Bone & Joint Research
Vol. 11, Issue 2 | Pages 82 - 90
7 Feb 2022
Eckert JA Bitsch RG Sonntag R Reiner T Schwarze M Jaeger S

Aims

The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation.

Methods

Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 6 - 6
1 Apr 2017
Kretzer J Sonntag R Kiefer H Reinders J Porporati AA Streicher R
Full Access

Background

The CoCrMo large bearings had shown a high failure rate, because of metal ion and particle release. Alumina matrix composite (AMC) ball heads have shown to mitigate such phenomena. The aim of this study was to investigate the leaching properties of AMC clinically as well as experimentally.

Methods

Two patient groups were compared: a control group (n=15) without any implant (Controls) and 15 Patients with unilateral treatment with Biolox delta ceramic-on-ceramic (CoC). Whole-blood samples of Controls and Patients (after 3 and 12 months from treatment with CoC) were measured by means of trace element analysis using a HR-ICPMS. The leaching behaviour of BIOLOX delta was also analysed in-vitro: five Biolox delta heads and five CoCrMo heads were immersed in serum for seven days at 37°C. Aluminium, cobalt, chromium and strontium were detected based on HR-ICPMS.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 13 - 13
1 Mar 2017
Sonntag R Al-Salehi L Braun S Mueller U Reinders J Kretzer J
Full Access

Introduction

Wear plays a key role in the clinical outcome of total hip replacements (THR). In addition, increased frictional moment can stress the implant interfaces which may lead to high torsional loadings in the intermodular taper junction (fretting) and cup loosening and to the development of noise (squeaking). Against the background of larger head diameters (increased range of motion and decreased risk of dislocation), the friction induced by the joint articulation is of particular interest. As of now, the investigation of friction with the use of relevant joint kinematics and loadings are limited to numerical studies. Experimental approaches use simplified models which do not take into consideration complex activities. Thus, with the aim of this study is the identification of articular frictional moments that consider critical in vivo loading conditions and kinematics as well as the clinical cup inclination, head size and clearance of ceramic-on-ceramic hip bearings.

Materials and Methods

A standard hip simulator (Minibionix 852 with 4 DOF Hip setup, MTS, Eden Prairie, USA) was modified in order to allow for high-precision friction measurements during head-insert articulation in all 6 DOF (MC2.5D-500, AMTI, Boston, USA). Disturbing systemic effects have been minimized by using quasi frictionless aerostatic lateral force compensation (Eitzenberger, Wessobrunn, Germany) and cross talk compensation. Beside the standard protocoll for in vitro wear assessment (ISO 14242-1), more complex profiles from in vivo patient data (Heidelberg Motion Lab and Orthoload database) have been used: normal walking with different walking speeds and patient's weights, stairs up/down and start-stop conditions. All-ceramic bearings (Biolox delta, Ceramtec, Plochingen, Germany) have been orientated in clinically relevant cup inclinations (30, 45, 60 and 75 deg). For each head diameter (28, 36 and 48 mm) n=8 specimens have been devided in two groups: small and large clearance according to the manufacturer's specification. All tests were run at 37°C in diluted bovine serum (20 g/l protein content).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 85 - 85
1 Feb 2017
Kretzer J Schroeder M Mueller U Sonntag R Braun S
Full Access

The numbers of anatomic total shoulder joint replacements (ATSR) is increasing during the past years with encouraging clinical results. However, the survivorship of ATSR is lower as compared to total knee and hip replacements. Although the reasons for revision surgery are multifactorial, wear-associated problems like loosening are well-known causes for long-term failure of ATSR. Furthermore there is lack of valid experimental wear tests for ATSR. Therefore the purpose of this study was to define experimental wear testing parameters for ATSR and to perform a wear study comparing ceramic and metallic humeral heads.

Kinetic and kinematic data were adopted from in-vivo loading measurements of the shoulder joint (orthoload.com) and from several clinical studies on shoulder joint kinematics. As activity an ab/adduction motion of 0 to 90° in combination with an ante/retroversion while lifting a load of 2 kg has been chosen. Also a superior-inferior translation of the humeral head has been considered. The wear assessment was performed using a force controlled AMTI joint simulator for 3×106 cycles (Fig. 1) and polyethylene wear has been assed gravimetrically.

The studied ATSR (TuronTM, DJO Surgical, USA) resulted in a polyethylene wear rate of 62.75 ± 1.60 mg/106 cycles in combination with metallic heads. The ceramic heads significantly reduced the wear rate by 26.7 % to 45.99 ± 1.31 mg/106 (p<0.01). The wear scars dimensions were in good agreement to clinical retrievals.

This study is the first that experimentally studied the wear behavior of ATSR based on clinical and biomechanical data under load controlled conditions. In term of wear the analyzed ATSR could clearly benefit from ceramic humeral heads. However, in comparison to experimental wear studies of total knee and hip replacements the wear rate of the studied ATSR was relatively high. Therefore further research may focus on optimized wear conditions of ATSR and the hereby described method may serve as a tool to evaluate a wear optimization process.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 122 - 122
1 Jan 2016
Kretzer JP Sonntag R Kiefer H Reinders J
Full Access

Introduction

Metal-on-metal bearings (MoM) have been reported to release metal ions that are potentially leading to adverse tissue reactions. Alternatively, ceramic-on-ceramic bearings (CoC) are an attractive treatment for young and active patients and composite materials like zirconia toughened alumina (ZTA) have been successfully introduced clinically. One of the most common ZTA-material in CoC is the Biolox® delta, manufactured by Ceramtec. Along with alumina and zirconia, this material also contains traces of chromium, strontium and yttrium. The aim of this study was to analyse the ion release for these materials clinically as well as experimentally.

Material and Methods

Within a clinical trial, three different patient groups were compared: a) a control group without any implants, b) patients, three months after unilateral treatment with Biolox® delta CoC and c) patients, twelve months after unilateral treatment with Biolox® delta CoC. Whole-blood samples were collected and analysed in regards to the trace elements using high-resolution-ICP-MS. In the experimental setup, the leaching behaviour of five Biolox® delta ceramic heads and five CoCr-heads was analysed. The heads were immersed in serum for seven days at 37°C. The ion-release of aluminium, zirconium, cobalt, chromium, strontium and yttrium were detected based on high-resolution-ICP-MS.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 133 - 133
1 Dec 2013
Nadorf J Thomsen M Sonntag R Reinders J Kretzer JP
Full Access

INTRODUCTION:

Good survival rates of cementless hip stems serve as motivation for further development, just like modular implant systems or short stems. New aims are worth striving for, e.g. soft tissue or bone sparing options with similar survival rates in case of short stems. Even minimal design modifications might result in complications, e.g. missing osseointegration, loosening of the implant or painful stem, as shown in the past.

One of these developments is the Biomet – GTS™ stem [Fig. 1], a hybrid between conventional cementless straight stem and potentially sparing short stem.

Aim of this biomechanical study was to analyze, if the biomechanical behavior of the stem is comparable to a clinically proofed design with respect to the stem fixation in the bone and to the mechanical behavior of the stem itself. That's why the primary stability of the GTS™ stem has been determined and subsequently was compared to the Zimmer – CLS® stem.

MATERIAL & METHODS

Four GTS™ stems and four CLS® stems were implanted standardized in eight synthetic femurs. Micromotions of the stem and the bone were measured at different sites. A high precision measuring device was used to apply two different cyclic load situations: 1. Axial torque of +/−7 Nm around the longitudinal stem axis to determine the rotational implant stability. 2. Varus-valgus-torque of +/−3, 5 Nm to determine the bending behavior of the stem. Comparing the motions of the stem and femur at different sites allowed the calculation of relative micromotions at the bone-implant-interface.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 380 - 380
1 Dec 2013
Kretzer JP Pohl V Zeifang F Buelhoff M Sonntag R Reinders J
Full Access

Hemi shoulder arthroplasty is an attractive treatment for shoulder arthritis in particular if the natural glenoid is still intact. However, comparing the clinical results of hemi and total shoulder arthroplasty clearly shows lower survival for the hemi arthroplasty. One of the most common reasons for revision surgery is gleniod erosion, where the cartilage or bone is worn of. Aim of the current study was to analyse if the metallic articular surface of retrieved hemi shoulder arthroplasty is different from new implants. We hypothesized that the surface roughness will increased due the articulation and that metallic wear is detectable on the implants. Twelve retrieved and three brand new hemi shoulder arthroplasty were included. The surface roughness (Ra, Rz, Rmax, Rsk) was measured on different sites of the surface (center of the head and at the edge). The implants were further measured using a coordinate measuring machine to gain information on volumetric wear and geometrical alterations.

Compared to new implants the surface roughness on the retrievals was significantly increased (Tab. 1), except for skewness.

Although the roughness parameters within the retrieval group were generally higher at the center of the head compared to the edge, this difference was not significant. Apart from form deviations no volumetric wear was detectable on the heads (Fig. 1).

The current results indicate that the metallic articular implant surface changes in vivo and that the material is hurt due to the articulation against the softer cartilage or bone. Although it can't be finally clarified by that study, to what extend the higher roughness is taking part in the process of the clinically observed erosion of the gleniod, it can be assumed that an increased roughness is disadvantageous. Possibly, the observed surface alterations won't occur clinically with harder materiel (e.g. ceramic), but this even needs to be validated.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 173 - 173
1 Dec 2013
Sonntag R Koch S Merziger J Rieger JS Reinders J Reiner T Kretzer JP
Full Access

Background

Migration analysis after total joint arthroplasty are performed using EBRA analysis (Krismer et al., 1997) or - more accurate but also much more cost-intensive and time-consuming – via radiostereometric analysis (RSA). For the latter, additional radiographs from two inclined perspectives are needed in regular intervals in order to define the position of the implant relative to tantalum bone markers which have been implanted during surgery of the artificial joint (Fig. 1). Modern analysis software promises a migration precision along the stem axis of a hip implant of less than 100 μm (Witvoet-Brahm et al., 2007). However, as the analysis is performed semi-automatically, the results are still dependent on the subjective evaluation of the X-rays by the observer. Thus, the present phantom study aims at evaluating the inter- and intra-observer reliability, the repeatability as well as the precision and gives insight into the potential and limits of the RSA method.

Materials and Methods

Considering published models, an RSA phantom model has been developed which allows a continuous and exact positioning of the prostheses in all six degrees of freedom (Fig. 2). The position sensitivities of the translative and rotative positioning components are 1 μm and 5 to 24, respectively. The roentgen setup and Model-Based RSA software (3.3, Medis specials bv, Leiden, Netherlands) was evaluated using the SL-PLUS® standard hip stem (size 7, Smith & Nephew, Baar, Switzerland). The inter-observer (10 repetitions) and intra-observer (3 observers) reliability have been considered. Additionally, the influences of the model repositioning and inclination as well as the precision after migration and rotation along the stem axis are investigated.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 159 - 159
1 Dec 2013
Reinders J Sonntag R Kretzer JP
Full Access

Background:

Standardized preclinical wear testing cannot replicate the variations of wear rates and wear mechanisms seen in-vivo [1]. Therefore, a lot of studies focused on testing scenarios which replicate a wider range of patient-specific conditions such as different activities or malalignment of components. However, differences between the in-vivo and in-vitro situation regarding the environmental conditions are often neglected.

Considerable differences between the in-vivo and the in-vitro situation are related to the surrounding synovial fluid and its in-vitro substitute (bovine serum). For the in-vivo knee only small volumes (1–4 ml) of synovial fluid are reported [2]. However, for in-vitro testing several hundreds of milliliters of bovine serum are typically used. Therefore, the hypothesis of this study is that fluid volume significantly influences the wear rates in simulator tests.

Material and Methods:

For wear testing an established implant system (Sigma® PFC, DePuy, Warsaw, USA) has been used.

Four wear tests with four different testing volumes of bovine serum were carried out: 250 ml; 150 ml; 75 ml; 45 ml. The testing volume of the original chamber (250 ml) was reduced using solids (Fig. 1). Care was taken, that the contact surfaces of the implants were completely immersed and the fluid level was kept constant in all cases (Fig. 1). The testing fluid was maintained at a temperature of 37 ± 1°.

All wear tests were run displacement controlled according to ISO 14243-3: 2004 on an AMTI knee simulator.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 498 - 498
1 Dec 2013
Reinders J Von Stillfried F Sonntag R Heisel C Kretzer JP
Full Access

Background:

Total ankle replacements (TAR) are not as successful as total hip or total knee replacements. A three-time increased revision rate is reported in registry data [1]. Therefrom, wear associated revisions are frequent [2]. However, there is little knowledge about the wear behavior of TAR. This may be partly related to the fact, that currently no standard for wear testing of TAR exists.

The aim of this study is to define a biomechanical valid, force-controlled test specification for level walking of TAR.

Material and Methods:

Basic requirements for force-controlled testing of TAR is the definition joint flexion, as well as active forces and torques acting on the joint and the definition of the ligamental stabilization of the joint.

To specify flexion of the ankle, gait analysis was performed on patients treated with a TAR (HINTEGRA, Smith & Nephew) using skin mounted markers. Data about in-vivo forces is missing for TAR. Hence, determination of active forces and torques was based on mathematical models as described in the literature.

A new testing device (figure 1) has been developed to measure ligamental stabilization of the ankle joint. Measurements were performed on 10 paired cadaver feet (n = 20). Measurements were performed in different flexion angles when applying anterior-posterior forces (± 160N) and internal-external torques (± 2,5 Nm) between the talus and the tibia.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 499 - 499
1 Dec 2013
Reinders J Von Stillfried F Sonntag R Heisel C Kretzer JP
Full Access

Background:

There is little knowledge about wear performance of total ankle arthroplasties (TAR). However, revisions rates are high for TAR [1] and wear associated revisions are frequent [2].

Therefore, the aim of this study is

To test the wear behavior of a TAR using a biomechanically valid testing scenario.

To test the influence of an alternative ceramic tibial component.

To test the long term wear performance of TAR.

Material and Methods:

In the first part of this study the HINTEGRA (Smith & Nephew) TAR has been used for wear testing. Wear testing was performed on a modified AMTI knee simulator. Level walking according to a previous described testing standard [see abstract: Development of a force controlled testing scenario for total ankle replacements] has been used. Level walking was simulated in three clinical relevant situations, first simulating the reduced loading after implantation, secondly simulating an increasing range of motion and at last a loading pattern orientating at the loadings in the native/healthy joint. Every simulation was run for 3 million cycles, resulting in 9 million total cycles.

In the second part of this study the metal tibial plateau was replaced by a ceramic tibial component (Biolox® Delta, CeramTec). Simulation was run, as described above, for additional 9 million cycles. Termed as a long term test, in total 18 million cycles of testing are performed.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 379 - 379
1 Dec 2013
Kretzer JP Reinders J Sonntag R Merle C Omlor G Streit M Gotterbarm T Aldinger P
Full Access

Corrosion in modular taper connections of total joint replacement has become a hot topic in the orthopaedic community and failures of modular systems have been reported. The objective of the present study was to determine in vivo titanium ion levels following cementless total hip arthroplasty (THA) using a modular neck system.

A consecutive series of 173 patients who underwent cementless modular neck THA and a ceramic on polyethylene bearing was evaluated retrospectively. According to a standardized protocol, titanium ion measurements were performed on 67 patients using high-resolution inductively coupled plasma-mass spectrometry. Ion levels were compared to a control group comprising patients with non-modular titanium implants and to individuals without implants.

Although there was a higher range, modular-neck THA (unilateral THA: 3.0 μg/L (0.8–21.0); bilateral THA: 6.0 μg/L (2.0–20.0)) did not result in significant elevated titanium ion levels compared to non-modular THA (unilateral THA: 2.7 μg/L (1.1–7.0), p = 0.821; bilateral THA: 6.2 μg/L, (2.3–8.0), p = 0.638). In the modular-neck THA group, patients with bilateral implants had significantly higher titanium ion levels than patients with an unilateral implant (p < 0.001). Compared to healthy controls (0.9 μg/L (0.1–4.5)), both modular THA (unilateral: p = 0.029; bilateral p = 0.003) and non-modular THA (unilateral: p < 0.001; bilateral: p < 0.001) showed elevated titanium ion levels.

The data suggest that the present modular stem system does not result in elevated systemic titanium ion levels in the medium term when compared to non-modular stems. However, more outliner were seen in modular-neck THA. Further longitudinal studies are needed to evaluate the use of systemic titanium ion levels as an objective diagnostic tool to identify THA failure and to monitor patients following revision surgery.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 276 - 276
1 Mar 2013
Nadorf J Jakubowitz E Heisel C Reinders J Sonntag R Kretzer JP
Full Access

Introduction

Concerning biomechanical research, human specimens are preferred to achieve conditions that are close to the clinical situation. On the other hand, synthetic femurs are used for biomechanical testing instead of fresh-frozen human femurs, to create standardized and comparable conditions. A new generation of synthetic femurs is currently available aiming to substitute the validated traditional one. Structural femoral properties of the new generation have already been validated, yet a biomechanical validation is missing.

The aim of our study was to analyse potential differences in the biomechanical behaviour of two different synthetic femoral designs by measuring the primary rotational stability of a cementless femoral hip stem.

Methods

The cementless SL-PLUS® standard stem (size 6, Smith&Nephew Orthopaedics AG, Rotkreuz, Swizerland) was implanted in two groups of synthetic femurs. Group A consists of three 2nd generation femurs and group B consists of three 4th generation femurs (both: size large, composite bone, Sawbones® Europe, Malmö, Sweden).

Using an established method to analyse the rotational stability, a cyclic axial torque of ±7.0 Nm along the longitudinal stem axis was applied. Micromotions were measured at defined levels of the bone and the implant. The calculation of relative micromotions at the bone-implant interface allowed classifying the rotational implant stability.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 337 - 337
1 Mar 2013
Sonntag R Reinders J Rieger JS Jaeger S Kretzer JP
Full Access

Background

Titanium, in particular Ti6Al4V, is the standard material used in cementless joint arthroplasty. Implants are subjected to cyclic loading where fracture is the reason for re-operation in 1.5–2.4% of all revisions in total hip arthroplasty. In order to strengthen critical regions, surface treatments such as shot peening may be applied.

A superficial titanium oxide layer is naturally formed on the surface as a protective film at ambient conditions. However, as its thickness is only in the range of several nanometers, it is prone to be destroyed by high loads - as present at the surface during bending - leading to an ‘oxidative wear’ in a corrosive environment [1]. The present study aims to evaluate the shot peening treatment on Ti6Al4V regarding its potential for cyclically loaded parts under a dry and a corrosive testing medium.

Materials and Methods

Hour-glass shaped titanium specimens (Ti6Al4V) with a minimal diameter of 10 mm have been subjected to an annealing treatment at 620°C for 10h to remove initial residual stresses introduced during machining. Subsequently, a high-intensity shot peening treatment with cut wire followed by a low-intensity cleaning process with glass beads have been performed (Metal Improvement, Germany). Arithmetic mean roughness Ra of the treated surfaces was measured (Mahr Perthometer M2, Germany). Residual stress depth profiles prior to and after shot peening have been measured by a Fe-filtered Co-K(alpha) radiation (GE Measurement&Control, USA) and calculated using the sin2(psi) method. Fatigue strength has been determined by two servo-hydraulic hydropulsers (Bosch Rexroth, Germany) at 10 Hz and a load ratio of R=0.1 either under dry conditions (8 specimens) or surrounded by a 0.9-% saline solution (6 specimens) (BBraun, Germany) (Fig. 1). Testing has been performed until fracture occurred or the total number of 10 × 106 cycles has been reached. All fracture surfaces have been analyzed after testing using FEG-SEM (Zeiss LEO 1530 VP Gemini, Germany).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 97 - 97
1 Sep 2012
Kretzer JP Moradi B Reinders J Jaeger S Bitsch R Sonntag R
Full Access

Introduction

Failure of total knee arthroplasty (TKA) is mainly caused by biological reactions against wear particles generated at the implant. So far, wear has been mainly attributed to polyethylene (PE) and much effort has been put into understanding and optimizing the wear mechanism of PE in recent years. However, evaluation of metal wear particles and ion release in TKR has been neglected so far although the implants present large metal surface areas. In the present study we aimed to analyse the wear performance of TKA and to study the kinetics of metal ion and particle release. We hypnotized that due to abrasion and corrosion TKA will release relevant levels of Cobalt (Co), Chromium (Cr), Molybdenum (Mo) and Titanium (Ti).

Methods

Implants were subjected to an in-vitro simulation applying physiological loadings and motions for 5 million walking cycles. Wear processes were determined gravimetrically and by measuring the release of Co, Cr, Mo and Ti ions using HR-ICP-MS. Surface alterations were determined through surface roughness measurements.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 152 - 152
1 Sep 2012
Reinders J Sonntag R Nadorf J Bitsch R Rieger JS Kretzer JP
Full Access

Background

Polyethylene (PE) as a bearing material for total joint replacements (TJR) represents the golden standard for the past forty years. However, over the past decade it becomes apparent that PE wear and the biological response to wear products are the limiting factor for the longevity of TJRs. For this reason research has focused onto PE wear particle analysis. A particle analysis highly depends on the methodological work and results often show discrepancies between different research groups. From there, our hypothesis was, that an often unattended influencing factor is the optical magnification which has been used for particle analyses.

Material and Methods

In the present study samples of a previous conducted knee wear simulator test were used. Wear particles were isolated from the bovine serum using an established method1. Briefly the serum was digested with hydrochloric acid and a continuous stirring and heating. Particles were filtered onto 20nm alumina filters and analyzed using high resolution field emission gun scanning electron microscopy (FEG-SEM). Filters were analyzed on the same points using three different magnifications: 5000, 15000 and 30000. To describe the size and morphology of the particles the equivalent circle diameter (ECD), aspect ratio (AR), roundness (R) and form factor (FF) were specified according to ASTM F 1877-05. The estimated total number (ETN) of particles was calculated based on the number of particles recovered on the filter, the analyzed area, the dilution, evaporation and the total serum volume.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 153 - 153
1 Sep 2012
Reinders J Sonntag R Bitsch R Jaeger S Rieger JS Kretzer JP
Full Access

Background

Polyethylene (PE) wear is known as a limiting factor for total knee replacements (TKR). Thus, preclinical wear testing is an important tool to assess the suitability of new designs and new materials. However, standardized testing (e.g. according to ISO 14243) does not cover the individual situation in the patient. Consequentially, this study investigates the following two parameters:

Testing-Frequency: Patients with TKR's show a humiliated walking frequency (down to 0,5Hz) compared to standardized testing (1Hz±0.1). In the first part of this study, the influence of a decreased test frequency on the PE wear behavior is investigated

Interval of lubricant replacement: For in-vitro testing bovine serum is used as a substitute for the synovial fluid. Physiologically a continuous regeneration and removement of destructed components is taking place. In contrast, for simulator testing the bovine serum is typically changed completely every 500.000 cycles/steps. Therefore the goal of the second part of this study was to test if the serum replacing interval affects the PE wear behavior.

Material and Methods

Wear tests were conducted on an AMTI force controlled knee simulator. A cruciate substituting (ultracongruent) implant design (TC Plus, Smith & Nephew, Rotkreuz, Switzerland) was used.

First, a reference wear study with a test frequency of 1Hz and a lubricant replacement interval (RI) of 500.000 cycles according to ISO 14243-1:2009 was carried out. Tests were run to a total of 5 million cycles.

A second wear test was run with a reduced frequency of 0.5 Hz. The reduced frequency resulted in an extended testing period for the same number of cycles. To exclude an influence of the extended time period, the lubricant was changed, in the first half of testing every 500.000 cycles corresponding to 12 days (cycle depending (CD)), and in the second part every 250.000 cycles corresponding to 6 days (time depending (TD)). Tests were run to a total of 3 million cycles.

A third test was run with a frequency of 1 Hz. For this test a reduced serum RI of 150.000 cycles was choosen. This test was run to a total of 1.500.000 cycles.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 118 - 118
1 Jun 2012
Kretzer JP Jakubowitz E Sonntag R Reinders J Heisel C
Full Access

Introduction

Osteolysis and aseptic loosening in total hip replacement (THR) is often associated with polyethylene (PE) wear. This caused interest in alternative bearing surfaces. Since the mid nineties, research focused on hard-hard bearings like metal-on-metal (MOM) or ceramic-on-ceramic (COC). However, concerns remain about biological reactions to metallic wear debris or failure of the ceramic components. A new approach to reduce wear with a minimized risk of failure may be the use of a metallic cup in combination with a ceramic head, the so called ceramic-on-metal bearing (COM). The aim of this study was to estimate the wear behaviour at an early stage of this COM bearing type in comparison to COC bearings using a hip simulator.

Material and Methods

Simulator studies were carried out on a single station hip simulator (MTS 858 Mini Bionix II, Eden Prairie, USA) in accordance to ISO 14242-1. Bovine serum was used as the test medium. Four COM and four COC bearings were used, both 36mm in diameter. The heads were made of a mixed-oxid ceramic (Biolox Delta(r)) paired with a high carbon wrought CoCrMo cup in the COM group whereas both components were made of Biolox Delta(r) in the COC group. Simulation was run to a total of 2.4×106 cycles. Wear measurements were performed in intervals of 0.2x106 cycles using a gravimetric method (Sartorius Genius ME235S, measuring solution: 15 μg, Sartorius, Göttingen, Germany).


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 453 - 453
1 Nov 2011
Kretzer J Jakubowitz E Hofmann K Heisel C Sonntag R Lietz E
Full Access

Knee wear simulator studies are performed to evaluate wear behavior of implants.

Simulation of the human gait cycle is often carried out continuously, without considering resting periods as they are part of patient’s daily live. In addition to dynamic activities like walking, daily activities also consist of static periods like standing, sitting or lying. During the day dynamic activities alternate continuously with static periods and most of the day is spent in passive periods, where no joint motion occurs. Such resting periods have not yet been considered in prosthetic knee wear tests. Implementing resting periods may cause an increase in friction and thus increased wear of the implant. The aim of the current study was to determine if the implementation of resting periods would increase polyethylene (PE) wear in total knee replacement (TKR).

Two wear studies were conducted using a force controlled AMTI knee simulator on a conventional bicondylar TKR. For the first study, simulation was carried out continuously according to ISO 14243-1. For the second test, four active gait cycles according to ISO 14243-1 were followed by one resting period cycle. In both tests 5x10E6 active load cycles at a frequency of 1 Hz (resulting in additional 1.25x10E6 pause cycles for the second test) were applied. Wear was measured gravimetrically and wear scars were documented photographically.

The mean wear rates measured 2.85 ± 0.27 mg/10E6 cycles for the ISO test without considering resting periods and 2.27 ± 0.23 mg/10E6 cycles for the test with resting periods implemented. There was no significant difference (p=0.22) in wear rate between both tests.

The inserts showed similar wear scars in both tests and no relevant differences in dimension and localization on the surface. Therefore the wear behavior after the two tests was similar.

Since wear is one of the most limiting factors for implant longevity, proper preclinical wear studies are essential. Based on the results of this experimental wear study, a continuous simulation without additional resting periods seems to be valid in wear simulation of TKR.