21.2 months (interquartile range (IQR) 9.9 to 41.8). The Kaplan-Meier survivorship estimate at seven years was 95.9% (95% confidence interval (CI) 91.7 to 98; 35 RTSAs at risk) for aseptic mechanical failure of the glenoid and 94.8% (95% CI 77.5 to 96.3; 35 RTSAs at risk) for all-cause revision. Minimum five-year clinical and radiological outcomes were available for 103 and 98 RTSAs respectively with a median follow-up time of six years (IQR 5.2 to 7.0). Median postoperative Oxford Shoulder Score was 38 (IQR 31 to 45); median Constant and Murley score was 60 (IQR 47.5 to 70); median forward flexion 115° (IQR 100° to 125°); median abduction 95° (IQR 80° to 120°); and external rotation 25° (IQR 15° to 40°) Scapular notching was seen in 62 RTSAs (63.2%). Conclusion. We present the largest and longest-term series of TM-backed glenoid implants demonstrating 94.8% all-cause survivorship at seven years. Specifically pertaining to glenoid loosening, survival of the implant increased to 95.9%. In addition, we report satisfactory minimum five-year clinical and radiological outcomes. Cite this article: Bone Joint J 2021;103-B(8):1333–1338


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 55 - 63
1 Jan 2020
Hagberg K Ghassemi Jahani S Kulbacka-Ortiz K Thomsen P Malchau H Reinholdt C

Aims. The aim of this study was to describe implant and patient-reported outcome in patients with a unilateral transfemoral amputation (TFA) treated with a bone-anchored, transcutaneous prosthesis. Methods. In this cohort study, all patients with a unilateral TFA treated with the Osseointegrated Prostheses for the Rehabilitation of Amputees (OPRA) implant system in Sahlgrenska University Hospital, Gothenburg, Sweden, between January 1999 and December 2017 were included. The cohort comprised 111 patients (78 male (70%)), with a mean age 45 years (17 to 70). The main reason for amputation was trauma in 75 (68%) and tumours in 23 (21%). Patients answered the Questionnaire for Persons with Transfemoral Amputation (Q-TFA) before treatment and at two, five, seven, ten, and 15 years’ follow-up. A prosthetic activity grade was assigned to each patient at each timepoint. All mechanical complications, defined as fracture, bending, or wear to any part of the implant system resulting in removal or change, were recorded. Results. The Q-TFA scores at two, five, seven, and ten years showed significantly more prosthetic use, better mobility, fewer problems, and an improved global situation, compared with baseline. The survival rate of the osseointegrated implant part (the fixture) was 89% and 72% after seven and 15 years, respectively. A total of 61 patients (55%) had mechanical complications (mean 3.3 (SD 5.76)), resulting in exchange of the percutaneous implant parts. There was a positive relationship between a higher activity grade and the number of mechanical complications. Conclusion. Compared with before treatment, the patient-reported outcome was significantly better and remained so over time. Although osseointegration and the ability to transfer loads over a 15-year period have been demonstrated, a large number of mechanical failures in the external implant parts were found. Since these were related to higher activity, restrictions in activity and improvements to the mechanical properties of the implant system are required. Cite this article: Bone Joint J 2020;102-B(1):55–63


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 610 - 614
1 May 2019
Aibinder WR Bartels DW Sperling JW Sanchez-Sotelo J

Aims. Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and mechanical failure rates with implantation of a short cementless humeral component in anatomical (TSA) and reverse shoulder arthroplasty (RSA). Patients and Methods. A total of 100 shoulder arthroplasties (35 TSA and 65 RSA) were evaluated at a mean of 3.8 years (3 to 8.3). The mean age at the time of surgery was 68 years (31 to 90). The mean body mass index was 32.7 kg/m. 2. (17.3 to 66.4). Results. Greater tuberosity stress shielding was noted in 14 shoulders (two TSA and 12 RSA) and was graded as mild in nine, moderate in two, and severe in three. Medial calcar resorption was noted in 23 shoulders (seven TSA and 16 RSA), and was graded as mild in 21 and moderate in two. No humeral components were revised for loosening or considered to be loose radiologically. Nine shoulders underwent reoperation for infection (n = 3), fracture of the humeral tray (n = 2), aseptic glenoid loosening (n = 1), and instability (n = 3). No periprosthetic fractures occurred. Conclusion. Implantation of this particular short cementless humeral component at the time of TSA or RSA was associated with a low rate of adverse radiological findings on the humeral side at mid-term follow-up. Our data do not raise any concerns regarding the use of a short stem in TSA or RSA. Cite this article: Bone Joint J 2019;101-B:610–614


The Bone & Joint Journal
Vol. 95-B, Issue 6 | Pages 825 - 830
1 Jun 2013
Abram SGF Pollard TCB Andrade AJMD

The Gamma nail is frequently used in unstable peri-trochanteric hip fractures. We hypothesised that mechanical failure of the Gamma nail was associated with inadequate proximal three-point fixation. We identified a consecutive series of 299 Gamma nails implanted in 299 patients over a five-year period, 223 of whom fulfilled our inclusion criteria for investigation. The series included 61 men and 162 women with a mean age of 81 years (20 to 101). Their fractures were classified according to the Modified Evans’ classification and the quality of fracture reduction was graded. The technical adequacy of three points of proximal fixation was recorded from intra-operative fluoroscopic images, and technical inadequacy for each point was defined. All patients were followed to final follow-up and mechanical failures were identified. A multivariate statistical analysis was performed, adjusting for confounders. A total of 16 failures (7.2%) were identified. The position of the lag screw relative to the lateral cortex was the most important point of proximal fixation, and when inadequate the failure rate was 25.8% (eight of 31: odds ratio 7.5 (95% confidence interval 2.5 to 22.7), p < 0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral fractures is rare (< 1%) when three-point proximal fixation is achieved. However, when proximal fixation is inadequate, failure rates increase. The strongest predictor of failure is positioning the lateral end of the lag screw short of the lateral cortex. Adherence to simple technical points minimises the risk of fixation failure in this vulnerable patient group. Cite this article: Bone Joint J 2013;95-B:825–30


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1592 - 1599
1 Dec 2018
Galea VP Connelly JW Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen MB Muratoglu O Bragdon C Malchau H

Aims. The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure. Patients and Methods. A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery. Results. We observed similar bedding in one year postoperatively and wear two years postoperatively between the two types of liner. However, there was significantly more penetration of the femoral head in the ModXLPE cohort compared with the VEPE cohort five years postoperatively (p < 0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (β = 0.22, p = 0.010) and metal (vs ceramic) femoral head (β = 0.21, p = 0.013). There was no association between increased wear and the development of radiolucency (p = 0.866) or PROMs. No patient had evidence of osteolysis. Conclusion. Five years postoperatively, patients with VEPE (vs ModXLPE) and ceramic (vs metal) femoral heads had decreased wear. The rates of wear for both liners were very low and have not led to any osteolysis or implant failure due to aseptic loosening


The Bone & Joint Journal
Vol. 100-B, Issue 3 | Pages 370 - 377
1 Mar 2018
Gilg MM Gaston CL Jeys L Abudu A Tillman RM Stevenson JD Grimer RJ Parry MC

Aims. The use of a noninvasive growing endoprosthesis in the management of primary bone tumours in children is well established. However, the efficacy of such a prosthesis in those requiring a revision procedure has yet to be established. The aim of this series was to present our results using extendable prostheses for the revision of previous endoprostheses. Patients and Methods. All patients who had a noninvasive growing endoprosthesis inserted at the time of a revision procedure were identified from our database. A total of 21 patients (seven female patients, 14 male) with a mean age of 20.4 years (10 to 41) at the time of revision were included. The indications for revision were mechanical failure, trauma or infection with a residual leg-length discrepancy. The mean follow-up was 70 months (17 to 128). The mean shortening prior to revision was 44 mm (10 to 100). Lengthening was performed in all but one patient with a mean lengthening of 51 mm (5 to 140). Results. The mean residual leg length discrepancy at final follow-up of 15 mm (1 to 35). Two patients developed a deep periprosthetic infection, of whom one required amputation to eradicate the infection; the other required two-stage revision. Implant survival according to Henderson criteria was 86% at two years and 72% at five years. When considering revision for any cause (including revision of the growing prosthesis to a non-growing prosthesis), revision-free implant survival was 75% at two years, but reduced to 55% at five years. Conclusion. Our experience indicates that revision surgery using a noninvasive growing endoprosthesis is a successful option for improving leg length discrepancy and should be considered in patients with significant leg-length discrepancy requiring a revision procedure. Cite this article: Bone Joint J 2018;100-B:370–7


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 905 - 911
1 Aug 2023
Giannicola G Amura A Sessa P Prigent S Cinotti G

Aims

The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance.

Methods

A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.