Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 35 - 35
1 May 2016
McTighe T Brazil D Keggi J Keppler L McPherson E
Full Access

Over the past 10 years, the orthopedic community has witnessed an increased interest in more conservative surgical techniques for hip arthroplasty. During this time, second-generation hip resurfacing and minimally invasive surgery enjoyed extensive marketing attention. After a decade of this renewed interest, both of these methods have met with serious concerns. As hip resurfacing numbers decline, both patients and surgeons are looking for other potentially successful conservative treatments to THA. This search has focused surgeon interest toward short-stem designs.

Today, a variety of short-stem implants are available with very little clarification of design rationale, fixation features, surgical technique, and clinical outcomes. Virtually every major implant company now offers a “short stem,” and now there are a plethora of different designs. It is important to note, however, that not all short stems achieve initial fixation at the same bone interface region. Furthermore, surgical techniques vary greatly, and postoperative radiographic interpretation of short-stem position and fixation need to be carefully scrutinized.

The purpose of this paper is to review past, present, and potential future developments of short femoral stems and to present a classification system that can offer guidance when reporting on the many different stem variations.

Short Curved neck-sparing stem (JISRF classification 2a). Recently, new designs are following Pipino's Biodynamic stem style of saving the femoral neck. These designs feature a short curved stem that finds its stabilization contact region in the femoral neck and saves considerable bone in the medial calcar region. In addition, the curvature of the stem prevents violation of the lateral trochanteric region. The shorter stem also reduces blood loss by not reaming the femoral canal distally. These style stems generally have a variable stem length between 90 and 135 mm. This might not appear much shorter than conventional cementless stems (110 to 150 mm). However, the shorter curved neck sparing stems penetrate on average 1 to 2 cm less distally in the femoral canal.

Short stems have a definite role in modern THA, as greater emphasis is being placed on soft tissue and bone sparing techniques and as refinements continue in the understanding of proximal femoral fixation.

Metaphyseal short stems have significantly less surface contact area compared with conventional length stems and as a result, they might have less torsional and axial resistance.

Neck-retaining short stems provide additional axial and torsional stability and reduced stress at the implant– bone interface and may be a consideration in the more active patient profile. Bone quality and the patient's physical activity should be considered prior to the selection of short-stem devices. Many short-stem designs have considerably different style features that may alter bone remodeling. Knowing the design and the required technique is vital to fit the device properly to the patient. The variations of short stems available call for caution in their overall use until there is better understanding of how dependent these stems are on individual stem features, bone quality, and surgical techniques. Overall, the authors are cautiously optimistic and continue advocating their selective use.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 397 - 397
1 Dec 2013
Levy J Keppler L Verborgt O Declercq G Frankle M
Full Access

Background and Motivation

Accurate placement of glenoid components in reverse and total shoulder arthroplasty has been shown to reduce the risk of implant failure (1, 2, 6). Surgical techniques and literature describe methods to determine favorable positions for implant placement (3, 4, 5) but achieving that position surgically remains a challenge. Placement of glenoid components is faced with the challenge of variable glenoid morphology on which conventional instrumentation does not always provide a reliable reference (6, 7, 8). Limited surgical exposure is another challenge since many anatomic landmarks are not visible to the surgeon to use as spacial reference. Anatomic landmarks and angles can be more reliabily selected on CT scans with 3-dimentional reconstruction (9,10) yet few methods allow for the reproducible translation of these plans to surgery. Navigation has produced better accuracy and lower variability than conventional instrumentation (11), yet its regular usage remains limited, especially in the shoulder.

Methods

A patient specific planning and guiding system has been developed for glenoid implant placement of total and reverse shoulder arthoplasty procedures. This method allows for preoperative planning on a patient specific virtual 3D model of the scapula derived from CT images (Figure 1), and guided placement of a pin which which serves as the central axis for determining proper implant position. An initial implant position was presented on the virtual model based on the methods described by the surgical technique of the corresponding procedure. These plans were either approved or adapted to a desired position within the planning software by the surgeons. Using this planned position as input, patient specific surgical guides were created which fit onto the exposed anatomy and guide the drilling of the pin (Figure 1). This method was tested on 14 cadavers, with attention directed to translation of the starting point from the original plan, the ability to reproduce the intended degree of inferior tilt, and the ability to reproduce the glenoid version angle.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 259 - 259
1 Mar 2013
McTighe T Keggi J Keppler L Aram T Bryant C Ponder C Vaughn BK McPherson E
Full Access

Introduction

Architectural changes occurring in the proximal femur after THA continues to be a problem. Stress shielding occurs regardless of fixation method. The resultant bone loss can lead to implant loosening and breakage of the implant. A new novel tissue sparing neck-stabilised stem has been designed to address these concerns.

Methods

Over 1,200 stems have been implanted since April 2010 and 2012. Patient profile showed two-thirds being female with an age range between 17 to early 90s. 90% were treated for OA. This stem has been used in all Dorr bone classification (A, B, & C). Two surgical approaches were utilised (single anterior incision and standard posterior incision). All were used with a variety of cementless acetabular components and a variety of bearing surfaces (CoC, CoP, MoM, MoP). Complications were track by surgeon Members of the Tissue Sparing Study Group of the Joint Implant Surgery and Research Foundation. Complications include first year of limited clinical release. No surgeon was permitted usage without specific cadaver / surgical training. No head diameters below 32 mm were used.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 260 - 260
1 Mar 2013
McTighe T Stulberg SD Keppler L Keggi J Kennon R Aram T McPherson E
Full Access

Introduction

The use of short stems has been growing in THA for the past five years. As a result, a large number of short stem designs are available in the market place. However, fixation points differ for many of the designs resulting in different radiographic modeling creating confusion when trying to collate to clinical findings. We have created a classification system in an attempted to provide clarity in analyzing radiographic and clinical findings.

Method

Femoral implants described as “short stems” were evaluated. The range of lengths for stem type and the method of achieving initial implant stability was determined. The optimal radiographic position of each of these implants and type of bone remodeling associated with this placement was evaluated. Stems were defined as “short” if the tip reached or was proximal to the metaphyseal-diaphyseal junction. This location on the proximal femur was defined as the place at which the medial-lateral metaphyseal flare became parallel. Stems were then classified as: 1.) Metaphyseal Stabilized; 2.) Neck Stabilized; 3.) Head Stabilized. An analysis of radiographic with a minimum of one year follow up were reviewed and posted as to the classification system


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 22 - 22
1 Oct 2012
Victor J Premanathan A Keppler L Deprez P Bellemans J
Full Access

Eight consecutive patients with significant malalignment of the lower limb were included in the study. Pre-operative CT scans of the affected limb and the normal contra-lateral side were obtained and 3D models of the patient's anatomy were created, using dedicated software. The healthy contralateral limb was mirrored and geometrically matched to the distal femur or proximal tibia of the healthy side. A virtual opening wedge correction of the affected bone was used to match the geometry of the healthy contralateral bone. Standard lower limb axes measurements confirmed correction of the alignment. Based on the virtual plan, surgical guides were designed to perform the planar osteotomy and achieve the planned wedge opening and hinge axis orientation. The osteotomy was fixed with locking plates and screws. Post-operative assessment included planar X-rays, CT-scan and full leg standing X-rays.

One three-planar, three bi-planar and four single-plane osteotomies were performed. Maximum weightbearing mechanical femoro-tibial coronal malalignment varied between 7° varus and 14° valgus (mean 7.6°, SD 3.1). Corrective angles varied from 7°–15°(coronal), 0°–13°(sagittal) and 0°–23°(horizontal). The maximum deviation between the planned pre-operative wedge angle and the executed post-operative wedge angle was 1° in the coronal, sagittal and horizontal plane. The desired mechanical femorotibial axis on full-leg standing X-rays was achieved in 6 patients. Two patients were undercorrected by 1° and 2° respectively.

Conclusion

3D planning and guided correction of multi-planar deformity of femur or tibia is a feasible and accurate novel technique.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 146 - 146
1 Sep 2012
Premanathan A Victor J Keppler L Deprez P Bellemans J
Full Access

Background

Osteotomies around the knee have been used to correct lower limb mal-alignment for over 50 years. The procedure is technically demanding and carries specific risks of neurovascular injury, incorrect planning and execution, and insufficient fixation. In recent years, with the advent of locking plates, fixation techniques have improved significantly but the correct planning and execution of the operation remains difficult. Despite the availability of CT and MRI 3D imaging, surgical planning is still traditionally performed on 2D plain X-rays [1]. Especially with multi-planar deformities, this technique is prone to error. The aim of this clinical pilot study is to evaluate the feasibility of virtual pre-operative three-dimensional planning and correct execution of osteotomies around the knee with the aid of patient specific surgical guides and locking plates.

Patients and methods

Eight consecutive patients, presenting with significant malalignment of the lower limb were included in the study. Pre-operative CT scans of the affected limb and the normal contra-lateral side were obtained and 3D models of the patient's anatomy were created, using dedicated software (Mimics® 3-matic®, Materialise, Leuven Belgium) [2]. These models were used to evaluate the required surgical correction. The healthy contralateral limb was mirrored and geometrically matched to the distal femur or proximal tibia of the healthy side. A virtual opening wedge correction of the affected bone was used to match the geometry of the healthy contralateral bone. Standard lower limb axes measurements confirmed correction of the alignment [3]. Based on the virtual plan, surgical guides were designed to perform the planar osteotomy and achieve the planned wedge opening and hinge axis orientation (see figure 1). Apart from guiding the osteotomy, the patient specific surgical guide also guided drilling of the planned screw holes. Post-operative assessment of the correction was obtained through planar X-rays, CT-scan and full leg standing X-ray.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 208 - 208
1 Sep 2012
Victor J Premanathan A Keppler L Deprez P Bellemans J
Full Access

Introduction

Osteotomies around the knee are traditionally templated on 2D plain X-rays. Results are often inaccurate and inconsistent and multiplanar ostetomies are hard to perform. The aim of this study is to evaluate the feasibility and accuracy of virtual three-dimensional CT-based planning and correct execution of osteotomies around the knee with the aid of patient specific surgical guides and locking plates.

Methods

Eight consecutive patients with significant malalignment of the lower limb were included in the study. Pre-operative CT scans of the affected limb and the normal contra-lateral side were obtained and 3D models of the patient's anatomy were created, using dedicated software. The healthy contralateral limb was mirrored and geometrically matched to the distal femur or proximal tibia of the healthy side. A virtual opening wedge correction of the affected bone was used to match the geometry of the healthy contralateral bone. Standard lower limb axes measurements confirmed correction of the alignment. Based on the virtual plan, surgical guides were designed to perform the planar osteotomy and achieve the planned wedge opening and hinge axis orientation. The osteotomy was fixed with locking plates and screws. Post-operative assessment included planar X-rays, CT-scan and full leg standing X-rays.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 69 - 69
1 May 2012
Keppler L McTighe T
Full Access

THA continues to improve but complications still occur. Improper restoration of hip mechanics can lead to a number of clinical problems: increase in leg length, soft tissue laxity, weakness of the abductors, mechanical impingement, increase of wear and improper implant sizing can lead to thigh pain, subsidence and hip dislocation.

Six-hundred-and-fifty-five primary cementless THA were performed over the past twenty-four years by the senior author at two hospitals. Three different stems were used, two being modular and one being monoblock. A variety of cups head sizes and bearing material were used. All cups were implanted cementless. All surgeries were performed with the posterior approach. Sixty percent of patients were female forty percent males. Majority of cases were for OA.

Cup revisions have been the biggest problem to-date with excessive wear of the poly material. This is more than likely due to the first and second generation designs that had poor locking mechanics. Over the last four years since going to MOM technology cup revisions have not been seen. On the femoral side there have been no femoral lysis, five dislocations two treated closed and three open reductions treated with constrained sockets. Four stem revisions, all for late sepsis. There has been two recent aseptic loosening, and only one traumatic dislocation since going to large MOM heads.

One was one post-op with an ASR MOM cup that had spun out of position of function and the second a week later that was only six weeks post-op and came in for her first post-operative visit.

Routine use of intra-operative x-rays has resulted in +80% decision on fine-tuning of implant sizing by either increasing stem size and or femoral offset. Intra- operative x-rays provide valuable assistance and allow full advantage of the features and benefits of stem modularity reducing post-operative complications.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 221 - 221
1 May 2012
Keppler L McTighe T
Full Access

Total hip arthroplasty has a very high success rate, as measured by pain relief, improved function and patient satisfaction. However, on occasion, complications do happen. This paper will review three cases that required surgical intervention and design features of a modular stem making revision surgery considerably easier. Three patients received cementless THA within the past two years. All three had a ‘Dual Pressâ’ proximal modular stem design. Two patients had metal-on-metal (MOM) bearings that shifted position and one had a cementless porous cup with a 36 mm poly bearing with metal head that dislocated. All were performed at the same hospital by the senior author using a small posterior surgical approach.

All three required revision surgery that was made considerably easier by the design feature of proximal Dual Pressâ modularity.

The Dual Pressâ modular junction has a novel design that allows for retrievability in just these types of cases. A proximal setscrew is removed allowing access of a slap hammer, making the proximal neck disengage from the stem body. This greatly improves exposure for removal and implantation of the acetabular component. In both cup revisions, the explant cup removal system was used without difficulty, as a result of increased exposure achieved by removal of the neck portion of the stem. Then a new proximal neck segment with head was attached with no disruption of the stem/bone interface.

The dislocation case was addressed by removal of the proximal modular neck body, then increased femoral offset and 13° angle added to the new neck position. The cup, liner and stem body were not changed at all. Patient was stable (hip closed) and discharged the next day.

Modern day designs in implants and instrumentation have made THA more reproducible and in cases of revision surgery provide design features and benefits that reduce operative time, complications and as a result offer significant cost savings as compared to traditional monoblock stem designs.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 67 - 67
1 May 2012
Bowsher J Nelson P Clarke PI McTighe T Woodgate I Turnbull A Keppler L Donaldson T
Full Access

Hip simulator studies on MOM bearings have historically involved ‘custom’ cetabular cups. I.e. having neither beaded layers nor biological coatings.

The aim of this study was to investigate wear using such MOM bearings and evaluate the potential wear and evaluate the potential for error in the gravimetric assessment.

Six x 38 mm HC Co-Cr bearings were supplied (Global and IO International Orthopaedics). The cups were received in ‘off-the-shelf’ condition with a cast Co-Cr beaded/HA-coated backing. To remove the HA-coating, the cups were pre-soaked in lemon juice for 4 days (articular surfaces shielded). Custom plastic fixtures were machined to fit the beaded contours of the cups. Test duration was 5Mc inorbital hip simulator (Shore-Western). MOM wear was estimated from serum ion contamination. Serum samples were digested and assessed using ICP/MS (Weck Labs Inc, CA).

The majority of the HA-coating was removed from the cups after four days of soaking inlemon juice after 21 days of soaking all cup weights appeared atable (within 1 mg). Reflected-light microscopy (RLM) showed no descernible signs of HA and the total weight loss due to HA remval averaged∼400mg.

During hip simulator there was no visual evidence of lost or broken beads, 3rd body abrasion etc (Sa<30nm). Both gravimetric and metal ion analysis showed consistent wear trends for all MOM cups. The MOM with the highest wear (predicted by ion analysis) demonstrated 1.2 mm (3)/Mc)OWR) at 5Mc. In comparsion, gravimetric analysis predicted an OWR of 1.3 mm (3)Mc for the same MOM, a difference of only 8%. Soaking beaded-HA cups in lemon juice and BCS proved effective in removing the coating.

The beaded cups remained stable in weight during the wear study and caused little discrepancy in gravimetric analysis (8%).

The method described did not lead to breaking of beads, elevated 3rd-body abrasion, cup damage or distorted wear scars.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 105 - 105
1 Mar 2010
Cameron HU McTighe T Woodgate I Turnbull A Harrison J Keggi J Kennon R Keppler L Brazil D
Full Access

Introduction: Architectural changes in the proximal femur after THA continue to be a problem. In an attempted, to reduce these changes some surgeon designers have advocated the concept of neck sparing stem designs.

To-date neck-sparing stems have been disappointing in their ability to maintain the calcar. A new approach was undertaken to improve load transfer and to create a tissue-sparing stem that would be simple in design, reproducible in technique and provide for fine-tuning joint mechanics while maintaining compressive loads to the calcar.

Methods: Review of previous published work was evaluated along with FEA modeling in creating a new approach to neck sparing stems for THA. The MSA Stem is a simple curved stem with a unique lateral T-back designed for torsional stability, ease of preparation and insertion. The proximal design has a novel proximal conical shape designed to transfer compressive forces to the calcar.

A modular neck provides for fine-tuning joint mechanics.

Results: FEA modeling will be reviewed. Strain patterns for the MSA stem demonstrated better patterns vs. long stems or the short Biodynamic stem.

Discussion: In theory neck retaining devices provide or:

Bone and Tissue sparring

Restoration of joint mechanics

Minimal blood loss

Potential reduction in rehabilitation

Ease of revision

Simple surgical technique

Options for bearing surface

Selection of femoral head diameter

Standard surgical approach to the hip

We are encouraged and believe there are advantages in the concept of neck sparing stems. Clinical/surgical evaluation is now underway and will be reported on in the future.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_II | Pages 321 - 321
1 May 2006
McTighe T Keppler L Cameron H
Full Access

Concern was expressed that the use of a modular stem might produce fretting leading to osteolysis, and component failure. The goal of this study was to document the variability of this design by looking at the long-term i.e. 5–17 year follow-up of the use of a Proximal Modular Stem in primary cases.

A cohort of 955 (S-Rom) primary cases have been followed prospectively and rated clinically using the Harris Hip Score and radiologically after the fashion of Gruen. The mean age was 53. Follow-up was 5–17 years (mean 8.5).

Aseptic loosening requiring revision occurred in three cases (0.3%). One a non-union of a subtrochanteric osteotomy. Two others, one for fracture at the stem tip and one for fracture of the proximal part of a subtrochanteric osteotomy. Harris rating was 78.2% excellent, 16.4% good, 2.3% fair and 3.1% poor. Gruen rating, no lucency in 98.8%, low grade in 1.1% and high grade in 0.1%. Distal osteolysis occurred in two cases. Six patients had persistent thigh pain (type C bone) that was treated by onlay-strut graphs.

There have been no cases of device failures. Other than in the two loose cases distal osteolysis has not been seen. It would appear therefore that the sleeve does act as an adequate seal. There have been no cases of late aseptic loosening and limited thigh pain in type C bone. The authors concluded that this modular device is safe, effective and continue to recommend its use in primary THA.