Total knee replacement is mostly done with alignment rods in order to achieve a proper Varus / Valgus alignement. Other techniques are computer assisted navigation or MRI based preoperative planning. iASSIST™ is a computer assisted stereotaxic surgical instrument system to assist the surgeon in the positioning of the orthopaedic implant system components intra-operatively. It is imageless and the communication between the PC and the “Pod's” does not require any direct camera view, it is a bluethooth comunication system. This study presents preliminary results utilizing iASSIST™. The aim of this study was to test and compare radiographic alignment, functional outcomes, and perioperative morbidity of the iASSIST™ Knee system versus conventional total knee arthroplasty. In a prospective randomized trial we investigated 60 patients with osteoarthritis of the knee joint. Each surgical procedure was conducted by highly experienced surgeons. In both groups the implant Legacy LPS-Flex Fixed Bearing Knee was used (Zimmer®, Warsaw, Indiana). The groups were equally divided and randomized by hazard. For clinical evaluation, the Short Form-36 and Knee Society Score were obtained. For the radiological assessment mediCAD® Classic, a digital measurement system, was used. The aim of the study was the comparison of results after 3 months.INTRODUCTION
METHODS