Advocates of debridement, antibiotics and implant retention (DAIR)
in hip periprosthetic joint infection (PJI) argue that a procedure
not disturbing a sound prosthesis-bone interface is likely to lead
to better survival and functional outcome compared with revision.
This case-control study aims were to compare outcome of DAIRs for
infected primary total hip arthroplasty (THA) with outcomes following
primary THA and two-stage revision of infected primary THAs. We retrospectively reviewed all DAIRs, performed for confirmed
infected primary hip arthropasty (n = 82) at out institution, between
1997 and 2013. Data recorded included full patient information and
type of surgery. Outcome measures included complications, mortality,
implant survivorship and functional outcome. Outcome was compared with
two control groups matched for gender and age; a cohort of primary
THAs (n = 120) and a cohort of two-stage revisions for infection
(n = 66).Aims
Patients and Methods
Advocates of Debridement-Antibiotics-and-Implant-Retention (DAIR) in hip peri-prosthetic joint infection (PJI) argue that a procedure not disturbing a sound prosthesis-bone interface is likely to lead to better survival and functional outcome compared to revision. However, no evidence supports this. This case-control study's aims were to compare outcome of DAIRs for infected 1° total hip arthroplasty (THA) with outcomes following 1° THA and 2-stage revisions of infected 1° THAs. We retrospectively reviewed all DAIRs, performed for confirmed infected 1° THR (DAIR-Group, n=80), in our unit between 1997–2013. Data recorded included patient demographics, medical history, type of surgery and organism identified. Outcome measures included complications, mortality, implant survivorship and functional outcome using the Oxford Hip Score (OHS). Outcome was compared with 2 control groups matched for gender and age; a cohort of 1° THA (1°-THA-Group, n=120) and a cohort of 2-stage revisions for infection (2-Stage-Revision-Group, n=66).Aim
Method
Native hip joint infection can result in hip arthrosis as a complication requiring the need for subsequent arthroplasty. There is little evidence to support single or staged THA. We present the results of patients who have undergone total hip arthroplasty (THA) following hip joint sepsis in our institution. Patients receiving a THA following previous hip joint sepsis between 2003 and 2015 were identified from the operative records database. Further clinical information was collected from the medical records, including timing of infection and surgery, as well as culture results. Functional outcome (Oxford Hip Score, OHS) and patient satisfaction scores (PSS) were recorded.Introduction
Methods
The value of Debridement-Antibiotics-and-Implant-Retention (DAIR) in prosthetic-joint-infection (PJI) is still a matter of debate as most studies to-date are underpowered with variable end-points. In our, tertiary referral, bone infection unit we consider DAIR to be a suitable option in all PJIs with soundly fixed prostheses, despite chronicity. The aims of this study were to define the long-term outcome following DAIR in hip PJI and identify factors that influence it. This is a retrospective consecutive case series of DAIRs performed between 1997 and 2013. Only infected cases confirmed by established criteria were included. Data recorded included patient demographics, medical history (ASA grade, Charlson and KLICC scores), type of surgery performed (DAIR or DAIR + exchange of modular components) and organism grown. Outcome measures included complications, implant survivorship and functional outcome (Oxford Hip Score, OHS).Introduction
Methods
Debridement, antibiotics and implant retention (DAIR) is a surgical option in the treatment of prosthetic joint infection (PJI). It is thought to be most appropriate in the treatment of early (≤6 weeks post-op) PJI. Most studies to-date reporting on DAIRs in hip PJI have been underpowered by reporting on small cohorts (n= <45), or report on registry data with associated biases and limitations. In our, tertiary referral, bone infection unit we consider DAIR to be a suitable option in all cases of PJI with a soundly fixed prosthesis, with early or late presentation, especially in patients who are too elderly or infirm to undergo major surgery. Aim: To define the 10-year outcome following DAIR in hip PJI and identify factors that influence it. We retrospectively reviewed all DAIRs performed in our unit between 1997 and 2013 for hip PJI. Only infected cases confirmed by histological and microbiological criteria were included. Data recorded included patient demographics and medical history, type of surgery performed (DAIR or DAIR + exchange of modular components), organism identified and type/duration of antibiotic treatment. Outcome measures included complications, mortality rate, implant survivorship and functional outcome. 121 DAIRs were identified with mean age of 71 years (range: 33–97). 67% followed an index procedure of 1° arthroplasty. 53% included exchange of modular components. 60% of DAIRs were for early onset PJI. Isolated staphylococcus was present in 50% of cases and 25% had polymicrobial infection. At follow-up (mean:7 years, range: 0.3 – 18), 83 patients were alive; 5- and 10- year mortality rates were 15% and 35% respectively. 45% had a complication (persistence of infection: 27%, dislocation: 10%) and 40% required further surgery. Twenty hips have been revised to-date (17%). Performing a DAIR and not exchanging the modular components was associated with an almost 3× risk (risk ratio: 2.9) of subsequent implant failure (p=0.04). 10-yr implant survivorship was 80% (95%CI: 70 – 90%). Improved 10-year implant survivorship was associated with DAIR performed for early PJI (85% Vs 68%, p=0.04). Functional outcome will be discussed. DAIR is a particularly valuable option in the treatment of hip PJI, especially in the early post-operative period. Whenever possible, exchange of modular implants should be undertaken, however DAIRs are associated with increased morbidity even in early PJI. Factors that predict success of DAIR in late PJI need to be identified.
Joint preserving procedures have gained popularity in an attempt to delay arthroplasty in young, dysplastic hips. Excellent results can be achieved with peri-acetabular osteotomy (PAO) in congruent non-arthritic hips. The role of salvage procedures such as the Shelf acetabuloplasty remains undefined. This study aims to determine the long-term survival and functional outcome following Shelf acetabuloplasty and to identify factors that influence outcome. This is a retrospective, consecutive, multi-surgeon, case series from a UK referral centre. 125 Shelf procedures were performed between 1987–2013 on 117 patients for symptomatic hip dysplasia. Mean age was 33 years (15–53). Mean follow-up was 10 years (1–27). Radiographic parameters measured included pre-operative arthritis, acetabular-index, centre-edge-angle, joint congruency, subluxation and femoral sphericity. Oxford Hip and UCLA scores were collected at follow-up. Failure was defined as conversion to arthroplasty or OHS<20.Introduction
Patients/Materials & Methods
There is great variability in acetabular component
orientation following hip replacement. The aims of this study were
to compare the component orientation at impaction with the orientation
measured on post-operative radiographs and identify factors that
influence the difference between the two. A total of 67 hip replacements
(52 total hip replacements and 15 hip resurfacings) were prospectively
studied. Intra-operatively, the orientation of the acetabular component
after impaction relative to the operating table was measured using
a validated stereo-photogrammetry protocol. Post-operatively, the
radiographic orientation was measured; the mean inclination/anteversion
was 43° ( This study demonstrated that in order to achieve a specific radiographic
orientation target, surgeons should implant the acetabular component
5° less inclined and 8° more anteverted than their target. Great
variability (2 Cite this article:
The orientation of the acetabular component is
influenced not only by the orientation at which the surgeon implants
the component, but also the orientation of the pelvis at the time
of implantation. Hence, the orientation of the pelvis at set-up
and its movement during the operation, are important. During 67
hip replacements, using a validated photogrammetric technique, we
measured how three surgeons orientated the patient’s pelvis, how
much the pelvis moved during surgery, and what effect these had
on the final orientation of the acetabular component. Pelvic orientation
at set-up, varied widely (mean (± 2, standard deviation ( Cite this article:
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the BHR from an independent centre. In this cohort of 554 patients (646 BHRs) with a mean age of
51.9 years (16.5 to 81.5) followed for a mean of eight years (1
to 12), the survival and patient-reported functional outcome depended
on gender and the size of the implant. In female hips (n = 267)
the ten-year survival was 74% (95% confidence interval (CI) 83 to
91), the ten-year revision rate for pseudotumour was 7%, the mean
Oxford hip score (OHS) was 43 ( This study supports the ongoing use of resurfacing in young active
men, who are a subgroup of patients who tend to have problems with
conventional THR. In contrast, the results in women have been poor
and we do not recommend metal-on-metal resurfacing in women. Continuous
follow-up is recommended because of the increasing incidence of
pseudotumour with the passage of time.
Since the introduction of 3rd generation Metal-on-Metal-Hip-Resurfacing-Arthroplasty (MoMHRA), thousands of such prostheses have been implanted worldwide in younger patients with end-stage hip osteoarthritis. However, no independent centre has reported their medium-to-long term outcome. The aim of this study is to report the ten year survival and outcome of the Birmingham Hip Resurfacing (BHR), the most commonly used MoMHRA worldwide. Since 1999, 648 BHRs were implanted in 555 patients, the majority of which were male (326). The mean age at surgery was 52.1years (range: 17–82), with primary OA as most common indication (85%). Mean follow up was 7.1years (range: 1–11). The Oxford Hip Score (OHS) and UCLA questionnaires were sent to all patients. Implant survival was established, with revision as the end point. Sub-analysis was performed by gender, femoral component size (small: <45mm, standard: 46–52, large: >53mm) and age at surgery (young:<50yrs, old:>50yrs).Introduction
Methods
Femoral neck narrowing (NN) following Metal-on-Metal Hip Resurfacing Arthroplasty (MoMHRA) is a well-recognised clinical phenomenon. The incidence of resurfaced hips with NN > 10% is reported to be up to 27%. Its pathogenesis is thought to be multi-factorial secondary to stress shielding, impingement, osteolysis secondary to wear/ion particles and as a result of reduced vascularity and pressure effect on cancellous bone secondary to the presence of a soft-tissue mass around the resurfaced hip. Recognised risk factors for its development include: female gender and the presence of a pseudotumour. Serum Chromium (Cr) and Cobalt (Co) are recognised surrogate markers of in-vivo wear of MoMHRA. The aims of this study were to establish whether NN is associated with increased wear. A cohort of 214 patients with unilateral MoMHRA (139M: 75F) was included in this study. Primary osteoarthritis was the diagnosis leading to surgery for the majority of patients (208). The average age at surgery was 54.1 years old (13–73). Six different implants were used; BHR (116), Conserve plus (92), Recap (2), ASR (2), Adept (1) and Cormet (1). The average femoral component size was 49.2mm (range: 38–59). The average follow up was 4.3 years (range: 2–10). Patients were subdivided into 3 groups as per implant size. Small size component group had implants <45mm, average size group had components 45–51mm and large component size group had components >51mm. All patients had Prosthesis-Junction-Ratio (PJR) measured from postoperative (PJRpost) and at latest follow up (PJRfollow) radiographs. Measurements were made using the method described by Lilikakis1. Metal ion levels (Cr/Co) were measured at last follow-up for all patients. Cr level >5.1g/ml and Co levels >4.4 g/ml were considered high2 and patients with such levels formed the high ion group.INTRODUCTION
METHODS
Studies have suggested that there is a reduction in head-neck-ratio (HNR) associated with MoMHRA. A reduction in HNR at operation would decrease range of movement and increase impingement risk. Impingement could lead to 20 edge loading, increasing wear. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Although acetabular component orientation has been shown to contribute to wear and PT development, the role of a decrease in HNR has only been highlighted in PT development. This study aimed to measure changes in HNR that occur at resurfacing and determine any gender- and component size-specific differences. In addition it aimed to determine whether changes in HNR could be associated with increased wear. 84 patients (56M: 28F) with unilateral MoMHRA were included. The mean age at surgery was 57 years. The mean femoral component was 49mm. Components were considered small if <45mm, average if between 45–50mm and large if >50mm. Three designs were implanted; BHR, C+ and Recap. The average follow up was 4 years. All patients had Cr/Co levels measured at follow up. Patients were considered to have high ions if Cr and Co levels were 5.1ppb and 4.4ppb respectively. Pre-operative HNR (HNRpre) and the post-operative HNR (HNRpost) were made from the respective pelvic radiographs. Assuming a 2mm thick cartilage layer, the HNR based on the diameter of the articular cartilage pre-operatively (HNRart) was calculated. The immediate changes in HNR as a result of the operation were expressed relative to articular HNR pre-op: HNRartpost=HNRpost–HNRartINTRODUCTION
METHODS
Although simulation studies have shown superior wear properties of metal-on-metal articulations, increased concern exists regarding the excess in-vivo wear of a small number of Metal-on-Metal-Hip-Resurfacing (MoMHRA) implants. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Risk factors associated with increased wear include female gender, small components, dysplasia, cup orientation outside safe zone and femoral head downsize during surgery with an associated decrease in Head-Neck-Ratio (HNR). However, these factors are interlinked. This study aims to identify the factors that are most important for subsequent wear of MoMHRA, by performing a multivariate analysis. 206 patients (124M: 82F) with unilateral MoMHRA were included in this study. The average follow up was 3.3 years. All patients had Cr/Co levels measured at follow up. Inclination and anteversion of each cup were measured using EBRA. Cups were analysed as being within or outside the previously defined optimum-zone. HNR measurements were made from pre-operative (HNRpre) and post-operative (HNRpost) radiographs. The immediate changes in HNR (downsize/upsize of femoral head) as a result of the operation were expressed as: HNRprepost=HNRpost–HNRpre Multivariate linear regression modelling was used to explore the association between measures of ions with the following predictor variables (gender, age, diagnosis, femoral component size, orientation of the acetabular component, head/neck ratio and position of femoral stem). Analyses were carried out separately for each outcome (Cr and Co). Classification and Regression Tree (CART) models were fitted as a complimentary approach to regression modelling.INTRODUCTION
METHODS
This study aims to investigate femoral blood flow during Metal-on-Metal Hip Resurfacing (MMHR) by monitoring oxygen concentration during the operative procedure. Patients undergoing MMHR using the posterior approach were evaluated. Following division of fascia lata, a calibrated gas-measuring electrode was inserted into the femoral neck, aiming for the supero-lateral quadrant of the head. Baseline oxygen concentration levels were detected after electrode insertion 2-3cm below the femoral head surface and all intra-operative measures were referenced against these. Oxygen levels were continuously monitored throughout the operation. Data from ten patients are presented. Oxygen concentration dropped most noticeably during the surgical approach and was reduced by 62% (Std.dev +/-26%) following dislocation and capsulectomy. Insertion of implants resulted in a further oxygenation decrease by 18% (Std.dev +/-28%). The last obtained measure before wound closure detected 22% (Std.dev +/-31%) of initial baseline oxygen levels. Variation between subjects was observed and three patients demonstrated a limited recovery of oxygen levels during implant insertion and hip relocation. Intra-operative measurement of oxygen concentration in blood perfusing the femoral head is feasible. Results in ten patients undergoing MMHR showed a dramatic effect on the oxygenation in the femoral head during surgical approach and implant fixation. This may increase the risk of avascular necrosis and subsequent femoral neck fracture. Future experiments will determine if less invasive procedures or specific positioning of the limb can protect the blood supply to femoral neck and head.
Metal on metal hip resurfacing (MMHR) is a popular procedure for the treatment of osteoarthritis in young patients. Several centres have observed masses, arising from around these devices, we call these inflammatory pseudotumours. They are locally invasive and may cause massive soft tissue destruction. The aim of this study was to determine the incidence and risk factors for pseudotumours that are serious enough to require revision surgery. In out unit, 1,419 MMHRs were performed between June 1999 and November 2008. All revisions were identified, including all cases revised for pseudotumour. Pseudotumour diagnosis was made by histological examination of samples from revision. A Kaplan-Meier survival analysis was performed, Cox regression analysis was used to estimate the independent effects of different factors. The revision rate for pseudotumour increased with time and was 4% (95% CI: 2.2% to 5.8%) at eight years. Female gender was a strong risk factor: at eight years the revision rate for pseudotumours in men was 0.5% (95% CI 0% to 1.1%), in women over 40 it was 6% (95% CI 2.3% to 10.1%) and in women under 40 it was 25% (95% CI 7.3% to 42.9%) (p<
0.001). Other factors associated with an increase in revision rate were, small components (p=0.003) and dysplasia (p=0.019), whereas implant type was not (p=0.156). We recommend that resurfacings are undertaken with caution in women, especially those younger than 40 years of age, but they remain a good option in men. Further work is required to understand the patho-aetiology of pseudotumours so that this severe complication can be avoided.
8 MoMHRA implants revised due to pseudotumour; 22 MoMHRA implants revised due to other reasons of failure (femoral neck fracture and infection). The linear wear of retrieved implants was measured using a Taylor-Hobson Roundness machine. The average linear wear rate was defined as the maximum linear wear depth divided by the duration of the implant in vivo.
significantly higher median linear wear rate of the femoral component: 8.1um/year (range 2.75–25.4um/year) vs. 1.79um/year (range 0.82–4.15um/year), p=0.002; and significantly higher median linear wear rate of the acetabular component: 7.36um/year (range1.61–24.9um/year) vs. 1.28um/year (range 0.18–3.33um/year), p=0.001. Similarly, differences were also measured in absolute wear values. The median absolute linear wear was significantly higher in the pseudotumour implant group:
21.05um (range 2.74–164.80um) vs. 4.44um (range 1.50–8.80um) for the femoral component, p=0.005; and 14.87um (range 1.93–161.68um) vs. 2.51um (range 0.23–6.04um) for the acetabular component, p=0.008. Wear on the acetabular cup components in the pseudotumour group always involved the edge, indicating edge-loading of the bearing. In contrast, edge-loading was observed in only one acetabular component in the non-pseudotumour group of implants. The deepest wear was observed well within the bearing surface for the rest of the non-pseudotumour group. The difference in the incidence of edge-loading between the two groups was statistically significant (Fisher’s exact test, p=0.03).