Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 139 - 139
1 Apr 2019
Nambu S Chang D
Full Access

Objective

Clinical wear depends on several factors such as implant specific factors (material, design, and sterilization), surgical factors/techniques, and patient-specific factors (weights and activities). The load magnitude for wear testing in the standard protocols (i.e., 2 kN as per ASTM F1714 or 3 kN as per ISO 14243-3) represent an average patient weight and does not address the other “what-if”’ scenarios (i.e., wear vs. patient weights, activities, duration, etc.,). The results from in-vitro testing report the data in wear (mg) or wear rate (mg/Mc) and are only applicable to the parameters (i.e., loads, bearing diameter, thickness, etc.,) used for the testing and not suitable to the variations seen in clinical scenarios. Therefore, it is essential to present the wear summary that can normalize the parameters and which is relevant in both in-vitro and in-vivo conditions. The goal of the current study is an attempt to present wear as a parameter (i.e., wear factor that combines the wear test data and established- theoretical relationship) and is thus applicable in both in-vivo and in-vitro scenarios.

Methods

Wear factor was first evaluated using actual wear testing conducted on metal on cross-linked polyethylene bearings along with well-established Dowson's wall bridge equation.

As per Dowson-Wallbridge, volumetric wear is V=2.376·KNWR+C or K=V/(2.376·NWR) where V is the volumetric wear in mm3, K is the wear factor in mm3/Nmm, N is the number of cycles, W is the load in Newtons, R is the bearing radius in mm, and C is the creep (assumed to be negligible, i.e., C=0 in this model.

28 mm simulator wear was first used to evaluate wear factor, but since simulator wear presented as a mass loss, these results were converted to volumetric wear using the equation

V = m / ρ ,

(m is the wear in mg and r is the density of XLPE in mg/mm3 (=0.923).

The Dowson-Wallbridge equation was then validated for predictive accuracy against actual wear testing on the predecessor THR system. The wear factor thus obtained was used to compute the theoretical-wear for other sizes (i.e., 42 and 46 mm bearings). The theoretical-wear was then compared to simulator wear for predictive accuracy.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 49 - 49
1 Mar 2017
Nambu S Hines G Timmerman I
Full Access

Background

Published simulator studies for metal/UHMWPE bearings couples showed that increasing the femoral head diameter by 1 mm increases wear by approximately 10% due to increased contact area. Therefore, there are concerns about increased wear with dual mobility hip bearings.

Purpose of the study

The purpose of the study was to compare wear from dual mobility hip bearings to that with traditional fixed bearings. In addition, for the dual mobility bearings, the effect of femoral head material type on the liner wear was also evaluated.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 119 - 119
1 Mar 2017
Roark M Nambu S
Full Access

Introduction

Modular acetabular liners offer surgeons the flexibility of using various bearing materials and sizes to accommodate the patient's needs. The need for a robust locking mechanism to ensure the long term successful performance of the implant is critical due to the hardship a revision surgery would have on the patient. The traditional method to evaluate torque resistance by using epoxy to affix a roughed femoral head to the acetabular liner has been acceptable to this point. However, efforts to design an acetabular liner that is resistant to high torque failures have shown this method to be inadequate to evaluate the performance of the lock detail, as failures only occur between the femoral head and the liner. Therefore a test method that would ensure failure of the lock detail was needed.

Materials and Methods

In the current study the performance of prototype vs. production acetabular liners and shells were evaluated. Aluminum test shells were provided and a combination of production acetabular liners and prototype liners were provided by the Prototype Department at MicroPort Orthopedics. The traditional method was followed. A custom holding fixture was attached to the load cell plate of the test machine, and a roughed femoral head was attached to the actuator. The appropriate shell and liner combination was selected and assembled using three firm hammer blows with a two pound surgical hammer. Once assembled, the test construct was affixed to the holding fixture mounted to the test machine. Devcon 5 minute epoxy was mixed per the instruction and approximately 10 cc was placed into the cup of the liner. The femoral head was then brought into place using load control until contact was made. After the epoxy had cured torque was applied via the femoral head at a rate of 0.417° per second until failure of the epoxy or the lock detail was observed. In every trial the epoxy failed before the lock detail. A new method was devised. A paddle fixture was fabricated and attached to the actuator of the test frame (See Figure 1). The interior of the cups were modified to receive the paddle fixture. The test was repeated using the new fixation method and failure of the lock detail was achieved.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 50 - 50
1 Mar 2017
Nambu S Timmerman I Ewing M
Full Access

Introduction

Laser marking of implants surfaces is necessary in order to provide traceability during revisions which will help identify product problems more quickly, better execute product recalls and improve patient safety. There are several methods of marking employed within the medical field such as chemical etching, electro pencil marking, mechanical imprinting, casting of markings, marking with vibratory type contact, ink jet, hot foil and screen printing. However, these methods have various drawbacks including marking durability or addition of potentially toxic chemical compounds. As a result laser marking has become the preferred identification process for orthopedic implants. Laser marking is known for its high visual quality, good reproducibility and precision. However there are concerns about the laser marking potential to affect fatigue life of a device. There is a limited number of research papers that studied the effect of laser marking on fatigue life of implants. The objective of the current study is to investigate the effects of laser marking on the fatigue life of titanium alloy material.

Material and Methods

Two groups of four point bend specimens were used to investigate the effect of laser marking on the fatigue life. The first group comprised of the specimens without laser marking while the second group comprised of specimens with laser marking currently utilized for the implant surfaces. Prior to conducting the fatigue testing, a non-destructive X-ray diffraction (XRD) residual stress analysis was conducted to determine if the laser marking had introduced any residual stresses. Imaging analysis was also conducted to examine any potential surface damage on the test sample's surface. A servo-hydraulic test machine was used for the fatigue four point bend testing regime where the inner and outer spans were 30 mm and 90 mm respectively. All testing was conducted at a frequency of 10 Hz, a stress ratio R=0.1, and sine-wave loading in air. Testing was stopped at 10 Million cycles or at failure of the specimen.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 48 - 48
1 Mar 2017
Nambu S Ewing M Timmerman I Roark M Fitch D
Full Access

INTRODUCTION

Recently there have been case reports of component fractures and elevated metal ion levels potentially resulting from the use of cobalt-chrome modular necks in total hip arthroplasty. One potential cause that has been suggested is fretting corrosion caused by micromotion at the taper junction between the modular neck and the femoral stem. The objective of the current study was to investigate the effects of various impaction and loading methods on micromotion at the modular neck-femoral stem interface in a total hip replacement system.

METHODS

A femoral stem was potted using dental acrylic and displacement transducers were inserted to measure micromotion in the modular neck pocket (Figure 1a). An 8° varus, long, cobalt-chrome, modular neck and 28 mm XXL cobalt-chrome femoral head were inserted in the femoral stem using various assembly techniques (a) hand assembly, (b) impaction loads: 2, 3, 4, 6, 16.4 kN and (c) in- vivo simulated impaction loads (constructs were placed on top of a block of ballistic gel (Clear Ballistic LLC, Fort Smith AR) and impacted): 2, 4, and 16.4 kN (Figure 1b). Impaction was obtained by placing the construct in a drop tower and impacting them. All constructs were oriented in 10/9 as per ISO 7206-6 and tested in an MTS machine with a sinusoidal load of 2.3 kN for 1,000 cycles in air at frequency of 10 Hz (Figure 1a). Micromotion data was recorded. To simulate the loading experienced with heavier patients and/or higher impact activities, selected constructs (as shown in Table 1) were sinusoidally loaded with 5.34 Kn load. Three samples were tested for all methods described above.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 11 - 11
1 Feb 2017
Parker A Ali A Nambu S Obert R
Full Access

Objectives

Modularity in total knee arthroplasty, particularly in revisions, is a common method to fit the implants to a patient's anatomy when additional stability or fixation is needed. In such cases, it may be necessary to employ multiple points of modularity to better match the anatomy. Taper junction strength at each of these levels is critical to maintain the mechanical stability of the implant and minimize micromotion. This effect of distributed assembly loads through multiple tapers and the resulting strength of the construct have not been previously evaluated on this revision tibial implant. The purpose of this study was to evaluate the possible dissipation of impaction force through multiple taper connections as compared to a single connection.

Methods

Two different constructs representative of modular implants were studied: a construct with a single axial taper connection (Group A; representing implant-stem) was compared to a construct with an adaptor that included two, offset, modular taper connections (Group B; representing implant-adapter-stem). For Group A, the stem taper was assembled and impacted through the stem. For Group B, the two tapers of the adapter and stem were hand assembled with the mating components and impacted simultaneously through the stem. Assembly load for each construct was recorded. As shown in Figure 1, the constructs were then fixed in a mechanical test frame and an axial distraction force was applied to the end of the stem at a constant displacement rate of 0.075 mm/sec until taper separation or mechanical failure occurred. Force and displacement data were recorded at 50 Hz. Disassembly force was normalized to assembly force for each component. Minitab software was used to analyze the data using a t-test.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 120 - 120
1 May 2016
Parker A Fitch D Nambu S Timmerman I
Full Access

Introduction

Total knee replacement (TKR) implant designs and materials have been shown to have a significant impact on tibial insert wear. A medial-pivot (MP) design theoretically should generate less wear due to a large contact area in the medial compartment and lower contact stresses. Synovial fluid aspiration studies have confirmed that a first generation MP TKR system (ADVANCE®, MicroPort Orthopedics Inc., Arlington, TN, USA) generates less wear debris than is seen with other implant designs articulating against conventional polyethylene (CP).

Objectives

The objective of this study was to evaluate the in vitro wear rate of a second generation MP TKR system (EVOLUTION® Cruciate-Sacrificing, MicroPort Orthopedics Inc., Arlington, TN, USA) using CP tibial inserts and compare to previously published values for other TKR designs with CP and first or second generation crosslinked polyethylene (XLPE) tibial inserts.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 449 - 449
1 Dec 2013
Nambu S Obert R Roark M Linton D Bible S Moseley J
Full Access

Introduction:

Modular necks allow intra-operative adjustment of neck length, offset, and version, enabling the surgeon to better match leg length and accommodate anatomical differences. However, there have been recent reports of early fatigue failures of the neck initiating from the neck/stem taper, and some retrieved components exhibit severe fretting corrosion.1 Fatigue testing according to ISO 7206-6 (10/9 orientation) has been shown to replicate the clinical fatigue failures, but results in relatively minor fretting and corrosion. The purpose of this pilot study was to evaluate techniques for accelerating fretting corrosion with the goal of replicating the most severely corroded clinical retrieval cases.

Methods:

Constructs tested in this study consisted of a single stem and neck design (PROFEMUR modular, Wright Medical Technology). The worst case long varus neck design was evaluated in two materials: Ti6Al4V and wrought CoCr. In vitro fatigue testing in the 10/9 configuration was conducted at 10 Hz in unbuffered, aerated saline. Fretting mass loss, distraction force, and assessment of taper corrosion via SEM/EDS were measured. Methods used to exacerbate fretting corrosion are shown in Figure 2.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 451 - 451
1 Nov 2011
Blaha JD DeBoer D Barnes CL Obert R Nambu S Stemniski P Carroll M
Full Access

Introduction: Many attempts have been made to describe the flexion axis of the knee based on landmarks or simple geometric representations of the anatomy. An alternative approach is to use kinematic data to describe the axis of motion of the joint. The helical axis is one kinematic parameter that can accomplish this. The purpose of this study was to compare the correlation between kinematic and anatomic axes of motion.

Methods: Six cadaver lower extremities were skeletonized except for the knee joint. Passive navigation markers were implanted, and CT scans obtained. The limbs were then placed in an open-chain lower extremity rig that allows full range of knee motion. Threedimensional kinematic data were recorded using a camera and the helical axis of motion was calculated. Anatomic landmarks were placed on CT derived CAD models of the extremities consisting of spherical and cylindrical fits of the femoral condyles and a trans-epicondylar axis. Data for the normal knee was processed, by comparison of the helical axis to the landmark axes over varying ranges of flexion and the variation in helical axis direction within that range was also calculated.

Results: The flexion range with the minimum variation of anatomic parameters to the helical axis was 30–100°. Helical axis variation in this range was 5.489 ± 1.173, while variation between the helical axis and those axis defined by spherical, TEA, and cylindrical landmarks were 5.115 ± 2.129°, 3.127 ± 2.029°, and 5.111 ± 1.710°, respectively. A students t-test was performed on each data set with the null hypothesis that the angular difference between the anatomically defined axes and the helical axis is zero. All axes were found to be significantly different from the average helical axis in the range of 30–100° (P= 0.002, 0.013, and 0.001, respectively). The tightest variation in the helical axis occurred at 40–50° of flexion 2.89 ± 0.722.

Conclusion/Discussion/Summary: None of the anatomic landmarks considered in this study represent a consistently valid approximation of the kinematic flexion axis of the knee. The TEA represents the closest approximation of the three with a 95% CI between 0.998 and 5.256°. The range of 30–100° represented the tightest variation over the largest range of flexion. Extension was defined at approximately 30° based on kinematic profiles of internal/external rotation which show a “screw-home” tendency beginning at 30° through extension. This behavior is consistent with an increase in helical axis variation in ranges that were less than 30° of flexion. In a previous open-chain model, both compartments of the joint were spinning around 45 degrees of flexion, which is consistent with the smallest helical axis variation observed in the 40–50° range.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 178 - 178
1 Mar 2008
NAMBU S CARROLL M SEYER S TIMMERMAN I
Full Access

Large diameter metal-metal bearings possess several clinical advantages over small bearings including greater joint stability, improved range of motion, and lower wear due to improved lubrication. Simulator wear tests were conducted to assess the effects of thermal processing on the wear behavior of large diameter metal-metal hip bearings.

Three groups of high carbon, cast 54 mm hip bearings with different thermal processing histories were tested. Two groups of bearings were manufactured to identical specifications and subjected to either no heat treatment(as-cast) or to typicall thermal processing prior to testing. The third group was comprised of commercially available as castbearing systems. Wear tests were performed on a Shore Western orbital bearing wear test machine. A simulated gait profile (triple-peak Paulprofile) with a maximum force of 2000N was applied to the bearings at a frequency of 1 Hz. The bearings were tested in the inverted position (headabove, shell below).

The general wear behavior of all three groups of bearing couples was similar to that previously reported for metal-metal bearings. All couples exhibited a run-in wear phase followed by a low-wear steady-state phase. For all bearing couples tested the heads demonstrated more wear than the shells. The appearance of the worn surfaces of all the bearing couples tested in this study were consistent with that of previously reported in-vitro wear testing as well as metal-metal hip bearing retrieval studies. There was no statistical difference among the three groups tested in the run-in or steady-state wear rates, although the heat treated bearings tended to wear less on average.

The results of this study indicate that thermal processing has no adverse effect on the wear of large diameter metal-metal hip bearings.