Abstract
This study utilized data from the NJR dataset on all Corail/Pinnacle total hip replacements (THR) to determine (a) the level of unit variation of the Corail/Pinnacle 36mm Metal On Metal THR within England and Wales; (b) patient, implant and surgeon factors that may be associated with higher revision rates; (c) Account for the influence of the MHRA announcement in 2010.
The national Revision Rate (RR) for the Corail / Pinnacle MOM THR was 10.77% (OR:1.46; CI:1.17–1.81). This was significantly greater than other articulation combinations (MOP 1.72%, COP 1.36%, COC 2.19%).
The 2010 MHRA announcement did not increase rate of revision (X2=1649.63, df=13, p<.001).
Patient factors associated with significantly increased revision rates included, female gender (OR 1.38 (CI 1.17–1.63, p<.001) and younger age OR 0.99 (CI 0.98–0.99), p<.001). Implant factor analysis demonstrated an inverse relationship between cup size and revision. As head length increased RR increased – highest risk of revision +12.5 (OR 1.69 (CI 1.12–2.55), p=0.13). Coxa vara, high offset stems had a higher risk of revision compared to standard offset stems (OR:1.41 (CI 1.15–1.74; p<.001). As stem size increased risk of revision decreased (OR 0.89 (CI 0.85–0.93); p<.001). Surgeon grade did not influence RR.
There was significant variation in RR between hospitals with 7 units (7/61 excluding low volume centres, <50 implants) identified as having significant higher rates of revision. However, for each of these units there was a greater proportion of higher risk patients (female, cup size 50–54, stem type).
This study has provided insight into unit variation, risk factors and the long term outcome of the Corail/Pinnacle 36mm MOMTHR. Future aims are to use these results to develop a risk stratified algorithm for the long term follow of these patients to minimize patient inconvenience and excess use of limited NHS resources.