Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 98 - 98
1 Mar 2013
Kiyotomo D Sugamoto K Murase T Tomita T Kunugiza Y Kawashima K Futai K Kuramoto K Yamamoto K
Full Access

Introduction

Regarding TKA, patient specific cutting guides (PSCG), which have the same fitting surface with patient's bones or cartilages and uniquely specify the resection plane by fitting guides with bones, have been developed to assist easy, low cost and accurate surgery. They have already been used clinically in Europe and the USA. However little has been reported on clinical positioning accuracy of PSCG. Generally, the methods of making PSCG can be divided into 3 methods; construct 3D bone models with Magnetic Resonance (MR) images, construct 3D bone models with Computed Tomography (CT) images, and the last is to construct 3D bone models with both MR and CT images. In the present study, PSCG were made based on 3D bone models with CT images, examined the positioning accuracy with fresh-frozen cadavers.

Materials and Methods

Two fresh-frozen cadavers with four knees were scanned by CT. Image processing software for 3D design (Mimics Ver. 14, Marialise Inc.) was used to construct 3D bone model by image thresholding. We designed femoral cutting guides and tibial cutting guides by CAD software (NX 5.0, Siemens PLM Software Co.). CT free navigation system (VectorVision Knee, BrainLab, Inc.) was used to measure positioning error. Average absolute value of positioning error for each PSCG was derived.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 82 - 82
1 Mar 2013
Iwamoto K Tomita T Yamazaki T Shimizu N Kurita M Futai K Kunugiza Y Yoshikawa H Sugamoto K
Full Access

Background

Various postoperative evaluations using fluoroscopy have reported in vivo knee flexion kinematics under weight bearing conditions. This method has been used to investigate which design features are more important for restoring normal knee function. The objective of this study is to evaluate the kinematics of a Low Contact Stress total knee arthroplasty (LCS TKA) in weight bearing deep knee flexion using 2D/3D registration technique.

Patients and methods

We investigated the in vivo knee kinematics of 6 knees (4 patients) implanted with the LCS meniscal bearing TKA (LCS Mobile-Bearing Knee System, Depuy, Warsaw, IN). Mean period between operation and surveillance was 170.7±14.2 months. Under fluoroscopic surveillance, each patient did a deep knee flexion under weight-bearing condition. Femorotibial motion was analyzed using 2D/3D registration technique, which uses computer-assisted design (CAD) models to reproduce the spatial position of the femoral, tibial components from single-view fluoroscopic images. We evaluated the knee flexion angle, femoral axial rotation, and antero-posterior translation of contact positions.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 255 - 255
1 Mar 2013
Masahiro K Tomita T Yamazaki T Kunugiza Y Shimizu N Iwamoto K Sugamoto K
Full Access

INTRODUCTION

The outcome after total knee arthroplasty is influenced by the postoperative orientation of the component. For accurate implantation, the surgeon performs a three dimensional preoperative planning and performs the surgery with reference to the anatomical bony landmarks. However, the assessment of orientation after TKA is generally performed on two dimensional radiographs. Despite the accurate implantation, radiographic assessment may not able to accurately evaluate the orientation of the component. CT images obtain a three dimensional information after TKA, but reliable identification of the anatomical bony landmarks remains the problem due to artifacts of metal components. In this study, we evaluate the three dimensional orientation of the component relative to the bone axis of anatomical landmarks using pre- and post-operative CT scanning.

PATIENTS AND METHODS

Two knees after primary TKA were assessed by one observer using preoperative and postoperative CT images. 3D models of pre-operative bone and post-operative bone with the exclusion of component data were constructed. Surface-based registration was performed by independently implementing the iterative closest point algorithm with the least-squares method to match the pre-operative bone model with the post-operative bone model. 3D surface model of the metal component from postoperative CT images was constructed. 3D surface model of the metal component was superimposed on original computer-aided design (CAD) data of the component using surface-based registration. The registration of the metal component was performed three times. Intra-observer reliability of the superimposed CAD models was evaluated. The orientation of the component was measured in euler angle between the axis of the superimposed CAD model and the bone axis of anatomical landmark.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 173 - 173
1 Sep 2012
Shimizu N Tomita T Yamazaki T Kurita M Kunugiza Y Sugamoto K
Full Access

Background

Various postoperative evaluations using fluoroscopy have reported in vivo knee flexion kinematics under weight bearing conditions. This method has been used to investigate which design features are more important for restoring normal knee function. The objective of this study is to evaluate the kinematics of a Posterior-Stabilized TKA in weight bearing deep knee flexion using 2D/3D registration technique.

Patients and methods

We investigated the in vivo knee kinematics of 9 knees (9 patients) implanted with a Posterior Stabilized TKA (Triathlon PS, Stlyker Orthopedics, Mahwah, NJ). Under fluoroscopic surveillance, each patient did a deep knee flexion under weight-bearing condition. Femorotibial motion including tibial polyethylene insert were analyzed using 2D/3D registration technique, which uses computer-assisted design (CAD) models to reproduce the spatial position of the femoral, tibial components from single-view fluoroscopic images. We evaluated the knee flexion angle, femoral axial rotation, antero-posterior translation of contact points, and post-cam engagement were evaluated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 111 - 111
1 Jun 2012
Kawashima K Tomita T Yamazaki T Futai K Shimizu N Tamaki M Kurita M Kunugiza Y Sugamoto K
Full Access

Background

Mobile-bearing (MB) total knee prostheses have been developed to achieve lower contact stress and higher conformity compared to fixed-bearing total knee prostheses. However, little is known about the in vivo kinematics of MB prostheses especially the motion of the polyethylene insert (PE) during various daily performances. And the in vivo motion of the PE during stairs up and down has not been clarified. The objective of this study is to clarify the in vivo motion of MB total knee arthroplasty including the PE during stairs up and down.

Patients and methods

We investigated the in vivo knee kinematics of 11 knees (10 patients) implanted with PFC-Sigma RP-F (DePuy). Under fluoroscopic surveillance, each patient did stairs up and down motion. And motion between each component was analyzed using two- to three-dimensional registration technique, which used computer-assisted design (CAD) models to reproduce the spatial position of the femoral, tibial components, and PE (implanted with four tantalum beads intra-operatively) from single-view fluoroscopic images. We evaluated the range of motion between the femoral and tibial components during being grounded, axial rotation between the femoral component and PE, the femoral and tibial component, and the PE and tibial component during being grounded.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 122 - 122
1 Jun 2012
Kunugiza Y Tomita T
Full Access

We report long term outcomes of the Rotaglide mobile bearing total knee arthroplasty (RTK).

Method

Between 1995 and 1998, 61 RTK prostheses were implanted at our institution consecutively. Of 34 knees with a median duration to end of follow up of 13.0 years (range 11.4 to 14.2 years), the clinical result using Knee Society Score and radiological analysis using Knee Society Roentgenographic Sytstem was evaluated.

Result

The prosthesis had an estimated survival probability of 94.1% at 13 years. There was one case of deep infection and one case of meniscal component failure.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 123 - 123
1 Jun 2012
Kurita M Tomita T Fujii M Yamazaki T Kunugiza Y Futai K Kawashima K Shimizu N Sugamoto K
Full Access

Background

Mobile-bearing (MB) total knee prostheses have been developed to achieve lower contact stress and higher conformity compared to fixed-bearing total knee prostheses. However, little is known about the in vivo kinematics of MB prostheses especially about the kinematics of polyethylene insert (PE). In vivo motion of PE during squatting still remains unclear. The objective of this study is to investigate the in vivo motion of MB total knee arthroplasty including PE during squatting.

Patients and methods

We investigated the in vivo knee kinematics of 11 knees (10 patients) implanted with Vanguard Rotationg Platform High Flex (Biomet(r)). Under fluoroscopic surveillance, each patient did a wight-bearing deep knee bending motion. Motion between each component was analyzed using two- to three-dimensional registration technique, which uses computer-assisted design (CAD) models to reproduce the spatial position of the femoral, tibial components, and PE (implanted with five tantalum beads intra-operatively) from single-view fluoroscopic images. We evaluated the range of motion between the femoral and tibial components, axial rotation between the femoral component and PE, the femoral and tibial component, and the PE and tibial component, and AP translation of the nearest point between the femoral and tibial component and between the femoral component and PE.