Research studies have established mathematical correlations between the lengths of bone segments and the possible biomechanical implications of these correlations. The Lucas sequence comprises a series of integers that adhere to the same recurrence relation as the Fibonacci sequence; it differs in that it can start with any two initial integers. The purpose of this study is to determine whether segmental lengths of the foot height, tibia, femur, and upper body follow a Lucas sequence pattern. This was a retrospective radiographic review of patients who underwent full-body EOS scans. The AP scan was used to measure standing foot height (Ft), tibial length (T), femoral length (Fe), upper body length (UB), and full body length. A linear regression test was performed to determine whether a Lucas sequence-based relationship exists between Ft + T and Fe, and between T + Fe and UB.Introduction
Method
The purpose of this study was to compare the results and complications of tibial lengthening over an intramedullary nail with treatment using the traditional Ilizarov method. In this matched case study, 16 adult patients underwent 19 tibial lengthening over nails (LON) procedures. For the matched case group, 17 patients who underwent 19 Ilizarov tibial lengthenings were retrospectively matched to the LON group.Objectives
Methods
External fixation is widely used in orthopaedic
and trauma surgery. Infections around pin or wire sites, which are usually
localised, non-invasive, and are easily managed, are common. Occasionally,
more serious invasive complications such as necrotising fasciitis
(NF) and toxic shock syndrome (TSS) may occur. We retrospectively reviewed all patients who underwent external
fixation between 1997 and 2012 in our limb lengthening and reconstruction
programme. A total of eight patients (seven female and one male)
with a mean age of 20 years (5 to 45) in which pin/wire track infections
became limb- or life-threatening were identified. Of these, four
were due to TSS and four to NF. Their management is described. A
satisfactory outcome was obtained with early diagnosis and aggressive
medical and surgical treatment. Clinicians caring for patients who have external fixation and
in whom infection has developed should be aware of the possibility
of these more serious complications. Early diagnosis and aggressive
treatment are required in order to obtain a satisfactory outcome. Cite this article:
Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along
the anatomical axis of the femur results in a lateral shift of the
mechanical axis by approximately 1 mm for each 1 cm of lengthening.
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.