We have investigated the long-term (minimum follow-up period; 10 years) clinical results of the total hip arthroplasty (THA) using K-MAX HS-3 tapered stem. In K-MAX HS-3 THA (Kyocera Medical, Kyoto, Japan), cemented titanium alloy stem and all polyethylene cemented socket are used. This stem has the double tapered symmetrical stem design, allowing the rotational stability and uniform stress distribution. The features of this stem are; 1. Vanadium-free high-strength titanium alloy (Ti-15Mo-5Zr-3Al), 2. Double-tapered design, 3. Smooth surface (Ra 0.4µm), 4. Broad proximal profile, 5. Small collar. Previous type stem, which was made of the same smooth-surface titanium alloy, has the design with cylindrical stem tip, allowing the maximum filling of the femoral canal. Osteolysis at the distal end of the stem had been reported in a few cases in previous type with cylindrical stem tip, probably due to the local stress concentration. Therefore the tapered stem was designed, expecting better clinical results. 157 THAs using HS-3 taper type stem were performed at Kitano Hospital between March 2004 and March 2008. And 101 THAs, followed for more than 10 years, were investigated (follow-up rate; 64.3%). The average age of the patients followed at the operation was 61.7 years and the average follow-up period was 10.9 years. The all-polyethylene socket was fixed by bone cement, and the femoral head material was CoCr (22mm; 5 hips, 26 mm; 96 hips).Introduction
Materials and Methods
We have compared the middle-term (average follow-up period; 10 years) clinical results of the K-MAX HS-3 tapered stem with those of the previous type having cylindrical tip. In K-MAX HS-3 THA (Kyocera Medical, Kyoto, Japan), cemented titanium alloy stem and all polyethylene cemented socket are used. This stem has the double tapered symmetrical stem design, allowing the rotational stability and uniform stress distribution (Type T) (Fig. 1). The features of this stem are; 1. Vanadium-free high-strength titanium alloy (Ti-15Mo-5Zr-3Al), 2. Double-tapered design, 3. Smooth surface (Ra 0.4μm), 4. Broad proximal profile, 5. Small collar. In contrast, previous type stem, which was made of the same smooth-surface titanium alloy, has the design with cylindrical stem tip, allowing the maximum filling of the femoral canal (Type C) (Fig. 2). Osteolysis at the distal end of the stem had been reported in a few cases in Type C, probably due to the local stress concentration. Therefore the tapered stem was designed, expecting better clinical results. All surgery was performed at Kitano Hospital between September 2003 and June 2006. 72 THA were performed (Type T; 52 hips, Type C; 20 hips). The average age of the patients at the operation was 61 and 69 years and the average follow-up period was 10.1 and 10.4 years for the Type T and C, respectively. The all-polyethylene socket was fixed by bone cement, and the femoral head material was alumina or CoCr (22 or 26 mm).Introduction
Materials and Methods
IBBC (interfacial bioactive bone cement method, Oonishi) (1) is an excellent technique for augmenting cement-bone fixation in the long term. However, the technique is difficult and there are concerns over some points, such as bleeding control, disturbance of cement intrusion to anchoring holes by granules, difficulty of the uniform granular dispersion to the acetabular bone (Zone 1 in particular). To improve this technique, we have modified IBBC (M-IBBC), and investigated the short-term clinical results and radiographic changes. K-MAX HS-3 THA (Kyocera Medical, Japan), with cemented stem and all polyethylene cemented socket, was used for THA implants. Basically the third generation cementing technique was used for THA using bone cement. The socket fixation was performed with bone cement (Endurance, DePuy) and hydroxyapatite (HA) granules (Ca10(PO4)6(OH)2, Boneceram P; G-2, Olympus, Japan). In original IBBC technique, HA granules were dispersed on reamed acetabulum before cementing. In M-IBBC technique, HA granules were attached to bone cement on plastic plate, then inserted to reamed acetabulum and pressurized. HA granules (G-2) are 0.3–0.6mm in size, with 35–38% porosity and sintered at 1150Introduction
Materials and Methods
We have compared the short-term clinical results of total hip arthroplasty (THA) using PMMA bone cement and hydroxyapatite (HA) granules (interfacial bioactive bone cement method; IBBC) with the results of conventional method using PMMA bone cement. K-MAX HS-3 THA (JMM, Japan), with cemented titanium alloy stem and all polyethylene cemented socket, was used for THA implants. The third generation cement technique was used for the conventional THA (Group C) using bone cement (Endurance, DePuy). In the IBBC group (Group BC), the socket fixation was performed by the third generation cement technique with HA granules (Boneceram P; G-2, Olympus, Japan) according to the Ohnishi's method. In both groups, the stems were fixed by conventional cementing technique using cement gun. 76 hip joins (69 cases) were operated between April 2005 and August 2007, and followed. The group C (22 hips, 19 cases, average follow-up; 5.6 years, average age at operation; 64 years) and the group BC (54 hips, 50 cases, 5.4 years, 65 years) were investigatedPurpose
Materials and Methods