Background

A stem sitting proud (SP) or that above the final rasp position remains in some patients who undergo hip replacement using proximally coated tapered wedge stems. Surgeons may face challenges providing the best fit due to unpredictable stem seating. Zimmer Inc. introduced a new rasp to solve this issue but the clinical results of this rasp have not yet been published. Therefore, we aimed to address the following: 1) What is SP incidence using a proximally coated cementless tapered wedge stem? 2) Does the new rasp system improve seating height? 3) What are the risk factors of SP?

Methods

We performed a retrospective study with 338 hips, in which Tri-Lock Bone Preservation Stem (BPS) was used in 181 and M/L Taper stem was used in 157 hips (82 hips before and 75 hips after the new rasp). A positive stem SP was defined as a stem proud height of >2 mm. We analyzed and compared SP incidence in two stems and in M/L Taper stems before and after the new rasp use.