Advertisement for orthosearch.org.uk
Results 1 - 6 of 6
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 97 - 97
19 Aug 2024
Ilo K Van Duren BH Higgins MA Manktelow ARJ Bloch BV
Full Access

We report the impact of implementing a new short-stay hip and knee arthroplasty pathway in a National Health Service (NHS) hospital. This was enacted due to existing concerns with a long length of stay (LOS) and reduced elective operating capacity each winter due to emergency bed pressures. The overnight introduction of this pathway was aimed to reduce LOS, alleviate bed pressures, minimise readmission rates and generate financial savings, all combining to facilitate full elective activity during the winter.

We conducted a prospective study at a regional tertiary arthroplasty centre. The new pathway was introduced across the service overnight. It includes rigorous preoperative optimisation, specific anaesthetic protocols and uniform changes in surgical practice to allow a focus on early mobilization and discharge on the day of surgery where possible.

Data collection spanned 17 months, including the initial six months post implementation of the short-stay pathway. LOS data was collected for the full period and data was compared pre- and post-implementation of the new pathway. Patient satisfaction and 30-day readmission data were also collected.

There was an immediate and significant decrease in median LOS from 4 days pre-implementation to 1 day post-implementation. Patient satisfaction was high, and the 30-day readmission rate was unchanged (5.95%), with no readmissions directly related to decreased inpatient stay. Financial analyses revealed substantial cost savings due to reduced LOS and the elimination of routine post-operative blood tests, estimated at over £1.6m per year. Elective activity over winter was significantly higher (79%) than in the same time period in the previous year.

An acute introduction of a carefully planned and coordinated short-stay hip and knee pathway is safe, cost-effective and popular with patients, but also contributes to increased efficiency in the delivery of elective healthcare in the context of increasing demand and financial constraints in the NHS.


Bone & Joint Open
Vol. 4, Issue 5 | Pages 357 - 362
17 May 2023
Naathan H Ilo K Berber R Matar HE Bloch B

Aims

It is common practice for patients to have postoperative blood tests after total joint replacement (TJR). However, there have been significant improvements in perioperative care with arthroplasty surgery, and a drive to reduce the length of stay (LOS) and move towards day-case TJR. We should reconsider whether this intervention is necessary for all patients.

Methods

This retrospective study included all patients who underwent a primary unilateral TJR at a single tertiary arthroplasty centre during a one-year period. Electronic medical records of 1,402 patients were reviewed for patient demographics, LOS, and American Society of Anesthesiologists (ASA) grade. Blood tests were examined to investigate the incidence of postoperative anaemia, electrolyte abnormalities, and incidence of acute kidney injury (AKI).


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 33 - 33
1 Jul 2022
Ilo K Lodge C Berber R Matar H Bloch B
Full Access

Abstract

Introduction

It is common practice for all patients to have postoperative bloods tests after total knee arthroplasty (TKA). However there have been significant improvement in perioperative care with arthroplasty surgery. As healthcare systems and providers strive towards a more efficient service we must re-consider whether such an intervention is necessary for all patients.

Methodology

This retrospective study included all patients who underwent a primary unilateral TKA at a single tertiary arthroplasty centre during a 1 year period. Electronic medical records of 776 patients were reviewed for patient demographics, length of stay and American Society of Anaesthesiologists grade. Blood tests were examined to investigate the incidence of post operative anaemia requiring transfusion, electrolyte abnormalities that required treatment and the incidence of acute kidney injury.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 547 - 548
1 Nov 2011
Underwood R Cann P Ilo K Wagner C Skinner J Cobb J Porter M Muirhead-Allwood S Hart A
Full Access

Introduction: The London Implant Retrieval Centre (LIRC) was founded to investigate the high incidence of unexplained failures of Metal-on-Metal (MoM) hips. A multidisciplinary team analyse the failed hips, investigations include CT and MRI scans, blood and synovial fluid tests, wear measurements, X-rays and clinical data from the explanting surgeons.

Wear measurements of 100 explanted hips have been carried out on a Taylor Hobson 365 Roundness Machine using the LIRC Wear Protocol. It was found that 50% of explanted cups were wearing less than 5 μm/year and 60% of components were wearing less than 10 μm/year. Wear tests on hip joint simulators predict wear rates between 2 and 8 μm/year. However, 6% of cups are wearing faster than 100 μm/year, with 16% of cups have wear patches deeper than 100 μm and that 4% have a wear patch deeper than 300 μm.

Discussion: This paper considers the common characteristics of components in this very high wearing category. Engineering parameters such as head/cup clearance, surface finish, form errors and head cup contact conditions are investigated. This is correlated with clinical data and other results from the LIRC.

Cup position is an important factor, all of the high wearing components are outside the Lewinick’s Box, however it is shown that mal position is does not always lead to extreme wear. Further analysis is taking place to calculate the size of the contact patch between head and cup (based on patient data and biomechanics) and the proximity of the contact patch to the edge of the cup.

Conclusion: The study of explanted components shows that 6% exhibit extreme wear, and although several “risk” factors can be identified, it is not clear why only a proportion of these components show extremely high wear rates. This is the subject of current investigation.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 315 - 320
1 Mar 2011
Hart AJ Ilo K Underwood R Cann P Henckel J Lewis A Cobb J Skinner J

We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine.

We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The acetabular inclination angle was strongly positively correlated with the rate of wear. Femoral version was weakly negatively correlated with the rate of wear. Application of a threshold of > 5 μm/year for the rate of wear in order to separate the revisions into low or high wearing groups showed that more high wearing components were implanted outside Lewinnek’s safe zone, but that this was mainly due to the inclination of the acetabular component, which was the only parameter that significantly differed between the groups.

We were unable to show that excess version of the acetabular component alone or combined with femoral version was associated with an increase in the rate of wear based on our assessment of version using CT.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 307 - 314
1 Mar 2011
Matthies A Underwood R Cann P Ilo K Nawaz Z Skinner J Hart AJ

This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR).

There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared.

Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.

Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur.

These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.