header advert
Results 1 - 4 of 4
Results per page:
Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_9 | Pages 15 - 15
1 May 2018
Bhangoo N White J Howard P
Full Access

Aims

Thromboprophylaxis following Total Hip Replacement (THR) surgery remains controversial, balancing VTE prevention against wound leakage and subsequent deep infection.

We analysed the 90 day cause of death post THR in our institution after the implementation of new thromboprophylactic policy of low dose aspirin for low risk patients, as part of a multimodal regime. Those at high risk were anticoagulated.

Patients and methods

The PAS database was used to identify patients undergoing primary THR between January 2012 and June 2017 at The Royal Derby Hospital, and all deaths within 90 days. Trauma cases were excluded. Case note review and Coroner's verdict were utilised to ascertain cause of death. Results were compared to a previous study at the same institution prior to the introduction of the new policy, where thromboprophylaxis was decided upon by surgeon preference for either LMWH, aspirin or warfarin.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 33 - 41
1 Jan 2018
Matharu GS Hunt LP Murray DW Howard P Pandit HG Blom AW Bolland B Judge A

Aims

The aim of this study was to determine whether the rates of revision for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle components varied according to the year of the initial operation, and compare these with the rates of revision for other designs of MoM THA.

Patients and Methods

Data from the National Joint Registry for England and Wales included 36 mm MoM THAs with Pinnacle acetabular components which were undertaken between 2003 and 2012 with follow-up for at least five years (n = 10 776) and a control group of other MoM THAs (n = 13 817). The effect of the year of the primary operation on all-cause rates of revision was assessed using Cox regression and interrupted time-series analysis.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 50 - 50
1 Jun 2017
Bolland B Cook E Tucker K Howard P
Full Access

This study utilized data from the NJR dataset on all Corail/Pinnacle total hip replacements (THR) to determine (a) the level of unit variation of the Corail/Pinnacle 36mm Metal On Metal THR within England and Wales; (b) patient, implant and surgeon factors that may be associated with higher revision rates; (c) Account for the influence of the MHRA announcement in 2010.

The national Revision Rate (RR) for the Corail / Pinnacle MOM THR was 10.77% (OR:1.46; CI:1.17–1.81). This was significantly greater than other articulation combinations (MOP 1.72%, COP 1.36%, COC 2.19%).

The 2010 MHRA announcement did not increase rate of revision (X2=1649.63, df=13, p<.001).

Patient factors associated with significantly increased revision rates included, female gender (OR 1.38 (CI 1.17–1.63, p<.001) and younger age OR 0.99 (CI 0.98–0.99), p<.001). Implant factor analysis demonstrated an inverse relationship between cup size and revision. As head length increased RR increased – highest risk of revision +12.5 (OR 1.69 (CI 1.12–2.55), p=0.13). Coxa vara, high offset stems had a higher risk of revision compared to standard offset stems (OR:1.41 (CI 1.15–1.74; p<.001). As stem size increased risk of revision decreased (OR 0.89 (CI 0.85–0.93); p<.001). Surgeon grade did not influence RR.

There was significant variation in RR between hospitals with 7 units (7/61 excluding low volume centres, <50 implants) identified as having significant higher rates of revision. However, for each of these units there was a greater proportion of higher risk patients (female, cup size 50–54, stem type).

This study has provided insight into unit variation, risk factors and the long term outcome of the Corail/Pinnacle 36mm MOMTHR. Future aims are to use these results to develop a risk stratified algorithm for the long term follow of these patients to minimize patient inconvenience and excess use of limited NHS resources.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 19 - 19
1 Jun 2017
Howard D Wall P Fernandez M Parsons H Howard P
Full Access

Ceramic on ceramic (CoC) bearings in total hip arthroplasty (THA) are commonly used but concerns exist regarding ceramic fracture. This study aims to report the risk of revision for fracture of modern CoC bearings and identify factors that might influence this risk, using data from the National Joint Registry.

We analysed data on 111,681 primary CoC THA's and 182 linked revisions for bearing fracture recorded in the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR). We used implant codes to identify ceramic bearing composition and generated Kaplan-Meier estimates for implant survivorship. Logistic regression analyses were performed for implant size and patient specific variables to determine any associated risks for revision.

99.8% of bearings were CeramTec Biolox® products. Revisions for fracture were linked to 7 of 79,442 (0.009%) Biolox® Delta heads, 38 of 31,982 (0.119%) Biolox® Forte heads, 101 of 80,170 (0.126%) Biolox® Delta liners and 35 of 31,258 (0.112%) Biolox® Forte liners. Regression analysis of implant size revealed smaller heads had significantly higher odds of fracture (χ2=68.0, p<0.0001). The highest fracture risk were observed in the 28mm Biolox® Forte subgroup (0.382%). There were no fractures in the 40mm head group for either ceramic type. Liner thickness was not predictive of fracture (p=0.67). BMI was independently associated with revision for both head fractures (OR 1.09 per unit increase, p=0.031) and liner fractures (OR 1.06 per unit increase, p=0.006).

We report the largest registry study of CoC bearing fractures to date. Modern CoC bearing fractures are rare events. Fourth generation ceramic heads are around 10 times less likely to fracture than third generation heads, but liner fracture risk remains similar between these generations.