Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE
Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries. Results. The distribution of liner geometries were neutral (39.4%; 79,822), 10° (34.5%; 69,894), 15° (21.6%; 43,722), offset reorientating (2.8%; 5705), offset neutral (0.9%; 1,767), and 20° (0.8%; 1,601). There were 690 (0.34%) revisions for instability. Compared to neutral liners, the adjusted subhazard ratios of revision for instability were: 10°, 0.64 (p < 0.001); 15°, 0.48 (p < 0.001); and offset reorientating, 1.6 (p = 0.010). No association was found with other geometries. 10° and 15° liners had a time-dependent lower risk of revision for instability within the first 1.2 years. In posterior approaches, 10° and 15° liners had a lower risk of revision for instability, with no significant difference between them. The protective effect of lipped over neutral liners was not observed in laterally approached THAs. There were 604 (0.3%) revisions for loosening, but no association between liner geometry and revision for loosening was found. Conclusion. This registry-based study confirms a lower risk of revision for instability in posterior approach THAs with 10° or 15° lipped liners compared to neutral liners, but no significant difference between these lip sizes. A higher revision risk is seen with
The aim of this study was to determine whether there is a difference
in the rate of wear between acetabular components positioned within
and outside the ‘safe zones’ of anteversion and inclination angle. We reviewed 100 hips in 94 patients who had undergone primary
total hip arthroplasty (THA) at least ten years previously. Patients
all had the same type of acetabular component with a bearing couple
which consisted of a 28 mm cobalt-chromium head on a highly crosslinked
polyethylene (HXLPE) liner. A supine radiostereometric analysis
(RSA) examination was carried out which acquired anteroposterior
(AP) and lateral paired images. Acetabular component anteversion
and inclination angles were measured as well as total femoral head
penetration, which was divided by the length of implantation to
determine the rate of polyethylene wear.Aims
Patients and Methods
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
We present the medium-term clinical results of a reverse total
shoulder arthroplasty with a trabecular metal glenoid base plate. We reviewed 125 consecutive primary reverse total shoulder arthroplasties
(RTSA) implanted in 124 patients for rotator cuff arthropathy. There
were 100 women and 24 men in the study group with a mean age of
76 years (58 to 89). The mean follow-up was 32 months (24 to 60).
No patient was lost to follow-up.Aim
Patients and Methods
We report the kinematic and early clinical results
of a patient- and observer-blinded randomised controlled trial in which
CT scans were used to compare potential impingement-free range of
movement (ROM) and acetabular component cover between patients treated
with either the navigated ‘femur-first’ total hip arthroplasty (THA) method
(n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional
THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75).
The Hip Osteoarthritis Outcome Score, the Harris hip score, the
Euro-Qol-5D and the Mancuso THA patient expectations score were
assessed at six weeks, six months and one year after surgery. A
total of 48 of the patients (84%) in the navigated ‘femur-first’
group and 43 (65%) in the conventional group reached all the desirable
potential ROM boundaries without prosthetic impingement for activities
of daily living (ADL) in flexion, extension, abduction, adduction
and rotation (p = 0.016). Acetabular component cover and surface
contact with the host bone were >
87% in both groups. There was
a significant difference between the navigated and the conventional
groups’ Harris hip scores six weeks after surgery (p = 0.010). There
were no significant differences with respect to any clinical outcome
at six months and one year of follow-up. The navigated ‘femur-first’
technique improves the potential ROM for ADL without prosthetic
impingement, although there was no observed clinical difference
between the two treatment groups. Cite this article:
We assessed the orientation of the acetabular
component in 1070 primary total hip arthroplasties with hard-on-soft, small
diameter bearings, aiming to determine the size and site of the
target zone that optimises outcome. Outcome measures included complications,
dislocations, revisions and ΔOHS (the difference between the Oxford
Hip Scores pre-operatively and five years post-operatively). A wide
scatter of orientation was observed (2 This study demonstrated that with traditional technology surgeons
can only reliably achieve a target zone of ±15°. As the optimal
zone to diminish the risk of dislocation is also ±15°, surgeons
should be able to achieve this. This is the first study to demonstrate
that optimal orientation of the acetabular component improves the
functional outcome. However, the target zone is small (± 5°) and
cannot, with current technology, be consistently achieved. Cite this article:
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the risk of revision was related
to patient, surgeon and implant covariates. The overall rate of
revision at five years was 2.4% (99% confidence interval 2.02 to
2.79). In the final adjusted model, we found that the risk of revision
was significantly higher in patients receiving metal-on-metal (MoM:
hazard ratio (HR) 1.93, p <
0.001) and ceramic-on-ceramic bearings
(CoC: HR 1.55, p = 0.003) compared with the best performing bearing
(metal-on-polyethylene). The risk of revision was also greater for
smaller femoral stems (sizes 8 to 10: HR 1.82, p <
0.001) compared
with mid-range sizes. In a secondary analysis of only patients where body
mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m. 2. significantly
increased the risk of revision (HR 1.55, p = 0.002). The influence
of the bearing on the risk of revision remained significant (MoM:
HR 2.19, p <
0.001; CoC: HR 2.09,
p = 0.001). The risk of revision was independent of age, gender,
head size and
We report the results of 62 hips in 62 patients
(17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent
revision of the acetabular component of a total hip replacement
due to aseptic loosening between May 2003 and November 2007. All
hips had a Paprosky type IIIa acetabular defect. Acetabular revision
was undertaken using a Procotyl E cementless oblong implant with
modular side plates and a hook combined with impaction allografting. At a mean follow-up of 60.5 months (36 to 94) with no patients
lost to follow-up and one died due to unrelated illness, the complication
rate was 38.7%. Complications included aseptic loosening (19 hips),
deep infection (3 hips), broken hook and side plate (one hip) and
a femoral nerve palsy (one hip). Further revision of the acetabular component
was required in 18 hips (29.0%) and a further four hips (6.4%) are
currently loose and awaiting revision. We observed unacceptably high rates of complication and failure
in our group of patients and cannot recommend this implant or technique.
Dislocation is a common reason for revision following total hip replacement. This study investigated the relationship between the bearing surface and the risk of revision due to dislocation. It was based on 110 239 primary total hip replacements with a diagnosis of osteoarthritis collected by the Australian Orthopaedic Association National Joint Replacement Registry between September 1999 and December 2007. A total of 862 (0.78%) were revised because of dislocation. Ceramic-on-ceramic bearing surfaces had a lower risk of requiring revision due to dislocation than did metal-on-polyethylene and ceramic-on-polyethylene surfaces, with a follow-up of up to seven years. However, ceramic-on-ceramic implants were more likely to have larger prosthetic heads and to have been implanted in younger patients. The size of the head of the femoral component and age are known to be independent predictors of dislocation. Therefore, the outcomes were stratified by the size of the head and age. There is a significantly higher rate of revision for dislocation in ceramic-on-ceramic bearing surfaces than in metal-on-polyethylene implants when smaller sizes (≤ 28 mm) of the head were used in younger patients (<
65 years) (hazard ratio = 1.53, p = 0.041) and also with larger (>
28 mm) and in older patients (≥ 65 years) (hazard ratio = 1.73, p = 0.016).
Ensuring the accuracy of the intra-operative orientation of the acetabular component during a total hip replacement can be difficult. In this paper we introduce a reproducible technique using the transverse acetabular ligament to determine the anteversion of the acetabular component. We have found that this ligament can be identified in virtually every hip undergoing primary surgery. We describe an intra-operative grading system for the appearance of the ligament. This technique has been used in 1000 consecutive cases. During a minimum follow-up of eight months the dislocation rate was 0.6%. This confirms our hypothesis that the transverse acetabular ligament can be used to determine the position of the acetabular component. The method has been used in both conventional and minimally-invasive approaches.