Acetabular component orientation is an important determinant of outcome following total hip arthroplasty (THA). Although surgeons aim to achieve optimal cup orientation, many studies demonstrate their inability to consistently achieve this. Factors that contribute are pelvic orientation and the surgeon's ability to correctly orient the cup at implantation. The goal of this study was to determine the accuracy with which surgeons can achieve cup orientation angles. In this Introduction
Methods
In total hip arthroplasty, a high radiographic inclination angle (RI) of the acetabular component has been linked to short- and long-term complications. There are several factors that lead to RI outliers including cup version, pelvic orientation and angle of the cup introducer relative to the floor. The primary aim of this study was to analyse what increases the risk of having a cup with an RI outside the target zone when controlling cup orientation with a digital inclinometer. In this prospective study, we included 200 consecutive patients undergoing uncemented primary THA in the lateral decubitus position using a posterior approach. Preoperatively, the surgeon determined the target intraoperative inclination (IOItarget). The intra-operative inclination of the cup (IOIcup) was measured with the aid of a digital inclinometer after seating of the acetabular component. Anteroposterior pelvic radiographs were made to measure the RI of the acetabular component. The target zones were defined as 30°-45° and 35°-45° of RI. The operative inclination relative to the sagittal plane of the pelvis (OImath) was calculated based on the radiographic inclination and anteversion angle. The difference between two outcome measures was expressed as Δ.Introduction
Methods
The orientation of the acetabular component is influenced by the orientation at which the surgeon implants the component and the orientation of the pelvis at the time of implantation. When operating with the patient in the lateral decubitus position, pelvic orientation can be highly variable. The goal of this study was to examine the effect of two different pelvic supports on cup orientation. In this prospective study, 200 consecutive patients undergoing uncemented primary THA in the lateral decubitus position were included. In the control group a single support over the pubic symphysis (PS) was used. In the study group, a single support over the ipsilateral anterior superior iliac spine (ASIS) was used. In every patient, the cup was inserted and the angle of the cup introducer relative to the floor (apparent operative inclination; OIa) was measured with the aid of a digital inclinometer. The radiographic inclination (RI) was measured on anteroposterior pelvic radiographs at 6 weeks postoperatively. The target zone for cup inclination was 35–45°. In both cohorts the cups were implanted close to the target OIa with an absolute difference with the OIa of 0.86° SD 0.82 in the PS cohort and 1.03° SD 0.99 in the ASIS cohort (p=0.18). The difference between the RI and OIa was higher in the PS cohort 12.2° SD 4.1 compared with 7.5° SD 3.7 in the ASIS cohort (p<0.0001) with also a bigger variance (p=0.04) in the PS cohort. The mean RI was 38.5° SD 4.4 compared with 39.2° SD 4.1 (p=0.26) respectively. There were more cups outside the RI target zone in the PS cohort compared with the ASIS cohort (respectively 26 versus 15; p<0.05). In this study the mean difference between the RI and OIa (the angle of the cup introducer during surgery) was significantly less when using a support over the ASIS compared with a support over the pubic symphysis. Apparently using a support over the ASIS causes less pelvic motion during surgery compared with a support over the pubic symphysis. This resulted in less variance and inclination outliers when using a tight target zone of 35–45°.
The most effective surgical approach for total hip arthroplasty
(THA) remains controversial. The direct anterior approach may be
associated with a reduced risk of dislocation, faster recovery,
reduced pain and fewer surgical complications. This systematic review
aims to evaluate the current evidence for the use of this approach
in THA. Following the Cochrane collaboration, an extensive literature
search of PubMed, Medline, Embase and OvidSP was conducted. Randomised
controlled trials, comparative studies, and cohort studies were
included. Outcomes included the length of the incision, blood loss,
operating time, length of stay, complications, and gait analysis.Aims
Materials and Methods
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
In total hip arthroplasty (THA), a high radiographic inclination angle (RI) of the acetabular component has been linked to an increased dislocation rate, liner fracture, and increased wear. In contrast to version, we have more proven boundaries when it comes to a safe zone for angles of RI. Although intuitively it seems easier to achieve a target RI, most studies demonstrate a lack of accuracy and the trend towards a high RI with all surgical approaches when using a freehand technique or a mechanical guide. This is due to pelvic motion during surgery, which can be highly variable. The current study had two primary aims, each with a different primary outcome. The first aim was to determine how accurate a surgeon could obtain the target operative inclination (OI) during THA when using a cementless cup using a digital protractor. The second aim was to determine how accurate a surgeon can estimate the target OI to obtain a RI of 40° based on the patient's hip circumference as demonstrated in a previous study. In this prospective study, we included 200 consecutive patients undergoing uncemented primary THA in the lateral decubitus position using a posterior approach. Preoperatively, the surgeon determined the target OI based on the patient's hip circumference (22.5°, 25°, 27.5° or 30°). Intraoperatively, the effective OI was measured with the aid of a digital inclinometer after seating of the acetabular component. Six weeks postoperatively anteroposterior pelvic radiographs were made and two evaluators, blinded to the effective OI, measured the RI of the acetabular component. The safe zone for inclination was defined as 30°-45° of inclination.Introduction
Methods
A high inclination angle has been linked to an increased dislocation rate, liner fracture, and increased wear. The aim of this study was to compare the operative (OI) with the radiological inclination (RI) angle and determine the influence of patient morphology on pelvic tilt and cup inclination angle. In the first cohort of 100 patients undergoing uncemented primary total hip arthroplasty, the cup was inserted freehand. In the second cohort of 100 patients, the OI was measured with the aid of a digital inclinometer. RI and pelvic tilt in lateral decubitus were measured.Introduction
Methods
A high radiographic inclination angle (RI) contributes
to accelerated wear and has been associated with dislocation after
total hip arthroplasty (THA). With freehand positioning of the acetabular
component there is a lack of accuracy, with a trend towards a high
radiographic inclination angle. The aim of this study was to investigate
whether the use of a digital protractor to measure the operative
inclination angle (OI) could improve the positioning of the acetabular
component in relation to a ‘safe zone’. We measured the radiographic inclination angles of 200 consecutive
uncemented primary THAs. In the first 100 the component was introduced
freehand and in the second 100 a digital protractor was used to
measure the operative inclination angle. The mean difference between the operative and the radiographic
inclination angles
(∆RI–OI) in the second cohort was 12.3° (3.8° to 19.8°). There was
a strong correlation between the circumference of the hip and ∆RI–OI.
The number of RI outliers was significantly reduced in the protractor
group (p = 0.002). Adjusting the OI, using a digital protractor and taking into
account the circumference of the patient’s hip, improves the RI
significantly (p <
0.001) and does not require additional operating
time. Cite this article:
The orientation of the acetabular component can
influence both the short- and long-term outcomes of total hip replacement
(THR). We performed a prospective, randomised, controlled trial
of two groups, comprising of 40 patients each, in order to compare
freehand introduction of the component with introduction using the transverse
acetabular ligament (TAL) as a reference for anteversion. Anteversion
and inclination were measured on pelvic radiographs. With respect to anteversion, in the freehand group 22.5% of the
components were outside the safe zone With respect to inclination, in the freehand group 37.5% of the
components were outside the safe zone The transverse acetabular ligament may be used to obtain the
appropriate anteversion when introducing the acetabular component
during THR, but not acetabular component inclination. Cite this article:
Biomechanical studies have shown that a long, centrally placed screw is favourable in scaphoid fracture fixation. A volar percutaneous transtrapezial approach was developed to facilitate central screw placement. The purpose of this study was to evaluate radiographic changes at the ST-joint at long term follow up in patients where this approach was used. Results were graded with use of the visual analogue scale (VAS) and modified Mayo wrist score. Radiographs of both hands, comprising an anteroposterior, lateral and 45 degrees pronated oblique view were obtained. Degenerative changes at the ST-joint were staged according to the modified Eaton & Glickel classification.Introduction
Method
Cemented femoral stems have an excellent long-term outcome. Modern cement techniques should be used to optimize femoral stem fixation. Bleeding from the bone surface during cemented hip arthroplasty compromises the bone-cement interface. However, no studies have examined this bleeding in vivo nor the effect the different cleaning methods used. In the present study we evaluated bleeding patterns and efficacy of cleaning methods used in third generation cementing techniques. We prospectively performed a medulloscopy with a 10 mm laparoscope in 200 primary hip arthroplasties. Intramedullary bleeding was evaluated after femoral canal preparation and use of the different cleaning methods. The femoral canal was divided into three areas to facilitate comparison. The intramedullary bleeding was standardized on a four point scale. A non-parametric repeated measures ANOVA was used for statistical analysis.Background
Methods
In recent years, there has been a significant advancement in our understanding of femoro-acetabular impingement and associated labral and chondral pathology. Surgeons worldwide have demonstrated the successful treatment of these lesions via arthroscopic and open techniques. The aim of this study is to validate a simple and reproducible classification system for acetabular chondral lesions. In our classification system, the acetabulum is first divided into 6 zones as described by Ilizalithurri VM et al [Arthroscopy 24(5) 534-539]. The cartilage is then graded as 0 to 4 as follows: Grade 0 – normal articular cartilage lesions; Grade 1 softening or wave sign; Grade 2 - cleavage lesion; Grade 3 - delamination and Grade 4 –exposed bone. The site of the lesion is further typed as A, B or C based on whether the lesion is 1/3 distance from acetabular rim to cotyloid fossa, 1/3 to 2/3 distance from acetabular rim to cotyloid fossa and > 2/3 distance from acetabular rim to cotyloid fossa. For validating the classification system, six surgeons reviewed 14 hip arthroscopy video clips. All surgeons were provided with written explanation of our classification system. Each surgeon then individually graded the cartilage lesion. A single observer then compared results for observer variability using kappa statistics.Introduction
Methods
For validating the classification system, six surgeons reviewed 14 hip arthroscopy video clips. All surgeons were provided with written explanation of our classification system. Each surgeon then individually graded the cartilage lesion. A single observer then compared results for observer variability using kappa statistics.
There have been considerable recent advances in the understanding and management of femoroacetabular impingement and associated labral and chondral pathology. We have developed a classification system for acetabular chondral lesions. In our system, we use the six acetabular zones previously described by Ilizaliturri et al. The cartilage is then graded on a scale of 0 to 4 as follows: grade 0, normal articular cartilage lesions; grade 1, softening or wave sign; grade 2, cleavage lesion; grade 3, delamination; and grade 4, exposed bone. The site of the lesion is further classed as A, B or C based on whether the lesion is less than one-third of the distance from the acetabular rim to the cotyloid fossa, one-third to two-thirds of the same distance and greater than two-thirds of the distance, respectively. In order to validate the classification system, six surgeons graded ten video recordings of hip arthroscopy. Our findings showed a high intra-observer reliability of the classification system with an intraclass correlation coefficient of 0.81 and a high interobserver reliability with an intraclass correlation coefficient of 0.88. We have developed a simple reproducible classification system for lesions of the acetabular cartilage, which it is hoped will allow standardised documentation to be made of damage to the articular cartilage, particularly that associated with femoroacetabular impingement.
The difference in the mean values regarding inclination was greater than would be expected by chance; there was a statistically significant difference (P = 0,010).
Navigation technique was discussed to equalize the drawback of MIS. However, tools like imageless navigation may further improve the cup position even in traditional approach.
The aim of our study was to determine the usefulness of preoperative digital templating of cementless total hip arthroplasty (THA). 60 consecutive cementless THA (synergy stem &
reflection cup) were templated digitally by two senior hip arthroplasty fellows (GM, YG) independently. A metallic marker ball of known diameter was used in all images to help scale for magnification. A blinded observer then collated information on the actual implant sizes, size of head component, offset, and level of neck cut intraoperatively. This was used to statistically analyse the correlation (Interclass correlation coefficient) between the digitally templated implant sizes and actual implant sizes used and the reliability of digital templating. A high rate of coincidence between digitally templated estimates and actual implant sizes was noted for both groups of templates. A high intraclass correlation coefficient (ICC) for the acetabular cup, stem and head were noted (ICC of 0.825, 0.794, and 0.884 respectively). Moderate agreement was noted for neck cut (ICC of 0.567) and leg length (ICC of 0.612). In conclusion, digital templating can reliably estimate implant sizes in cementless total hip arthroplasty. Valuable information on neck cut and leg length can be obtained by preoperative templating.