Abstract
Background
Osteotomies around the knee have been used to correct lower limb mal-alignment for over 50 years. The procedure is technically demanding and carries specific risks of neurovascular injury, incorrect planning and execution, and insufficient fixation. In recent years, with the advent of locking plates, fixation techniques have improved significantly but the correct planning and execution of the operation remains difficult. Despite the availability of CT and MRI 3D imaging, surgical planning is still traditionally performed on 2D plain X-rays [1]. Especially with multi-planar deformities, this technique is prone to error. The aim of this clinical pilot study is to evaluate the feasibility of virtual pre-operative three-dimensional planning and correct execution of osteotomies around the knee with the aid of patient specific surgical guides and locking plates.
Patients and methods
Eight consecutive patients, presenting with significant malalignment of the lower limb were included in the study. Pre-operative CT scans of the affected limb and the normal contra-lateral side were obtained and 3D models of the patient's anatomy were created, using dedicated software (Mimics® 3-matic®, Materialise, Leuven Belgium) [2]. These models were used to evaluate the required surgical correction. The healthy contralateral limb was mirrored and geometrically matched to the distal femur or proximal tibia of the healthy side. A virtual opening wedge correction of the affected bone was used to match the geometry of the healthy contralateral bone. Standard lower limb axes measurements confirmed correction of the alignment [3]. Based on the virtual plan, surgical guides were designed to perform the planar osteotomy and achieve the planned wedge opening and hinge axis orientation (see figure 1). Apart from guiding the osteotomy, the patient specific surgical guide also guided drilling of the planned screw holes. Post-operative assessment of the correction was obtained through planar X-rays, CT-scan and full leg standing X-ray.
Results
One three-planar, three bi-planar and four single-plane osteotomies were performed. All guides could be used during surgery and served accurate guidance of the osteotomy plane and screwholes. The guides matched the bone very well in all cases without remaining toggle. The maximum deviation between the planned pre-operative wedge angle and the executed post-operative wedge angle was 1° in the coronal, sagittal and horizontal plane. The desired mechanical femorotibial axis on full-leg standing X-rays was achieved in 6 patients. Two patients were undercorrected by 1° and 2° respectively. No significant peri-operative complications occurred.
Conclusion
3D planning and guided correction of multi-planar deformity of femur or tibia is a feasible and accurate novel technique.